• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Exactech

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Exactechsee related information
Date Initiated by FirmAugust 11, 2022
Create DateSeptember 09, 2022
Recall Status1 Open3, Classified
Recall NumberZ-1728-2022
Recall Event ID 90279
510(K)NumberK051556 
Product Classification Prosthesis, hip, semi-constrained, metal/polymer, porous uncemented - Product Code LPH
ProductGXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH GXL 0 DEGREE LINER 28MM SZ E, REF 130-28-25 b. ACUMATCH GXL 0 DEGREE LINER 28MM SZ F, REF 130-28-26 c. ACUMATCH GXL 0 DEGREE LINER 28MM SZ G, REF 130-28-27 d. ACUMATCH GXL 0 DEGREE LINER 28MM SZ H, REF 130-28-28 e. ACUMATCH GXL 0 DEGREE LINER 28MM SZ J, REF 130-28-29 f. ACUMATCH GXL 0 DEGREE LINER 32MM SZ G, REF 130-32-27 g. ACUMATCH GXL 0 DEGREE LINER 32MM SZ H, REF 130-32-28 f. ACUMATCH GXL 0 DEGREE LINER 32MM SZ J, REF 130-32-29 g. ACUMATCH GXL 0 DEGREE LINER 36MM SZ H, REF 130-36-28 h. ACUMATCH GXL 0 DEGREE LINER 36MM SZ J, REF 130-36-29 i. ACUMATCH GXL 0 DEGREE LINER 36MM SZ K, REF 130-36-30 j. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ E, REF 138-28-25 k. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ F, REF 138-28-26 l. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ G, REF 138-28-27 m. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ H, REF 138-28-28 n. ACUMATCH GXL 15 DEG +5 LAT LINER 28MM SZ J, REF 138-28-29 o. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ H, REF 138-36-28 p. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ J, REF 138-36-29 q. ACUMATCH GXL 15 DEG +5 LAT LINER 36MM SZ K, REF 138-36-30 r. ACUMATCH GXL 15 DEG +5 LAT LINER SZG 38MM, REF 138-36-27 s. ACUMATCH GXL 15DEG LINER 28MM SZ D, REF 132-28-24 t. ACUMATCH GXL 15DEG LINER 28MM SZ E, REF 132-28-25 u. ACUMATCH GXL 15DEG LINER 28MM SZ F, REF 132-28-26 v. ACUMATCH GXL 15DEG LINER 28MM SZ G, REF 132-28-27 w. ACUMATCH GXL 15DEG LINER 28MM SZ H, REF 132-28-28 x. ACUMATCH GXL 15DEG LINER 28MM SZ J, REF 132-28-29 y. ACUMATCH GXL 15DEG LINER 28MM SZ K, REF 132-28-30 z. ACUMATCH GXL 15DEG LINER 32MM SZ G, REF 132-32-27 aa. ACUMATCH GXL 15DEG LINER 32MM SZ H, REF 132-32-28 bb. ACUMATCH GXL 15DEG LINER 32MM SZ J, REF 132-32-29 cc. ACUMATCH GXL 15DEG LINER 32MM SZ K, REF 132-32-30 dd.ACUMATCH GXL 15DEG LINER 36MM SZ H, REF 132-36-28 ee. ACUMATCH GXL 15DEG LINER 36MM SZ J, REF 132-36-29 ff. ACUMATCH GXL 15DEG LINER 36MM SZ K, REF 132-36-30 gg. ACUMATCH GXL EXT COV LINER 28MM SZ E, REF 134-28-25 hh. ACUMATCH GXL EXT COV LINER 28MM SZ F, REF 134-28-26 ii. ACUMATCH GXL EXT COV LINER 28MM SZ G, REF 134-28-27 jj. ACUMATCH GXL EXT COV LINER 28MM SZ H, REF 134-28-28 kk. ACUMATCH GXL EXT COV LINER 28MM SZ J, REF 134-28-29 ll. ACUMATCH GXL EXT COV LINER 28MM SZ K, REF 134-28-30 mm. MCS GXL LINER 5/15 DEG 4448X28, REF 104-28-43 nn. MCS GXL LINER 5/15 DEG 5052X28, REF 104-28-44 oo. MCS GXL LINER 5/15 DEG 5458X28, REF 104-28-45 pp. MCS GXL LINER 5/15 DEG 5458X32, REF 104-32-45 qq. MCS GXL LINER 5/15 DEG 6064X32, REF 104-32-46 rr. MCS GXL LINER 5/15 DEG 6064X36, REF 104-36-46 ss. MCS GXL LINER 5/15 DEG 6670X36, REF 104-36-47
Code Information a. REF 130-28-25, GUDID 10885862021984, Serial Numbers: 802593, 802599, 803717, 803737, 841318, 841321, 1572107, 1579369, 1580046, 1583363, 1613871, 1613872, 1613873, 1618998, 1638490, 1638492, 1638494, 1990100, 2282432, 2282433, 2282434; b. REF 130-28-26, GUDID 10885862021991, Serial Numbers: 801865, 801867, 801870, 801872, 801882, 801883, 801886, 801902, 803686, 803690, 803691, 803694, 803695, 803701, 848210, 848213, 848217, 848221, 848222, 848223, 848226, 848232, 848233, 848236, 1677105, 1677106, 1677108, 1677109, 1677110, 1677112, 1677113, 1677114, 1677115, 1677116, 1677118, 1677119, 1677121, 1677122, 1677123, 1677124, 1677126, 1677127, 1677128, 1677129, 1677130, 1677131, 2515330, 2787553, 3871934, 3871935, 3871943; c. REF 130-28-27, GUDID 10885862022004, Serial Numbers: 801542, 801544, 801546, 801547, 801554, 801571, 801572, 801578, 801580, 804033, 804034, 804035, 804036, 862139, 862144, 862146, 1681578, 1681579, 1681580, 1681581, 1681583, 1703643, 1727491, 1727492, 1727493, 1727495, 1727496, 1727497, 1727498, 1727499, 1727500, 1727501, 1727502, 1727503, 1727504, 1727505, 1727506, 1727507, 1727508; d. REF 130-28-28, GUDID 10885862022011, Serial Numbers: 801765; e. REF 130-28-29, GUDID 10885862022028, Serial Numbers: 801654; f. REF 130-32-27, GUDID 10885862022097, Serial Numbers: 966203, 966204, 966205, 966207, 966208, 966209, 966210, 966211, 966212, 966213, 966215, 966216, 966217, 966218, 966219, 966220, 966221, 966222, 966223, 966224, 966225, 966226, 966227, 966228, 966229, 966230, 966231, 966232, 966233, 1042377, 1042378, 1042379, 1042381, 1042382, 1042383, 1042384, 1042385, 1042387, 1042388, 1042391, 1042392, 1042394, 1042395, 1042396, 1042397, 1042398, 1042399, 1042400, 1042401, 1042402, 1042405, 1042406, 1042407, 1042408, 1095431, 1095432, 1095433, 1095434, 1095435, 1095436, 1095437, 1095438, 1095439, 1095440, 1095441, 1095443, 1095444, 1095445, 1095446, 1095447, 1095448, 1095449, 1095537, 1095538, 1095539, 1095541, 1095542, 1095545, 1095546, 1095547, 1095548, 1095550, 1308088, 1308090, 1308092, 1308093, 1308094, 1308095, 1308096, 1308098, 1308102, 1308103, 1308104, 1308105, 1308106, 1308109, 1308111, 1308113, 1308115, 1308116, 1308117, 1308118, 1308119, 1308120, 1984610, 1984611, 1984612, 1984613, 1984614, 1984616, 2053081, 2075035, 2075037, 2075038, 2075039, 2075040, 2075041, 2075042, 2075043, 2075044, 2075045, 2075046, 2075048, 2075049, 2075050, 2075051, 2398494, 2398496, 2398499, 2517474, 2517476, 2517478, 3818434, 3818437, 3818439, 3818444, 3818447; g. REF 130-32-28, GUDID 10885862022103, Serial Numbers: 966299, 966302, 966305, 966307, 966308, 966312, 966313, 966316, 966317, 966318, 966321, 966322, 966327, 1090743, 1090745; f. REF 130-32-29, GUDID 10885862022110, Serial Numbers: 969025, 969026, 969031, 969043; g. REF 130-36-28, GUDID 10885862022202, Serial Numbers: 967669, 967670, 967671, 967672, 967673, 967674, 967675, 967676, 967677, 967678, 967680, 967681, 967683, 967684, 967685, 967686, 967687, 967688, 967689, 967690, 967691, 967692, 967693, 967694, 967695, 967698, 967699, 967700, 967701, 1051661, 1051662, 1051663, 1051664, 1051665, 1051666, 1051667, 1051669, 1051670, 1085957, 1085959, 1085960, 1085961, 1085962, 1085963, 1085964, 1085965, 1085966, 1116002, 1116003, 1116005, 1116006, 1116007, 1116008, 1116009, 1116010, 1116011, 1116012, 1160079, 1160080, 1160081, 1160082, 1160083, 1160084, 1160085, 1160086, 1160087, 1160089, 1206218, 1206219, 1206223, 1206224, 1206225, 1206226, 1668596, 1668597, 1668598, 1668599, 1668600, 1668601, 1668602, 1668603, 1668604, 1668605, 2112376, 2112377, 2112381, 2112382, 2112383, 2112384, 2140027, 2140028, 2504302, 2537040, 2710341, 3871865, 4984351; h. REF 130-36-29, GUDID 10885862022219, Serial Numbers: 975034, 975035, 975036, 975037, 975038, 975040, 975041, 975042, 975043, 975044, 975046, 975047, 975048, 975049, 975050, 975053, 975054, 975055, 975057, 975059, 975063, 1111355, 1111357, 1111358, 1111359, 1111361, 1121576, 1121578, 1121584, 1300472, 1300476, 1300477, 1984741, 1984743, 1984744, 3877076, 3877077, 3877094; i. REF 130-36-30, GUDID 10885862022226, Serial Numbers: 1880259, 1880260, 1880261, 1880262, 1880263, 1880264, 1880265, 1880266, 1880267, 1880268; j. REF 138-28-25, GUDID 10885862024589, Serial Numbers: 805443, 805445, 830281, 830311, 1093756, 1093757, 1632712, 1632713, 1632714, 1632715, 1632716, 1632717, 1662735, 1662736, 1662737, 1667712, 1667715, 1667716, 1667717, 1667718, 1667723, 1667725, 1667726, 1667727, 1667728, 3873889; k. REF 138-28-26, GUDID 10885862024596, Serial Numbers: 805410, 805413, 805421, 805431, 830169, 830177, 830179, 830183, 830184, 830203, 830207, 830209, 830214, 1063813, 1063818, 1063821, 1063830, 1063831, 1737026, 1737027, 1737028, 1737029, 1737030, 1737031, 1737032, 1737033, 2042361, 2042363, 2042364, 2042366, 2042369, 2042372, 2042373, 3840141; l. REF 138-28-27, GUDID 10885862024602, Serial Numbers: 805797, 805799, 805802, 805805, 805806, 805807, 805809, 805810, 805811, 805812, 805816, 805818, 805819, 805820, 805821, 825944, 825949, 825951, 825956, 825961, 825967, 825969, 825970, 825973, 825977, 825978, 825982, 825987, 1063928, 1063941, 2077678, 2077685, 2089964, 3641016, 3873936, 3949986, 3949987; m. REF 138-28-28, GUDID 10885862024619, Serial Numbers: 805355, 805363, 805370, 826263, 826270, 826272, 826275, 1063978; n. REF 138-28-29, GUDID 10885862024626, Serial Numbers: 805827, 805831, 826312, 1063872, 1063889; o. REF 138-36-28, GUDID 10885862024824, Serial Numbers: 807111, 807112, 807113, 807114, 807116, 807119, 807120, 807121, 807122, 807123, 807124, 807125, 807126, 807127, 807128, 807129, 832731, 832732, 832733, 832734, 832735, 832736, 832737, 832738, 832739, 832740, 832742, 832745, 832747, 832748, 832750, 832752, 832754, 832756, 832757, 832758, 832759, 832761, 832763, 832764, 832765, 832767, 832768, 832769, 832770, 832771, 832772, 832773, 832774, 832775, 1019498, 1019499, 1019500, 1019501, 1062910, 1062911, 1062912, 1062913, 1062915, 1062917, 1062918, 1062919, 1062925, 1062927, 1062928, 1223016, 1223017, 1223018, 1223019, 1223022, 1223024, 1223025, 1223026, 1223030, 1223032, 1223033, 1223035, 1223038, 1223039, 1223041, 1900707, 1900708, 1900709, 1900710, 1900711, 1900712, 1920463, 1920464, 1920466, 1920467, 1920468, 1920469, 1920470, 1920471, 1973173, 1973174, 1973175, 1973176, 1973177, 2042392, 2042394, 2313822, 2515477, 2515478, 2800310, 2914514, 2914515, 2914516, 2914517, 2914519, 2914520, 4127080, 4127085, 4127087, 4127089; p. REF 138-36-29, GUDID 10885862024831, Serial Numbers: 807004, 807006, 807013, 807014, 807015, 807017, 807018, 807022, 832683, 832684, 832685, 832686, 832687, 832688, 832690, 832692, 832693, 832694, 832695, 832697, 832698, 832699, 832700, 832706, 832715, 832716, 832717, 832718, 832722, 832723, 832724, 832730, 1086653, 1086658, 1086660, 1086662, 1086665, 1086666, 1086668, 1086672, 1086673, 1636898, 1636900, 1636905, 1636906, 1636908, 1636912, 2102752, 2102757, 2102768, 4127096; q. REF 138-36-30, GUDID 10885862024848, Serial Numbers: 1601226, 1619050, 1619053, 1619054, 1619055. 1619056, 1781527; r. REF 138-36-27, GUDID 10885862024817, Serial Numbers: 1803304; s. REF 132-28-24, GUDID 10885862023063, Serial Numbers: 774713, 775770, 780332, 780345, 780350, 780361, 832888, 832889, 832892, 832893, 832895, 832896, 832897, 832898, 832899, 832907, 832910, 832911, 832913, 832916, 832917, 832918, 832919, 832920, 832921, 832927, 832928, 1425136, 1425137, 1425138, 1425139, 1425140, 1425141, 1425142, 1425143, 1425144, 1461022, 1461023, 1461024, 1461025, 1461026, 1461027, 1461028, 1461029, 1461030, 1461031, 1468006, 1468007, 1468008, 1468011, 1468013, 1500600, 1500601, 1500604, 1500605, 1500606, 1500607, 1506186, 1506187, 1506189, 1506191, 1506193, 1506709, 1506711, 1506712, 1506713, 1506715, 1539850, 1539851, 1547645, 1547646, 1547647, 1547650, 1553669, 1553670, 1553671, 1553673, 1583122, 1583123, 1583124, 1606151, 1606152, 1606155, 1606156, 1606162, 1606163, 1606164, 1606165, 1606166, 1606167, 1606169, 1606170, 1606173, 1606174, 1606175, 1726208, 1726209, 1726211, 1726212, 1726213, 1726214, 1726216, 1726217, 1726218, 1726219, 1726220, 1726221, 1726222, 1726224, 1726225, 1726226, 1726227, 1726228, 1726229, 1726230, 1726232, 1726233, 1726234, 1797506, 1797509, 1797510, 1797511, 1797513, 1797515, 1797516, 1797519, 1797520, 1797521, 1797522, 1797523, 1797524, 1797525, 1797527, 1797528, 1857368, 1857369, 1857370, 1945042, 1945043, 1945044, 1945046, 1945047, 1945048, 1945049, 1945052, 1945054, 1945055, 1945056, 1945057, 1945058, 1945059, 1945061, 1945062, 2420338, 2420339, 2420340, 2420341, 2420342, 2420343, 2574499, 2574500, 2574501, 2574502, 2574503, 2574504, 2574505, 2673229, 2673231, 2673233, 2673235, 2673236, 2673237, 2673238, 2673239, 2673240, 2767967, 2767981, 3863516; t. REF 132-28-25, GUDID 10885862023070, Serial Numbers: 772272, 772273, 772274, 772275, 772277, 772278, 772280, 772281, 772283, 772284, 772285, 774527, 774530, 774531, 774534, 774535, 774536, 780857, 780858, 780859, 780860, 780861, 780862, 780863, 780867, 780870, 781762, 781763, 781765, 781767, 781768, 781769, 795983, 795984, 795987, 795988, 795989, 795990, 795991, 795994, 795996, 795998, 795999, 796001, 796002, 796003, 796005, 796006, 796007, 796008, 796009, 796011, 796012, 796013, 796014, 796015, 796016, 796017, 796019, 796020, 848240, 848242, 848243, 848244, 848245, 848246, 848248, 848252, 848254, 848256, 848258, 848260, 848526, 848527, 848529, 848530, 848534, 848536, 848537, 848538, 848539, 848541, 848542, 848543, 848545, 848546, 848547, 848549, 848550, 848551, 848554, 848555, 848556, 848558, 848560, 848562, 848563, 848565, 848566, 848567, 848568, 848569, 848570, 848571, 848572, 848573, 848575, 848576, 848577, 1061548, 1061549, 1061550, 1061551, 1061552, 1061553, 1061554, 1061555, 1061556, 1061557, 1061558, 1061559, 1061561, 1061562, 1061567, 1061568, 1061569, 1061570, 1061571, 1061572, 1061573, 1061574, 1061575, 1061576, 1061577, 1061578, 1061579, 1061580, 1160004, 1160005, 1160006, 1160007, 1160008, 1160009, 1160010, 1160011, 1160012, 1160013, 1160014, 1160015, 1160016, 1160017, 1160018, 1160019, 1160020, 1160021, 1160022, 1160023, 1160024, 1160025, 1160026, 1160027, 1160028, 1160029, 1160030, 1160031, 1160032, 1160033, 1160034, 1160035, 1160036, 1202519, 1202520, 1202521, 1202522, 1202523, 1202524, 1202525, 1202526, 1202527, 1202528, 1202529, 1202530, 1202531, 1202532, 1202533, 1202534, 1202535, 1202536, 1202537, 1202538, 1202539, 1202540, 1202541, 1202542, 1202543, 1202544, 1202545, 1202546, 1202547, 1202548, 1202549, 1202550, 1202551, 1202552, 1202553, 1202554, 1202555, 1202556, 1377964, 1377965, 1377966, 1377967, 1377968, 1377969, 1377970, 1377971, 1377972, 1377973, 1380135, 1380136, 1380137, 1380138, 1380139, 1431888, 1431889, 1431890, 1431891, 1431892, 1431893, 1431894, 1431895, 1431896, 1431897, 1433056, 1433057, 1433058, 1433059, 1433060, 1433061, 1433062, 1458382, 1458383, 1458384, 1458385, 1458386, 1458387, 1458388, 1458389, 1458390, 1458391, 1458392, 1458393, 1458394, 1458395, 1458396, 1458847, 1458848, 1458849, 1458850, 1458851, 1458852, 1458853, 1489124, 1489125, 1489126, 1489128, 1489129, 1489130, 1489131, 1489132, 1489133, 1489134, 1489135, 1489136, 1489137, 1489138, 1489140, 1489141, 1489142, 1489143, 1489144, 1489145, 1489146, 1489147, 1489148, 1489149, 1489150, 1502678, 1502679, 1502683, 1502684, 1502685, 1502686, 1502687, 1502688, 1502689, 1502690, 1502693, 1502694, 1502695, 1502696, 1502698, 1502699, 1502700, 1502701, 1527446, 1527447, 1527448, 1527449, 1527450, 1527452, 1527453, 1527454, 1527455, 1527456, 1527457, 1527458, 1527459, 1527460, 1527461, 1527462, 1527463, 1527464, 1527465, 1527466, 1527468, 1527469, 1527470, 1586106, 1586108, 1586109, 1586110, 1586111, 1586112, 1586113, 1586114, 1586115, 1586116, 1586117, 1586118, 1586119, 1586120, 1586121, 1586122, 1586123, 1586125, 1586127, 1586128, 1586129, 1586130, 1586131, 1586133, 1586134, 1586135, 1586136, 1586137, 1586138, 1586139, 1586140, 1586141, 1586142, 1586143, 1586144.1586145, 1586146, 1587461, 1587462, 1587463, 1587464, 1587465, 1587466, 1587467, 1587468, 1587471, 1587472, 1587474, 1587475, 1587476, 1587477, 1587478, 1587479, 1587480, 1587481, 1587482, 1587483, 1587486, 1587488, 1587489, 1587490, 1587491, 1587492, 1587493, 1587494, 1587496, 1587497, 1713627, 1713628, 1713629, 1713630, 1713631, 1713632, 1713633, 1713634, 1713635, 1713636, 1713637, 1713638, 1713639, 1713640, 1713641, 1713642, 1713643, 1713644, 1713645, 1713646, 1713647, 1713648, 1713649, 1713650, 1713651, 1713652, 1779520, 1779521, 1779522, 1779524, 1779525, 1779527, 1779528, 1779529, 1779530, 1779531, 1779532, 1779533, 1779534, 1779536, 1779537, 1779539, 1779540, 1779541, 1779542, 1779543, 1779545, 1779546, 1781473, 1781474, 1781476, 1781477, 1781478, 1781479, 1781480, 1781481, 1781482, 1781483, 1781484, 1781485, 1781486, 1781487, 1781488, 1781489, 1781490, 1781491, 1781492, 1781493, 1781494, 1781495, 1781496, 1781497, 1781498, 1781499, 1781500, 1781501, 1781503, 1781504, 1781505, 1781506, 1781507, 1781508, 1781509, 1781510, 1781512, 1781514, 1781515, 1781516, 1781517, 1781518, 1781519, 1781520, 1781521, 1781522, 1781523, 1781524, 1873693, 1873694, 1873696, 1873697, 1873698, 1873699, 1873700, 1873701, 1873702, 1873703, 1873704, 1873705, 1873706, 1873707, 1873708, 1873709, 1873710, 1873711, 1873712, 1873713, 1873714, 1873715, 1873716, 1873717, 1873718, 1873719, 1873720, 1873721, 1873722, 1873723, 1873724, 2244826, 2244827, 2244828, 2244829, 2244830, 2244831, 2244832, 2244833, 2244834, 2244835, 2244836, 2244837, 2244838, 2244839, 2244840, 2244841, 2244842, 2244843, 2244844, 2244845, 2244846, 2244847, 2244848, 2244849, 2244850, 2244851, 2244852, 2396247, 2396248, 2396249, 2396250, 2396251, 2396252, 2396253, 2396254, 2396255, 2396256, 2396257, 2396258, 2396259, 2396260, 2396261, 2396262, 2396263, 2396264, 2396265, 2396266, 2396267, 2396268, 2396269, 2396270, 2396271, 2396272, 2396273, 2575428, 2575429, 2575430, 2575431, 2575432, 2575433, 2575434, 2575435, 2575436, 2575437, 2575438, 2575439, 2575440, 2575441, 2575442, 2575443, 2575444, 2575445, 2575446, 2575447, 2575449, 2575450, 2575451, 2575452, 2575453, 2575454, 2687702, 2687703, 2687704, 2687705, 2687706, 2687707, 2687709, 2687710, 2687711, 2687712, 2687713, 2687716, 2687717, 2687720, 2687722, 2687723, 2687724, 2687725, 2687726, 2847932, 2847934, 2847935, 2847936, 2847937, 2847938, 2847939, 2847941, 2847942, 2847943, 2993028, 2993030, 2993031, 2993033, 2993034, 2993035, 2993036, 2993037, 2993040, 2993041, 2993042, 3558245, 3558247, 3558249, 3558262, 3558265, 3665411, 3689528, 3689529, 3689535, 3689536, 3852374, 3852376, 4127030, 4127032, 4127037, 4127044, 5246611, 5246612, 5246613, 5246614, 5246615, 6230836, 6230837; u. REF 132-28-26, GUDID 10885862023087, Serial Numbers: 768376, 768377, 768378, 768379, 768380, 768381, 768382, 768383, 768384, 768385, 768386, 768387, 775736, 775739, 775740, 775741, 775742, 775743, 775744, 775745, 775747, 775748, 775749, 775750, 775751, 775752, 775753, 775755, 775756, 775757, 775759, 775760, 775761, 775762, 775763, 779211, 779212, 779213, 779214, 779215, 779217, 779219, 779220, 779221, 779222, 779223, 779224, 779225, 779226, 779227, 779229, 779230, 779231, 779232, 779233, 779234, 779235, 779236, 779237, 779238, 779239, 779241, 779242, 779244, 779245, 779246, 779247, 779250, 779251, 779252, 779253, 779254, 779255, 779256, 779257, 779258, 779259, 779260, 779261, 779262, 779263, 779264, 779266, 779267, 779268, 779269, 779271, 779272, 779273, 779274, 779275, 779276, 779277, 779278, 779279, 779280, 779281, 779282, 779283, 779284, 779285, 779286, 779287, 779290, 779291, 779293, 779294, 779295, 779296, 779297, 779298, 779299, 779300, 779301, 779302, 779303, 779304, 779305, 779306, 779307, 779308, 779309, 779310, 779311, 779312, 779313, 779314, 779315, 779316, 779317, 779318, 779319, 779320, 779321, 779322, 780895, 780896, 780897, 841129, 841130, 841131, 841132, 841133, 841134, 841135, 841136, 841137, 841138, 841141, 841142, 841143, 841144, 841145, 841146, 841147, 841148, 841149, 841150, 841151, 841152, 841153, 841154, 841155, 841156, 841157, 841158, 841159, 841160, 841161, 841162, 841163, 841164, 841167, 841168, 841169, 841170, 841171, 841172, 841173, 841174, 841175, 841176, 841177, 841178, 841939, 841940, 841941, 841942, 841943, 841944, 841945, 841946, 841947, 841948, 841950, 841951, 841952, 841953, 841954, 841955, 841956, 841957, 841958, 841959, 841960, 841961, 841962, 841963, 841964, 841966, 841967, 841968, 841969, 841970, 841971, 841972, 841973, 841974, 841976, 841978, 841979, 841980, 841982, 841984, 841985, 841986, 841987, 841988, 841989, 841990, 841991, 850008, 850009, 850010, 850013, 850014, 850017, 850019, 850020, 850021, 850022, 850023, 850024, 850025, 850026, 850027, 850028, 850029, 850030, 850031, 850032, 850033, 850034, 850035, 850036, 850037, 850038, 850039, 850040, 850041, 850042, 850044, 850045, 850046, 850047, 850048, 850049, 850050, 850051, 850052, 850054, 1051615, 1051616, 1051617, 1051618, 1051619, 1051620, 1051621, 1051622, 1051623, 1051624, 1051625, 1051626, 1051627, 1051628, 1051630, 1051631, 1051632, 1051633, 1051634, 1051635, 1051636, 1051638, 1051640, 1051641, 1051642, 1051643, 1051644, 1051645, 1051646, 1087087, 1087088, 1087089, 1087090, 1087091, 1087094, 1087095, 1087096, 1087097, 1087099, 1087100, 1087101, 1087102, 1087103, 1087104, 1087105, 1087106, 1087108, 1087110, 1087111, 1087112, 1087113, 1087114, 1087115, 1087116, 1087117, 1087118, 1087119, 1145083, 1145084, 1145085, 1145086, 1145087, 1145088, 1145089, 1145090, 1145091, 1145092, 1145093, 1145094, 1145095, 1145096, 1145097, 1145098, 1145099, 1145100, 1145101, 1145102, 1145103, 1145104, 1145105, 1145106, 1145107, 1145108, 1145109, 1202558, 1202559, 1202560, 1202561, 1202562, 1202563, 1202564, 1202565, 1202566, 1202567, 1202568, 1202569, 1202570, 1202571, 1202572, 1202573, 1202574, 1202575, 1202576, 1202577, 1202578, 1202579, 1202580, 1202581, 1202582, 1202583, 1202584, 1202585, 1202586, 1202587, 1202588, 1202589, 1342362, 1342363, 1342364, 1342365, 1342366, 1342367, 1342368, 1342369, 1342370, 1342371, 1342372, 1342373, 1342374, 1342375, 1342376, 1342377, 1342378, 1342379, 1342380, 1342381, 1358274, 1358275, 1358276, 1358277, 1358279, 1358280, 1358281, 1358282, 1358283, 1358284, 1358285, 1358286, 1377442, 1377443, 1377444, 1377445, 1377446, 1377447, 1377448, 1377449, 1377450, 1377451, 1377452, 1377453, 1377454, 1377455, 1377456, 1377457, 1377458, 1377459, 1377460, 1377461, 1377462, 1377463, 1377464, 1377465, 1377466, 1377467, 1377468, 1377469, 1377470, 1377471, 1377472, 1377473, 1377474, 1404991, 1404992, 1404993, 1404994, 1404995, 1404996, 1404997, 1404998, 1404999, 1405000, 1405001, 1405002, 1405003, 1405004, 1405005, 1405006, 1405007, 1405008, 1405009, 1405010, 1405011, 1405012, 1405013, 1405014, 1405015, 1405016, 1405017, 1405018, 1405019, 1405020, 1405021, 1405022, 1405023, 1405024, 1405025, 1405026, 1405027, 1405028, 1405029, 1405030, 1405031, 1405032, 1405033, 1405034, 1405035, 1405036, 1405037, 1405038, 1405039, 1405040, 1405041, 1405042, 1405043, 1405044, 1405045, 1405046, 1405047, 1405048, 1405049, 1405050, 1405051, 1405052, 1405053, 1405054, 1407084, 1407085, 1407086, 1407087, 1407088, 1407089, 1407090, 1407091, 1407092, 1407093, 1407094, 1407095, 1407096, 1407097, 1407098, 1407099, 1407100, 1407101, 1407102, 1407103, 1407104, 1407105, 1407106, 1407107, 1407108, 1407109, 1407110, 1407111, 1407112, 1407113, 1407114, 1407115, 1407116, 1407117, 1407118, 1407119, 1407120, 1407121, 1407122, 1407123, 1407124, 1407125, 1407126, 1407127, 1407128, 1407129, 1407130, 1407131, 1407132, 1407133, 1407134, 1407135, 1407136, 1407137, 1407138, 1407139, 1407140, 1407141, 1407142, 1407143, 1407144, 1407145, 1407146, 1407147, 1407148, 1479380, 1479381, 1479382, 1479383, 1479384, 1479385, 1479386, 1479387, 1479388, 1479389, 1479390, 1479391, 1479392, 1479393, 1479394, 1479395, 1479396, 1479397, 1479398, 1479399, 1479400, 1479401, 1479402, 1479403, 1479404, 1479405, 1479406, 1479407, 1479408, 1479409, 1479410, 1479411, 1479412, 1479413, 1479414, 1479415, 1479416, 1479417, 1479418, 1479419, 1479420, 1479421, 1479422, 1479423, 1479424, 1479425, 1479426, 1479427, 1479428, 1479429, 1489416, 1489417, 1489418, 1489419, 1489420, 1489421, 1489422, 1489423, 1489424, 1489425, 1489426, 1489427, 1489428, 1489429, 1489430, 1489431, 1489432, 1489433, 1489434, 1489435, 1489436, 1489437, 1489438, 1489439, 1489440, 1489441, 1489442, 1489443, 1489444, 1489445, 1489446, 1489447, 1489448, 1489449, 1489450, 1489451, 1489452, 1489453, 1489454, 1489455, 1489456, 1489457, 1489458, 1489459, 1489460, 1489461, 1489462, 1489463, 1489464, 1489465, 1489466, 1489467, 1489468, 1489469, 1489470, 1500514, 1500515, 1500516, 1500517, 1500518, 1500519, 1500520, 1500521, 1500522, 1500523, 1500524, 1500525, 1500526, 1500527, 1500528, 1500529, 1500530, 1500531, 1500532, 1500533, 1500534, 1500535, 1500536, 1500537, 1500538, 1506059, 1506060, 1506061, 1506062, 1506063, 1506064, 1506065, 1506066, 1506067, 1506068, 1506069, 1506070, 1506071, 1506072, 1506073, 1506074, 1506075, 1506076, 1506077, 1506078, 1506079, 1506080, 1506081, 1506082, 1506083, 1506084, 1506085, 1506086, 1506087, 1506088, 1506089, 1506090, 1506091, 1506092, 1506093, 1506094, 1506095, 1506096, 1506097, 1506098, 1506099, 1506100, 1506101, 1506102, 1506103, 1506104, 1506105, 1506106, 1506107, 1506108, 1511460, 1511461, 1511462, 1511463, 1511464, 1511465, 1511466, 1511467, 1511468, 1511469, 1511470, 1511471, 1511472, 1511473, 1511474, 1511475, 1511476, 1511477, 1511478, 1511479, 1511480, 1511481, 1511482, 1511483, 1511484, 1511485, 1511486, 1511487, 1511488, 1511489, 1511490, 1511491, 1511492, 1511493, 1511494, 1511495, 1511496, 1511497, 1511498, 1511499, 1511500, 1511501, 1511502, 1511503, 1511504, 1511505, 1511506, 1511507, 1511508, 1511509, 1543436, 1543437, 1543438, 1543439, 1543440, 1543441, 1543442, 1543443, 1543444, 1543445, 1543446, 1543447, 1543448, 1543449, 1543450, 1543451, 1543452, 1543453, 1543454, 1543455, 1543456, 1543457, 1543458, 1543459, 1543460, 1543461, 1543462, 1543463, 1543464, 1543465, 1543466, 1543467, 1543468, 1543469, 1543470, 1543471, 1543472, 1543473, 1543474, 1543475, 1543476, 1543477, 1543478, 1543479, 1543480, 1543481, 1543482, 1543483, 1543484, 1543485, 1571042, 1571043, 1571044, 1571045, 1571046, 1571047, 1571048, 1571049, 1571050, 1571051, 1571052, 1571053, 1571054, 1571055, 1571056, 1571057, 1571058, 1571059, 1571060, 1571061, 1571062, 1571063, 1571064, 1571065, 1571066, 1571067, 1571068, 1571069, 1571070, 1571071, 1571072, 1571073, 1571074, 1583131, 1583132, 1583133, 1583134, 1583135, 1583136, 1583137, 1583138, 1583139, 1583140, 1583141, 1583142, 1583143, 1583144, 1583145, 1583146, 1583147, 1583148, 1583149, 1583150, 1583151, 1583152, 1583153, 1583154, 1583155, 1583156, 1583157, 1583158, 1583159, 1583160, 1583161, 1583162, 1583163, 1583164, 1583165, 1583166, 1583167, 1583168, 1583169, 1583170, 1583171, 1583172, 1583173, 1583174, 1583175, 1583176, 1583177, 1583178, 1583179, 1583180, 1583181, 1583183, 1619735, 1619736, 1619737, 1619738, 1619739, 1619740, 1619741, 1619742, 1619743, 1619744, 1619745, 1619746, 1619747, 1619748, 1619749, 1619750, 1619751, 1619752, 1619753, 1619754, 1619755, 1619756, 1619757, 1619758, 1619759, 1619760, 1619761, 1619762, 1619763, 1619764, 1619765, 1619766, 1619767, 1619768, 1621143, 1621144, 1621145, 1621146, 1621147, 1621148, 1621149, 1621150, 1621151, 1621152, 1621153, 1621154, 1621155, 1621156, 1621157, 1621158, 1621159, 1621160, 1621161, 1621162, 1621163, 1621164, 1621165, 1621166, 1621167, 1621168, 1621169, 1621170, 1621171, 1621172, 1621173, 1621174, 1621175, 1621176, 1621177, 1621178, 1621179, 1621180, 1621181, 1621182, 1621183, 1621184, 1621185, 1621186, 1621187, 1621188, 1621189, 1621190, 1621191, 1621192, 1621193, 1621194, 1621195, 1626080, 1626081, 1626082, 1626083, 1626084, 1626085, 1626086, 1626087, 1626088, 1626089, 1626090, 1626091, 1626092, 1626093, 1626094, 1626095, 1626096, 1626097, 1626098, 1626099, 1626100, 1626101, 1626102, 1626103, 1626104, 1626105, 1626106, 1626107, 1626108, 1626109, 1626110, 1626111, 1626112, 1626113, 1626114, 1626115, 1626116, 1626117, 1626118, 1626119, 1626120, 1626121, 1626123, 1626124, 1626125, 1626126, 1626127, 1626128, 1626129, 1668977, 1668978, 1668979, 1668980, 1668981, 1668982, 1668983, 1668984, 1668985, 1668986, 1668987, 1668988, 1668989, 1668990, 1668991, 1668992, 1668993, 1668994, 1668995, 1668996, 1668997, 1668998, 1668999, 1669000, 1669001, 1669002, 1669003, 1669004, 1669005, 1669006, 1669007, 1669008, 1669009, 1669010, 1669011, 1669012, 1669013, 1669015, 1669016, 1669017, 1669018, 1669020, 1669021, 1669022, 1669023, 1669024, 1669025, 1669026, 1669027, 1677061, 1677062, 1677063, 1677064, 1677065, 1677066, 1677067, 1677068, 1677069, 1677070, 1677071, 1677072, 1677073, 1677074, 1677075, 1677076, 1677077, 1677078, 1677079, 1677080, 1677081, 1677082, 1677083, 1677084, 1677085, 1677086, 1677087, 1677088, 1677089, 1677090, 1677091, 1677092, 1677093, 1677094, 1677095, 1677096, 1677097, 1677098, 1677099, 1677100, 1677101, 1677102, 1677103, 1677104, 1683338, 1683339, 1683340, 1683341, 1683342, 1683343, 1683344, 1683345, 1683346, 1683347, 1683348, 1683349, 1683350, 1683351, 1683352, 1683353, 1683354, 1683355, 1683356, 1683357, 1683358, 1683359, 1683360, 1683361, 1683362, 1683363, 1683364, 1683365, 1683366, 1683367, 1683368, 1683369, 1683370, 1683371, 1683372, 1683373, 1683374, 1683375, 1683376, 1683377, 1683378, 1683379, 1683380, 1683381, 1683382, 1683383, 1683384, 1683385, 1761552, 1761553, 1761554, 1761555, 1761556, 1761557, 1761558, 1761559, 1761560, 1761561, 1761562, 1761563, 1761564, 1761565, 1761566, 1761567, 1761568, 1761569, 1761570, 1761571, 1761572, 1761573, 1761574, 1761575, 1761576, 1761577, 1762315, 1762316, 1762317, 1762318, 1762319, 1762320, 1762321, 1762322, 1762323, 1762324, 1762325, 1762326, 1762327, 1762328, 1762329, 1762330, 1762331, 1762332, 1762333, 1762334, 1762335, 1762336, 1762337, 1762338, 1762339, 1762907, 1762908, 1762909, 1762910, 1762911, 1762912, 1762913, 1762914, 1762915, 1762916, 1762917, 1762918, 1762919, 1762920, 1762921, 1762922, 1762923, 1762924, 1762925, 1762926, 1762927, 1762928, 1762929, 1762930, 1762931, 1762932, 1762933, 1762934, 1762935, 1762936, 1762937, 1762938, 1762939, 1765130, 1765131, 1765132, 1765133, 1765134, 1765135, 1765136, 1765137, 1765138, 1765139, 1765140, 1765141, 1765142, 1765143, 1765144, 1765145, 1765146, 1765147, 1765148, 1765149, 1765150, 1765151, 1765152, 1765153, 1765154, 1765155, 1765156, 1765157, 1765158, 1765159, 1765160, 1765161, 1765162, 1803258, 1803259, 1803260, 1803261, 1803262, 1803263, 1803264, 1803265, 1803266, 1803267, 1803268, 1803269, 1803270, 1803271, 1803272, 1803273, 1803274, 1803275, 1803276, 1803277, 1803278, 1803279, 1803280, 1803281, 1803282, 1803283, 1803284, 1803285, 1803286, 1803287, 1803288, 1803289, 1803290, 1803291, 1803292, 1803293, 1803294, 1803295, 1803296, 1803297, 1803298, 1803299, 1803300, 1803301, 1803302, 1804007, 1804008, 1804009, 1804010, 1804011, 1804012, 1804013, 1804014, 1804015, 1804016, 1804017, 1804018, 1804019, 1804020, 1804021, 1804022, 1804023, 1804024, 1804025, 1804026, 1804027, 1804028, 1807087, 1807088, 1807089, 1807090, 1807091, 1807092, 1807093, 1807094, 1807095, 1807096, 1807097, 1807098, 1807099, 1807100, 1807101, 1807102, 1807103, 1807104, 1807105, 1807106, 1807107, 1807108, 1807109, 1807110, 1807111, 1807112, 1807113, 1807114, 1814932, 1814933, 1814934, 1814935, 1814936, 1814937, 1814938, 1814939, 1814940, 1814941, 1814942, 1814943, 1814944, 1814945, 1814946, 1814947, 1814948, 1814949, 1814950, 1814951, 1814952, 1814953, 1814954, 1814955, 1814956, 1814957, 1814958, 1814959, 1814960, 1814961, 1814962, 1814963, 1814964, 1814965, 1814966, 1814967, 1814968, 1814969, 1814970, 1814971, 1814972, 1814973, 1814974, 1814975, 1814976, 1814977, 1814978, 1814979, 1814980, 1814981, 1814982, 1814983, 1814984, 1814985, 1815284, 1815285, 1815286, 1815287, 1815288, 1815289, 1815290, 1815291, 1815292, 1815293, 1815294, 1815295, 1815296, 1815297, 1815298, 1815299, 1815300, 1815301, 1815302, 1815303, 1815304, 1815305, 1815306, 1815307, 1815308, 1815309, 1815310, 1815311, 1815312, 1815313, 1815314, 1815315, 1815316, 1815317, 1815318, 1815319, 1815320, 1815321, 1815322, 1815323, 1815324, 1815325, 1815326, 1815327, 1815328, 1815329, 1815330, 1815331, 1815332, 1815333, 1815334, 1815335, 1815336, 1815337, 1870519, 1870520, 1870521, 1870522, 1870523, 1870524, 1870525, 1870526, 1870527, 1870528, 1870529, 1870530, 1870531, 1870532, 1870533, 1870534, 1870535, 1870536, 1870537, 1870538, 1870539, 1870540, 1870541, 1870542, 1870543, 1870544, 1870545, 1870546, 1870547, 1870548, 1870549, 1870550, 1870551, 1870552, 1870553, 1870554, 1870555, 1870556, 1870557, 1870558, 1870559, 1870560, 1870561, 1870562, 1870563, 1870564, 1870565, 1870566, 1870567, 1870568, 1870569, 1870570, 1870571, 1870572, 1870573, 1870574, 1870591, 1870592, 1870593, 1870594, 1870595, 1870596, 1870597, 1870598, 1870599, 1870600, 1870601, 1870602, 1870603, 1870604, 1870605, 1870606, 1870607, 1870608, 1870609, 1870610, 1870611, 1870613, 1870614, 1870615, 1870616, 1870617, 1870618, 1870619, 1870620, 1870621, 1870622, 1870623, 1870624, 1870625, 1870626, 1870627, 1870628, 1870629, 1870630, 1870631, 1870632, 1870633, 1870634, 1870635, 1870636, 1870637, 1870638, 1870639, 1870640, 1870673, 1870674, 1870675, 1870676, 1870677, 1870678, 1870679, 1870680, 1870681, 1870682, 1870683, 1870684, 1870685, 1870686, 1870687, 1870688, 1870689, 1870690, 1870691, 1870692, 1870693, 1870694, 1870695, 1870696, 1870697, 1870698, 1870699, 1870700, 1870701, 1870702, 1870703, 1870704, 1870705, 1870706, 1870707, 1870708, 1870709, 1870710, 1870711, 1870712, 1870713, 1870714, 1870715, 1870716, 1870717, 1870718, 1870719, 1870720, 1870721, 1870722, 1872812, 1872813, 1872814, 1872815, 1872816, 1872817, 1872818, 1872819, 1872820, 1872821, 1872822, 1872823, 1872824, 1872825, 1872826, 1872827, 1872828, 1872829, 1872830, 1872831, 1872832, 1872833, 1872834, 1872835, 1872836, 1872837, 1872838, 1872839, 1872840, 1872841, 1872842, 1872843, 1872844, 1872845, 1872846, 1872847, 1872848, 1872849, 1872850, 1872851, 1872852, 1923017, 1923018, 1923019, 1923020, 1923021, 1923022, 1923023, 1923024, 1923025, 1923026, 1923027, 1923028, 1923029, 1923030, 1923031, 1923032, 1923033, 1923034, 1923035, 1923036, 1923037, 1923038, 1923039, 1923040, 1923041, 1923042, 1923043, 1923044, 1923045, 1923046, 1923047, 1923048, 1923049, 1923050, 1923051, 1923052, 1923053, 1923054, 1923055, 1923056, 1923057, 1923058, 1923059, 1923060, 1923061, 1923062, 1923063, 1923064, 1923065, 1923066, 1923067, 1923068, 1923069, 1923070, 1923071, 1923072, 1923073, 1923074, 1923075, 1923076, 1923077, 1923078, 1923079, 1923080, 1923081, 1923082, 1923083, 1923084, 1923085, 1923086, 1923087, 1923088, 1923089, 1923090, 1923091, 1923092, 1923093, 1923094, 1923095, 1923096, 1923097, 1923098, 1923099, 1923100, 1923101, 1923102, 1923103, 1923104, 1923105, 1923106, 1923107, 1923108, 1923109, 1923110, 1923111, 1923112, 1923113, 1923114, 1923115, 1923116, 1923117, 1923118, 1923119, 1923120, 1923121, 1923122, 1923123, 1923124, 1923125, 1952208, 1952209, 1952210, 1952211, 1952212, 1952213, 1952214, 1952215, 1952216, 1952217, 1952218, 1952219, 1952220, 1952221, 1952222, 1952223, 1952224, 1952225, 1952226, 1952227, 1952228, 1952229, 1952230, 1952231, 1952232, 1952233, 1952234, 1952235, 1952236, 1952237, 1952238, 1952239, 1952240, 1952241, 1952242, 1952243, 1952244, 1952245, 1952246, 1952247, 1952248, 1952249, 1952250, 1952251, 1952252, 1952253, 1952254, 1952255, 1952256, 1952257, 1952258, 1952259, 1952260, 1952261, 1952262, 1952263, 1952264, 1952265, 1952266, 1952267, 1953747, 1953748, 1953749, 1953750, 1953751, 1953752, 1953753, 1953754, 1953755, 1953756, 1953757, 1953758, 1953759, 1953760, 1953761, 1953762, 1953763, 1953764, 1953765, 1953766, 1953767, 1953768, 1953769, 1953770, 1953771, 1953772, 1953773, 1953774, 1953775, 1953776, 1953777, 1953778, 1953779, 1953780, 1953781, 1953782, 1953783, 1953784, 1953785, 1953786, 1953787, 1953788, 1953789, 1953790, 1953791, 1953792, 1953793, 1953794, 1953795, 1953796, 1953797, 1953798, 1953799, 1953800, 1953801, 1953802, 1953803, 1953804, 1953805, 1953806, 2056430, 2056431, 2056432, 2056433, 2056434, 2056435, 2056436, 2056437, 2056438, 2056439, 2056440, 2056441, 2056442, 2056443, 2056444, 2056445, 2056446, 2056447, 2056448, 2056449, 2056450, 2056451, 2056452, 2056453, 2056454, 2056455, 2056456, 2056457, 2056458, 2056459, 2056460, 2056461, 2056462, 2056463, 2056464, 2056465, 2056466, 2056467, 2056468, 2056469, 2056470, 2056471, 2056472, 2056473, 2056474, 2056475, 2056476, 2056477, 2056478, 2056480, 2056481, 2056482, 2056483, 2056484, 2056887, 2056888, 2056889, 2056890, 2056891, 2056892, 2056893, 2056894, 2056895, 2056896, 2056897, 2056898, 2056899, 2056900, 2056901, 2056902, 2056903, 2056904, 2056905, 2056906, 2056907, 2056908, 2056909, 2056910, 2056911, 2056912, 2056913, 2056914, 2056915, 2056916, 2056917, 2056918, 2056919, 2056920, 2056921, 2056922, 2056923, 2056924, 2056925, 2056926, 2056927, 2056928, 2056929, 2056930, 2056931, 2056932, 2056933, 2056934, 2056936, 2056937, 2056938, 2056939, 2056940, 2056941, 2056992, 2056993, 2056994, 2056995, 2056996, 2056997, 2056998, 2056999, 2057000, 2057001, 2057002, 2057003, 2057004, 2057005, 2057006, 2057007, 2057008, 2057009, 2057010, 2057011, 2057012, 2057013, 2057014, 2057015, 2057016, 2057017, 2057018, 2057019, 2057020, 2057021, 2057022, 2057023, 2057024, 2057025, 2057026, 2057027, 2057028, 2057029, 2057030, 2057031, 2057032, 2057033, 2057034, 2057035, 2057036, 2057037, 2057038, 2057039, 2057040, 2057041, 2057109, 2057110, 2057111, 2057112, 2057113, 2057114, 2057115, 2057116, 2057117, 2057118, 2057119, 2057120, 2057122, 2057123, 2057124, 2057125, 2057126, 2057127, 2057128, 2057129, 2057130, 2057131, 2057132, 2057133, 2057134, 2057135, 2057136, 2057137, 2057138, 2057139, 2057140, 2057141, 2057142, 2057143, 2057144, 2057145, 2057146, 2057147, 2057148, 2057149, 2057150, 2057151, 2057152, 2057153, 2057154, 2057155, 2057156, 2057157, 2057158, 2107011, 2107012, 2107013, 2107014, 2107015, 2107016, 2107017, 2107018, 2107019, 2107020, 2107021, 2107022, 2107023, 2107024, 2107025, 2107026, 2107027, 2107028, 2107029, 2107030, 2107031, 2107032, 2107033, 2107034, 2107035, 2107036, 2107037, 2107038, 2107039, 2107040, 2107041, 2107043, 2107044, 2107045, 2107046, 2107047, 2107048, 2107049, 2107050, 2107051, 2107052, 2107053, 2107054, 2107055, 2107056, 2107057, 2107058, 2107059, 2107060, 2112458, 2112459, 2112460, 2112461, 2112462, 2112463, 2116108, 2116109, 2116110, 2116111, 2116112, 2116113, 2127985, 2127986, 2127987, 2127988, 2127989, 2127990, 2140640, 2140642, 2140643, 2140644, 2140645, 2140646, 2140647, 2141862, 2141863, 2141864, 2141865, 2141866, 2141867, 2141868, 2141869, 2141870, 2141871, 2141872, 2141873, 2141874, 2141875, 2162771, 2162772, 2162773, 2162774, 2162776, 2162777, 2162778, 2162779, 2162780, 2162781, 2162782, 2162783, 2162784, 2162785, 2162786, 2162787, 2162788, 2162789, 2162791, 2162792, 2162793, 2162794, 2162795, 2162797, 2162798, 2162800, 2162801, 2162802, 2162803, 2162804, 2162805, 2162806, 2162807, 2162808, 2162809, 2162810, 2162811, 2162812, 2162813, 2162814, 2162815, 2162816, 2162817, 2162827, 2162828, 2162829, 2162830, 2162832, 2162833, 2162834, 2162835, 2162836, 2162837, 2162838, 2162839, 2162841, 2162842, 2162843, 2162845, 2162846, 2162847, 2162849, 2162850, 2162852, 2162855, 2162856, 2162859, 2162860, 2162861, 2162862, 2162863, 2162864, 2162865, 2162867, 2162868, 2162869, 2162870, 2162871, 2162872, 2162873, 2162875, 2162876, 2162878, 2162879, 2162880, 2162881, 2162882, 2162883, 2162884, 2162885, 2186883, 2186884, 2186885, 2186886, 2186887, 2186888, 2186889, 2186890, 2186891, 2186892, 2186893, 2186894, 2186895, 2186896, 2186897, 2186898, 2186899, 2186900, 2186901, 2186902, 2219932, 2219933, 2219934, 2219935, 2219936, 2219937, 2219938, 2219939, 2219940, 2219941, 2224130, 2224131, 2224132, 2224133, 2224134, 2224135, 2224136, 2224137, 2224138, 2224139, 2224140, 2224141, 2224142, 2224143, 2224144, 2224145, 2224146, 2224147, 2224148, 2224149, 2224150, 2224151, 2224152, 2224154, 2224155, 2257902, 2257903, 2257904, 2257905, 2257906, 2257907, 2257908, 2257909, 2257910, 2257911, 2257912, 2257913, 2257914, 2257915, 2257916, 2257917, 2257918, 2257919, 2257920, 2257921, 2257922, 2257923, 2257924, 2257925, 2257926, 2257927, 2257928, 2262631, 2262632, 2262633, 2262634, 2262635, 2262636, 2262637, 2262638, 2262639, 2262640, 2262641, 2262642, 2262643, 2262644, 2262645, 2262646, 2262647, 2262648, 2262649, 2262650, 2262651, 2262652, 2262653, 2262654, 2262655, 2262656, 2262657, 2299275, 2299276, 2299277, 2299278, 2299281, 2299283, 2299284, 2299286, 2299287, 2299288, 2299468, 2299469, 2299472, 2299474, 2299475, 2299476, 2299480, 2299483, 2299484, 2299486, 2299488, 2299489, 2471885, 2471888, 2471895, 2471896, 2471897, 2471898, 2471902, 2471905, 2471906, 2471907, 2471908, 2471910, 2471911, 2524743, 2524745, 2524746, 2524747, 2524748, 2524749, 2524751, 2524752, 2524753, 2524754, 2524757, 2524763, 2524764, 2524765, 2524766, 2524767, 2524768, 2730616, 2741964, 2741965, 2741966, 2741967, 2741968, 2899985, 2899986, 2899989, 2899990, 2899991, 2993117, 2993118, 2993120, 2993121, 2993122, 2993123, 2993124, 2993125, 2993130, 2993131, 2993132, 2993133, 2993134, 2993136, 2993137, 2993138, 2993140, 2993141, 2993142, 2993143, 2993144, 2993145, 2993146, 2993148, 2993149, 2993150, 2993151, 2993152, 2993153, 2993155, 2993156, 3871842, 3871846, 3871847, 3871851, 3871854, 3871860, 3871862, 3871863, 3871864, 4127131, 4127133, 4127134, 4127135, 4127136, 4127138, 4127139, 4127140, 4127142, 4127143, 4127144, 4127146, 4127147, 4127148, 4127149, 4127151, 4127152, 4127153, 4127155, 4127156, 4127157, 5245407, 5346897, 5346899, 5346902, 5346903, 5346904, 5346905, 5346906, 5346907, 5346908, 5346910, 5346911, 5346912, 5346915, 5346917, 5346918, 5346920, 5456098, 6144087, 6144096, 6144098; v. REF 132-28-27, GUDID 10885862023094, Serial Numbers: 772244, 772246, 772247, 772248, 772249, 772250, 772251, 772253, 772254, 772255, 772256, 774696, 774697, 774698, 774700, 774701, 774702, 774703, 774704, 774708, 774709, 780838, 780839, 780840, 780846, 780847, 780848, 780849, 780850, 780851, 780852, 780853, 780855, 780856, 781740, 781741, 781742, 781746, 781747, 781750, 781751, 781752, 781753, 781754, 796093, 796094, 796096, 796097, 796098, 796099, 796100, 796101, 796104, 796105, 796106, 796107, 796108, 796109, 796110, 823296, 823298, 823299, 823300, 823301, 823302, 823303, 823305, 823306, 823308, 823309, 823313, 823315, 823316, 823317, 823318, 823319, 823320, 823321, 823322, 823323, 823324, 823325, 823327, 823329, 823331, 823333, 823334, 823336, 823337, 823338, 823339, 823340, 848673, 848674, 848675, 848676, 848677, 850056, 850057, 850058, 850061, 850062, 850063, 850064, 850066, 850067, 850069, 850070, 850071, 850073, 850074, 850076, 850077, 850079, 850080, 850082, 850083, 850084, 850086, 850087, 850088, 850091, 850092, 850093, 850094, 850095, 850096, 850098, 850099, 850100, 850103, 850104, 850107, 850402, 850403, 850404, 850406, 850407, 850409, 850410, 850412, 850413, 850414, 850415, 850418, 850420, 850421, 850423, 850424, 850425, 850426, 850429, 850430, 850433, 853707, 853708, 853709, 853711, 853712, 853713, 853715, 853716, 853717, 853719, 853720, 853721, 853722, 853723, 853724, 853725, 853726, 853728, 853729, 853730, 853731, 853732, 853733, 853734, 853735, 853737, 853739, 853740, 853741, 853742, 853743, 853744, 853746, 853747, 853748, 853749, 853750, 853751, 853752, 853753, 853755, 853757, 853759, 1093448, 1093449, 1093450, 1093451, 1093452, 1093454, 1093455, 1093456, 1093457, 1093458, 1093459, 1093460, 1093461, 1093462, 1093463, 1093464, 1093465, 1093466, 1093467, 1093468, 1093469, 1093470, 1093471, 1093472, 1093473, 1093474, 1093475, 1093476, 1093477, 1093478, 1093479, 1093480, 1384936, 1384937, 1384938, 1384939, 1384940, 1384941, 1384942, 1384943, 1389960, 1389961, 1389962, 1389963, 1389964, 1389965, 1389966, 1389967, 1389968, 1389969, 1389970, 1389971, 1389972, 1389973, 1389974, 1389975, 1389976, 1389977, 1389978, 1389979, 1389980, 1389981, 1389982, 1389983, 1389984, 1389985, 1389986, 1389987, 1389988, 1389989, 1389990, 1389991, 1389992, 1389993, 1389994, 1435171, 1435172, 1435173, 1435174, 1435175, 1435176, 1435177, 1435178, 1435179, 1435180, 1435181, 1435182, 1435183, 1435184, 1435185, 1435186, 1435187, 1435188, 1435189, 1435190, 1435191, 1435192, 1435193, 1435194, 1435195, 1435196, 1435197, 1435198, 1435199, 1435200, 1435201, 1435202, 1435203, 1485806, 1485807, 1485808, 1485809, 1485810, 1485811, 1485812, 1485813, 1485814, 1485815, 1485816, 1485817, 1485818, 1485819, 1485820, 1485821, 1485822, 1485823, 1485824, 1485825, 1485826, 1485827, 1485828, 1485829, 1485830, 1485831, 1485832, 1485833, 1485834, 1485835, 1485836, 1485837, 1485838, 1485839, 1485840, 1485841, 1485842, 1485843, 1485844, 1485845, 1489863, 1489864, 1489865, 1489866, 1489867, 1489868, 1489869, 1489870, 1489871, 1489872, 1489873, 1489874, 1489875, 1489876, 1489877, 1489878, 1489879, 1489880, 1489881, 1489882, 1489883, 1489884, 1489885, 1489886, 1489887, 1489888, 1489889, 1489890, 1489891, 1489892, 1489893, 1489894, 1489895, 1489896, 1489897, 1489898, 1489899, 1489900, 1489901, 1489902, 1511208, 1511209, 1511210, 1511211, 1511212, 1511213, 1511214, 1511215, 1511216, 1511217, 1511218, 1511219, 1511220, 1511221, 1511222, 1511223, 1511224, 1511225, 1511226, 1511227, 1511228, 1511229, 1511230, 1511231, 1511232, 1511234, 1511235, 1511236, 1511237, 1511238, 1511239, 1511240, 1511241, 1511242, 1511243, 1511244, 1511245, 1511246, 1511247, 1511560, 1511561, 1511562, 1511563, 1511564, 1511565, 1511566, 1511567, 1511568, 1511569, 1511570, 1511571, 1511572, 1511573, 1511574, 1511575, 1511576, 1511577, 1511578, 1511579, 1511580, 1511581, 1511582, 1511584, 1511585, 1511586, 1511587, 1511588, 1511589, 1511590, 1511591, 1511592, 1511593, 1511594, 1511595, 1511596, 1511597, 1511598, 1511599, 1526076, 1526077, 1526078, 1526079, 1526080, 1526081, 1526082, 1526083, 1526084, 1526085, 1526086, 1526087, 1526088, 1526089, 1526090, 1526091, 1526092, 1526093, 1526094, 1526095, 1526096, 1526097, 1526098, 1526099, 1526100, 1526101, 1526102, 1526103, 1526104, 1526105, 1534093, 1534094, 1534095, 1534096, 1534097, 1534098, 1534099, 1534100, 1534101, 1534102, 1534103, 1534104, 1534105, 1534106, 1534107, 1534108, 1534109, 1534110, 1534111, 1534112, 1534113, 1585120, 1585121, 1585122, 1585123, 1585124, 1585125, 1585126, 1585127, 1585128, 1585129, 1585130, 1585131, 1585132, 1585133, 1585134, 1585135, 1585136, 1585137, 1585138, 1585139, 1585140, 1585141, 1585142, 1585143, 1585144, 1585145, 1585146, 1585147, 1585148, 1585149, 1585150, 1585151, 1585509, 1585510, 1585512, 1585513, 1585514, 1585515, 1585516, 1585517, 1585518, 1585519, 1585520, 1585521, 1585522, 1585523, 1585524, 1585525, 1585526, 1585527, 1585528, 1585529, 1585530, 1585531, 1585532, 1585533, 1585534, 1585535, 1585536, 1585537, 1585538, 1585539, 1585540, 1585541, 1585542, 1585543, 1585544, 1585545, 1585546, 1585547, 1585548, 1585549, 1585550, 1585551, 1585552, 1605234, 1605235, 1605236, 1605237, 1605238, 1605239, 1605240, 1605241, 1605242, 1605243, 1605244, 1605245, 1605246, 1605247, 1605248, 1605249, 1605250, 1605251, 1605252, 1605253, 1605254, 1605255, 1605256, 1605257, 1605258, 1605259, 1605260, 1605261, 1605262, 1605263, 1605264, 1605265, 1605266, 1605267, 1605268, 1605269, 1605270, 1605271, 1605272, 1605273, 1605274, 1605275, 1627901, 1627902, 1627903, 1627905, 1627906, 1627907, 1627908, 1627909, 1627910, 1627911, 1627912, 1627913, 1627914, 1627915, 1627916, 1627917, 1627918, 1627919, 1627920, 1627921, 1627923, 1627924, 1627925, 1627927, 1627928, 1627929, 1627930, 1627931, 1627932, 1627933, 1627934, 1627935, 1627936, 1627937, 1627938, 1627939, 1627940, 1652731, 1652732, 1652733, 1652734, 1652736, 1652737, 1652738, 1652739, 1652740, 1652741, 1652742, 1652743, 1652744, 1652745, 1652746, 1652747, 1652748, 1652749, 1652750, 1652751, 1652752, 1652753, 1652754, 1652755, 1652756, 1652757, 1652758, 1652759, 1652760, 1652761, 1652763, 1652764, 1652765, 1652766, 1652767, 1652768, 1652769, 1652770, 1652771, 1652772, 1652773, 1668459, 1668460, 1668461, 1668462, 1668463, 1668464, 1668465, 1668466, 1668467, 1668468, 1668469, 1668470, 1668471, 1668472, 1668473, 1668474, 1668475, 1668476, 1668477, 1668478, 1668479, 1668480, 1668481, 1668482, 1668483, 1668484, 1668485, 1668486, 1668487, 1668488, 1668490, 1668491, 1668492, 1668493, 1668494, 1668495, 1668496, 1668497, 1668498, 1702046, 1702047, 1702048, 1702049, 1702050, 1702051, 1702052, 1702053, 1702054, 1702055, 1702056, 1702057, 1702058, 1702059, 1702060, 1702061, 1702062, 1702063, 1702064, 1702065, 1702066, 1702067, 1702068, 1702069, 1702070, 1702071, 1702072, 1702073, 1702074, 1702075, 1702076, 1702077, 1702078, 1702079, 1702080, 1713504, 1713505, 1713506, 1713507, 1713508, 1713509, 1713510, 1713511, 1713512, 1713513, 1713514, 1713515, 1713516, 1713517, 1713518, 1713519, 1713520, 1713521, 1713522, 1713523, 1713524, 1713525, 1713526, 1713527, 1713528, 1713529, 1713530, 1713531, 1713532, 1713533, 1780122, 1780123, 1780124, 1780125, 1780126, 1780127, 1780128, 1780129, 1780130, 1780131, 1780132, 1780133, 1780134, 1780135, 1780136, 1780137, 1780138, 1780139, 1780140, 1780141, 1780142, 1780143, 1780144, 1780145, 1780146, 1780147, 1780148, 1780149, 1780150, 1780151, 1780152, 1780153, 1780154, 1780155, 1780156, 1780157, 1780158, 1780159, 1780160, 1780161, 1780162, 1780163, 1780164, 1780165, 1780166, 1780167, 1780168, 1780169, 1780170, 1780171, 1780172, 1780173, 1780174, 1780175, 1788137, 1788138, 1788139, 1788140, 1788141, 1788142, 1788143, 1788144, 1788145, 1788146, 1788147, 1788148, 1788149, 1788150, 1788151, 1788152, 1788153, 1788154, 1788155, 1788156, 1788157, 1788158, 1788159, 1788160, 1788161, 1788162, 1821809, 1821810, 1821811, 1821812, 1821813, 1821814, 1821815, 1821816, 1821817, 1821818, 1821819, 1821820, 1821821, 1821822, 1821823, 1821824, 1821825, 1821826, 1821827, 1821828, 1821829, 1821830, 1821831, 1821832, 1821833, 1821834, 1821835, 1821836, 1821837, 1821838, 1821839, 1821840, 1821841, 1821842, 1821843, 1821844, 1821845, 1821846, 1821847, 1821848, 1845150, 1845151, 1845152, 1845153, 1845154, 1845155, 1845156, 1845157, 1845158, 1845159, 1845160, 1845161, 1845162, 1845163, 1845164, 1845165, 1845166, 1845167, 1845168, 1845169, 1845170, 1845171, 1845172, 1845173, 1845174, 1845175, 1845176, 1845177, 1845178, 1845179, 1845180, 1845181, 1845182, 1845183, 1845184, 1845185, 1845186, 1845187, 1845188, 1845189, 1845237, 1845238, 1845239, 1845240, 1845241, 1845242, 1845243, 1845244, 1845245, 1845246, 1845247, 1845248, 1845249, 1845250, 1845251, 1845252, 1845253, 1845254, 1845255, 1845256, 1852674, 1852675, 1852676, 1852677, 1852678, 1852679, 1852680, 1852681, 1852682, 1873667, 1873668, 1873669, 1873670, 1873671, 1873672, 1873673, 1873674, 1873675, 1873676, 1873677, 1873678, 1873679, 1873680, 1873681, 1873682, 1873683, 1873684, 1873685, 1873686, 1873687, 1873688, 1873689, 1873690, 1873691, 1885150, 1885151, 1885152, 1885153, 1885154, 1885155, 1885156, 1885157, 1885158, 1885159, 1885160, 1885161, 1885162, 1885163, 1885164, 1885165, 1885166, 1885167, 1885168, 1885169, 1885170, 1885171, 1885172, 1885173, 1885174, 1885175, 1885176, 1885177, 1885178, 1885179, 1885180, 1885181, 1885182, 1885183, 1885184, 1885185, 1885186, 1885187, 1885188, 1885189, 1885190, 1885650, 1885651, 1885652, 1885653, 1885654, 1885655, 1885656, 1885657, 1885658, 1885659, 1885660, 1885661, 1885662, 1885663, 1885664, 1885665, 1885666, 1885667, 1885668, 1885669, 1885670, 1885671, 1885672, 1885673, 1885674, 1885675, 1885676, 1885677, 1885678, 1885679, 1885680, 1885681, 1885682, 1885683, 1885684, 1885685, 1885686, 1885687, 1885688, 1885689, 1928382, 1928383, 1928384, 1928385, 1928386, 1928387, 1928388, 1928389, 1928390, 1928391, 1928392, 1928393, 1928394, 1928395, 1928396, 1928397, 1928398, 1928399, 1928400, 1928401, 1928402, 1928403.1928404, 1928405, 1928406, 1928407, 1928408, 1928409, 1928410, 1928411, 1928412, 1928413, 1928414, 1945180, 1945181, 1945182, 1945183, 1945184, 1945185, 1945186, 1945187, 1945188, 1945189, 1945190, 1945191, 1945192, 1945193, 1945194, 1945195, 1945196, 1945197, 1945198, 1945199, 1945200, 1945201, 1945202, 1945203, 1945204, 1945205, 1945206, 1945207, 1945208, 1945209, 1945210, 1945211, 1945212, 1945213, 1945214, 1945215, 1945216, 1945217, 1945218, 1945219, 1945220, 1945221, 1945222, 1945223, 1945224, 1945225, 1945226, 1945227, 1945228, 1945229, 1945230, 1945231, 1945232, 1945233, 1945234, 1945235, 1945236, 1945237, 1945238, 1945239, 1945240, 1945241, 1945242, 1945243, 1945244, 1945245, 2021459, 2021460, 2021461, 2021462, 2021463, 2021464, 2021465, 2021466, 2021467, 2021468, 2021470, 2021471, 2021472, 2021473, 2021474, 2021475, 2021476, 2021477, 2021478, 2021479, 2021480, 2021481, 2021482, 2021483, 2021484, 2021485, 2021486, 2021487, 2021488, 2021489, 2021490, 2021491, 2021492, 2021493, 2021494, 2021495, 2021496, 2021497, 2021498, 2021499, 2021500, 2021501, 2021502, 2026993, 2026994, 2026995, 2026996, 2026997, 2026998, 2026999, 2027000, 2027001, 2027002, 2027003, 2027004, 2027005, 2027006, 2027007, 2027008, 2027009, 2027010, 2027011, 2027012, 2027013, 2027014, 2027015, 2027016, 2027017, 2027018, 2027019, 2027020, 2027021, 2027022, 2027023, 2027024, 2027025, 2027026, 2027027, 2027028, 2027029, 2027030, 2027031, 2027032, 2030202, 2030203, 2030204, 2030205, 2030206, 2030207, 2030208, 2030209, 2030210, 2030211, 2030212, 2030213, 2030214, 2030215, 2030216, 2030217, 2030218, 2030219, 2030220, 2030221, 2030222, 2030223, 2030224, 2030225, 2030226, 2030227, 2030228, 2030229, 2030230, 2056323, 2056324, 2056325, 2056326, 2056327, 2056328, 2056329, 2056330, 2056331, 2056332, 2056333, 2056334, 2056335, 2056336, 2056337, 2056338, 2056339, 2056340, 2056341, 2056342, 2056343, 2056344, 2056345, 2056346, 2056347, 2056348, 2056349, 2056350, 2056351, 2056352, 2056353, 2056354, 2056355, 2056356, 2056357, 2056358, 2056359, 2056360, 2056361, 2056362, 2056363, 2056364, 2119985, 2119986, 2119987, 2119988, 2119989, 2119990, 2119991, 2119992, 2119993, 2119994, 2119995, 2119996, 2119997, 2119998, 2119999, 2120000, 2120001, 2120002, 2120003, 2120004, 2120005, 2120006, 2120007, 2120008, 2120009, 2120010, 2127756, 2127757, 2127758, 2127759, 2127760, 2127761, 2127762, 2127763, 2127764, 2127765, 2127766, 2127767, 2127768, 2127769, 2127770, 2127771, 2127772, 2127773, 2127774, 2127775, 2127776, 2127777, 2127778, 2127779, 2127780, 2140040, 2140041, 2140042, 2142864, 2142865, 2142866, 2142867, 2142868, 2142869, 2142870, 2142871, 2142872, 2142873, 2142874, 2142875, 2142876, 2142877, 2142878, 2142879, 2142880, 2142881, 2142882, 2142883, 2142884, 2142885, 2142886, 2142887, 2142888, 2142889, 2142890, 2142891, 2142892, 2142893, 2142894, 2142895, 2142896, 2142897, 2142898, 2142900, 2142901, 2142902, 2142903, 2183342, 2183343, 2183344, 2183345, 2183346, 2183347, 2183348, 2183349, 2183350, 2183351, 2183352, 2183353, 2183354, 2183355, 2183356, 2183357, 2183358, 2183359, 2183360, 2183361, 2183362, 2183363, 2187120, 2187121, 2187122, 2187123, 2187124, 2187125, 2187126, 2187127, 2187128, 2215755, 2215756, 2215757, 2215758, 2215759, 2215760, 2215761, 2215762, 2215763, 2215764, 2215765, 2215766, 2215767, 2215768, 2215769, 2215770, 2215771, 2215772, 2215773, 2215774, 2215775, 2215776, 2215777, 2219740, 2219741, 2219742, 2219743, 2219744, 2219745, 2219746, 2219747, 2219748, 2219749, 2219750, 2219751, 2219752, 2219753, 2219754, 2219755, 2219756, 2219757, 2219758, 2219759, 2219760, 2219761, 2219762, 2219763, 2254552, 2254553, 2254554, 2254555, 2254556, 2254557, 2254558, 2254559, 2254560, 2254561, 2254562, 2254563, 2254564, 2254565, 2254566, 2258031, 2258032, 2258033, 2258034, 2258035, 2258036, 2258037, 2258038, 2258039, 2258040, 2276557, 2276558, 2276559, 2276560, 2276561, 2276562, 2276563, 2276564, 2276565, 2276566, 2276567, 2276568, 2276569, 2276570, 2276571, 2276572, 2276573, 2276574, 2276575, 2276576, 2276577, 2276578, 2276579, 2276580, 2276581, 2276582, 2276583, 2276584, 2329261, 2329262, 2329263, 2329264, 2329265, 2329266, 2329267, 2329268, 2329269, 2329270, 2329271, 2329272, 2329273, 2329274, 2329275, 2329276, 2329277, 2329278, 2329279, 2329280, 2329281, 2329282, 2405298, 2405299, 2405300, 2405301, 2405302, 2405305, 2405309, 2405312, 2405313, 2405315, 2405316, 2405317, 2405318, 2405319, 2405320, 2405321, 2405322, 2405323, 2405325, 2405326, 2405327, 2405328, 2405329, 2405330, 2405331, 2504221, 2504222, 2504223, 2504224, 2504225, 2504226, 2504227, 2504228, 2504229, 2504230, 2504231, 2504232, 2504233, 2504234, 2504235, 2504236, 2504237, 2504238, 2504239, 2504240, 2504241, 2504242, 2504244, 2504246, 2504247, 2574496, 2574498, 2935324, 3852282, 3852296, 3852301, 3852303, 3852306; w. REF 132-28-28, GUDID 10885862023100, Serial Numbers: 771906, 771907, 771908, 771910, 771913, 771914, 771916, 774277, 774278, 774279, 774280, 774281, 774284, 774285, 774286, 774289, 780795, 780797, 780799, 780802, 780803, 780805, 780807, 782096, 782099, 782100, 782102, 796021, 796022, 796024, 796026, 796030, 796031, 796033, 796034, 796035, 796038, 796039, 796040, 796041, 796042, 796043, 796044, 796046, 796047, 796048, 796049, 796050, 796053, 796054, 796055, 796056, 796057, 796058, 796059, 796060, 796061, 796062, 796063, 796065, 844347, 844348, 844349, 844350, 844353, 844354, 844355, 844357, 844358, 844361, 844363, 844364, 844365, 844368, 844371, 844372, 844373, 844374, 844375, 844376, 844377, 844378, 848161, 848162, 848163, 848164, 848169, 848170, 848171, 848174, 848176, 848178, 848181, 848185, 848186, 848189, 848190, 848191, 848195, 848196, 848197, 848198, 848201, 848202, 848206, 848207, 848209, 848513, 848514, 848517, 848518, 848519, 848520, 848521, 848522, 848524, 1385662, 1385663, 1385664, 1385665, 1385666, 1385667, 1385668, 1385669, 1385670, 1385671, 1385672, 1385673, 1385674, 1385675, 1385676, 1385677, 1385678, 1385679, 1385680, 1385681, 1385682, 1385683, 1385684, 1385685, 1385686, 1385687, 1385688, 1385689, 1385690, 1385691, 1385692, 1435454, 1435455, 1435456, 1435457, 1435458, 1435459, 1435460, 1435461, 1435462, 1435463, 1435464, 1435465, 1435466, 1435467, 1435468, 1435469, 1435470, 1435471, 1435472, 1435473, 1435474, 1435475, 1435476, 1435477, 1435478, 1435479, 1435480, 1435481, 1435482, 1435483, 1435484, 1435485, 1435486, 1485714, 1485715, 1485716, 1485717, 1485718, 1485719, 1485720, 1485721, 1485722, 1485723, 1485724, 1485725, 1485726, 1485727, 1485728, 1485729, 1485730, 1485731, 1485732, 1485733, 1485734, 1485735, 1485736, 1485737, 1485738, 1485739, 1489822, 1489823, 1489824, 1489825, 1489826, 1489827, 1489828, 1489829, 1489830, 1489831, 1489832, 1489833, 1489834, 1489835, 1489836, 1489837, 1489838, 1489839, 1489840, 1489841, 1489842, 1489843, 1489844, 1489845, 1489846, 1525971, 1525972, 1525973, 1525974, 1525975, 1525976, 1525977, 1525978, 1525979, 1525980, 1525981, 1525982, 1525983, 1525984, 1525985, 1525986, 1525987, 1525988, 1525989, 1525990, 1525992, 1525993, 1525994, 1525995, 1525996, 1525997, 1569538, 1569539, 1569540, 1569541, 1569542, 1569543, 1569544, 1569545, 1569546, 1569547, 1569548, 1569549, 1569550, 1569551, 1569552, 1569553, 1569554, 1569555, 1569556, 1569557, 1569558, 1569559, 1569560, 1569561, 1569562, 1569563, 1569564, 1583280, 1583281, 1583282, 1583283, 1583284, 1583285, 1583286, 1583287, 1583288, 1583289, 1583290, 1583292, 1583293, 1583294, 1583295, 1583296, 1583297, 1583298, 1583299, 1583300, 1583301, 1583302, 1583303, 1583304, 1583305, 1583306, 1610083, 1610085, 1610086, 1610089, 1610090, 1610091, 1610092, 1610093, 1610094, 1610095, 1610097, 1610098, 1610099, 1610100, 1610102, 1610103, 1610104, 1610105, 1662782, 1662783, 1662784, 1662785, 1662786, 1662787, 1662788, 1662789, 1662790, 1662791, 1662793, 1662794, 1662795, 1662796, 1662797, 1662798, 1662799, 1662800, 1662801, 1662802, 1662803, 1662804, 1662805, 1662806, 1662807, 1701868, 1701869, 1701870, 1701871, 1701872, 1701873, 1701874, 1701875, 1701876, 1701877, 1701878, 1701879, 1701880, 1701881, 1701882, 1701883, 1701884, 1701885, 1701886, 1701887, 1701888, 1701889, 1701890, 1701891, 1701892, 1701893, 1761515, 1761516, 1761517, 1761518, 1761519, 1761520, 1761521, 1761522, 1761523, 1761524, 1761525, 1761526, 1761527, 1761528, 1761529, 1761530, 1761531, 1761532, 1761533, 1761534, 1761535, 1761536, 1761537, 1761538, 1761539, 1761540, 1761541, 1798578, 1798579, 1798580, 1798581, 1798582, 1798583, 1798584, 1798585, 1798586, 1798587, 1798588, 1798589, 1798590, 1798591, 1798592, 1798593, 1798594, 1798595, 1798596, 1798597, 1798598, 1798599, 1803997, 1803998, 1803999, 1804000, 1804001, 1843425, 1843426, 1843427, 1843428, 1843429, 1843430, 1843431, 1843432, 1843433, 1843434, 1843435, 1843436, 1843437, 1843438, 1843439, 1843440, 1843441, 1843442, 1843443, 1843444, 1843445, 1843446, 1843447, 1843448, 1843449, 1843450, 1859888, 1859889, 1859890, 1859891, 1859892, 1859893, 1859894, 1859895, 1859896, 1859897, 1859898, 1859899, 1859900, 1859901, 1859902, 1859903, 1859904, 1859905, 1859906, 1859907, 1859908, 1859909, 1859910, 1859911, 1859912, 1892606, 1892607, 1892608, 1892609, 1892610, 1892611, 1892612, 1892613, 1892614, 1892615, 1892616, 1892617, 1892618, 1892619, 1892620, 1892621, 1892622, 1892623, 1892624, 1892625, 1892626, 1892627, 1892628, 1892629, 1892630, 1892631, 1926010, 1926011, 1926012, 1926013, 1926014, 1926015, 1926016, 1926017, 1926018, 1926019, 1926020, 1926021, 1926022, 1926023, 1926024, 1926025, 1926026, 1926027, 1926028, 1926029, 1926030, 1926031, 1926032, 1926033, 1926034, 1926035, 1945246, 1945247, 1945248, 1945249, 1945250, 1945251, 1945252, 1945253, 1945255, 1945256, 1945257, 1945258, 1945259, 1945260, 1945261, 1945262, 1945263, 1945264, 2057082, 2057083, 2057084, 2057085, 2057086, 2057087, 2057088, 2057089, 2057091, 2057092, 2057093, 2057094, 2057095, 2057097, 2057098, 2057100, 2057101, 2057104, 2057105, 2057106, 2057107, 2058151, 2058152, 2058153, 2058154, 2058155, 2058156, 2058157, 2058158, 2058159, 2058160, 2058161, 2058163, 2058164, 2058165, 2058166, 2058167, 2058168, 2058169, 2058170, 2058171, 2058172, 2058173, 2058174, 2058175, 2091707, 2091708, 2091709, 2091711, 2091712, 2091713, 2091714, 2091716, 2091717, 2091718, 2091719, 2091720, 2091721, 2091722, 2091723, 2091724, 2091725, 2091726, 2091727, 2091728, 2091729, 2091730, 2091731, 2091732, 2091734, 2091735, 2162183, 2162184, 2162185, 2162186, 2162187, 2162188, 2162190, 2162191, 2162192, 2162193, 2162194, 2162195, 2170192, 2170193, 2170194, 2170195, 2170196, 2170197, 2170198, 2170199, 2170200, 2170201, 2170202, 2170203, 2170204, 2170205, 2216569, 2216570, 2216571, 2216572, 2216573, 2216574, 2216575, 2216576, 2216577, 2216578, 2216579, 2216580, 2216581, 2216582, 2216583, 2216584, 2216585, 2216586, 2216587, 2216588, 2216589, 2216590, 2216591, 2216592, 2238209, 2238210, 2238211, 2238213, 2238214, 2238215, 2238216, 2238217, 2238219, 2238220, 2263812, 2263813, 2263814, 2263815, 2263816, 2263817, 2263818, 2263819, 2263820, 2263821, 2263822, 2263823, 2263824, 2263825, 2263826, 2333442, 2333444, 2333445, 2333446, 2333447, 2333448, 2333450, 2333451, 2333452, 2333453, 2333454, 2333455, 2333456, 2333457, 2333458, 2907626, 3866618, 3866624, 3866625, 3866632, 3866633, 3866635; x. REF 132-28-29, GUDID 10885862023117, Serial Numbers: 771946, 771947, 776479, 780813, 795566, 795569, 795580, 795582, 795583, 795586, 795600, 795601, 795606, 795614, 841913, 841914, 841915, 841918, 841922, 841923, 841924, 841925, 841927, 841928, 841931, 843835, 843836, 843841, 843842, 843843, 843847, 843848, 843849, 843850, 843851, 843853, 843857, 843858, 843860, 843861, 843863, 843864, 843865, 843866, 843868, 843869, 843870, 843875, 1093689, 1093690, 1093692, 1093693, 1093694, 1093695, 1093697, 1093698, 1093699, 1093700, 1093701, 1093702, 1093703, 1093704, 1093705, 1093706, 1093707, 1093708, 1093710, 1093711, 1093712, 1093713, 1093714, 1093715, 1093716, 1093717, 1093718, 1093719, 1511104, 1511105, 1511106, 1511107, 1511108, 1511109, 1511110, 1511111, 1511112, 1511387, 1511388, 1511389, 1511390, 1511391, 1511392, 1517840, 1517841, 1517843, 1517844, 1517845, 1517846, 1517847, 1517848, 1517849, 1517850, 1517852, 1517853, 1517854, 1517855, 1584993, 1584995, 1584996, 1584999, 1585000, 1585002, 1585003, 1585005, 1585007, 1585009, 1585010, 1585011, 1585012, 1585013, 1585014, 1585016, 1585017, 1664617, 1664618, 1664619, 1664620, 1664621, 1664623, 1664624, 1664625, 1664626, 1664627, 1664628, 1664629, 1664630, 1664631, 1664632, 1664633, 1664634, 1664635, 1664636, 1664637, 1664638, 1664639, 1664640, 1664641, 1664642, 1664643, 1810066, 1810067, 1810068, 1810069, 1810070, 1810071, 1810072, 1810073, 1810074, 1810075, 1810076, 1810077, 1810078, 1810079, 1810080, 1810081, 1810082, 1810083, 1810084, 1810085, 1810086, 1810087, 1810088, 1810089, 1810090, 1810091, 1810092, 1895604, 1895606, 1895607, 1895608, 1895609, 1895610, 1895611, 1895612, 1895613, 1895614, 1895615, 1895617, 1895618, 1895619, 1895621, 1895622, 1895625, 1895626, 1895627, 1895628, 1895629, 1954016, 1954017, 1954019, 1954020, 1954021, 1954022, 2033602, 2033604, 2033606, 2033607, 2033609, 2033610, 2033611, 2042551, 2042552, 2042553, 2042554, 2107338, 2107340, 2156295, 2156297, 2156298, 2156300, 2156301, 2156302, 2231337, 2231338, 2231339, 2231340, 2231341, 2231342, 2231343, 2231344, 2231345, 2237058, 2237059, 2237060, 2237061, 2237062, 2237063, 2237064, 2237065, 2237066, 2237067, 2237068, 2237069, 2327122, 2327123, 2327124, 2327125, 2327126, 2327127, 2327128, 2327129, 2327130, 2392867, 2392868, 2392869, 2392870, 2392871, 2392872, 2392873, 2392874, 2392875, 2392876, 2392877, 2392878, 2392879, 2392880, 2392881, 2392882, 2691753, 2691754, 2691755, 2691756, 2691758, 2691764, 2691765, 2691766, 2730614, 2730615; y. REF 132-28-30, GUDID 10885862023124, Serial Numbers: 796073, 1093659, 1093661, 1093662, 1093663, 1093664, 1093666, 1093668, 1093670, 1093671, 1093673, 1093674, 1093676, 1093678, 1093680, 1093683, 1093685, 1093687, 1093688, 1845709, 1845710, 1845711, 1857273, 1972634, 1972635, 2057346, 2057347, 2077334, 2077336, 2077337, 2077340, 2077341, 2077342, 2077343, 2077344, 2077345, 2129235, 2129236, 2129237, 2129238, 2129239, 2129240, 2129241, 2129242, 2129243, 2129244, 2129245, 2131276, 2131279, 2131280, 2626073, 2626074, 2626075, 2626076, 2626077, 2626078, 2697712, 2730613; z. REF 132-32-27, GUDID 10885862023247, Serial Numbers: 771278, 771279, 771280, 771281, 771283, 771284, 771285, 771286, 771287, 771288, 771289, 771290, 771291, 774249, 774250, 774251, 774252, 774253, 774254, 774255, 774256, 774257, 774258, 774259, 774260, 774261, 774262, 781697, 781698, 781699, 781700, 781815, 781816, 781817, 781818, 781819, 781820, 781821, 781823, 781824, 781825, 781826, 781827, 781828, 781829, 781830, 781831, 781832, 781834, 781835, 781836, 781837, 781838, 781839, 781840, 781841, 781842, 798998, 798999, 799000, 799001, 799002, 799003, 799004, 799005, 799006, 799007, 799008, 799009, 799010, 799011, 799012, 799013, 799014, 799015, 799016, 799017, 799018, 799019, 799020, 799021, 799022, 799023, 799024, 799025, 823271, 823272, 823273, 823274, 823275, 823276, 823277, 823278, 823279, 823280, 823281, 823282, 823283, 823284, 823285, 823286, 823287, 823288, 823289, 823290, 823291, 823292, 823293, 823294, 823295, 840362, 840363, 840364, 840365, 840366, 840367, 840368, 840369, 840370, 840371, 840372, 840373, 840374, 840375, 840376, 840377, 840378, 840379, 840380, 840381, 840382, 840383, 840384, 843628, 843629, 843630, 843631, 843632, 843633, 843634, 843635, 843636, 843637, 843638, 843639, 843640, 843641, 843642, 843643, 843644, 843645, 843646, 843647, 843648, 843649, 843650, 843651, 843652, 862284, 862285, 862286, 862287, 862288, 862289, 862290, 862291, 862292, 862293, 862294, 862295, 862296, 862297, 862298, 862299, 862300, 862301, 862302, 868754, 868755, 868756, 868757, 868758, 868760, 868761, 868763, 868764, 868765, 868766, 868767, 868768, 868769, 868770, 868771, 868772, 868773, 868774, 868775, 868776, 868777, 868778, 868779, 868780, 868781, 868782, 868783, 868784, 868785, 868786, 868787, 868788, 868789, 868790, 868791, 868792, 868793, 868794, 868795, 868796, 868797, 868798, 868799, 868800, 868801, 868802, 868803, 868804, 868805, 868806, 887724, 887725, 887726, 887727, 887728, 887729, 887730, 887731, 887732, 887733, 887734, 887735, 887736, 887737, 887738, 887739, 887740, 887741, 887742, 887743, 887744, 887745, 887746, 887747, 887748, 887749, 887750, 887751, 887752, 887753, 887754, 887755, 887756, 896336, 896337, 896338, 896339, 896340, 896341, 896342, 896343, 896344, 896345, 896346, 896347, 896348, 896350, 896351, 896352, 896353, 896355, 896356, 896357, 896358, 896359, 896360, 896361, 896362, 896363, 896364, 896365, 896366, 896367, 896368, 908859, 908860, 908861, 908862, 908863, 908864, 908865, 908866, 908867, 908868, 908869, 908870, 908871, 908872, 908873, 908874, 908875, 908876, 908877, 908878, 908879, 908880, 908881, 908882, 908883, 908884, 908885, 908886, 908887, 908888, 908889, 908890, 908891, 917763, 917764, 917765, 917766, 917767, 917768, 917769, 917770, 917771, 917772, 917773, 917774, 917775, 917776, 917777, 917778, 917779, 917780, 917781, 917782, 917783, 917784, 917785, 917786, 917787, 917788, 917789, 917791, 917792, 917793, 917794, 917795, 919096, 919097, 919098, 919099, 919100, 919101, 919102, 919103, 919104, 919105, 919106, 919107, 919108, 919109, 919110, 919111, 919112, 919113, 919114, 919115, 919116, 919117, 919118, 919119, 919120, 919121, 919122, 919123, 919124, 919125, 919126, 919127, 921810, 921811, 921812, 921813, 921814, 921815, 921816, 921817, 921818, 921819, 921820, 921821, 921822, 921823, 921824, 921825, 921826, 921827, 921828, 921829, 921830, 921831, 921832, 921833, 921834, 921835, 921836, 921837, 921839, 921840, 921841, 921842, 928002, 928003, 928004, 928005, 928006, 928007, 928008, 928009, 928010, 928011, 928012, 928013, 928014, 928015, 928016, 928017, 928018, 928019, 928020, 928021, 928022, 928023, 928024, 928025, 928026, 928027, 928028, 928029, 928031, 928032, 928033, 928034, 935552, 935553, 935554, 935555, 935556, 935557, 935558, 935559, 935560, 935561, 935562, 935563, 935564, 935565, 935566, 935567, 935568, 935569, 935570, 935571, 935572, 935573, 935574, 935576, 935577, 935578, 935579, 935580, 935581, 935582, 935583, 935584, 935585, 935586, 935587, 935588, 935589, 935590, 935591, 935592, 935593, 935594, 935595, 935596, 935597, 935598, 935599, 935600, 935601, 935602, 935603, 935604, 935606, 965950, 965951, 965952, 965953, 965954, 965955, 965956, 965957, 965958, 965959, 965960, 965961, 965962, 965963, 965964, 965965, 965966, 965967, 965968, 965969, 965970, 965971, 965972, 965973, 965974, 965975, 965976, 965977, 965978, 965979, 965980, 965981, 965982, 965983, 965984, 965985, 965986, 965987, 965988, 965989, 965990, 965991, 965992, 965993, 965994, 965995, 965996, 965997, 965998, 965999, 966000, 966001, 966002, 966003, 966004, 985651, 985652, 985653, 985654, 985655, 985656, 985657, 985658, 985659, 985660, 985661, 985662, 985663, 985664, 985665, 985666, 985667, 985668, 985669, 985670, 985671, 985672, 985673, 985674, 985675, 985676, 985677, 985678, 985679, 985680, 985681, 985682, 985683, 985684, 985685, 985687, 985688, 985689, 985690, 985691, 985692, 985693, 985694, 985695, 985696, 985697, 985698, 985699, 985700, 985701, 985702, 1007068, 1007069, 1007070, 1007071, 1007072, 1007073, 1007074, 1007075, 1007076, 1007077, 1007078, 1007079, 1007080, 1007081, 1007082, 1007083, 1007084, 1007085, 1007086, 1007087, 1007088, 1007089, 1007090, 1007091, 1007092, 1007093, 1007094, 1007095, 1007096, 1007097, 1007098, 1007099, 1007100, 1007101, 1007102, 1007103, 1007104, 1007105, 1007106, 1007107, 1007108, 1007109, 1007110, 1007111, 1007112, 1007113, 1007114, 1007115, 1007116, 1007117, 1007118, 1007119, 1007120, 1007121, 1007122, 1007123, 1007124, 1007125, 1007126, 1007127, 1007128, 1007129, 1007130, 1007131, 1007132, 1007133, 1007134, 1007135, 1007136, 1007137, 1007138, 1007139, 1007140, 1007141, 1007142, 1007143, 1007144, 1007145, 1007146, 1007147, 1007148, 1007149, 1007150, 1007151, 1007152, 1007153, 1007154, 1007155, 1007156, 1007157, 1007158, 1007159, 1007160, 1007161, 1007162, 1007163, 1007164, 1007165, 1007166, 1007167, 1007168, 1007169, 1007170, 1007171, 1007172, 1054248, 1054249, 1054250, 1054251, 1054252, 1054253, 1054254, 1054255, 1054256, 1054257, 1054258, 1054259, 1054260, 1054261, 1054262, 1054263, 1054264, 1054265, 1054266, 1054267, 1054268, 1054269, 1054270, 1054271, 1054272, 1054273, 1054274, 1054275, 1054276, 1054277, 1054278, 1054279, 1054280, 1054281, 1054282, 1054283, 1054284, 1054285, 1054286, 1054287, 1054288, 1054289, 1054290, 1054291, 1054292, 1054293, 1054294, 1054295, 1054296, 1054297, 1054298, 1054299, 1054300, 1054301, 1054302, 1058305, 1058306, 1058307, 1058308, 1058309, 1058310, 1058311, 1058312, 1058313, 1058314, 1058315, 1058316, 1058317, 1058318, 1058320, 1058321, 1058322, 1058323, 1058324, 1058325, 1058326, 1058327, 1058328, 1058329, 1058330, 1058331, 1058332, 1058333, 1058334, 1058335, 1058336, 1058337, 1058338, 1058339, 1058340, 1058341, 1058342, 1058343, 1058344, 1058345, 1058346, 1058347, 1058348, 1058349, 1058350, 1058351, 1058352, 1058353, 1058354, 1058355, 1058356, 1058357, 1058358, 1087027, 1087028, 1087029, 1087030, 1087031, 1087032, 1087033, 1087034, 1087035, 1087036, 1087037, 1087038, 1087039, 1087040, 1087041, 1087042, 1087043, 1087044, 1087045, 1087046, 1087047, 1087048, 1087049, 1087050, 1087051, 1087052, 1087053, 1087054, 1087055, 1087056, 1087057, 1087058, 1087059, 1087060, 1087061, 1087062, 1087063, 1087064, 1087065, 1087066, 1087067, 1087068, 1087069, 1087070, 1087071, 1087072, 1087073, 1087074, 1087075, 1087077, 1087078, 1087079, 1087080, 1087081, 1113104, 1113105, 1113106, 1113107, 1113108, 1113109, 1113110, 1113111, 1113112, 1113113, 1113114, 1113115, 1113116, 1113117, 1113118, 1113119, 1113120, 1113121, 1113122, 1113123, 1113124, 1113125, 1113126, 1113127, 1113128, 1113129, 1113130, 1113131, 1113132, 1113133, 1113134, 1113135, 1113136, 1113137, 1113138, 1113139, 1113140, 1113141, 1113142, 1113143, 1113144, 1113145, 1113146, 1113147, 1113148, 1113149, 1113150, 1113151, 1113152, 1113153, 1113154, 1113155, 1113156, 1113157, 1177343, 1177344, 1177345, 1177346, 1177347, 1177348, 1177349, 1177350, 1177351, 1177352, 1177353, 1177354, 1177355, 1177356, 1177357, 1177358, 1177359, 1177360, 1177361, 1177362, 1177363, 1177364, 1177365, 1177366, 1177367, 1177368, 1177369, 1177370, 1177371, 1177372, 1177374, 1177375, 1177376, 1177377, 1177378, 1177379, 1177380, 1177381, 1177382, 1177383, 1177384, 1177385, 1177386, 1177387, 1177388, 1177389, 1177390, 1177391, 1177392, 1177393, 1177394, 1177395, 1177396, 1219592, 1219593, 1219594, 1219595, 1219596, 1219597, 1219598, 1219599, 1219600, 1219602, 1219603, 1219604, 1219605, 1219606, 1219607, 1219608, 1219609, 1219610, 1219611, 1219612, 1219613, 1219614, 1219615, 1219616, 1219617, 1219618, 1219619, 1219620, 1219621, 1219622, 1219623, 1219624, 1219625, 1219626, 1219627, 1219628, 1219629, 1219630, 1219631, 1219632, 1219633, 1219634, 1219635, 1219636, 1219637, 1219638, 1219639, 1219640, 1219641, 1220598, 1220599, 1220600, 1220601, 1220602, 1220603, 1220604, 1220605, 1220606, 1220607, 1220608, 1220609, 1220610, 1220611, 1220612, 1220613, 1220614, 1220615, 1220616, 1220617, 1220618, 1220619, 1220620, 1220621, 1220622, 1220623, 1220624, 1220625, 1220626, 1220627, 1220628, 1220629, 1220630, 1220631, 1220632, 1220633, 1220634, 1220635, 1220636, 1220638, 1220639, 1220641, 1220643, 1220644, 1220645, 1220646, 1220647, 1220648, 1220649, 1220650, 1220651, 1339716, 1339717, 1339718, 1339719, 1339720, 1339721, 1339722, 1339723, 1339724, 1339725, 1339726, 1339727, 1339728, 1339730, 1339731, 1339732, 1339733, 1339734, 1339735, 1339736, 1339737, 1339738, 1339739, 1339741, 1339742, 1341257, 1341258, 1341259, 1341260, 1341261, 1341262, 1341263, 1341264, 1341265, 1341266, 1341267, 1341268, 1341269, 1341270, 1341271, 1341272, 1341273, 1341274, 1341275, 1341276, 1341277, 1341278, 1341279, 1341280, 1341281, 1341282, 1341283, 1464875, 1464876, 1464877, 1464878, 1464879, 1464880, 1464881, 1464882, 1464883, 1467974, 1467975, 1467976, 1467977, 1467978, 1467979, 1467980, 1467981, 1467982, 1467983, 1467984, 1467985, 1481578, 1481579, 1481580, 1481581, 1481582, 1481583, 1481584, 1481585, 1481586, 1481587, 1481588, 1481589, 1481590, 1481591, 1481592, 1481593, 1481594, 1481595, 1481596, 1481597, 1481598, 1481599, 1481600, 1481601, 1481602, 1570961, 1570962, 1570963, 1570964, 1570965, 1570966, 1570967, 1570968, 1581785, 1581786, 1581787, 1581788, 1581789, 1581790, 1581791, 1581792, 1581793, 1581794, 1581795, 1581796, 1581797, 1581798, 1581799, 1581800, 1581801, 1581802, 1581803, 1581804, 1581805, 1581806, 1581807, 1581808, 1581809, 1581810, 1581811, 1709291, 1709292, 1709293, 1709294, 1709295, 1709296, 1709297, 1709298, 1709299, 1709300, 1709301, 1709302, 1709303, 1709304, 1709305, 1709306, 1709307, 1709308, 1709309, 1709310, 1709311, 1709312, 1709313, 1709314, 1709315, 1709316, 1709317, 1709318, 1709319, 1709320, 1709321, 1709322, 1709323, 1713582, 1713583, 1713584, 1713585, 1713586, 1713587, 1713588, 1713589, 1713590, 1713591, 1713592, 1713593, 1713594, 1713595, 1713596, 1713597, 1713598, 1713599, 1713600, 1713601, 1713602, 1713604, 1713605, 1713606, 1804315, 1804316, 1804317, 1804318, 1804319, 1804320, 1804321, 1804322, 1804323, 1804324, 1804325, 1804326, 1804327, 1804328, 1804329, 1804330, 1804331, 1804332, 1804333, 1804334, 1804335, 1804336, 1804337, 1804338, 1804339, 1804340, 1804341, 1804342, 1804343, 1804344, 1933844, 1933845, 1933846, 1933847, 1933848, 1933849, 1933850, 1933851, 1933852, 1933853, 1933854, 1933855, 1933856, 1933857, 1933858, 1933859, 2026979, 2026980, 2026981, 2026982, 2026983, 2026984, 2026985, 2026986, 2026987, 2026988, 2079729, 2079730, 2079731, 2079732, 2079733, 2079734, 2079735, 2079736, 2079737, 2079738, 2079739, 2079740, 2079741, 2079742, 2079743, 2079744, 2079745, 2079746, 2079747, 2079748, 2079749, 2079750, 2079752, 2079753, 2123412, 2123413, 2123414, 2123415, 2123416, 2123417, 2123418, 2123419, 2123420, 2123421, 2123422, 2123423, 2123424, 2123425, 2123426, 2123427, 2123428, 2123429, 2123430, 2123431, 2123433, 2123434, 2123435, 2123436, 2123437, 2123438, 2156182, 2156183, 2156184, 2156185, 2156186, 2156187, 2156188, 2156189, 2156190, 2156191, 2156192, 2156193, 2156194, 2186601, 2186602, 2186603, 2186604, 2275414, 2275415, 2275416, 2275417, 2275418, 2314353, 2314354, 2314355, 2314356, 2384035, 2384036, 2384037, 2384038, 2384039, 2384040, 2384041, 2384042, 2384043, 2384044, 2384045, 2384046, 2384047, 2384048, 2384049, 2384050, 2384051, 2384052, 2384053, 2384054, 2384055, 2384056, 2384057, 2384058, 2384059, 2384060, 2384061, 2461733, 2461734, 2461735, 2461736, 2461737, 2461738, 2461739, 2461740, 2461741, 2461742, 2461743, 2499109, 2499110, 2499111, 2499112, 2499113, 2499114, 2499115, 2499116, 2499117, 2500893, 2500894, 2500895, 2500896, 2500897, 2500898, 2500899, 2547197, 2547198, 2547199, 2547200, 2547201, 2547202, 2662382, 2662383, 2662384, 2662385, 2662386, 2662387, 2662388, 2662389, 2662390, 2662391, 2662392, 2662393, 2710221, 2710222, 2710223, 2710224, 2710225, 2710226, 2710227, 2710228, 2753569, 2753570, 2753571, 2753572, 2753573, 2753575, 2753576, 2753577, 2753578, 2753579, 2753580, 2753581, 2753582, 2753583, 2753584, 2753585, 2753586, 2753588, 2753589, 2753590, 2753591, 2753593, 2753594, 2753595, 3565713, 3565714, 3565715, 3565716, 3565717, 3565718, 3668823, 3668824, 3712453, 3712454, 3719383, 3719385, 3719389, 3719391, 3719392, 3719393, 3719394, 3719395, 3866429, 3866432, 3866433, 3866439, 3866442, 3866445, 4304944, 4304945, 4304946, 4304947, 4304948, 4304950, 4304951, 4304952, 4304953, 4304955, 4304958, 4304960, 4304961, 4304962, 4304963, 4415072, 4415073, 4415074, 4415075, 4415077, 4415078, 4415079, 4415080, 4415081, 4734462, 4734464, 4746085, 4746086, 4746088, 4746089, 4746090, 4746091, 4750924, 4750925, 4750926, 4750927, 4750928, 4750929, 4750930, 4817285, 4817287, 4854315, 4854316, 4854317, 4854318, 4854319, 4854321, 4854323, 4854324, 4854325, 4854327, 4854330, 4854331, 4854332, 4854334, 5086198, 5086199, 5086200, 5086201, 5086202, 5086203, 5086204, 5086205, 5086206, 5086207, 5115719, 5115720, 5115721, 5115722, 5115723, 5115724, 5115725, 5115726, 5115727, 5115728, 5182945, 5182946, 5182947, 5182948, 5182949, 5182950, 5182951, 5182952, 5182953, 5182954, 5263793, 5263794, 5263795, 5263797, 5263798, 5263799, 5263800, 5263801, 5263802, 5263803, 5263804, 5263805, 5263806, 5263807, 5263808, 5263809, 5263810, 5263811, 5263812, 5263813, 5263814, 5263815, 5263816, 5263817, 5267042, 5400283, 5400284, 5400285, 5400286, 5400287, 5400288, 5400289, 5400290, 5400291, 5400292, 5400293, 5400294, 5400295, 5400296, 5400297, 5400298, 5400299, 5400300, 5400301, 5400302, 5491938, 5491939, 5491940, 5491941, 5491942, 5491943, 5491944, 5491945, 5491946, 5491947, 5491948, 5491949, 5491950, 5491951, 5491952, 5491953, 5491954, 5491955, 5491956, 5491957, 5575374, 5575375, 5575376, 5575377, 5575378, 5575379, 5575380, 5575381, 5575382, 5575383, 5575384, 5575385, 5575386, 5575387, 5575388, 5575389, 5575390, 5575391, 5844889, 5844891, 5844892, 5844893, 5844894, 5844896, 5844897, 5844898, 5844899, 5844900, 5844901, 5844902, 5844903, 5844904, 5844905, 5844906, 5844907, 5844908, 5844909, 5844910, 5844911, 5844913, 5844914, 5844916, 5844917, 5844918, 5875570, 5875571, 5875572, 5875573, 5875574, 5875575, 5875576, 5875577, 5875578, 5875579, 5875580, 5875581, 5875582, 5875583, 5875584, 5875585, 5875586, 5988719, 5988720, 5988721, 5988723, 5988724, 5988725, 5988726, 5988727, 5988728, 5988729, 5988730, 5988731, 5988732, 5988733, 5988735, 5988736, 5988737, 5988738, 5988739, 5988740, 5988741, 5988742; aa. REF 132-32-28, GUDID 10885862023254, Serial Numbers: 771551, 771552, 771554, 771555, 771556, 771557, 771558, 771559, 771560, 771561, 771563, 771564, 774231, 774233, 774234, 774235, 774236, 774238, 774239, 774240, 774241, 781770, 781771, 781772, 781773, 781774, 781775, 781776, 781777, 781778, 781779, 781780, 781781, 781782, 781783, 799026, 799027, 799028, 799030, 799031, 799032, 799033, 799034, 799035, 799036, 799037, 799038, 799039, 799040, 799041, 799042, 799043, 799044, 799045, 799046, 799047, 799048, 799049, 799050, 799051, 799052, 799053, 799054, 799055, 799056, 799057, 799058, 799059, 799060, 799061, 799062, 799064, 799065, 799066, 799067, 799068, 799070, 799072, 836200, 836201, 836202, 836203, 836204, 836205, 836208, 836210, 836212, 836213, 836214, 836215, 836216, 836217, 836218, 836219, 836220, 836221, 836222, 836223, 836225, 836226, 836228, 836231, 836234, 836235, 836236, 836237, 836238, 836239, 836240, 836241, 836242, 836243, 836245, 879301, 879302, 879304, 879305, 879306, 879307, 879308, 879309, 879310, 879311, 879312, 879313, 879314, 879315, 879316, 879317, 879318, 879320, 879321, 879322, 879323, 879324, 879325, 879326, 879327, 879328, 879329, 879330, 887868, 887869, 887870, 887871, 887872, 887873, 887874, 887875, 887876, 887877, 887878, 887879, 887880, 887881, 887882, 887883, 887884, 887887, 887888, 887889, 887890, 887891, 887892, 887893, 887894, 887895, 887896, 887897, 887898, 887899, 913143, 913144, 913145, 913147, 913148, 913149, 913150, 913152, 913154, 913155, 913157, 913159, 913160, 913161, 913162, 913163, 913164, 913165, 913166, 913167, 913168, 913169, 913170, 913171, 913172, 913173, 913174, 913175, 917034, 917035, 917036, 917037, 917038, 917040, 917041, 917042, 917043, 917044, 917046, 917048, 917049, 917050, 917051, 917052, 917053, 917054, 917057, 917062, 917064, 921777, 921778, 921779, 921780, 921781, 921783, 921784, 921785, 921786, 921787, 921788, 921789, 921790, 921791, 921792, 921793, 921794, 921796, 921797, 921798, 921800, 921801, 921803, 921805, 921806, 921807, 921808, 921809, 1039219, 1039223, 1039224, 1039225, 1039226, 1039227, 1039228, 1039231, 1039236, 1039237, 1039238, 1039240, 1039241, 1039244, 1039245, 1039248, 1080429, 1080430, 1080432, 1080433, 1080434, 1080440, 1080445, 1080446, 1080449, 1080450, 1080451, 1080452, 1080453, 1202469, 1202471, 1202472, 1202473, 1202474, 1202475, 1202476, 1202477, 1202481, 1202483, 1202484, 1202486, 1202494, 1202495, 1202497, 1736644, 1736645, 1736646, 1736647, 1736649, 1736650, 1736651, 1736652, 1736653, 1736654, 1736655, 1736656, 1736658, 1736659, 1736660, 1736661, 1736662, 1736663, 1928639, 1928640, 1928642, 1990221, 1990222, 1990223, 1990224, 1990225, 1990227, 1990228, 1990229, 1990230, 1990231, 1990232, 1990234, 1990235, 1990237, 1990238, 1990239, 1990241, 1990242, 1990243, 1990244, 1990245, 1990246, 1990247, 1990248, 1990249, 1990250, 2033597, 2033598, 2033599, 2033600, 2033601, 2042570, 2112094, 2112095, 2112096, 2112097, 2112098, 2112099, 2335465, 2335466, 2335467, 2335468, 2335469, 2335470, 2383788, 2383789, 2383790, 2383791, 2383792, 2383793, 2383794, 2383795, 2383796, 2383797, 2383798, 2383799, 2383800, 2383801, 2383802, 2383803, 2383804, 2383805, 2383806, 2383807, 2383808, 2383809, 2383810, 2383811, 2383812, 2383813, 2383814, 2383815, 2383816, 2383817, 2383818, 2383819, 2383820, 2383821, 2383822, 2383823, 2581902, 2581903, 2581904, 2581905, 2581906, 2581907, 2673754, 2673755, 2673757, 2673759, 2673760, 2673761, 2673762, 2673763, 2673764, 2673765, 2673767, 2673768, 2673769, 2673771, 2673772, 2673773, 2673774, 2786750, 2786751, 2786752, 2786754, 3569825, 3569827, 3569828, 3569830, 3866450, 3866465, 3866467, 3866470, 4673782, 4673783, 4746060, 4746062, 4746064, 4781984, 4781986, 4781987, 4781988, 4781989, 4781990, 4781991, 4781992, 4854336, 4854339, 4854340, 4854341, 4854342, 5245445, 5346893, 5346894, 5346895, 5346896, 5862911, 5862912, 5862913, 5862914, 5862915, 5862916, 5862917, 6019323, 6019325, 6019334, 6019335, 6019337; bb. REF 132-32-29, GUDID 10885862023261, Serial Numbers: 771191, 771192, 771193, 771194, 771195, 771196, 771197, 771199, 771200, 771201, 771202, 771203, 774266, 774270, 774272, 774274, 774275, 774276, 781707, 781708, 781709, 781711, 781712, 781713, 781714, 781715, 799079, 799080, 799082, 799084, 799085, 799086, 799087, 799088, 799090, 799093, 799099, 799101, 799102, 799103, 799104, 799105, 799106, 799107, 799108, 799111, 799112, 799114, 799115, 799116, 799118, 799120, 799121, 799122, 799123, 799124, 799125, 799126, 799127, 799128, 799129, 799130, 799131, 799133, 799134, 799136, 799137, 799138, 799139, 799141, 844057, 844058, 844060, 844061, 844065, 844066, 844067, 844068, 844069, 844070, 844071, 844074, 844075, 844101, 887788, 887792, 887799, 887800, 887802, 887803, 887805, 887807, 887809, 887811, 887815, 887817, 887818, 887819, 918693, 918694, 918696, 918697, 918698, 918699, 918700, 918703, 918707, 918709, 918711, 918713, 918715, 918717, 918718, 918719, 918720, 918721, 918722, 918723, 918724, 928298, 928310, 928319, 928321, 928324, 1276586, 1276587, 1276592, 1276598, 1276599, 1276600, 1276602, 1276603, 1276605, 1276607, 1276611, 1276619, 1276620, 1276621, 1276623, 1276629, 1896636, 1896637, 1896638, 1896642, 1896643, 1896644, 1896645, 1896646, 1896651, 1896652, 1896653, 1896654, 1896656, 1896657, 1896658, 1896659, 1896661, 2323665, 2347657, 2347658, 2347659, 2384062, 2384063, 2384064, 2384065, 2384066, 2384067, 2384068, 2384069, 2384070, 2384071, 2384072, 2384073, 2384074, 2384075, 2384076, 2384077, 2384078, 2384079, 2384080, 2384081, 2384082, 2667203, 2667204, 2667205, 2667208, 2667210, 2667211, 2667212, 2667214, 2667215, 2667217, 2667218, 2667219, 2667220, 2667222, 2667224, 2667227, 2667228, 2667229, 3822478, 5870368, 5870369, 5870370, 5870371, 5870372; cc. REF 132-32-30, GUDID 10885862023278, Serial Numbers: 771240, 771242, 793763, 798611, 1058424, 1058428, 1058430, 1058455, 2083749, 2083755, 2083757, 2083758, 2083759, 2083761, 2089577, 2089578, 2089580, 2089582, 2089585, 2089586, 2089647, 2089648, 2089650, 2089654, 2089655, 2089656, 2089658, 2089661, 2089663, 2089967, 2089969, 3565702, 3565703, 3565704, 3565705, 3861721; dd. REF 132-36-28, GUDID 10885862023391, Serial Numbers: 772757, 772758, 772759, 772760, 772761, 772763, 772764, 772765, 772766, 772767, 772768, 772769, 772770, 776296, 776297, 776298, 776299, 776300, 776301, 776302, 776303, 776304, 776305, 776306, 776307, 776308, 776309, 781986, 781987, 781988, 781989, 781990, 781991, 807044, 807045, 807046, 807047, 807048, 807049, 807050, 807051, 807052, 807053, 807054, 807055, 807056, 830320, 830321, 830322, 830323, 830324, 830325, 830326, 830327, 830328, 830329, 830330, 830331, 830332, 830333, 830334, 830335, 830336, 830337, 830338, 830339, 830340, 830341, 830342, 862147, 862148, 862149, 862150, 862151, 862152, 862153, 862154, 862155, 862156, 862157, 862158, 862159, 862160, 862161, 862162, 862163, 862164, 862165, 862166, 862167, 862168, 862169, 862170, 862171, 862172, 862173, 869050, 869051, 869052, 869053, 869054, 869055, 869056, 869057, 869058, 869059, 869060, 869061, 869062, 869063, 869064, 869065, 869066, 869067, 869068, 869069, 869070, 869071, 869072, 869073, 869074, 869075, 869076, 869077, 869078, 896221, 896222, 896223, 896224, 896225, 896226, 896227, 896228, 896229, 896230, 896231, 896232, 896233, 896235, 896236, 896237, 896238, 896239, 896240, 896241, 896242, 896243, 896244, 896246, 896247, 896248, 896249, 896250, 896251, 918658, 918659, 918660, 918661, 918662, 918663, 918664, 918665, 918666, 918667, 918668, 918669, 918670, 918671, 918672, 918673, 918674, 918675, 918676, 918677, 918678, 918679, 918680, 918681, 918682, 918683, 918684, 918685, 918686, 918687, 918688, 918689, 935607, 935608, 935609, 935610, 935611, 935612, 935613, 935614, 935615, 935616, 935617, 935618, 935619, 935620, 935621, 935622, 935623, 935624, 935625, 935626, 935627, 935628, 935629, 935630, 935631, 935632, 935633, 935634, 935635, 935636, 935637, 935638, 935639, 953240, 953241, 953242, 953243, 953244, 953245, 953246, 953247, 953248, 953249, 953250, 953251, 953252, 953253, 953254, 953255, 953256, 953257, 953258, 953259, 953260, 953261, 953262, 953263, 953264, 953265, 953266, 953267, 953268, 953269, 953270, 953271, 953272, 953273, 953274, 953275, 953276, 953277, 953278, 953279, 953280, 953281, 953282, 953283, 953284, 953285, 953286, 953287, 953288, 953289, 953290, 953292, 991797, 991798, 991799, 991800, 991801, 991802, 991803, 991804, 991805, 991806, 991807, 991808, 991809, 991810, 991811, 991812, 991813, 991814, 991815, 991816, 991817, 991818, 991819, 991820, 991821, 991822, 991823, 991824, 991825, 991826, 991827, 991828, 991829, 991830, 991831, 991832, 991833, 991834, 991835, 991836, 991837, 991838, 991839, 991840, 991841, 991842, 991843, 991844, 991845, 991846, 991847, 991848, 991849, 991850, 1026463, 1026464, 1026465, 1026466, 1026467, 1026468, 1026469, 1026470, 1026471, 1026472, 1026473, 1026474, 1026475, 1026476, 1026477, 1026478, 1026479, 1026480, 1026481, 1026482, 1026483, 1026484, 1026485, 1026486, 1026487, 1026488, 1026489, 1026490, 1026491, 1026492, 1026493, 1026494, 1026495, 1026496, 1026497, 1026498, 1026499, 1026500, 1026501, 1026502, 1026503, 1026504, 1026505, 1026506, 1026507, 1026508, 1026509, 1026510, 1026511, 1026512, 1026513, 1026514, 1026515, 1026516, 1062509, 1062510, 1062511, 1062512, 1062513, 1062514, 1062515, 1062516, 1062517, 1062518, 1062519, 1062520, 1062521, 1062522, 1062523, 1062524, 1062525, 1062526, 1062527, 1062528, 1062529, 1062530, 1062531, 1062532, 1062533, 1062534, 1062535, 1062536, 1062537, 1062538, 1062539, 1062540, 1062541, 1062542, 1062543, 1062682, 1062683, 1062684, 1062685, 1062686, 1062687, 1062688, 1062689, 1062690, 1062691, 1062692, 1062693, 1062694, 1062695, 1062696, 1062697, 1062698, 1068095, 1068096, 1068097, 1068098, 1068099, 1068100, 1068101, 1068102, 1068103, 1068104, 1068105, 1068106, 1068107, 1068108, 1068109, 1068110, 1068111, 1068112, 1068113, 1068114, 1068115, 1068116, 1068117, 1068118, 1068119, 1068120, 1068121, 1068122, 1068123, 1068124, 1068125, 1068126, 1068127, 1068128, 1068129, 1068130, 1068131, 1068132, 1068133, 1068134, 1068135, 1068136, 1068137, 1068138, 1068139, 1068140, 1068141, 1068142, 1068143, 1068144, 1068145, 1068146, 1068147, 1068148, 1068149, 1068150, 1068151, 1068152, 1068153, 1068154, 1144008, 1144009, 1144010, 1144011, 1144012, 1144013, 1144014, 1144015, 1144016, 1144017, 1144018, 1144019, 1144020, 1144021, 1144022, 1144023, 1144024, 1144025, 1144026, 1144027, 1144028, 1144029, 1144030, 1144031, 1144032, 1144033, 1144034, 1144035, 1144036, 1144037, 1144038, 1144039, 1144040, 1144042, 1144043, 1144045, 1144046, 1144047, 1144048, 1144049, 1144050, 1144051, 1144052, 1144053, 1144054, 1144055, 1144056, 1144057, 1144058, 1144059, 1144060, 1144061, 1193640, 1193641, 1193642, 1193643, 1193644, 1193645, 1193646, 1193647, 1193648, 1193649, 1193650, 1193651, 1193652, 1193653, 1193654, 1193655, 1193657, 1193658, 1193659, 1193660, 1193661, 1193662, 1193663, 1193664, 1193665, 1193667, 1193668, 1193669, 1193670, 1193671, 1193672, 1193673, 1193674, 1193675, 1193676, 1193677, 1193678, 1193679, 1193680, 1193681, 1193682, 1193683, 1193684, 1193685, 1193687, 1193688, 1193689, 1193690, 1193691, 1193692, 1193693, 1288457, 1288458, 1288459, 1288460, 1288461, 1288462, 1288463, 1288464, 1288465, 1288466, 1288467, 1288468, 1288469, 1288470, 1288471, 1288472, 1288473, 1288474, 1288475, 1288476, 1288477, 1288478, 1288479, 1288480, 1288481, 1288482, 1288483, 1288484, 1288485, 1288486, 1288487, 1288488, 1288489, 1288490, 1288491, 1288492, 1288493, 1288494, 1288495, 1288496, 1288497, 1288498, 1288499, 1288500, 1288501, 1288502, 1288503, 1288504, 1288505, 1288506, 1288507, 1288508, 1288509, 1288510, 1288511, 1385705, 1385706, 1385707, 1385708, 1385709, 1385710, 1385711, 1385712, 1385713, 1385714, 1385715, 1385716, 1385717, 1385718, 1385719, 1385720, 1385721, 1385722, 1385723, 1385724, 1385725, 1385726, 1439111, 1439112, 1439113, 1439114, 1439115, 1439116, 1439117, 1439118, 1439119, 1439120, 1439121, 1439122, 1439123, 1439124, 1439125, 1439126, 1439127, 1439128, 1439129, 1439130, 1540443, 1540444, 1540445, 1540446, 1540447, 1540448, 1540449, 1540450, 1540511, 1540512, 1540513, 1540514, 1540515, 1540516, 1540517, 1540518, 1540519, 1581338, 1581339, 1581340, 1581341, 1581342, 1581343, 1581345, 1581346, 1581347, 1581348, 1581349, 1581350, 1581351, 1581352, 1581353, 1581354, 1581355, 1581356, 1581357, 1581358, 1581359, 1581360, 1581361, 1581362, 1581363, 1581364, 1581365, 1581367, 1637470, 1637471, 1637472, 1637473, 1637474, 1637475, 1637476, 1637477, 1637478, 1637479, 1637480, 1637481, 1637482, 1637483, 1637484, 1637485, 1727011, 1727012, 1727013, 1727014, 1727015, 1727016, 1727017, 1727018, 1727019, 1727020, 1727021, 1727022, 1727023, 1727024, 1727025, 1727026, 1727027, 1727028, 1727029, 1727030, 1727031, 1727032, 1727033, 1727034, 1727035, 1727036, 1727173, 1727174, 1727175, 1727176, 1727177, 1727178, 1727179, 1727180, 1727181, 1727182, 1727183, 1727184, 1727185, 1727186, 1727187, 1727188, 1727189, 1727190, 1727191, 1727192, 1727193, 1727194, 1727195, 1727196, 1727197, 1727198, 1889749, 1889750, 1889751, 1889752, 1889753, 1889754, 1889755, 1889756, 1889757, 1889758, 1889759, 1889760, 1889761, 1889762, 1889763, 1889764, 1889765, 1889766, 1889767, 1889768, 1889769, 1889770, 1889771, 1900332, 1900333, 1900334, 1900335, 1900336, 1900337, 1900338, 1900339, 1900340, 1900341, 1900342, 1900343, 1900344, 1900345, 1900346, 1900347, 1900348, 1900349, 1900350, 1900351, 1900352, 1900353, 1900354, 1900355, 1900356, 1900357, 1900358, 1900359, 1900360, 1900361, 2019166, 2019167, 2019168, 2019169, 2019170, 2019171, 2019172, 2019173, 2019174, 2019175, 2019176, 2019177, 2019178, 2019179, 2019180, 2019181, 2019182, 2019183, 2019184, 2019185, 2019186, 2019187, 2019188, 2019189, 2019190, 2019191, 2019192, 2019193, 2019194, 2042572, 2042573, 2042574, 2042575, 2067975, 2067976, 2067977, 2067978, 2067979, 2067980, 2089034, 2089035, 2089036, 2089037, 2089038, 2089039, 2089040, 2089041, 2089042, 2089043, 2123593, 2123594, 2123595, 2123596, 2123597, 2123598, 2123599, 2123601, 2123602, 2123603, 2123604, 2123605, 2123606, 2123607, 2123608, 2123609, 2123610, 2123611, 2123612, 2123613, 2123614, 2123615, 2123616, 2123617, 2123618, 2162168, 2162169, 2162170, 2162171, 2162174, 2162175, 2162176, 2162177, 2162178, 2162179, 2162180, 2162182, 2170206, 2170207, 2170208, 2170210, 2170211, 2170214, 2170215, 2170216, 2170217, 2225473, 2225474, 2225475, 2225476, 2225477, 2225478, 2225479, 2225480, 2230644, 2244608, 2244609, 2263208, 2263209, 2263210, 2263211, 2263212, 2309398, 2309399, 2309400, 2380581, 2380582, 2380583, 2380584, 2380585, 2380587, 2380590, 2380592, 2380593, 2380594, 2380596, 2380597, 2380598, 2380600, 2380601, 2380602, 2380604, 2380605, 2380606, 2667859, 2667860, 2667861, 2667862, 2667863, 2667864, 2667865, 2667866, 2667867, 2667868, 2667869, 2667871, 2667872, 2667873, 2667874, 2667876, 2667877, 2667878, 2667879, 2667880, 2667881, 2667882, 2667883, 2667884, 2667885, 4447491, 4447492, 4447493, 4447494, 4447495, 4447496, 4447497, 4447498, 4447499, 4447500, 4447501, 4447502, 4447503, 4447504, 4447505, 4447506, 4447507, 4447508, 4447509, 4447510, 5670653, 5670654, 5670655, 5754417, 5754418, 5754419, 5931999, 5932000, 5932004, 5932007, 5932011, 5932012, 5932015; ee. REF 132-36-29, GUDID 10885862023407, Serial Numbers: 772775, 772776, 772777, 772778, 772779, 772780, 772781, 772782, 772783, 772784, 772785, 772786, 772787, 772788, 776324, 776325, 776326, 776327, 776328, 776329, 776330, 776331, 776332, 776333, 776334, 776335, 776336, 776337, 782065, 782066, 782067, 782068, 782069, 782072, 823528, 823529, 823530, 823531, 823533, 823534, 823535, 823536, 823537, 823538, 823539, 823540, 823542, 823543, 823544, 835823, 835824, 835825, 835826, 835827, 835828, 835829, 835830, 835831, 835832, 835834, 835835, 835836, 835838, 835839, 835840, 836696, 836697, 836699, 836700, 836701, 836702, 836703, 836704, 897630, 897631, 897632, 897633, 897634, 897635, 897636, 897638, 897639, 897640, 897641, 897642, 897643, 897645, 897646, 897647, 897648, 897649, 897650, 897651, 897652, 897653, 897654, 897655, 897657, 897659, 897660, 922706, 922707, 922708, 922709, 922710, 922711, 922712, 922713, 922714, 922715, 922716, 922717, 922718, 922719, 922720, 922721, 922722, 922723, 922725, 922726, 922727, 922728, 922729, 922730, 922731, 944883, 944884, 944885, 944886, 944887, 944888, 944889, 944890, 944892, 944893, 944894, 944895, 944896, 944898, 944899, 944900, 944901, 944903, 944905, 944906, 944907, 944908, 944909, 944912, 944913, 944914, 944915, 944916, 944917, 944918, 944919, 944920, 944921, 944922, 944923, 944924, 944925, 944926, 944928, 944929, 944932, 944934, 944936, 944937, 1039890, 1039891, 1039892, 1039893, 1039894, 1039895, 1039896, 1039897, 1039898, 1039899, 1039900, 1039901, 1039902, 1039903, 1039905, 1039906, 1039908, 1039909, 1039910, 1039911, 1039912, 1039914, 1039915, 1039917, 1039918, 1039920, 1039922, 1039923, 1039924, 1039925, 1039926, 1039927, 1039929, 1039930, 1039931, 1039932, 1039933, 1105465, 1105467, 1105468, 1105469, 1105472, 1105473, 1105475, 1105476, 1105477, 1105478, 1105479, 1105483, 1105486, 1105488, 1105489, 1105491, 1105492, 1105493, 1105494, 1105495, 1105498, 1105500, 1105501, 1105502, 1105503, 1105504, 1105507, 1105511, 1105513, 1105514, 1105515, 1105517, 1105518, 1105519, 1540463, 1540464, 1540465, 1540466, 1540467, 1540469, 1540470, 1540471, 1662593, 1662594, 1662595, 1662596, 1662597, 1662598, 1662599, 1662600, 1662601, 1662602, 1662603, 1662604, 1662605, 1662606, 1662607, 1662608, 1662609, 1662610, 1662611, 1662612, 1662613, 1662614, 1662615, 1662616, 1859567, 1859569, 1859571, 1859572, 1859573, 1859574, 1900217, 1900219, 1900220, 1900221, 1900222, 1900223, 1900225, 1900228, 1990694, 1990696, 1990699, 1990702, 1990706, 1990707, 1990708, 1990709, 1990710, 1990712, 1990714, 1990715, 1990716, 2042399, 2107199, 2107204, 2107205, 2107206, 2107211, 2107215, 2107223, 2139832, 2140741, 2140742, 2140743, 2140744, 2140745, 2140746, 2157100, 2667671, 2667672, 2667673, 2667674, 2667675, 2667676, 2667677, 2741959, 2741960, 2741961, 2741962, 2781222, 2781224, 2781225, 3873790, 3873795, 3873796, 3873800, 3873801, 3873807, 6300588; ff. REF 132-36-30, GUDID 10885862023414, Serial Numbers: 835868, 1097647, 1097672, 1097675, 1097680, 1633919, 1633920, 1633921, 1633922, 1633923, 1633924, 1633926, 1633927, 1633928, 1633929, 1633930, 2107805, 2140752, 2140758, 2140762, 2140766, 2140794, 2667727, 2667728, 2667729, 2667730, 2667731, 2667732, 2741948, 2741949, 2741953, 2741954, 2741955, 2741957; gg. REF 134-28-25, GUDID 10885862023919, Serial Numbers: 803943, 803945, 803958, 836308, 2127532, 2127558; hh. REF 134-28-26, GUDID 10885862023926, Serial Numbers: 804295, 804296, 804297, 804298, 804299, 804301, 804304, 804306, 804310, 804311, 804312, 804315, 804316, 830219, 830225, 830227, 830228, 830230, 830232, 830239, 830246, 830251, 830252, 830258, 830259, 830260, 830261, 830265, 830274, 1093286, 1093294, 1093295, 1093296, 1093297, 2112278, 2112279, 2112305, 2112312, 2112324, 2112327, 3792310, 3792311; ii. REF 134-28-27, GUDID 10885862023933, Serial Numbers: 804139, 804140, 804141, 804142, 804144, 804146, 804148, 804151, 804152, 804153, 804155, 804156, 804157, 804159, 804160, 804161, 804163, 823181, 823182, 823183, 823184, 823186, 823189, 823191, 823194, 823198, 823199, 823200, 823201, 823202, 823204, 839535, 839538, 839539, 839540, 839542, 839543, 839546, 839547, 839550, 839551, 839560, 839561, 839564, 1093483, 1093491, 1093503, 1093505, 2103929; jj. REF 134-28-28, GUDID 10885862023940, Serial Numbers: 805333, 805334, 805335, 805336, 805337, 805339, 805340, 805346, 805347, 805348, 805350, 805351, 805352, 823207, 823209, 823210, 823212, 823215, 823217, 823219, 823222, 823224, 823225, 823226, 839513, 839516, 839521, 839525, 839527, 839530, 1093772, 1093789; kk. REF 134-28-29, GUDID 10885862023957, Serial Numbers: 805240, 805241, 805243, 805246, 805247, 805252, 805254, 827206, 827212, 827215, 827216, 827217, 827224, 827225, 827228, 827234, 827241, 827247, 827248, 827253, 1093609, 2103873; ll. REF 134-28-30, GUDID 10885862023964, Serial Numbers: 823177 mm. REF 104-28-43, GUDID 10885862013491, Serial Numbers: 979925, 979932, 979935, 1453403, 1453406, 1453408, 1453409, 1453411, 1453412, 1973630, 1973632, 2669930, 2669931, 2669934, 3840144, 3871959, 3871960; nn. REF 104-28-44, GUDID 10885862013507, Serial Numbers: 972597, 972598, 972601, 972602, 972603, 972604, 972605, 972606, 1307760, 1307761, 1307762, 1307763, 1307764, 1307766, 1307767, 1307768, 1307769, 1307770, 1502626, 1502627, 1502628, 1502630, 1502631, 1502632, 1502633, 1502634, 1988966, 2018727, 2018728, 2018729, 2018730, 2018731, 2203229, 2203230, 2219876, 2275736, 2275737, 2697719, 2697720, 2697721, 2697722, 2730617, 2776599, 2776600, 2782325, 2782326, 2818614, 3622010, 3622011, 3622012, 4024781, 4024782, 4024783, 4024784, 4024785, 4024787, 4024788, 4024790, 4024791, 4024792, 4994819, 5091046, 5091047, 5543694, 5772778, 5772781; oo. REF 104-28-45, GUDID 10885862013514, Serial Numbers: 1511277, 1511279, 1511280, 1511281, 1668331, 1668334, 2684641, 5318065, 5334779; pp. REF 104-32-45, GUDID 10885862013606, Serial Numbers: 971675, 971676, 971677, 971678, 971679, 971680, 971682, 971683, 1097196, 1097197, 1097198, 1097199, 1097200, 1097201, 1097202, 1097203, 1097204, 1193427, 1193428, 1193429, 1193430, 1193432, 1193434, 1394007, 1394008, 1394009, 1394010, 1433160, 1433161, 1433162, 1433163, 1453971, 1453972, 1453973, 1453974, 1453975, 1453976, 1453977, 1453978, 1453979, 1453980, 1453981, 1920459, 1920460, 1920461, 1920462, 1935776, 1935777, 1935778, 1935779, 1935780, 1935781, 1935782, 1935783, 2131262, 2131263, 2131264, 2131265, 2131266, 2219836, 2219837, 2219838, 2219839, 2219840, 2667678, 2667679, 2667680, 2667681, 2667682, 2710324, 2710325, 2710326, 3511533, 3544589, 3634243, 3808040, 3808041, 3808042, 3808043, 3808044, 3808045, 3808046, 3808047, 3808048, 3808049, 3808050, 3808051, 3808052, 3808053, 3808054, 3808056, 3808057, 3808058, 3808059, 3808060, 3808061, 3808062, 5092213, 5092214, 5092215, 5092216, 5519138, 5519139, 5772773, 5772774, 5772776, 6237495; qq. REF 104-32-46, GUDID 10885862013613, Serial Numbers: 972592, 972596, 1306174, 1306177, 2669935, 2669936.3769239; rr. REF 104-36-46, GUDID 10885862013682, Serial Numbers: 973299, 973302, 973303, 973304, 973305, 1302613, 1302614, 1302615, 1302616, 1302617, 1302618, 1776716, 1776717, 1776718, 1776719, 1776721, 1776722, 1776727, 1776735, 1776737, 1776740, 3702614, 3735566; ss. REF 104-36-47, GUDID 10885862013699, Serial Numbers: 2669940
FEI Number 1038671
Recalling Firm/
Manufacturer
Exactech, Inc.
2320 Nw 66th Ct
Gainesville FL 32653-1630
For Additional Information Contact
888-912-0403
Manufacturer Reason
for Recall
Specific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.
FDA Determined
Cause 2
Process control
ActionThe firm sent an "URGENT MEDICAL DEVICE RECALL" DEAR HEALTHCARE PROFESSIONAL COMMUNICATION on 08/11/2022 by email. The purpose of the current letter is to update surgeons regarding the implications of the packaging non-conformity on performance of the Connexion GXL and other conventional UHMWPE acetabular liners. In addition, this letter provides updates to patient management and follow-up recommendations since issuance of the July 2021 DHCP letter. In addition to providing surgeons with lists of all their affected GXL and/or conventional polyethylene patients since 2004, Exactech is providing surgeons with two draft letters directed at patients who have been implanted with Exactech GXL hip devices. The Customers were instructed to take the following actions: CAREFULLY REVIEW THIS RECALL NOTIFICATION to ensure that you fully understand the issue identified, the recall strategy, and all actions required. IMMEDIATELY IDENTIFY AND QUARANTINE any of the subject devices in your inventory and/or customers inventory. EXTEND THE DESCRIPTION OF THE ISSUE AND CLINICAL IMPACT as described in the recall notification to your accounts that may have this product in their possession. COMPLETE AND RETURN the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com. Reach out to Customer Service at cust.svc@exac.com to request a Returned Goods (RG) Number for ALL affected product requested to be returned. WE ARE REQUIRING 100% EFFECTIVENESS FOR THIS RECALL. If you have any questions, call (888) 912-0403.
Quantity in Commerce11143 DEVICES
DistributionWorldwide distribution: US (nationwide) and OUS (foreign) countries of: ARGENTINA, BRAZIL, CHILE, CHINA, COLOMBIA, ECUADOR, GREECE, GUATEMALA, INDIA, LEBANON PORTUGAL,PUERTO RICO,SPAIN
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = LPH
-
-