• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall nanoDot

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall nanoDotsee related information
Date Initiated by FirmJuly 12, 2023
Date PostedSeptember 12, 2023
Recall Status1 Completed
Recall NumberZ-2569-2023
Recall Event ID 92647
510(K)NumberK192196 
Product Classification Accelerator, linear, medical - Product Code IYE
ProductQC (Cs-137) nanoDot D2DNS, Model Numbers: a) 18135-000; b) 18135-KIT; c) 18135-SET; radiation monitoring dosimeter used with the microSTAR readers
Code Information UDI/DI 0860003399903, all batch numbers
FEI Number 3008426232
Recalling Firm/
Manufacturer
Landauer
2 Science Rd
Glenwood IL 60425-1531
For Additional Information ContactBrian Malone
815-690-1231
Manufacturer Reason
for Recall
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
FDA Determined
Cause 2
Under Investigation by firm
ActionLandauer disseminated an URGENT - Medical Device Recall notice to its consignees on 07/12/2023 by US mail and email. The notice explained the problem with the devices, potential risk, and requested the following: Discontinue use of the affected products, destroy used products, and return unused products.
Quantity in Commerce10741 units
DistributionWorldwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
510(K) Database510(K)s with Product Code = IYE
-
-