• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Sensi Disc Amoxicillin with Clavulanic Acid

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Sensi Disc Amoxicillin with Clavulanic Acidsee related information
Date Initiated by FirmJanuary 08, 2024
Date PostedFebruary 08, 2024
Recall Status1 Open3, Classified
Recall NumberZ-1011-2024
Recall Event ID 93789
510(K)NumberK842253 
Product Classification Susceptibility test discs, antimicrobial - Product Code JTN
ProductSensi Disc Amoxicillin with Clavulanic Acid 20/10 g, Catalog No. 231628 and 231629. These discs are used for semi-quantitative in vitro susceptibility testing
Code Information Catalog No. 231628 UDI-DI 00382902316282 Lots 2090490 2146520 2210201 2276143 3062095; Catalog No. 231629¿¿ UDI-DI 30382902316290 Lots 3163306 3163307 3163310 3254307 3254309 3271840 3271853 2146520 2146526 2147225 2210201 2224295 2276143 2276144 3003894 3062095 3090649
FEI Number 1119779
Recalling Firm/
Manufacturer
Becton Dickinson & Co.
7 Loveton Cir
Sparks MD 21152-9212
Manufacturer Reason
for Recall
There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
FDA Determined
Cause 2
Process control
ActionOn January 8, 2024, the firm notified customers through URGENT Medical Device Product Correction letters. Users were instructed to cease use of catalog number 291270, and to refrain from conducting AST for H. influenzae with impacted Sensi-Discs. Customers should destroy any units of catalog number 291270 in their inventory. Customers should share and post the recall notification within their facility network and forward to any customers to whom they may have distributed the product. Future lots of affected Sensi-Disc products (excluding catalog number 291270) will be labeled with a sticker notifying customers "This product should not be used for the semi-quantitative in vitro susceptibility testing of Haemophilus influenzae." Identify impacted lot numbers: https://legacy.bd.com/alerts-notices/IDS-23-4851.asp For customer questions: Phone: 1-844-823-5433 Say "Recall when prompted Mon-Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
Quantity in Commerce2,363,168 total units
DistributionDomestic distribution nationwide. International distribution worldwide.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = JTN
-
-