• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Medline

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 1 Device Recall Medline see related information
Date Initiated by Firm November 15, 2023
Date Posted February 25, 2024
Recall Status1 Open3, Classified
Recall Number Z-1128-2024
Recall Event ID 93800
Product Classification Catheter and tip, suction - Product Code JOL
Product Medline Kits, trays, and packs labeled as follows:
a) KIT,SUCTION CATH,10FR W/WATER, REF DYND41470;
b) KIT,SUCTION CATH,12FR W/H20 12, REF DYND41471;
c) KIT,SUCTION CATH,14FR W/SALIN, REF DYND40442;
d) KIT,SUCTION CATH,14FR W/WATER, REF DYND41472;
e) KIT,SUCTION CATH,18FR W/H20, REF DYND41474;
f) KIT,SUCTION CATH,6FR,DELEE,W/H, REF DYND41476;
g) KIT,SUCTION CATH,8FR DELEE W/H, REF DYND41478;
h) KIT,SUCTION CATHETER,14FR W/S, REF DYND41442
Code Information a) REF DYND41470, UDI/DI 40080196033608(case), 10080196033607(unit), Lot Numbers: 2022042690, 2022070690, 2022090790, 2022102590, 2023051790; b) REF DYND41471, UDI/DI 40080196033622(case), 10080196033621(unit), Lot Numbers: 2022070190, 2022072890, 2022101090, 2022121390, 2022121990, 2023022890, 2023051890; c) REF DYND40442, UDI/DI 40080196032618(case), 10080196032617(unit), Lot Numbers: 2022040590, 2022051890, 2022072290, 2022090690, 2022100790, 2022093090, 2022101190, 2022122890, 2023051290, 2023060590, 2023080290, 2023092690; d) REF DYND41472, UDI/DI 40080196996538(case), 10080196996537(unit), Lot Numbers: 2022042590, 2022062390, 2022070590, 2022072590, 2022080890, 2022082590, 2023010690, 2023022790, 2023030890, 2023041090, 2023041990, 2023070590; e) REF DYND41474, UDI/DI 40080196033653(case), 10080196033652(unit), Lot Numbers: 2022082390, 2022101090, 2022121290, 2022122090, 2023030390, 2023070590, 2023092590; f) REF DYND41476, UDI/DI 40080196033660(case), 10080196033669(unit), Lot Numbers: 2022062290, 2022072890, 2022102490, 2022122090, 2023030190, 2023051990; g) REF DYND41478, UDI/DI 40080196033677(case), 10080196033676(unit), Lot Numbers: 2022042890, 2022070690, 2022082390, 2022100690, 2022121290, 2023010490, 2023041490, 2023080890, 2023092590; h) REF DYND41442, UDI/DI 40080196033578(case), 10080196033577(unit), Lot Numbers: 2022031590, 2022032290, 2022042790, 2022060290, 2022053190, 2022052790, 2022060690, 2022072690, 2022072790, 2022082490, 2022102490, 2023012490, 2023030390, 2023032390, 2023032490, 2023041790, 2023060190, 2023060290, 2023070590
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information Contact Karin Johnson
886-359-1704
Manufacturer Reason
for Recall
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
FDA Determined
Cause 2
Nonconforming Material/Component
Action MEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different sterile supply. Distributors or consignees who further distributed the affected products were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity in Commerce 1,012,700 units
Distribution Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-