• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Smiths Medical

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Smiths Medicalsee related information
Date Initiated by FirmMay 31, 2024
Date PostedSeptember 20, 2024
Recall Status1 Open3, Classified
Recall NumberZ-3091-2024
Recall Event ID 94974
510(K)NumberK123957 
Product Classification Ventilator, emergency, powered (resuscitator) - Product Code BTL
ProductparaPAC 300 ventilator kit without internal PEEP and CPAP, Item Number P300N
Code Information Serial Numbers: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37, 38, 39, 40, 41, 42, 43, 44, 45, 46, 47, 48, 49, 50, 51, 52, 53, 54, 55, 56, 57, 58, 59, 60, 61, 62, 63, 64, 65, 66, 67, 68, 69, 234, 235, 243, 244, 245, 246, 247, 248, 249, 250, 251, 252, 259, 260, 261, 262, 263, 264, 265, 266, 267, 268, 269, 270, 271, 272, 273, 274, 275, 276, 277, 278, 279, 280, 281, 282, 283, 284, 285, 286, 287, 288, 289, 290, 291, 292, 293, 294, 295, 296, 297, 298, 299, 300, 301, 302, 303, 304, 70085, 70091, 70096, 70097, 70101, 70103, 70111, 70112, 70114, 70119, 70142, 70171, 70299, 1007188, 1102432, 1102433, 1206210, 1206211, 1206212, 1206213, 1206214, 1206235, 1206236, 1206237, 1206238, 1210164, 1210165, 1210166, 1303220, 1303228, 1303229, 1303230, 1303231, 1303232, 1303233, 1303234, 1304223, 1304224, 1304225, 1304226, 1304227, 1304228, 1304229, 1304230, 1304231, 1304232, 1304233, 1304234, 1304235, 1304236, 1304237, 1305192, 1305193, 1305194, 1305195, 1305196, 1305197, 1305198, 1305199, 1305200, 1305201, 1310001, 1310129, 1311132, 1311133, 1311134, 1311135, 1311136, 1311137, 1311138, 1311139, 1311140, 1311141, 1401228, 1403142, 1406374, 1406375, 1406376, 1406377, 1406378, 1406379, 1406380, 1406381, 1406382, 1406383, 1407291, 1407292, 1407293, 1407294, 1407295, 1407296, 1407297, 1407298, 1407299, 1407300, 1407430, 1407431, 1407432, 1407433, 1407434, 1407441, 1407442, 1407443, 1407444, 1408134, 1408135, 1408136, 1408137, 1408138, 1408139, 1408140, 1408141, 1408142, 1408143, 1408180, 1408181, 1408182, 1408183, 1408184, 1408185, 1408186, 1408187, 1408188, 1408189, 1408205, 1408206, 1408207, 1408208, 1408209, 1408210, 1408211, 1408212, 1408213, 1408214, 1408225, 1408226, 1408227, 1408228, 1408254, 1502242, 1504129, 1504130, 1505175, 1505176, 1505177, 1505178, 1505179, 1505180, 1505181, 1505182, 1505183, 1505184, 1505185, 1505186, 1505187, 1505188, 1505189, 1505190, 1505191, 1505192, 1505193, 1505194, 1505195, 1505196, 1505197, 1505198, 1505199, 1505200, 1509136, 1509272, 1509273, 1509311, 1510151, 1510254, 1510255, 1511190, 1511215, 1511216, 1511217, 1511218, 1511219, 1511220, 1511221, 1511222, 1601236, 1601237, 1601238, 1601239, 1601240, 1601241, 1601242, 1601243, 1601244, 1601245, 1601366, 1601367, 1601368, 1601369, 1601370, 1601371, 1601372, 1601373, 1601374, 1601375, 1601376, 1601377, 1601378, 1601379, 1601380, 1601381, 1601382, 1601383, 1602226, 1603294, 1605121, 1605122, 1605123, 1605124, 1605125, 1605126, 1605154, 1605155, 1605156, 1605157, 1605158, 1605159, 1605160, 1605161, 1605162, 1605163, 1605164, 1605165, 1605166, 1605167, 1605168, 1605169, 1605170, 1605171, 1606166, 1606167, 1606168, 1606189, 1606190, 1607203, 1607204, 1607205, 1607206, 1607207, 1607208, 1607209, 1607210, 1607211, 1608230, 1608231, 1608232, 1608233, 1608234, 1608235, 1608236, 1608237, 1608238, 1608239, 1609169, 1609170, 1609171, 1609172, 1609173, 1609174, 1609175, 1609176, 1609177, 1609178, 1609179, 1609180, 1609181, 1609182, 1609183, 1609184, 1610240, 1610241, 1610242, 1610243, 1610244, 1610245, 1610246, 1610247, 1610248, 1610249, 1610259, 1610260, 1610261, 1610262, 1610263, 1610264, 1610265, 1610266, 1610267, 1610268, 1610269, 1610270, 1610271, 1610272, 1610273, 1610274, 1610275, 1610276, 1610277, 1610278, 1611232, 1611233, 1611234, 1611235, 1611236, 1611237, 1611238, 1611239, 1611240, 1611298, 1611299, 1611300, 1701211, 1701212, 1701213, 1701214, 1701215, 1701216, 1701217, 1701218, 1701219, 1701220, 1701221, 1701224, 1701225, 1701226, 1701227, 1701228, 1701229, 1701230, 1703061, 1703062, 1703082, 1703083, 1703084, 1703085, 1703086, 1703087, 1703088, 1703150, 1703151, 1703217, 1703218, 1703219, 1703220, 1703221, 1703222, 1703223, 1703224, 1703225, 1703226, 1703229, 1703230, 1703231, 1703232, 1703233, 1703234, 1703235, 1703236, 1703237, 1703238, 1703239, 1703240, 1703241, 1704100, 1704101, 1704102, 1704103, 1704104, 1704105, 1704106, 1704107, 1704108, 1704109, 1704122, 1704123, 1704124, 1704125, 1704126, 1704127, 1704128, 1704129, 1704130, 1704131, 1704132, 1705111, 1709166, 1709167, 1709168, 1709169, 1709170, 1709171, 1710098, 1710099, 1710100, 1710101, 1710102, 1710103, 1710104, 1710105, 1710106, 1710107, 1710109, 1710110, 1710111, 1710112, 1710113, 1710114, 1710115, 1710116, 1710117, 1710118, 1710121, 1710122, 1710123, 1710124, 1710125, 1710126, 1710127, 1710128, 1710129, 1710130, 1710139, 1710140, 1710141, 1710142, 1710143, 1710144, 1710145, 1710146, 1711120, 1711121, 1711122, 1711123, 1711124, 1711125, 1711126, 1711127, 1711128, 1711129, 1711130, 1711131, 1711132, 1711145, 1711146, 1711147, 1711148, 1711149, 1711150, 1711151, 1711152, 1711153, 1711154, 1711155, 1711156, 1711157, 1712113, 1712114, 1712115, 1712116, 1712117, 1712118, 1712119, 1712120, 1712121, 1712122, 1712123, 1712124, 1712125, 1712126, 1712127, 1712128, 1712129, 1712130, 1712131, 1712132, 1712133, 1712134, 1801124, 1801125, 1801126, 1801127, 1801128, 1801129, 1801130, 1801131, 1801132, 1801133, 1802054, 1802055, 1802056, 1802057, 1802058, 1802059, 1802060, 1802061, 1802062, 1802063, 1802069, 1802070, 1802071, 1802072, 1802073, 1802074, 1802075, 1802076, 1802077, 1802078, 1802113, 1802114, 1802115, 1802116, 1802117, 1802118, 1802119, 1802120, 1802121, 1802122, 1802125, 1802126, 1802127, 1802128, 1802129, 1802130, 1802131, 1802132, 1802133, 1802134, 1802176, 1802177, 1802178, 1802179, 1802180, 1802181, 1802182, 1802183, 1802184, 1802185, 1802233, 1802234, 1802235, 1802236, 1802237, 1802238, 1802239, 1802240, 1802241, 1802242, 1802244, 1802245, 1802246, 1802247, 1802248, 1802249, 1802250, 1802251, 1802252, 1802253, 1802254, 1804066, 1804067, 1809028, 1809029, 1809030, 1809031, 1809032, 1809033, 1809034, 1809035, 1809036, 1809037, 1809046, 1809047, 1809048, 1809049, 1809050, 1809056, 1809057, 1809058, 1809059, 1809060, 1809061, 1809062, 1809063, 1809064, 1809065, 1809069, 1809070, 1809071, 1809072, 1809073, 1809074, 1809075, 1809076, 1809077, 1809078, 1810001, 1810002, 1810003, 1810004, 1810005, 1810006, 1810007, 1810008, 1810009, 1810010, 1810011, 1810012, 1810013, 1811054, 1811055, 1811056, 1811057, 1811058, 1811059, 1811060, 1811061, 1811062, 1811063, 1811069, 1811070, 1811071, 1811072, 1811073, 1811074, 1811075, 1811076, 1811077, 1811078, 1812015, 1812016, 1812017, 1812018, 1812019, 1812020, 1812021, 1812022, 1812061, 1812062, 1812063, 1812064, 1812065, 1812066, 1812067, 1812068, 1812069, 1901001, 1901002, 1901003, 1901004, 1901005, 1901006, 1901007, 1901008, 1901009, 1901010, 1901012, 1901013, 1901014, 1901015, 1901016, 1901017, 1901018, 1901019, 1901020, 1901021, 1902081, 1902082, 1902083, 1902084, 1902085, 1902086, 1902087, 1902088, 1902089, 1902090, 1902101, 1902102, 1902103, 1902104, 1902105, 1902106, 1902107, 1902108, 1902109, 1902110, 1902115, 1902116, 1902117, 1902118, 1902119, 1902120, 1902121, 1902122, 1902123, 1902124, 1902145, 1902146, 1902147, 1902148, 1902149, 1902150, 1902151, 1903116, 1903117, 1903118, 1903119, 1903120, 1903121, 1903122, 1903123, 1903124, 1903125, 1903126, 1903127, 1903128, 1903129, 1903130, 1903131, 1903132, 1903133, 1903134, 1903135, 1903136, 1903137, 1903138, 1903139, 1903140, 1903141, 1903142, 1903143, 1903144, 1903145, 1904001, 1904002, 1904003, 1904004, 1904005, 1904006, 1904007, 1904008, 1904009, 1904010, 1904020, 1904021, 1904022, 1904023, 1904024, 1904025, 1904026, 1904027, 1904028, 1904029, 1904030, 1904031, 1904032, 1904033, 1904034, 1904035, 1904036, 1904037, 1904038, 1904039, 1904101, 1904102, 1904103, 1904104, 1904105, 1904106, 1904107, 1904108, 1904109, 1904110, 1904121, 1904122, 1904123, 1904124, 1904125, 1904126, 1904127, 1904128, 1904129, 1904130, 1905001, 1905002, 1905003, 1905004, 1905005, 1905006, 1905007, 1905008, 1905009, 1905010, 1905011, 1905012, 1905013, 1905014, 1905015, 1905016, 1905017, 1905018, 1905019, 1905020, 1905021, 1905022, 1905023, 1905024, 1905025, 1905026, 1905027, 1905028, 1905029, 1905030, 1905031, 1905032, 1905033, 1905034, 1905035, 1905036, 1905037.1905038, 1905039, 1905040, 1905041, 1905042, 1905043, 1905044, 1905045, 1905046, 1905047, 1905048, 1905049, 1905050, 1905051, 1905052, 1905053, 1905104, 1905105, 1905106, 1905107, 1905108, 1905109, 1905110, 1905111, 1905112, 1905113, 1905114, 1905115, 1905116, 1905117, 1905118, 1905119, 1905120, 1905121, 1905122, 1905123, 1905124, 1905125, 1905126, 1905127, 1905128, 1905129, 1905130, 1905131, 1905132, 1905133, 1905134, 1905135, 1905136, 1905137, 1905138, 1905139, 1905140, 1905141, 1905142, 1905143, 1905144, 1905145, 1905146, 1905147, 1905148, 1905149, 1905150, 1905151, 1905152, 1905153, 1905154, 1905155, 1905156, 1905157, 1905218, 1906011, 1906012, 1906013, 1906014, 1906015, 1908001, 1908002, 1908003, 1908004, 1908005, 1912128, 1912129, 1912130, 1912131, 1912132, 1912133, 1912134, 1912135, 1912190, 1912191, 1912192, 1912193, 1912194, 1912195, 1912196, 1912197, 1912198, 1912199, 1912210, 1912211, 1912212, 1912213, 1912214, 1912215, 1912216, 1912217, 1912218, 1912219, 1912230, 1912231, 1912232, 1912233, 1912234, 1912235, 1912236, 1912237, 1912238, 1912239, 1912240, 1912241, 1912242, 1912243, 1912244, 1912245, 1912246, 1912247, 1912248, 1912249, 1912260, 1912261, 1912262, 1912263, 1912264, 1912265, 1912266, 1912267, 1912268, 1912269, 1912280, 1912281, 1912282, 1912283, 1912284, 1912285, 1912286, 1912287, 1912288, 1912289, 1912302, 1912303, 1912304, 1912305, 1912306, 1912307, 1912308, 1912309, 2002051, 2002052, 2002053, 2002054, 2002055, 2002056, 2002057, 2002058, 2002059, 2002060, 2002061, 2002062, 2002063, 2002064, 2003005, 2003052, 2003053, 2003054, 2003055, 2003056, 2003142, 2003178, 2003179, 2003180, 2003181, 2003182, 2003183, 2003184, 2003185, 2003186, 2003187, 2003188, 2003189, 2003190, 2003191, 2003192, 2003193, 2003194, 2003195, 2003196, 2003197, 2003198, 2003199, 2003200, 2003201, 2003202, 2003203, 2003204, 2003205, 2003206, 2003207, 2003208, 2003209, 2003210, 2003211, 2003212, 2003213, 2003214, 2003215, 2003216, 2003217, 2003218, 2003219, 2003220, 2003221, 2003222, 2003223, 2003224, 2003225, 2003226, 2003227, 2003228, 2003229, 2003230, 2003231, 2003232, 2003233, 2003301, 2003302, 2004483, 2004484, 2004485, 2004486, 2004487, 2004488, 2006250, 2007410, 2007411, 2007412, 2007413, 2007414, 2007415, 2007416, 2007417, 2007418, 2007419, 2007430, 2007431, 2007432, 2007433, 2007434, 2007435, 2007436, 2007437, 2007438, 2007439, 2007440, 2007441, 2007442, 2007443, 2007444, 2007445, 2007446, 2007447, 2007448, 2007449, 2007450, 2007451, 2007452, 2007453, 2007454, 2007455, 2007456, 2007457, 2007458, 2007459, 2007460, 2007461, 2007462, 2007463, 2007464, 2007465, 2007466, 2007467, 2007468, 2007469, 2007570, 2007571, 2007572, 2007573, 2007574, 2007575, 2007576, 2007577, 2007578, 2007579, 2008045, 2008046, 2008047, 2008048, 2008049, 2008050, 2008051, 2008052, 2009004, 2009005, 2009006, 2009007, 2009008, 2009009, 2009010, 2009011, 2009012, 2009013, 2009021, 2009022, 2009023, 2009024, 2009025, 2009026, 2009027, 2009028, 2009029, 2009030, 2009034, 2009035, 2009036, 2009037, 2009038, 2009039, 2009040, 2009041, 2009042, 2009043, 2009044, 2009045, 2009046, 2009047, 2009048, 2009049, 2009050, 2009051, 2009052, 2009053, 2009054, 2009055, 2009056, 2009057, 2009058, 2010001, 2010002, 2010006, 2010007, 2010008, 2010009, 2010010, 2010011, 2010012, 2010013, 2010014, 2010015, 2010016, 2010017, 2010018, 2010019, 2010020, 2010021, 2010022, 2010023, 2010024, 2010025, 2010026, 2010027, 2010028, 2010029, 2010030, 2010031, 2010032, 2010033, 2010034, 2010035, 2010036, 2010037, 2010038, 2010039, 2010040, 2010041, 2010042, 2010043, 2010044, 2010045, 2010046, 2010047, 2010048, 2010049, 2010050, 2010051, 2010052, 2010053, 2010054, 2010055, 2010056, 2010057, 2010058, 2010059, 2011001, 2011038, 2011039, 2011040, 2011041, 2011042, 2011043, 2011044, 2011045, 2011046, 2011047, 2011048, 2011049, 2011050, 2011051, 2011052, 2011053, 2011054, 2011055, 2011056, 2012006, 2012007, 2012008, 2012009, 2012010, 2012011, 2012012, 2012013, 2012014, 2012015, 2012016, 2012017, 2012018, 2012019, 2012020, 2012021, 2012022, 2012023, 2012079, 2012080, 2012081, 2012082, 2012083, 2012084, 2012085, 2012086, 2012087, 2012088, 2012089, 2012090, 2012091, 2012092, 2012093, 2012094, 2012095, 2012096, 2012097, 2012098, 2012099, 2012102, 2012103, 2012104, 2012105, 2012106, 2101116, 2101117, 2101118, 2101119, 2101120, 2101121, 2101122, 2101123, 2101124, 2101125, 2101139, 2101140, 2101141, 2101142, 2101143, 2101144, 2101145, 2101146, 2101147, 2101148, 2102001, 2102002, 2102003, 2102004, 2102005, 2102006, 2102007, 2102008, 2102009, 2102010, 2102021, 2102022, 2102023, 2102024, 2102025, 2102026, 2102027, 2102028, 2102029, 2102030, 2102031, 2102032, 2102033, 2102034, 2102035, 2102036, 2102037, 2102038, 2102039, 2102040, 2102041, 2102042, 2102043, 2102044, 2102045, 2104061, 2105021, 2105022, 2105023, 2105038, 2105052, 2106013, 2106014, 2106015, 2106016, 2106017, 2106073, 2106074, 2106075, 2106076, 2106077, 2106078, 2106094, 2106095, 2106096, 2106097, 2106098, 2106165, 2106167, 2106168, 2106169, 2106170, 2106171, 2106172, 2106173, 2106174, 2106175, 2106176, 2106177, 2107076, 2107077, 2107078, 2107079, 2107080, 2107136, 2107137, 2107138, 2107139, 2108011, 2108012, 2108023, 2108024, 2108025, 2108026, 2108027, 2108028, 2108029, 2108030, 2108031, 2108032, 2108072, 2108073, 2108074, 2109024, 2109025, 2109026, 2109027, 2109028, 2109029, 2109030, 2109031, 2109032, 2109033, 2109034, 2109035, 2109036, 2109037, 2109135, 2109136, 2109137, 2109138, 2109139, 2109140, 2109141, 2109142, 2109143, 2109185, 2110054, 2110055, 2110056, 2110057, 2110058, 2110059, 2110060, 2110061, 2110062, 2110063, 2110064, 2110065, 2110066, 2110067, 2110068, 2110069, 2110070, 2110071, 2110072, 2110073, 2110074, 2110075, 2110076, 2110077, 2110078, 2110079, 2110080, 2110081, 2110082, 2110083, 2110084, 2110085, 2110086, 2110087, 2110088, 2110089, 2110090, 2110091, 2110092, 2110093, 2110094, 2110095, 2110096, 2110097, 2110098, 2110099, 2110101, 2110102, 2110103, 2110104, 2110105, 2110106, 2110107, 2110108, 2110109, 2110110, 2110111, 2110112, 2110113, 2110114, 2110115, 2110116, 2110117, 2110118, 2110119, 2110120, 2110121, 2110122, 2110123, 2110124, 2111001, 2111002, 2111003, 2111004, 2111005, 2111006, 2111007, 2111008, 2111009, 2111010, 2111011, 2111012, 2111013, 2111014, 2111015, 2111016, 2111017, 2111018, 2111019, 2111020, 2111021, 2111022, 2111023, 2111024, 2111025, 2111026, 2111027, 2111028, 2111029, 2111030, 2111031, 2111032, 2111033, 2111034, 2111035, 2111113, 2112050, 2112051, 2112052, 2112053, 2112054, 2112055, 2112056, 2112057, 2112058, 2112059, 2112060, 2112061, 2112062, 2201001, 2201002, 2201003, 2201014, 2201015, 2201016, 2201017, 2201058, 2201059, 2201060, 2201061, 2201062, 2201063, 2201064, 2201065, 2201066, 2201067, 2201078, 2201079, 2201080, 2201081, 2201082, 2201083, 2201084, 2201085, 2201086, 2201087, 2201203, 2201204, 2201205, 2201206, 2201207, 2201208, 2201209, 2201210, 2201211, 2201212, 2201213, 2201214, 2201215, 2201216, 2201217, 2201218, 2201219, 2201220, 2201221, 2201222, 2201223, 2201224, 2201225, 2201226, 2202016, 2202017, 2202018, 2202019, 2202020, 2202021, 2202022, 2202023, 2202024, 2202025, 2202026, 2202027, 2202028, 2202029, 2202030, 2202031, 2202032, 2202033, 2202034, 2202035, 2202110, 2202111, 2202112, 2202113, 2202114, 2202115, 2202116, 2202117, 2202118, 2202119, 2202120, 2203005, 2203006, 2203007, 2203008, 2203009, 2203010, 2203011, 2203012, 2203013, 2203014, 2203015, 2203016, 2203017, 2203018, 2203019, 2203020, 2203072, 2203076, 2203077, 2203078, 2203079, 2203080, 2203081, 2203082, 2203083, 2203084, 2203085, 2203204, 2203205, 2203206, 2203207, 2204049, 2204050, 2204071, 2204072, 2205061, 2205062, 2205063, 2205064, 2205065, 2205077, 2205078, 2206048, 2206049, 2206050, 2208135, 2208136, 2208137, 2208144, 2208145, 2208146, 2208147, 2208148, 2208149, 2208150, 2208151, 2208152, 2208153, 2208154, 2208195, 2208196, 2208197, 2208198, 2208199, 2208200, 2208201, 2208202, 2208203, 2208204, 2208205, 2208206, 2208207, 2208208, 2208209, 2208210, 2208211, 2209087, 2209088, 2209089, 2209090, 2209091, 2209092, 2209093, 2209094, 2209095, 2209096, 2209106, 2209107, 2209108, 2209109, 2209110, 2209111, 2209112, 2209113, 2209114, 2209115, 2209116, 2209117, 2209118, 2209119, 2209120, 2209121, 2209122, 2209123, 2209124, 2209125, 2209126, 2209127, 2209128, 2209129, 2209130, 2209131, 2209132, 2209133, 2209134, 2209135, 2209136, 2209137, 2209138, 2209139, 2209140, 2209141, 2209142, 2209143, 2209144, 2209145, 2210059, 2210060, 2210061, 2210062, 2210063, 2210064, 2210065, 2210066, 2210067, 2210068, 2210069, 2210070, 2210071, 2210072, 2210073, 2210074, 2210075, 2210076, 2210077, 2210078, 2210152, 2210153, 2210154, 2210155, 2210156, 2210157, 2210158, 2210159, 2210160, 2210161, 2211037, 2211038, 2211039, 2211040, 2211041, 2211042, 2211043, 2211044, 2211045, 2211046, 2211047, 2211048, 2211049, 2211050, 2211051, 2211052, 2211053, 2211054, 2211055, 2211056, 2211057, 2211058, 2211059, 2211060, 2211061, 2211062, 2211063, 2211064, 2211065, 2211066, 2211067, 2211068, 2211069, 2211070, 2211092, 2211104, 2211105, 2211106, 2211107, 2211119, 2211120, 2211121, 2211122, 2211123, 2211124, 2211125, 2211126, 2211127, 2211128, 2211137, 2211138, 2211139, 2211140, 2211141, 2211142, 2211143, 2211144, 2211145, 2211146, 2211147, 2211148, 2211149, 2211150, 2211151, 2211152, 2211153, 2211154, 2211155, 2211156, 2211157, 2211158, 2211159, 2211160, 2211161, 2211162, 2211163, 2211164, 2211165, 2211166, 2211167, 2211168, 2211169, 2301036, 2301037, 2301038, 2301039, 2301040, 2301041, 2301042, 2301043, 2301044, 2301045, 2301046, 2301047, 2301056, 2301061, 2301109, 2301110, 2301111, 2301117, 2301118, 2301119, 2301120, 2301121, 2301122, 2301123, 2301124, 2301125, 2301126, 2301127, 2301128, 2301129, 2301130, 2301131, 2301132, 2301133, 2301134, 2301135, 2301136, 2301137, 2301138, 2301139, 2301140, 2301141, 2301142, 2302031, 2302032, 2302033, 2302044, 2302045, 2302046, 2302047, 2302048, 2302049, 2302050, 2302051, 2302052, 2302053, 2302054, 2302055, 2302056, 2302057, 2302058, 2302059, 2302060, 2302061, 2302062, 2302063, 2302064, 2302065, 2302066, 2302067, 2302068, 2302069, 2302070, 2302071, 2302072, 2302073, 2302074, 2302075, 2302076, 2302077, 2302078, 2302079, 2302080, 2302081, 2302082, 2302083, 2302084, 2302085, 2302086, 2302087, 2302088, 2302089, 2302090, 2302091, 2302092, 2302093, 2302094, 2302095, 2302096, 2303001, 2303002, 2303003, 2303004, 2303005, 2303006, 2303007, 2303008, 2303009, 2303010, 2303011, 2303012, 2303013, 2303014, 2303015, 2303016, 2303017, 2303018, 2303019, 2303020, 2303021, 2303022, 2303023, 2303024, 2303025, 2303026, 2303027, 2303028, 2303029, 2303030, 2303031, 2303032, 2303033, 2303034, 2303035, 2303036, 2303037, 2304031, 2304032, 2304033, 2304034, 2304035, 2304036, 2304037, 2304038, 2304039, 2304040, 2304041, 2304042, 2304043, 2304044, 2304045, 2307001, 2307002, 2307007, 2307008, 2307009, 2307010, 2307011, 2307012, 2307013, 2307014, 2307015, 2307016, 2307017, 2307018, 2307019, 2308006, 2308007, 2308008, 2308009, 2308010, 3540877, 19012128, 21114024, 1605167A, 2106048, 2106049, 2106050, 2106051, 2106052, 2106053, 2106054, 2106055, 2106056, 2106057, 2106058, 2106059, G;10328, GC00001, GC00002, GC00003, GC00004, GC00005, GC00006, GC00007, GC00008, GC00009, GC00010, GC00011, GC00012, GC00013, GC00014, GC00015, GC00016, GC00017, GC00018, GC00019, GC00020, GC00021, GC00022, GC00023, GC00024, GC00025, GC00026, GC00027, GC00028, GC00029, GC00030, GC00031, GC00032, GC00033, GC00034, GC00035, GC00036, GC00037, GC00038, GC00039, GC00040, GC00041, GC00042, GC00043, GC00044, GC00045, GC00046, GC00047, GC00048, GC00049, GC00050, GC00051, GC00052, GC00053, GC00054, GC00055, GC00056, GC00057, GC00058, GC00059, GC00060, GC00061, GC00062, GC00063, GC00064, GC00065, GC00066, GC00067, GC00068, GC00069, GC00070, GC00071, GC00072, GC00073, GC00074, GC00076, GC00077, GC00078, GC00079, GC00080, GC00081, GC00082, GC00083, GC00085, GC00086, GC00087, GC00088, GC00089, GC00090, GC00091, GC00092, GC00093, GC00094, GC00095.GC00096, GC00097, GC00098, GC00099, GC00100, GC00101, GC00102, GC00103, GC00104, GC00105, GC00106, GC00107, GC00108, GC00109, GC00110, GC00111, GC00112, GC00113, GC00114, GC00115, GC00116, GC00117, GC00118, GC00119, GC00120, GC00121, GC00122, GC00123, GC00124, GC00125, GC00126, GC00127, GC00128, GC00129, GC00130, GC00131, GC00132, GC00133, GC00134, GC00135, GC00136, GC00137, GC00138, GC00139, GC00140, GC00141, GC00142, GC00143, GC00144, GC00145, GC00146, GC00147, GC00148, GC00149, GC00150, GC00151, GC00152, GC00153, GC00154, GC00155, GC00156, GC00157, GC00158, GC00159, GC00160, GC00161, GC00162, GC00163, GC00164, GC00165, GC00166, GC00167, GC00168, GC00169, GC00170, GC00171, GC00172, GC00173, GC00174, GC00175, GC00176, GC00177, GC00178, GC00179, GC00180, GC00181, GC00182, GC00183, GC00184, GC00185, GC00186, GC00187, GC00188, GC00189, GC00190, GC00191, GC00192, GC00193, GC00194, GC00195, GC00196, GC00197, GC00198, GC00199, GC00200, GC00201, GC00202, GC00203, GC00204, GC00205, GC00206, GC00207, GC00208, GC00209, GC00210, GC00211, GC00212, GC00213, GC00214, GC00215, GC00216, GC00217, GC00218, GC00219, GC00220, GC00221, GC00222, GC00223, GC00224, GC00225, GC00226, GC00227, GC00228, GC00229, GC00230, GC00231, GC00232, GC00233, GC00234, GC00235, GC00236, GC00237, GC00238, GC00239, GC00240, GC00241, GC00242, GC00243, GC00244, GC00245, GC00246, GC00247, GC00248, GC00249, GC00250, GL10006, GL10007, GL10008, GL10009, GL10010, GL10011, GL10012, GL10013, GL10014, GL10016, GL10017, GL10018, GL10019, GL10021, GL10022, GL10023, GL10024, GL10025, GL10026, GL10027, GL10028, GL10035, GL10036, GL10037, GL10038, GL10039, GL10040, GL10041, GL10042, GL10043, GL10044, GL10045, GL10046, GL10047, GL10048, GL10049, GL10050, GL10051, GL10052, GL10053, GL10054, GL10055, GL10056, GL10057, GL10058, GL10059, GL10060, GL10061, GL10062, GL10063, GL10064, GL10065, GL10066, GL10067, GL10068, GL10069, GL10070, GL10071, GL10072, GL10073, GL10074, GL10075, GL10076, GL10077, GL10078, GL10080, GL10081, GL10082, GL10083, GL10084, GL10085, GL10086, GL10087, GL10088, GL10089, GL10090, GL10091, GL10092, GL10093, GL10094, GL10095, GL10096, GL10097, GL10098, GL10099, GL10100, GL10101, GL10102, GL10103, GL10104, GL10105, GL10106, GL10107, GL10108, GL10109, GL10110, GL10111, GL10112, GL10113, GL10114, GL10115, GL10116, GL10117, GL10118, GL10119, GL10120, GL10121, GL10122, GL10123, GL10124, GL10125, GL10126, GL10127, GL10128, GL10129, GL10130, GL10131, GL10132, GL10133, GL10134, GL10135, GL10136, GL10137, GL10138, GL10139, GL10140, GL10141, GL10142, GL10143, GL10144, GL10145, GL10146, GL10147, GL10148, GL10149, GL10150, GL10151, GL10152, GL10153, GL10154, GL10155, GL10156, GL10157, GL10158, GL10159, GL10160, GL10161, GL10162, GL10163, GL10164, GL10165, GL10166, GL10167, GL10168, GL10169, GL10170, GL10171, GL10172, GL10173, GL10174, GL10175, GL10176, GL10177, GL10178, GL10179, GL10180, GL10181, GL10182, GL10183, GL10184, GL10185, GL10186, GL10187, GL10188, GL10189, GL10190, GL10191, GL10192, GL10193, GL10194, GL10195, GL10196, GL10197, GL10198, GL10199, GL10200, GL10201, GL10202, GL10203, GL10204, GL10205, GL10206, GL10207, GL10208, GL10209, GL10210, GL10211, GL10212, GL10213, GL10214, GL10215, GL10216, GL10217, GL10218, GL10219, GL10220, GL10221, GL10222, GL10223, GL10224, GL10225, GL10226, GL10227, GL10228, GL10229, GL10230, GL10231, GL10232, GL10233, GL10234, GL10235, GL10236, GL10237, GL10238, GL10239, GL10240, GL10241, GL10242, GL10243, GL10244, GL10245, GL10246, GL10247, GL10248, GL10249, GL10250, GL10251, GL10252, GL10253, GL10254, GL10255, GL10256, GL10257, GL10258, GL10259, GL10260, GL10261, GL10262, GL10263, GL10264, GL10265, GL10266, GL10267, GL10268, GL10269, GL10270, GL10271, GL10272, GL10273, GL10274, GL10275, GL10276, GL10277, GL10278, GL10279, GL10280, GL10281, GL10282, GL10283, GL10284, GL10285, GL10286, GL10287, GL10288, GL10289, GL10290, GL10291, GL10292, GL10293, GL10294, GL10295, GL10296, GL10297, GL10298, GL10299, GL10300, GL10301, GL10302, GL10303, GL10304, GL10305, GL10306, GL10307, GL10308, GL10309, GL10310, GL10311, GL10312, GL10313, GL10314, GL10315, GL10316, GL10317, GL10318, GL10319, GL10320, GL10321, GL10322, GL10323, GL10324, GL10325, GL10326, GL10327, GL10328, GL10329, GL10330, GL10331, GL10332, GL10333, GL10334, GL10335, GL10336, GL10337, GL10338, GL10339, GL10340, GL10341, GL10342, GL10343, GL10344, GL10345, GL10346, GL10347, GL10348, GL10349, GL10350, GL10351, GL10352, GL10353, GL10354, GL10355, GL10356, GL10357, GL10358, GL10359, GL10360, GL10361, GL10362, GL10363, GL10364, GL10365, GL10366, GL10367, GL10368, GL10369, GL10370, GL10371, GL10372, GL10373, GL10374, GL10375, GL10376, GL10377, GL10378, GL10379, GL10380, GL10381, GL10382, GL10383, GL10384, GL10385, GL10386, GL10387, GL10388, GL10389, GL10390, GL10391, GL10392, GL10393, GL10394, GL10395, GL10396, GL10397, GL10398, GL10399, GL10400, GL10401, GL10402, GL10403, GL10404, GL10405, GL10406, GL10407, GL10408, GL10409, GL10410, GL10411, GL10412, GL10413, GL10414, GL10415, GL10416, GL10417, GL10418, GL10421, GL10422, GL10423, GL10424, GL10425, GL10426, GL10427, GL10428, GL10429, GL10430, GL10431, GL10432, GL10433, GL10434, GL10435, GL10436, GL10437, GL10438, GL10439, GL10440, GL10441, GL10442, GL10443, GL10444, GL10445, GL10446, GL10447, GL10448, GL10449, GL10450, GL10451, GL10452, GL10453, GL10454, GL10455, GL10456, GL10457, GL10458, GL10459, GL10460, GL10461, GL10462, GL10463, GL10464, GL10465, GL10466, GL10467, GL10468, GL10469, GL10470, GL10471, GL10473, GL10474, GL10475, GL10476, GL10477, GL10478, GL10479, GL10480, GL10481, GL10482, GL10483, GL10484, GL10485, GL10486, GL10487, GL10488, GL10489, GL10491, GL10492, GL10493, GL10583, GL19479, RF20267, RF70003, RF70004, RF70005, RF70006, RF70007, RF70009, RF70010, RF70011, RF70012, RF70013, RF70014, RF70015, RF70016, RF70018, RF70019, RF70020, RF70021, RF70023, RF70024, RF70025, RF70026, RF70027, RF70028, RF70030, RF70031, RF70033, RF70034, RF70035, RF70036, RF70037, RF70038, RF70039, RF70040, RF70041, RF70042, RF70043, RF70044, RF70045, RF70046, RF70047, RF70048, RF70049, RF70050, RF70051, RF70052, RF70053, RF70054, RF70055, RF70056, RF70057, RF70058, RF70059, RF70060, RF70061, RF70062, RF70063, RF70064, RF70065, RF70066, RF70067, RF70068, RF70069, RF70070, RF70071, RF70072, RF70073, RF70074, RF70075, RF70076, RF70077, RF70078, RF70079, RF70080, RF70081, RF70082, RF70084, RF70085, RF70086, RF70087, RF70088, RF70089, RF70090, RF70091, RF70092, RF70093, RF70094, RF70095, RF70096, RF70097, RF70098, RF70099, RF70100, RF70101, RF70102, RF70103, RF70104, RF70105, RF70106, RF70107, RF70108, RF70109, RF70110, RF70111, RF70112, RF70113, RF70114, RF70116, RF70117, RF70118, RF70119, RF70121, RF70122, RF70123, RF70124, RF70128, RF70129, RF70130, RF70131, RF70132, RF70133, RF70134, RF70135, RF70136, RF70137, RF70138, RF70139, RF70140, RF70141, RF70142, RF70143, RF70144, RF70145, RF70146, RF70147, RF70148, RF70149, RF70150, RF70151, RF70152, RF70153, RF70154, RF70155, RF70156, RF70157, RF70158, RF70159, RF70160, RF70161, RF70162, RF70164, RF70165, RF70166, RF70167, RF70168, RF70169, RF70171, RF70173, RF70174, RF70175, RF70176, RF70177, RF70179, RF70180, RF70181, RF70182, RF70183, RF70184, RF70185, RF70186, RF70187, RF70188, RF70189, RF70190, RF70191, RF70192, RF70193, RF70194, RF70195, RF70196, RF70197, RF70199, RF70200, RF70203, RF70204, RF70206, RF70208, RF70209, RF70210, RF70211, RF70212, RF70213, RF70214, RF70215, RF70216, RF70217, RF70218, RF70219, RF70220, RF70221, RF70222, RF70223, RF70224, RF70225, RF70226, RF70227, RF70228, RF70229, RF70230, RF70231, RF70232, RF70233, RF70234, RF70235, RF70236, RF70237, RF70238, RF70239, RF70240, RF70241, RF70242, RF70243, RF70244, RF70245, RF70247, RF70248, RF70249, RF70250, RF70251, RF70252, RF70253, RF70254, RF70256, RF70257, RF70258, RF70259, RF70260, RF70261, RF70264, RF70267, RF70268, RF70269, RF70270, RF70271, RF70272, RF70273, RF70274, RF70275, RF70276, RF70278, RF70279, RF70280, RF70281, RF70282, RF70283, RF70284, RF70287, RF70289, RF70292, RF70293, RF70296, RF70298
Recalling Firm/
Manufacturer
Smiths Medical ASD Inc.
6000 Nathan Ln N
Minneapolis MN 55442-1690
For Additional Information Contact
763-383-3072
Manufacturer Reason
for Recall
Smiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function.
FDA Determined
Cause 2
Other
ActionSmiths Medical issued an URGENT MEDIC AL DEVICE CORRECTION notice to its consignees on05/31/2024 via letter. The notice explained the issue, potential risk, and requested the following: "Perform an inspection to determine if your devices are affected, per the instructions below: a. First perform a visual inspection to determine if the outlet connector is disconnected. b. Second evaluate the connector physically to determine if the outlet connector is loose or moves when placing a patient circuit on the connector or when removing it. c. If outlet connector remains tight after physical inspection, you can continue use of the device with heightened awareness and following all pre-use checks as per the user manual. d. If the outlet connector moves or feels loose, the device must be removed from use and repaired by Smiths Medical. Report the event to Global Complaint Management at globalcomplaints@icumed.com. Every use thereafter of every device, pre-use checks must be completed as described in the user manual and extra caution must be taken in inspecting the outlet connector prior to use and placing the patient circuit on the connector and during removal. Share this recall notification with all potential users of the devices to ensure they are aware of this recall and proposed mitigations." Those distributing the affected devices were directed to notify those to whom the devices were distributed. Device Correction Inquiries: contact - https://icumed.custhelp.com/app/marketaction
Quantity in Commerce3182 units
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = BTL
-
-