| | Class 2 Device Recall Access Intact PTH |  |
| Date Initiated by Firm | November 08, 2024 |
| Date Posted | December 11, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0673-2025 |
| Recall Event ID |
95738 |
| 510(K)Number | K232791 |
| Product Classification |
Radioimmunoassay, parathyroid hormone - Product Code CEW
|
| Product | Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma. |
| Code Information |
UDI-DI 15099590201937
Lots 338921
339071
372223
439138
439205
439206
439207
439650
439801
439889
440026
440090
472012
472092
472136
472154
|
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 250 S Kraemer Blvd Brea CA 92821-6232
|
| For Additional Information Contact | Courtney Walton 714-961-5321 |
Manufacturer Reason for Recall | The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition. |
FDA Determined Cause 2 | Process design |
| Action | On November 8, 2024, the firm notified affected customers via URGENT MEDICAL DEVICE RECALL letters that the affected assay may produce falsely decreased results if EDTA and lithium heparin plasma samples are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the test.
Access Intact PTH assay Routine Mode and Intraoperative Mode users can select any of the following options:
. Use serum as the sample type.
. Perform testing on an alternate methodology, such as the Access 2 or UniCel DxI 600 or 800 Immunoassay Analyzers.
. Continue to use EDTA and/or lithium heparin plasma sample types if the tests are completed within the timeframes listed below upon sample presentation to the DxI 9000 Access Immunoassay Analyzer (see page 3 for instructions):
o Access Intact PTH assay Routine Mode: 25 minutes.
o Access Intact PTH assay Intraoperative Mode: 20 minutes.
. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to perform a retrospective review of the patient results generated from Access Intact PTH Routine Mode plasma samples on the DxI 9000 to determine if retesting is necessary. |
| Quantity in Commerce | 31 US; 167 OUS |
| Distribution | Foreign distribution to Albania, Australia, Austria, Bahrain, Belgium, Brazil, C¿te
d'Ivoire, Croatia, Czechia, Egypt, France, Germany, Ghana,
Hungary, India, Ireland, Israel, Italy, Kazakhstan, Kuwait,
Malaysia, Netherlands, New Zealand, Poland, Portugal,
Slovakia, South Africa, S. Korea, Spain, Switzerland, Taiwan,
United Kingdom of Great Britain and Northern Ireland. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = CEW
|
|
|
|