• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Medline Convenience kits

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Medline Convenience kitssee related information
Date Initiated by FirmNovember 06, 2024
Date PostedDecember 20, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0721-2025
Recall Event ID 95806
Product Classification laparoscopy kit - Product Code FDE
ProductMedline brand, medical procedure convenience kits, labeled as: 1) MVFF/FIBU #54-RF , SKU DYNJ21927R; 2) DIEP FLAP, SKU DYNJ910120A
Code Information UDI-DI: DYNJ21927R, 10195327104733(each), 40195327104734(case); Lot Number: 23IBF547 23GBT501 23FBU360 23EBU929 23EBL624 23EBA114 23CBN730 23ABF393 22JBK203 22IBN858 22HBO301 22GBC794 22CBI774; UDI-DI: DYNJ910120A, 10195327689872 (each), 40195327689873 (case), Lot Numbers: 24HMK254
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactKarin Johnson
866-359-1704
Manufacturer Reason
for Recall
Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods
FDA Determined
Cause 2
Nonconforming Material/Component
ActionAn IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 11/7/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.comRecall Reference #: R-24-209-FGX1 Recall Code: Recall Code 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity in Commerce318 kits
DistributionUS Nationwide. Canada.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
-
-