 Class 2 Device Recall Medlinesee related information
Date Initiated by FirmNovember 21, 2024
Date PostedDecember 30, 2024
Recall Status1 Open3, Classified
Recall NumberZ-0784-2025
Recall Event ID 95852
Product Classification Endoscope introducer kit - Product Code NWU
ProductMedline custom medical procedure kits labeled as: 1) ENDOSCOPIC SKULL BASE, REF DYNJ58648B; 2) ENDOSCOPIC SKULL BASE, REF DYNJ58648C; 3) ENT ENDO SINUS WILMINGTON PACK, REF DYNJ66447; 4) ENDOSCOPIC SKULL BASE, REF DYNJ58648D; 5) ENDOSCOPIC SKULL BASE, REF DYNJ58648J.
Code Information 1) REF DYNJ58648B, UDI/DI 10193489624182 (EA) 40193489624183 (CS), Lot Numbers: 19BDB193; 2) REF DYNJ58648C, UDI/DI 10193489688542 (EA) 40193489688543 (CS), Lot Numbers: 19EDD585, 19IDA770, 20AMA033, 20AMC750, 20EMB449; 3) REF DYNJ66447, UDI/DI 10193489367379 (EA) 40193489367370 (CS), Lot Numbers: 20KMC595, 21AMB129; 4) REF DYNJ58648D, UDI/DI 10195327212124 (EA) 40195327212125 (CS), Lot Numbers: 22KMD675, 23DMD129, 23HMG766, 23JMD938; 5) REF DYNJ58648J, UDI/DI 10193489356793 (EA) 40193489356794 (CS), Lot Numbers: 20EBN856, 20JBG508, 20JBK503, 20JBT975, 20KBQ270, 21ABK829, 21EBP637, 21EBP646, 21JBO555, 21KBC098, 21KBV126, 21LBP144.
FEI Number 1417592
Recalling Firm/
Manufacturer
MEDLINE INDUSTRIES, LP - Northfield
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information ContactHaley Barclay
866-359-1704
Manufacturer Reason
for Recall
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionMedline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit.
Quantity in Commerce1072 units
DistributionWorldwide distribution - US Nationwide and the countries of UAE, and Canada.