| | Class 1 Device Recall Automated Impella Controller |  |
| Date Initiated by Firm | June 23, 2025 |
| Date Posted | July 26, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2164-2025 |
| Recall Event ID |
97078 |
| PMA Number | P140003 P170011 |
| Product Classification |
Temporary non-roller type left heart support blood pump - Product Code OZD
|
| Product | Automated Impella Controller (AIC), used in left heart support blood pump, including the following:
1. Impella Controller, Packaged, AU; Product Code: 0042-0000-AU.
2. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA.
3. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU.
4. Impella Controller, Packaged, EU, Loaner; Product Code: 0042-0000-EU-L.
5. Impella Controller, Packaged, IN; Product Code: 0042-0000-IN.
6. Impella Controller, Packaged, JP; Product Code: 0042-0000-JP.
7. Impella Controller, Japan, Loaner; Product Code: 0042-0000-JP-L.
8. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK.
9. Impella Controller, UK, Loaner; Product Code: 0042-0000-UK-L.
10. Impella Controller, Packaged, US; Product Code: 0042-0000-US.
11. Impella Controller, Packaged, US, Loaner; Product Code: 0042-0000-US-L.
12. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU.
13. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU.
14. Impella Optical Controller, EU, Loaner; Product Code: 0042-0010-EU-L.
15. Impella Optical Controller, Packaged, IN; Product Code: 0042-0010-IN.
16. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK.
17. Impella Optical Controller, UK, Loaner; Product Code: 0042-0010-UK-L.
18. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US.
19. Impella Optical Controller, US, Loaner; Product Code: 0042-0010-US-L.
20. Optical, AIC, Impella Connect, Packaged, AU; Product Code: 0042-0040-AU.
21. Optical AIC w/Impella Connect, Packaged, CA; Product Code: 0042-0040-CA.
22. AIC w/Remote Link, Packaged, Cln Trl, US; Product Code: 0042-0040-CT-L.
23. Optical AIC w/Impella Connect, Packaged, EU; Product Code: 0042-0040-EU.
24. Optical AIC w/Impella Connect, EU, Loaner; Product Code: 0042-0040-EU-L.
25. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP.
26. Optical AIC w/Impella Connect, Pkgd, JP; Product Code: 0042-0040-JP-L.
27. Optical, AIC, Impella Connect, Pkgd, UK; Product Code: 0042-0040-UK.
28. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.
29. Optical AIC Impella Connect, US, Loaner; Product Code: 0042-0040-US-L.
30. Dbl optical, AIC Impella Connect, Phg US; Product Code: 1000201.
31. AIC w/Impella Connect for ECP; Product Code: 1000432. |
| Code Information |
1. Product Code: 0042-0000-AU; GTIN Number: 813502010886; Serial Numbers: IC7849, IC7850, IC7287, IC7288, IC7289, IC7290, IC6526, IC6527, IC6528, IC6529, IC6530, IC6531, IC5622, IC5677, IC5752, IC5836, IC5341, IC5521, IC5522, IC5523, IC5524, IC5525, IC5237, IC5216, IC5217, IC5236, IC5238.
2. Product Code: 0042-0000-CA; GTIN Number: 813502011272; Serial Numbers: IC9416, IC9414, IC9415, IC9413, IC8283, IC7695, IC7716, IC7694, IC6633, IC6634, IC6635, IC6636, IC6082, IC6083, IC5821, IC5824, IC5680, IC5034, IC5050, IC4944, IC4945, IC3802, IC3803, IC3661, IC3662, IC2686, IC2687, IC2369, IC2370, IC2033, IC2035, IC2032, IC2034, IC1911, IC1912, IC1913, IC1790, IC1791, IC1715, IC1769, IC1664, IC1665, IC1595, IC1598, IC1572, IC1543, IC1544, IC1480, IC1481, IC1324, IC1346.
3. Product Code: 0042-0000-EU; GTIN Number: 813502011289; Serial Numbers: IC7809, IC7717, IC7718, IC7719, IC7720, IC7721, IC7748, IC7674, IC7675, IC7686, IC7687, IC7688, IC7689, IC7690, IC7691, IC7692, IC7693, IC7653, IC7654, IC7655, IC7666, IC7667, IC7646, IC7647, IC7648, IC7649, IC7650, IC7651, IC7652, IC7613, IC7614, IC7615, IC7421, IC7576, IC7577, IC7578, IC7579, IC7580, IC7581, IC7582, IC7583, IC7584, IC7455, IC7452, IC7453, IC7454, IC7378, IC7388, IC7389, IC7390, IC7392, IC7393, IC7394, IC7407, IC7395, IC7406, IC7263, IC7264, IC7265, IC7286, IC7258, IC7259, IC7260, IC7261, IC7262, IC7087, IC7092, IC7093, IC7094, IC7095, IC7096, IC7097, IC7098, IC7099, IC7100, IC7075, IC7076, IC7077, IC7078, IC6843, IC6844, IC6845, IC6876, IC6877, IC6878, IC6879, IC6880, IC6804, IC6805, IC6806, IC6807, IC6808, IC6809, IC6810, IC6812, IC6813, IC6814, IC6728, IC6738, IC6739, IC6740, IC6741, IC6742, IC6710, IC6711, IC6712, IC6713, IC6714, IC6715, IC6716, IC6717, IC6688, IC6689, IC6690, IC6691, IC6692, IC6693, IC6694, IC6695, IC6706, IC6707, IC6708, IC6709, IC6587, IC6588, IC6589, IC6590, IC6591, IC6592, IC6593, IC6594, IC6595, IC6577, IC6578, IC6579, IC6582, IC6580, IC6581, IC6534, IC6535, IC6536, IC6510, IC6449, IC6469, IC6474, IC6487, IC6488, IC6489, IC6437, IC6439, IC6440, IC6441, IC6443, IC6431, IC6432, IC6435, IC6436, IC6434, IC6370, IC6371, IC6372, IC6373, IC6374, IC6375, IC6386, IC6387, IC6270, IC6331, IC6205, IC6214, IC6216, IC6293, IC6294, IC6272, IC6273, IC6274, IC6275, IC6223, IC6224, IC6225, IC6236, IC6164, IC6165, IC6147, IC6148, IC6149, IC6150, IC6151, IC6078, IC6081, IC6079, IC6080, IC6064, IC6076, IC5817, IC5818, IC5774, IC5775, IC5777, IC5779, IC5781, IC5780, IC5754, IC5755, IC5756, IC5757, IC5611, IC5714, IC5715, IC5716, IC5717, IC5722, IC5681, IC5682, IC5683, IC5684, IC5685, IC5686, IC5688, IC5687, IC5559, IC5560, IC5558, IC5517, IC5518, IC5519, IC5520, IC5495, IC5496, IC5487, IC5488, IC5489, IC5490, IC5491, IC5492, IC5493, IC5494, IC5476, IC5477, IC5357, IC5465, IC5466, IC5467, IC5460, IC5461, IC5462, IC5463, IC5464, IC5440, IC5441, IC5442, IC5443, IC5433, IC5434, IC5435, IC5436, IC5331, IC5356, IC5421, IC5422, IC5391, IC5392, IC5251, IC5252, IC5193, IC5194, IC5195, IC5210, IC5212, IC5169, IC5171, IC5173, IC5174, IC5175, IC5176, IC5177, IC5178, IC5051, IC5053, IC5075, IC5086, IC5087, IC5088, IC5046, IC5047, IC5048, IC5049, IC5057, IC5055, IC4973, IC5007, IC5008, IC5009, IC4984, IC4986, IC4987, IC4988, IC4850, IC4928, IC4929, IC4930, IC4931, IC4932, IC4933, IC4934, IC4911, IC4912, IC4913, IC4914, IC4915, IC4916, IC4917, IC4845, IC4846, IC4833, IC4834, IC4835, IC4836, IC4837, IC4839, IC4840, IC4823, IC4824, IC4815, IC4816, IC4817, IC4818, IC4819, IC4820, IC4821, IC4822, IC4790, IC4791, IC4792, IC4793, IC4794, IC4795, IC4796, IC4797, IC4781, IC4748, IC4757, IC4758, IC4759, IC4760, IC4761, IC4762, IC4763, IC4723, IC4724, IC4733, IC4734, IC4735, IC4736, IC4737, IC4738, IC4725, IC4727, IC4728, IC4729, IC4677, IC4678, IC4679, IC4680, IC4681, IC4682, IC4645, IC4646, IC4647, IC4648, IC4649, IC4656, IC4657, IC4658, IC4585, IC4630, IC4631, IC4632, IC4633, IC4634, IC4635, IC4636, IC4623, IC4624, IC4626, IC4627, IC4628, IC4629, IC4609, IC4610, IC4611, IC4612, IC4613, IC4614.IC4594, IC4595, IC4597, IC4598, IC4557, IC4565, IC4566, IC4567, IC4568, IC4569, IC4559, IC4560, IC4561, IC4562, IC4563, IC4564, IC4500, IC4501, IC4502, IC4503, IC4505, IC4506, IC4507, IC4508, IC4464, IC4465, IC4466, IC4467, IC4468, IC4469, IC4470, IC4471, IC4442, IC4443, IC4444, IC4445, IC4446, IC4365, IC4366, IC4367, IC4370, IC4371, IC4356, IC4357, IC4358, IC4359, IC4360, IC4361, IC4362, IC4363, IC4322, IC4323, IC4324, IC4325, IC4326, IC4328, IC4329, IC4251, IC4252, IC4253, IC4254, IC4255, IC4256, IC4257, IC4258, IC4235, IC4236, IC4237, IC4238, IC4241, IC4242, IC4240, IC4233, IC4227, IC4228, IC4229, IC4230, IC4231, IC4232, IC4234, IC4217, IC4218, IC4207, IC4208, IC4209, IC4210, IC4211, IC4212, IC4175, IC4146, IC4147, IC4148, IC4149, IC4150, IC4151, IC4152, IC4153, IC4128, IC4129, IC4096, IC4097, IC4098, IC4099, IC4100, IC4081, IC4083, IC4068, IC4069, IC4054, IC4055, IC4036, IC4037, IC4033, IC4034, IC4035, IC4002, IC4006, IC4012, IC4013, IC4014, IC4015, IC3885, IC3886, IC3887, IC3888, IC3889, IC3891, IC3884, IC3890, IC3868, IC3869, IC3870, IC3871, IC3872, IC3874, IC3875, IC3873, IC3859, IC3861, IC3862, IC3863, IC3864, IC3866, IC3867, IC3860, IC3828, IC3829, IC3830, IC3831, IC3833, IC3834, IC3832, IC3815, IC3816, IC3817, IC3818, IC3812, IC3813, IC3814, IC3777, IC3778, IC3780, IC3781, IC3782, IC3783, IC3776, IC3779, IC3717, IC3724, IC3718, IC3719, IC3720, IC3701, IC3703, IC3706, IC3707, IC3708, IC3709, IC3704, IC3705, IC3696, IC3697, IC3698, IC3699, IC3700, IC3702, IC3657, IC3667, IC3668, IC3669, IC3670, IC3671, IC3672, IC3673, IC3663, IC3664, IC3665, IC3666, IC3492, IC3493, IC3495, IC3656, IC3658, IC3659, IC3494, IC3660, IC3477, IC3485, IC3486, IC3487, IC3488, IC3489, IC3491, IC3490, IC3484, IC3466, IC3467, IC3468, IC3469, IC3470, IC3471, IC3472, IC3458, IC3459, IC3460, IC3461, IC3462, IC3463, IC3313, IC3319, IC3323, IC3329, IC3330, IC3331, IC3332, IC3327, IC3328, IC3314, IC3315, IC3316, IC3317, IC3318, IC3321, IC3322, IC3324, IC3304, IC3305, IC3306, IC3307, IC3308, IC3309, IC3310, IC3311, IC3296, IC3297, IC3298, IC3299, IC3300, IC3303, IC3301, IC3302, IC3286, IC3287, IC3288, IC3289, IC3290, IC3291, IC3292, IC3184, IC3178, IC3179, IC3180, IC3181, IC3182, IC3183, IC3142, IC3143, IC3144, IC3145, IC3146, IC3147, IC3148, IC3149, IC3127, IC3128, IC3129, IC3130, IC3133, IC3134, IC3131, IC3132, IC3095, IC3096, IC3097, IC3098, IC3100, IC3101, IC3099, IC3102, IC3067, IC3068, IC3069, IC3070, IC3048, IC3049, IC3051, IC3052, IC3053, IC3054, IC3050, IC3015, IC3016, IC3017, IC3018, IC3019, IC3020, IC3021, IC3022, IC2996, IC2997, IC2984, IC2985, IC2987, IC2988, IC2989, IC2960, IC2961, IC2962, IC2963, IC2964, IC2873, IC2874, IC2875, IC2876, IC2877, IC2878, IC2855, IC2856, IC2858, IC2859, IC2862, IC2860, IC2861, IC2835, IC2836, IC2837, IC2838, IC2823, IC2824, IC2825, IC2826, IC2828, IC2829, IC2830, IC2827, IC2763, IC2766, IC2767, IC2765, IC2755, IC2756, IC2757, IC2758, IC2753, IC2754, IC2751, IC2752, IC2676, IC2677, IC2678, IC2679, IC2708, IC2709, IC2710, IC2712, IC2668, IC2669, IC2670, IC2652, IC2653, IC2654, IC2655, IC2656, IC2657, IC2658, IC2659, IC2606, IC2607, IC2609, IC2610, IC2559, IC2491, IC2512, IC2534, IC2560, IC2471, IC2481, IC2483, IC2484, IC2485, IC2486, IC2490, IC2494, IC2475, IC2476, IC2477, IC2478, IC2479, IC2469, IC2470, IC2472, IC2473, IC2474, IC2556, IC2557, IC2441, IC2442, IC2443, IC2444, IC2411, IC2412, IC2413, IC2414, IC2380, IC2325, IC2326, IC2327, IC2329, IC2330, IC2331, IC2332, IC2266, IC2267, IC2268, IC2239, IC2240, IC2241, IC2242, IC2243, IC2244, IC2245, IC2231, IC2232, IC2233, IC2234, IC2235, IC2236, IC2220, IC2221, IC2222, IC2219, IC2204, IC2205, IC2206, IC2207, IC2171, IC2172, IC2173, IC2174, IC2178, IC2179, IC2175, IC2095, IC2096, IC2097, IC2098, IC2099, IC2100, IC2101, IC2102, IC2021, IC2038, IC2039, IC2040, IC2005, IC2006, IC1971, IC1972, IC1973, IC1974, IC1975, IC1976, IC1977, IC1978, IC1901, IC1902, IC1903, IC1904, IC1905, IC1907, IC1908, IC1775, IC1780, IC1881, IC1886, IC1870, IC1872, IC1874, IC1871, IC1876.IC1821, IC1822, IC1823, IC1824, IC1825, IC1826, IC1761, IC1763, IC1764, IC1765, IC1766, IC1767, IC1768, IC1720, IC1721, IC1722, IC1723, IC1726, IC1727, IC1724, IC1725, IC1699, IC1700, IC1677, IC1678, IC1684, IC1685, IC1676, IC1653, IC1584, IC1585, IC1587, IC1588, IC1590, IC1591, IC1592, IC1593, IC1594, IC1586, IC1578, IC1579, IC1580, IC1577, IC1546, IC1547, IC1548, IC1549, IC1522, IC1523, IC1524, IC1525, IC1488, IC1490, IC1413, IC1415, IC1411, IC1416, IC1395, IC1396, IC1397, IC1398, IC1393, IC1381, IC1383, IC1384, IC1385, IC1386, IC1347, IC1349, IC1350, IC1340, IC1341, IC1343, IC1344, IC1345, IC1301, IC1302, IC1304, IC1305, IC1284, IC1288, IC1283, IC1285, IC1276, IC1277, IC1278, IC1280, IC1275, IC1265, IC1216, IC1217.
4. Product Code: 0042-0000-EU-L; GTIN Number: 813502011289; Serial Numbers: IC7810, IC6065, IC4910, IC4901, IC4891, IC4892, IC4875, IC4876, IC4879, IC4880, IC4881, IC4874, IC4877, IC4878, IC4867, IC4868, IC4869, IC4870, IC4871, IC4872, IC4873, IC4866, IC4855, IC4847, IC4848, IC4849, IC4851, IC4852, IC4853, IC4838, IC4622, IC4625, IC4364, IC4239, IC4032, IC3827, IC3320, IC2986, IC2857, IC2176, IC1906, IC1762, IC1520, IC1521, IC1487, IC1399, IC1303, IC1011.
5. Product Code: 0042-0000-IN; GTIN Number: 813502011418; Serial Numbers: IC7410, IC7411, IC7408, IC7409, IC7079, IC7080, IC6475, IC6476, IC6477, IC6478, IC6139, IC6252, IC6253, IC6254.
6. Product Code: 0042-0000-JP; GTIN Number: 813502010428; Serial Numbers: IC8276, IC8277, IC8278, IC8279, IC8280, IC8281, IC8282, IC8082, IC8083, IC8084, IC8078, IC8079, IC8080, IC8081, IC8047, IC8048, IC8049, IC8050, IC8051, IC8052, IC8053, IC8054, IC8055, IC8076, IC8077, IC8085, IC7981, IC7982, IC7983, IC7984, IC7985, IC8046, IC7933, IC7934, IC7935, IC7976, IC7977, IC7978, IC7979, IC7980, IC7928, IC7929, IC7902, IC7903, IC7904, IC7905, IC7926, IC7927, IC7899, IC7900, IC7901, IC7852, IC7853, IC7854, IC7855, IC7896, IC7897, IC7898, IC7851, IC7812, IC7813, IC7814, IC7815, IC7846, IC7808, IC7811, IC7784, IC7785, IC7806, IC7779, IC7780, IC7781, IC7782, IC7783, IC7746, IC7747, IC7749, IC7750, IC7751, IC7752, IC7753, IC7754, IC7755, IC7672, IC7673, IC7593, IC7594, IC7595, IC7606, IC7607, IC7608, IC7609, IC7610, IC7611, IC7612, IC7450, IC7451, IC7433, IC7434, IC7435, IC7439, IC7440, IC7412, IC7413, IC7414, IC7415, IC7416, IC7417, IC7420, IC7422, IC7423, IC7424, IC7425, IC7426, IC7427, IC7429, IC7430, IC7431, IC7376, IC7377, IC7379, IC7380, IC7381, IC7382, IC7383, IC7384, IC7385, IC7386, IC7387, IC7291, IC7293, IC7294, IC7295, IC7102, IC7103, IC7104, IC7105, IC7256, IC7257, IC7081, IC7082, IC7083, IC7084, IC7085, IC7089, IC7090, IC7091, IC7058, IC7059, IC7060, IC7061, IC7062, IC7063, IC7064, IC7065, IC7066, IC7067, IC7068, IC6985, IC7029, IC7031, IC7033, IC7034, IC7035, IC6882, IC6883, IC6885, IC6886, IC6887, IC6888, IC6855, IC6856, IC6857, IC6858, IC6859, IC6860, IC6861, IC6862, IC6863, IC6745, IC6746, IC6747, IC6748, IC6749, IC6750, IC6751, IC6752, IC6662, IC6642, IC6650, IC6651, IC6652, IC6653, IC6654, IC6656, IC6657, IC6658, IC6659, IC6660, IC6661, IC6519, IC6605, IC6606, IC6607, IC6610, IC6548, IC6549, IC6550, IC6551, IC6552, IC6553, IC6554, IC6555, IC6556, IC6533, IC6429, IC6430, IC6452, IC6490, IC6491, IC6492, IC6442, IC6451, IC6466, IC6467, IC6468, IC6405, IC6406, IC6407, IC6388, IC6389, IC6306, IC6307, IC6308, IC6309, IC6310, IC6311, IC6312, IC6313, IC6314, IC6315, IC6316, IC6202, IC6203, IC6204, IC6210, IC6211, IC6212, IC6084, IC6085, IC6086, IC6087, IC6088, IC6089, IC6090, IC6091, IC6046, IC6047, IC6048, IC6049, IC5929, IC5930, IC5935, IC5936, IC5937, IC5938, IC5928, IC5864, IC5865, IC5866, IC5867, IC5868, IC5819, IC5820, IC5860, IC5861, IC5862, IC5863, IC5727, IC5728, IC5729, IC5731, IC5732, IC5733, IC5734, IC5735, IC5556, IC5563, IC5651, IC5652, IC5653, IC5654, IC5656, IC5657, IC5658, IC5659, IC5660, IC5665, IC5700, IC5505, IC5506, IC5507, IC5510, IC5511, IC5512, IC5515, IC5516, IC5478, IC5447, IC5444, IC5445, IC5259, IC5260, IC5261, IC5241, IC5243, IC5245, IC5027, IC4732.
7. Product Code: 0042-0000-JP-L; GTIN Number: 813502010428; Serial Numbers: IC7448, IC7449, IC7391, IC7432, IC7436, IC7437, IC7438, IC7418, IC7419, IC7428, IC7292, IC7101, IC7086, IC7088, IC6881, IC6884, IC6655, IC6608, IC6609, IC6539, IC6532, IC6493, IC6494, IC6207, IC6209, IC5931, IC5932, IC5934, IC5869, IC5730, IC5508, IC5509, IC5513, IC5514, IC5258, IC5246, IC5029, IC5030, IC5031, IC4730.
8. Product Code: 0042-0000-UK; GTIN Number: 813502011296; Serial Numbers: IC7847, IC7848, IC7668, IC7669, IC7670, IC7671, IC7585, IC7591, IC7444, IC7445, IC7446, IC7447, IC7441, IC7442, IC7443, IC7072, IC7069, IC7070, IC7071, IC7073, IC7074, IC7056, IC7057, IC6934, IC6925, IC6924, IC6638, IC6669, IC6670, IC6671, IC6583, IC6584, IC6585, IC6586, IC6537, IC6538, IC6542, IC6543, IC6540, IC6286, IC6287, IC6289, IC6290, IC6291, IC6292, IC6001, IC5995, IC5996, IC5997, IC5998, IC5999, IC5914, IC5915, IC5911, IC5912, IC5926, IC5927, IC5910, IC5830, IC5831, IC5832, IC5833, IC5553, IC5554, IC5555, IC5557, IC5253, IC5254, IC5180, IC5211, IC5213, IC5247, IC5240, IC5218, IC5219, IC5220, IC5214, IC5215, IC5179, IC5181, IC5182, IC5184, IC4926, IC4927, IC4856, IC4857, IC4841, IC4842, IC4843, IC4844, IC4739, IC4741, IC4721, IC4722, IC4676, IC4675, IC4599, IC4600, IC4525, IC4438, IC4439, IC4440, IC4441, IC4330, IC4331, IC4082, IC4166, IC3784, IC3785, IC3674, IC3675, IC3475, IC3474, IC3464, IC3465, IC3333, IC3334, IC3325, IC3326, IC3312, IC3293, IC3294, IC3295, IC3174, IC2871, IC2872, IC2704, IC2705, IC2706, IC2707, IC2492, IC2493, IC2497, IC2496, IC2367, IC2368, IC2277, IC2278, IC2023, IC2024, IC2025, IC1909, IC1910, IC1868, IC1869, IC1792, IC1573, IC1581, IC1582, IC1583, IC1545, IC1542, IC1519, IC1494, IC1387, IC1388, IC1389, IC1391, IC1392.
9. Product Code: 0042-0000-UK-L; GTIN Number: 813502011296; Serial Numbers: IC7776, IC7777, IC7778, IC7592, IC3175, IC1576, IC1493.
10. Product Code: 0042-0000-US; GTIN Number: 813502010022; Serial Numbers: IC7919, IC7922, IC7028, IC7030, IC7032, IC6976, IC6977, IC6978, IC6979, IC6980, IC6981, IC6982, IC6983, IC6984, IC7026, IC7027, IC6926, IC6927, IC6928, IC6929, IC6930, IC6931, IC6932, IC6933, IC6935, IC6914, IC6915, IC6916, IC6917, IC6918, IC6919, IC6920, IC6921, IC6922, IC6923, IC6904, IC6905, IC6906, IC6907, IC6908, IC6909, IC6910, IC6911, IC6912, IC6913, IC6889, IC6893, IC6894, IC6895, IC6896, IC6897, IC6898, IC6899, IC6900, IC6901, IC6833, IC6836, IC6837, IC6838, IC6839, IC6840, IC6841, IC6842, IC6834, IC6835, IC6811, IC6829, IC6831, IC6832, IC6830, IC6815, IC6826, IC6827, IC6828, IC6796, IC6797, IC6798, IC6799, IC6800, IC6801, IC6802, IC6803, IC6864, IC6865, IC6850, IC6851, IC6852, IC6853, IC6854, IC6824, IC6847, IC6848, IC6849, IC6825, IC6846, IC6774, IC6775, IC6816, IC6817, IC6818, IC6819, IC6820, IC6822, IC6823, IC6732, IC6753, IC6754, IC6755, IC6766, IC6767, IC6768, IC6769, IC6770, IC6771, IC6772, IC6773, IC6743, IC6744, IC6726, IC6727, IC6729, IC6730, IC6731, IC6733, IC6734, IC6735, IC6736, IC6737, IC6718, IC6719, IC6720, IC6721, IC6722, IC6723, IC6724, IC6725, IC6678, IC6679, IC6680, IC6681, IC6682, IC6683, IC6684, IC6685, IC6686, IC6687, IC6672, IC6674, IC6675, IC6676, IC6677, IC6673, IC6663, IC6664, IC6665, IC6666, IC6667, IC6668, IC6639, IC6640, IC6641, IC6643, IC6644, IC6645, IC6646, IC6647, IC6648, IC6649, IC6637, IC6624, IC6625, IC6626, IC6627, IC6628, IC6629, IC6630, IC6631, IC6632, IC6611, IC6612, IC6613, IC6614, IC6615, IC6616, IC6617, IC6618, IC6619, IC6620, IC6621, IC6622, IC6623, IC6515, IC6598, IC6599, IC6600, IC6601, IC6602, IC6603, IC6604, IC6596, IC6597, IC6562, IC6563, IC6564, IC6565, IC6566, IC6567, IC6568, IC6569, IC6570, IC6571, IC6572, IC6573, IC6574, IC6575, IC6576, IC6557, IC6558, IC6559, IC6560, IC6561, IC6544, IC6545, IC6546, IC6547, IC6511, IC6512, IC6513, IC6514, IC6516, IC6517, IC6518, IC6520, IC6521, IC6522, IC6523, IC6524, IC6525, IC6500, IC6502, IC6503, IC6504, IC6505, IC6506, IC6507, IC6508, IC6509, IC6495, IC6496, IC6497, IC6498, IC6499, IC6501, IC6428, IC6445, IC6446, IC6447, IC6448, IC6433, IC6438, IC6479, IC6480, IC6481, IC6482, IC6483, IC6484, IC6485, IC6486, IC6470, IC6471, IC6472, IC6473, IC6444, IC6450, IC6453, IC6454, IC6455, IC6408, IC6409, IC6410, IC6411, IC6412, IC6413, IC6414, IC6415, IC6426, IC6427, IC6390, IC6391, IC6392, IC6393, IC6394, IC6395, IC6396, IC6397, IC6398, IC6399, IC6400, IC6401, IC6402, IC6404, IC6403, IC6361, IC6362, IC6363, IC6364, IC6365, IC6366, IC6367, IC6368, IC6369, IC6360, IC6288, IC6295, IC6352, IC6353, IC6354, IC6355, IC6356, IC6357, IC6358, IC6359, IC6332, IC6333, IC6334, IC6335, IC6346, IC6347, IC6348, IC6349, IC6350, IC6351, IC6323, IC6324, IC6325, IC6326, IC6327, IC6330, IC6321, IC6322, IC6328, IC6329, IC6095, IC6317, IC6318, IC6319, IC6320, IC6255, IC6266, IC6267, IC6268, IC6269, IC6271, IC6242, IC6243, IC6244, IC6245, IC6246, IC6247, IC6248, IC6249, IC6250, IC6251, IC5933, IC6237, IC6238, IC6239, IC6240, IC6241, IC6206, IC6208, IC6213, IC6215, IC6217, IC6218, IC6219, IC6220, IC6221, IC6222, IC6045, IC6192, IC6193, IC6194, IC6195, IC6196, IC6197, IC6198, IC6200, IC6201, IC6199, IC6184, IC6185, IC6186, IC6187, IC6188, IC6189, IC6190, IC6191, IC6176, IC6177, IC6178, IC6179, IC6180, IC6181, IC6182, IC6183, IC6152, IC6153, IC6154, IC6155, IC6156, IC6157, IC6158, IC6159, IC6160, IC6161, IC6115, IC6135, IC6137, IC6138, IC6140, IC6141, IC6144, IC6145, IC6146, IC6128, IC6129, IC6130, IC6131, IC6132, IC6133, IC6134, IC6136, IC6142, IC6143, IC6103, IC6104, IC6105, IC6106, IC6107, IC6092, IC6093, IC6094, IC6096, IC6097, IC6098, IC6099, IC6100, IC6101, IC6102, IC6077, IC6059, IC6050, IC6051, IC6052, IC6053, IC6054, IC6055, IC6056, IC6057, IC6058, IC6060, IC6061, IC6062, IC6063, IC6000, IC6034, IC6035, IC6036, IC6037, IC6038, IC6039, IC6040, IC6041, IC6043, IC6044, IC6042, IC6002, IC6003, IC6004, IC6005, IC6026, IC6027, IC6028, IC6029, IC6030, IC6031, IC6032, IC6033, IC5565, IC5983, IC5985, IC5986, IC5987, IC5988, IC5989, IC5990, IC5991, IC5992, IC5993, IC5994, IC5984, IC5953, IC5954, IC5955, IC5976, IC5977, IC5978, IC5979, IC5980, IC5981, IC5982, IC5945, IC5946, IC5947, IC5948, IC5949, IC5950, IC5951, IC5952, IC5939, IC5940, IC5941, IC5942, IC5943, IC5944, IC5843, IC5889, IC5890, IC5891, IC5892, IC5893, IC5894, IC5895, IC5906, IC5907, IC5908, IC5909, IC5881, IC5882, IC5883, IC5884, IC5885, IC5886, IC5887, IC5888, IC5776, IC5778, IC5794, IC5838, IC5880, IC5870, IC5871, IC5872, IC5873, IC5874, IC5875, IC5876, IC5877, IC5878, IC5879, IC5848, IC5849, IC5850, IC5851, IC5852, IC5853, IC5854, IC5855, IC5856, IC5857, IC5858, IC5859, IC5839, IC5840, IC5841, IC5842, IC5844, IC5845, IC5846, IC5847, IC5816, IC5822, IC5823, IC5825, IC5826, IC5827, IC5829, IC5828, IC5787, IC5788, IC5789, IC5790, IC5791, IC5792, IC5793, IC5795, IC5766, IC5767, IC5768, IC5769, IC5770, IC5772, IC5773, IC5771, IC5758, IC5759, IC5760, IC5761, IC5762, IC5763, IC5764, IC5765, IC5746, IC5747, IC5748, IC5745, IC5749, IC5750, IC5736, IC5737, IC5738, IC5739, IC5740, IC5741, IC5742, IC5743, IC5718, IC5719, IC5723, IC5724, IC5725, IC5726, IC5720, IC5721, IC5704, IC5705, IC5706, IC5707, IC5708, IC5709, IC5710, IC5711, IC5712, IC5713, IC5701, IC5702, IC5703, IC5692, IC5693, IC5694, IC5695, IC5696, IC5697, IC5698, IC5699, IC5609, IC5690, IC5691, IC5599, IC5621, IC5623, IC5663, IC5664, IC5676, IC5678, IC5661, IC5662, IC5598, IC5549, IC5605, IC5607, IC5608, IC5619, IC5620, IC5624, IC5625, IC5564, IC5617, IC5647, IC5648, IC5649, IC5650, IC5655, IC5562, IC5596, IC5602, IC5603, IC5613, IC5614, IC5615, IC5616, IC5561, IC5597, IC5600, IC5601, IC5604, IC5606, IC5610, IC5612, IC5536, IC5537, IC5538, IC5539, IC5540, IC5541, IC5542, IC5543, IC5544, IC5545, IC5546, IC5547, IC5548, IC5550, IC5551, IC5552, IC5498, IC5499, IC5501, IC5502, IC5503, IC5504, IC5500, IC5497, IC5244, IC5479, IC5480, IC5481, IC5482, IC5483, IC5484, IC5485, IC5486, IC5468, IC5469, IC5470, IC5471, IC5472, IC5473, IC5474, IC5475, IC5453, IC5454, IC5455, IC5456, IC5457, IC5458, IC5459, IC5425, IC5393, IC5394, IC5395, IC5416, IC5417, IC5418, IC5419, IC5420, IC5365, IC5366, IC5367, IC5368, IC5369, IC5370, IC5371, IC5372, IC5373, IC5374, IC5375, IC5386, IC5387, IC5388, IC5389, IC5390, IC5345, IC5358, IC5359, IC5360, IC5361, IC5362, IC5363, IC5364, IC5336, IC5337, IC5338, IC5339, IC5340, IC5342, IC5343, IC5344, IC5319, IC5320, IC5321, IC5322, IC5323, IC5324, IC5325, IC5326, IC5327, IC5328, IC5329, IC5330, IC5332, IC5333, IC5334, IC5335, IC5311, IC5312, IC5313, IC5314, IC5315, IC5316, IC5317, IC5318, IC5242, IC5304, IC5305, IC5306, IC5307, IC5308, IC5309, IC5310, IC5271, IC5297, IC5298, IC5299, IC5300, IC5301, IC5302, IC5303, IC5289, IC5290, IC5291, IC5292, IC5293, IC5294, IC5295, IC5296, IC5270, IC5272, IC5273, IC5274, IC5275, IC5286, IC5287, IC5288, IC5262, IC5263, IC5264, IC5265, IC5266, IC5267, IC5268, IC5269, IC5170, IC5172, IC5206, IC5221, IC5222, IC5223, IC5224, IC5225, IC5188, IC5189, IC5190, IC5191, IC5192, IC5203, IC5204, IC5205, IC5207, IC5208, IC5209, IC5056, IC5201, IC5183, IC5185, IC5196, IC5197, IC5198, IC5199, IC5200, IC5202, IC5003, IC5150, IC5166, IC5167, IC5168, IC5024, IC5052, IC5124, IC5125, IC5146, IC5147, IC5148, IC5149, IC5151, IC5152, IC5153, IC5154, IC5155, IC5095, IC5116, IC5117, IC5118, IC5119, IC5122, IC5123, IC5120, IC5069, IC5070, IC5072, IC5090, IC5091, IC5092, IC5093, IC5094, IC4807, IC5054, IC5058, IC5059, IC5060, IC5063, IC5068, IC5089, IC5061, IC5062, IC5064, IC5066, IC5067, IC5071, IC5073, IC5074, IC5019, IC5017, IC5018, IC5015, IC4997, IC4998, IC4999, IC5001, IC5002, IC5004, IC5005, IC5006, IC4989, IC4990, IC4991, IC4992, IC4993, IC4994, IC4995, IC4996, IC4965, IC4969, IC4970, IC4972, IC4974, IC4980, IC4981, IC4982, IC4956, IC4971, IC4975, IC4976, IC4977, IC4978, IC4979, IC4983, IC4941, IC4943, IC4962, IC4963, IC4964, IC4966, IC4967, IC4968, IC4854, IC4935, IC4936, IC4937, IC4938, IC4939, IC4940, IC4942, IC4918, IC4919, IC4920, IC4921, IC4922, IC4923, IC4924, IC4925, IC4902, IC4903, IC4904, IC4905, IC4906, IC4907, IC4908, IC4909, IC4861, IC4885, IC4886, IC4887, IC4888, IC4889, IC4890, IC4893, IC4894, IC4895, IC4896, IC4897, IC4898, IC4899, IC4900, IC4884, IC4858, IC4859, IC4860, IC4862, IC4863, IC4864, IC4865, IC4825, IC4826, IC4827, IC4828, IC4829, IC4830, IC4831, IC4832, IC4798, IC4799, IC4800, IC4801, IC4802, IC4803, IC4804, IC4805, IC4806, IC4808, IC4809, IC4810, IC4811, IC4812, IC4813, IC4814, IC4782, IC4783, IC4784, IC4785, IC4786, IC4787, IC4788, IC4789, IC4772, IC4773, IC4774, IC4775, IC4776, IC4777, IC4778, IC4779, IC4764, IC4765, IC4766, IC4767, IC4768, IC4769, IC4770, IC4771, IC4749, IC4750, IC4751, IC4752, IC4753, IC4754, IC4755, IC4756, IC4744, IC4705, IC4706, IC4707, IC4709, IC4710, IC4711, IC4712, IC4713, IC4714, IC4717, IC4718, IC4719, IC4720, IC4708, IC4715, IC4716, IC4697, IC4698, IC4699, IC4700, IC4701, IC4702, IC4703, IC4704, IC4596, IC4690, IC4691, IC4693, IC4694, IC4695, IC4696, IC4692, IC4588, IC4683, IC4684, IC4685, IC4686, IC4687, IC4688, IC4689, IC4667, IC4668, IC4669, IC4670, IC4671, IC4672, IC4673, IC4674, IC4659, IC4660, IC4661, IC4662, IC4663, IC4664, IC4665, IC4666, IC4637, IC4638, IC4639, IC4642, IC4643, IC4644, IC4640, IC4641, IC4615, IC4617, IC4618, IC4619, IC4620, IC4621, IC4616, IC4601, IC4602, IC4603, IC4604, IC4605, IC4606, IC4607, IC4608, IC4591, IC4593, IC4577, IC4578, IC4579, IC4580, IC4581, IC4582, IC4583, IC4584, IC4488, IC4570, IC4571, IC4572, IC4573, IC4574, IC4575, IC4576, IC4504, IC4549, IC4551, IC4554, IC4555, IC4556, IC4558, IC4534, IC4535, IC4536, IC4537, IC4538, IC4539, IC4540, IC4541, IC4542, IC4543, IC4544, IC4546, IC4547, IC4548, IC4550, IC4552, IC4545, IC4526, IC4527, IC4528, IC4529, IC4530, IC4531, IC4532, IC4533, IC4509, IC4510, IC4511, IC4512, IC4513, IC4514, IC4515, IC4516, IC4517, IC4518, IC4519, IC4520, IC4521, IC4522, IC4523, IC4524, IC4483, IC4484, IC4485, IC4486, IC4487, IC4489, IC4490, IC4491, IC4492, IC4493, IC4494, IC4495, IC4496, IC4497, IC4498, IC4499, IC4472, IC4473, IC4474, IC4475, IC4476, IC4477, IC4478, IC4479, IC4480, IC4481, IC4482, IC4456, IC4457, IC4458, IC4459, IC4460, IC4461, IC4462, IC4463, IC4451, IC4452, IC4453, IC4454, IC4455, IC4432, IC4433, IC4434, IC4435, IC4436, IC4437, IC4430, IC4431, IC4422, IC4423, IC4424, IC4425, IC4426, IC4427, IC4428, IC4429, IC4413, IC4414, IC4415, IC4416, IC4417, IC4418, IC4419, IC4420, IC4405, IC4406, IC4407, IC4408, IC4409, IC4410, IC4411, IC4412, IC4397, IC4398, IC4399, IC4400, IC4401, IC4402, IC4403, IC4404, IC4389, IC4390, IC4391, IC4392, IC4393, IC4394, IC4395, IC4396, IC4381, IC4382, IC4383, IC4386, IC4387, IC4380, IC4372, IC4373, IC4374, IC4375, IC4377, IC4378, IC4379, IC4376, IC4348, IC4349, IC4350, IC4351, IC4352, IC4353, IC4354, IC4340, IC4341, IC4342, IC4343, IC4344, IC4345, IC4346, IC4347, IC4333, IC4334, IC4336, IC4337, IC4339, IC4314, IC4315, IC4316, IC4317, IC4318, IC4320, IC4321, IC4306, IC4307, IC4308, IC4309, IC4310, IC4311, IC4312, IC4313, IC4298, IC4299, IC4300, IC4301, IC4302, IC4303, IC4304, IC4305, IC4290, IC4291, IC4292, IC4293, IC4294, IC4297, IC4295, IC4278, IC4279, IC4280, IC4281, IC4282, IC4283, IC4284, IC4285, IC4275, IC4276, IC4277, IC4268, IC4269, IC4270, IC4271, IC4272, IC4273, IC4274, IC4260, IC4261, IC4262, IC4263, IC4264, IC4265, IC4266, IC4267, IC4259, IC4243, IC4244, IC4245, IC4246, IC4247, IC4248, IC4249, IC4250, IC4219, IC4220, IC4221, IC4222, IC4223, IC4224, IC4225, IC4226, IC4199, IC4200, IC4201, IC4202, IC4203, IC4204, IC4205, IC4206, IC4191, IC4192, IC4193, IC4194, IC4195, IC4196, IC4197, IC4198, IC4167, IC4168, IC4169, IC4170, IC4171, IC4172, IC4173, IC4174, IC4158, IC4159, IC4160, IC4161, IC4162, IC4163, IC4164, IC4165, IC4130, IC4131, IC4132, IC4133, IC4134, IC4135, IC4136, IC4137, IC4109, IC4110, IC4111, IC4112, IC4113, IC4114, IC4115, IC4116, IC4101, IC4102, IC4103, IC4104, IC4105, IC4106, IC4107, IC4108, IC4084, IC4085, IC4087, IC4088, IC4089, IC4090, IC4091, IC4070, IC4071, IC4072, IC4073, IC4074, IC4075, IC4076, IC4077, IC4061, IC4062, IC4063, IC4064, IC4065, IC4066, IC4067, IC4060, IC4038, IC4039, IC4040, IC4043, IC4044, IC4041, IC4042, IC4045, IC4024, IC4025, IC4026, IC4027, IC4028, IC4029, IC4030, IC4031, IC4016, IC4018, IC4019, IC4020, IC4021, IC4022, IC4023, IC4003, IC4004, IC4005, IC4007, IC4008, IC4009, IC4010, IC4011, IC3994, IC3995, IC3996, IC3997, IC3998, IC3999, IC4000, IC4001, IC3766, IC3986, IC3988, IC3989, IC3990, IC3991, IC3992, IC3993, IC3979, IC3980, IC3981, IC3982, IC3983, IC3984, IC3985, IC3987, IC3969, IC3970, IC3971, IC3972, IC3973, IC3974, IC3975, IC3976, IC3961, IC3962, IC3963, IC3964, IC3965, IC3966, IC3967, IC3968, IC3865, IC3954, IC3955, IC3956, IC3957, IC3958, IC3959, IC3960, IC3946, IC3947, IC3948, IC3949, IC3950, IC3951, IC3952, IC3953, IC3939, IC3940, IC3941, IC3942, IC3943, IC3944, IC3945, IC3931, IC3932, IC3933, IC3934, IC3935, IC3936, IC3937, IC3938, IC3923, IC3924, IC3925, IC3926, IC3927, IC3928, IC3929, IC3930, IC3915, IC3916, IC3917, IC3918, IC3919, IC3920, IC3921, IC3922, IC3907, IC3908, IC3909, IC3910, IC3911, IC3912, IC3913, IC3914, IC3900, IC3901, IC3902, IC3903, IC3904, IC3906, IC3892, IC3893, IC3894, IC3895, IC3896, IC3897, IC3898, IC3899, IC3876, IC3877, IC3878, IC3879, IC3880, IC3881, IC3882, IC3883, IC3851, IC3852, IC3853, IC3854, IC3855, IC3856, IC3857, IC3858, IC3843, IC3844, IC3845, IC3846, IC3847, IC3848, IC3849, IC3850, IC3835, IC3836, IC3837, IC3838, IC3839, IC3840, IC3841, IC3842, IC3819, IC3820, IC3821, IC3824, IC3825, IC3826, IC3822, IC3823, IC3804, IC3805, IC3807, IC3808, IC3809, IC3810, IC3811, IC3806, IC3794, IC3795, IC3796, IC3797, IC3798, IC3799, IC3800, IC3801, IC3786, IC3787, IC3788, IC3789, IC3790, IC3791, IC3792, IC3793, IC3768, IC3769, IC3770, IC3771, IC3774, IC3775, IC3772, IC3773, IC3749, IC3750, IC3751, IC3752, IC3753, IC3754, IC3755, IC3756, IC3741, IC3742, IC3743, IC3744, IC3745, IC3746, IC3747, IC3748, IC3733, IC3734, IC3735, IC3736, IC3737, IC3738, IC3739, IC3740, IC3725, IC3726, IC3727, IC3728, IC3729, IC3730, IC3731, IC3732, IC3694, IC3695, IC3710, IC3711, IC3712, IC3713, IC3714, IC3715, IC3686, IC3687, IC3688, IC3689, IC3690, IC3691, IC3693, IC3679, IC3680, IC3681, IC3682, IC3683, IC3684, IC3685, IC3643, IC3644, IC3645, IC3646, IC3647, IC3648, IC3649, IC3650, IC3635, IC3636, IC3637, IC3638, IC3640, IC3641, IC3642, IC3627, IC3628, IC3629, IC3630, IC3631, IC3632, IC3633, IC3634, IC3564, IC3610, IC3622, IC3623, IC3624, IC3625, IC3626, IC3609, IC3599, IC3600, IC3601, IC3602, IC3603, IC3604, IC3589, IC3590, IC3591, IC3592, IC3594, IC3595, IC3596, IC3575, IC3576, IC3577, IC3578, IC3579, IC3580, IC3581, IC3582, IC3567, IC3568, IC3569, IC3570, IC3571, IC3572, IC3573, IC3574, IC3561, IC3562, IC3563, IC3565, IC3566, IC3542, IC3543, IC3544, IC3545, IC3546, IC3547, IC3548, IC3549, IC3550, IC3551, IC3552, IC3553, IC3554, IC3555, IC3556, IC3557, IC3523, IC3524, IC3525, IC3526, IC3527, IC3528, IC3529, IC3530, IC3515, IC3516, IC3517, IC3518, IC3519, IC3520, IC3521, IC3522, IC3507, IC3508, IC3509, IC3510, IC3511, IC3512, IC3513, IC3514, IC3499, IC3500, IC3501, IC3502, IC3503, IC3504, IC3505, IC3506, IC3451, IC3452, IC3453, IC3454, IC3455, IC3496, IC3497, IC3498, IC3443, IC3444, IC3445, IC3446, IC3447, IC3448, IC3449, IC3450, IC3436, IC3437, IC3438, IC3439, IC3440, IC3441, IC3427, IC3428, IC3429, IC3432, IC3433, IC3434, IC3419, IC3420, IC3421, IC3422, IC3423, IC3424, IC3425, IC3426, IC3349, IC3412, IC3413, IC3414, IC3415, IC3416, IC3417, IC3418, IC3404, IC3405, IC3406, IC3407, IC3408, IC3409, IC3410, IC3411, IC3346, IC3354, IC3378, IC3379, IC3347, IC3351, IC3394, IC3396, IC3397, IC3399, IC3400, IC3401, IC3384, IC3385, IC3386, IC3388, IC3390, IC3374, IC3375, IC3376, IC3377, IC3380, IC3381, IC3382, IC3383, IC3366, IC3367, IC3368, IC3369, IC3370, IC3371, IC3372, IC3373, IC3358, IC3359, IC3360, IC3361, IC3362, IC3363, IC3364, IC3365, IC3345, IC3348, IC3350, IC3352, IC3353, IC3355, IC3356, IC3357, IC3336, IC3337, IC3338, IC3339, IC3340, IC3341, IC3343, IC3342, IC3278, IC3279, IC3281, IC3282, IC3283, IC3284, IC3285, IC3280, IC3270, IC3271, IC3272, IC3273, IC3274, IC3275, IC3276, IC3277, IC3219, IC3261, IC3262, IC3263, IC3266, IC3267, IC3268, IC3269, IC3253, IC3254, IC3255, IC3257, IC3258, IC3259, IC3260, IC3214, IC3215, IC3216, IC3217, IC3218, IC3220, IC3221, IC3222, IC3245, IC3247, IC3248, IC3249, IC3250, IC3251, IC3252, IC3246, IC3229, IC3230, IC3231, IC3232, IC3233, IC3234, IC3235, IC3236, IC3237, IC3238, IC3239, IC3242, IC3243, IC3241, IC3240, IC3244, IC3212, IC3224, IC3225, IC3226, IC3227, IC3228, IC3213, IC3223, IC3204, IC3205, IC3206, IC3207, IC3208, IC3209, IC3210, IC3211, IC3196, IC3197, IC3198, IC3199, IC3200, IC3201, IC3202, IC3203, IC3166, IC3167, IC3186, IC3187, IC3188, IC3189, IC3191, IC3192, IC3193, IC3194, IC3185, IC3168, IC3169, IC3170, IC3171, IC3172, IC3173, IC3176, IC3177, IC3158, IC3159, IC3160, IC3161, IC3162, IC3163, IC3164, IC3165, IC3150, IC3151, IC3152, IC3153, IC3156, IC3157, IC3154, IC3155, IC2978, IC3135, IC3136, IC3137, IC3138, IC3139, IC3140, IC3141, IC3119, IC3120, IC3121, IC3122, IC3123, IC3124, IC3125, IC3126, IC3111, IC3112, IC3113, IC3114, IC3115, IC3116, IC3117, IC3118, IC3103, IC3104, IC3105, IC3106, IC3107, IC3108, IC3109, IC3110, IC3087, IC3088, IC3089, IC3090, IC3091, IC3092, IC3093, IC3094, IC3079, IC3080, IC3081, IC3082, IC3083, IC3084, IC3085, IC3086, IC3063, IC3064, IC3065, IC3066, IC3071, IC3072, IC3073, IC3074, IC3075, IC3076, IC3077, IC3078, IC3055, IC3056, IC3057, IC3058, IC3059, IC3060, IC3061, IC3062, IC3039, IC3040, IC3041, IC3042, IC3043, IC3044, IC3045, IC3046, IC3031, IC3032, IC3033, IC3034, IC3035, IC3036, IC3037, IC3038, IC3023, IC3024, IC3025, IC3026, IC3027, IC3028, IC3029, IC3030, IC3006, IC3007, IC3008, IC3011, IC3012, IC3013, IC3009, IC3010, IC2998, IC2999, IC3000, IC3001, IC3002, IC3003, IC3004, IC3005, IC2982, IC2983, IC2990, IC2991, IC2992, IC2993, IC2994, IC2995, IC2973, IC2974, IC2975, IC2979, IC2980, IC2981, IC2976, IC2977, IC2965, IC2966, IC2969, IC2970, IC2971, IC2972, IC2949, IC2950, IC2951, IC2952, IC2953, IC2954, IC2955, IC2956, IC2731, IC2942, IC2943, IC2944, IC2945, IC2946, IC2947, IC2948, IC2934, IC2935, IC2936, IC2937, IC2938, IC2939, IC2940, IC2941, IC2926, IC2927, IC2928, IC2929, IC2930, IC2931, IC2932, IC2933, IC2918, IC2919, IC2920, IC2921, IC2922, IC2923, IC2924, IC2925, IC2711, IC2914, IC2915, IC2916, IC2917, IC2913, IC2907, IC2908, IC2909, IC2910, IC2887, IC2888, IC2889, IC2890, IC2891, IC2892, IC2893, IC2895, IC2896, IC2897, IC2898, IC2899, IC2901, IC2879, IC2880, IC2881, IC2882, IC2883, IC2884, IC2885, IC2886, IC2863, IC2865, IC2866, IC2867, IC2868, IC2869, IC2870, IC2847, IC2848, IC2849, IC2850, IC2851, IC2852, IC2853, IC2854, IC2839, IC2842, IC2843, IC2844, IC2845, IC2846, IC2840, IC2841, IC2831, IC2832, IC2833, IC2834, IC2815, IC2816, IC2817, IC2818, IC2820, IC2821, IC2822, IC2807, IC2808, IC2809, IC2810, IC2811, IC2812, IC2813, IC2814, IC2764, IC2792, IC2793, IC2794, IC2795, IC2796, IC2797, IC2798, IC2799, IC2800, IC2801, IC2802, IC2803, IC2804, IC2805, IC2806, IC2784, IC2785, IC2786, IC2787, IC2788, IC2789, IC2790, IC2791, IC2776, IC2777, IC2778, IC2779, IC2780, IC2781, IC2782, IC2783, IC2759, IC2760, IC2761, IC2762, IC2768, IC2769, IC2770, IC2771, IC2772, IC2773, IC2774, IC2775, IC2740, IC2741, IC2742, IC2743, IC2744, IC2745, IC2746, IC2739, IC2730, IC2732, IC2733, IC2734, IC2735, IC2736, IC2737, IC2738, IC2747, IC2748, IC2749, IC2819, IC2722, IC2723, IC2724, IC2725, IC2726, IC2727, IC2728, IC2729, IC2717, IC2718, IC2719, IC2721, IC2720, IC2713, IC2714, IC2715, IC2716, IC2689, IC2690, IC2693, IC2694, IC2695, IC2696, IC2697, IC2698, IC2699, IC2700, IC2701, IC2702, IC2703, IC2691, IC2692, IC2675, IC2680, IC2681, IC2682, IC2684, IC2621, IC2650, IC2651, IC2660, IC2661, IC2662, IC2663, IC2664, IC2665, IC2666, IC2667, IC2671, IC2672, IC2674, IC2643, IC2644, IC2645, IC2646, IC2647, IC2648, IC2649, IC2575, IC2631, IC2635, IC2636, IC2637, IC2638, IC2639, IC2640, IC2641, IC2615, IC2616, IC2626, IC2627, IC2628, IC2629, IC2630, IC2632, IC2633, IC2634, IC2617, IC2618, IC2619, IC2620, IC2622, IC2623, IC2624, IC2625, IC2611, IC2612, IC2613, IC2614, IC2601, IC2602, IC2603, IC2604, IC2605, IC2608, IC2506, IC2594, IC2595, IC2596, IC2597, IC2598, IC2599, IC2600, IC2590, IC2591, IC2592, IC2593, IC2585, IC2587, IC2588, IC2589, IC2578, IC2579, IC2580, IC2581, IC2582, IC2583, IC2584, IC2586, IC2569, IC2570, IC2571, IC2572, IC2573, IC2574, IC2576, IC2577, IC2563, IC2564, IC2565, IC2566, IC2567, IC2568, IC2561, IC2562, IC2468, IC2480, IC2482, IC2495, IC2546, IC2547, IC2548, IC2549, IC2538, IC2539, IC2540, IC2541, IC2542, IC2543, IC2544, IC2545, IC2529, IC2530, IC2531, IC2532, IC2533, IC2535, IC2536, IC2537, IC2464, IC2465, IC2466, IC2507, IC2515, IC2516, IC2517, IC2518, IC2519, IC2520, IC2522, IC2523, IC2524, IC2525, IC2526, IC2527, IC2528, IC2521, IC2508, IC2509, IC2510, IC2511, IC2513, IC2514, IC2498, IC2499, IC2500, IC2501, IC2502, IC2503, IC2504, IC2505, IC2436, IC2439, IC2467, IC2487, IC2488, IC2489, IC2449, IC2458, IC2462, IC2463, IC2459, IC2460, IC2461, IC2450, IC2454, IC2456, IC2457, IC2451, IC2452, IC2453, IC2550, IC2551, IC2552, IC2553, IC2554, IC2555, IC2402, IC2430, IC2438, IC2440, IC2445, IC2446, IC2447, IC2448, IC2433, IC2434, IC2435, IC2437, IC2431, IC2432, IC2428, IC2429, IC2409, IC2421, IC2422, IC2423, IC2424, IC2425, IC2426, IC2427, IC2408, IC2410, IC2415, IC2416, IC2417, IC2418, IC2419, IC2420, IC2404, IC2405, IC2406, IC2407, IC2397, IC2398, IC2399, IC2400, IC2401, IC2403, IC2388, IC2389, IC2390, IC2391, IC2392, IC2393, IC2394, IC2395, IC2396, IC2381, IC2382, IC2383, IC2384, IC2385, IC2386, IC2387, IC2371, IC2372, IC2373, IC2374, IC2375, IC2376, IC2377, IC2378, IC2379, IC2359, IC2360, IC2361, IC2362, IC2363, IC2364, IC2365, IC2366, IC2351, IC2352, IC2353, IC2354, IC2355, IC2356, IC2357, IC2358, IC2347, IC2348, IC2349, IC2350, IC2339, IC2340, IC2341, IC2342, IC2343, IC2344, IC2345, IC2323, IC2324, IC2333, IC2334, IC2335, IC2336, IC2337, IC2338, IC2295, IC2297, IC2310, IC2311, IC2312, IC2313, IC2316, IC2317, IC2318, IC2319, IC2320, IC2322, IC2314, IC2315, IC2321, IC2302, IC2303, IC2304, IC2305, IC2306, IC2307, IC2308, IC2309, IC1794, IC2293, IC2294, IC2296, IC2298, IC2299, IC2300, IC2301, IC2279, IC2280, IC2281, IC2282, IC2283, IC2284, IC2285, IC2286, IC2287, IC2288, IC2289, IC2290, IC2291, IC2292, IC2269, IC2270, IC2271, IC2272, IC2273, IC2274, IC2275, IC2276, IC2248, IC2262, IC2263, IC2264, IC2265, IC2254, IC2255, IC2256, IC2257, IC2258, IC2259, IC2260, IC2261, IC2238, IC2246, IC2247, IC2249, IC2250, IC2251, IC2252, IC2253, IC2037, IC2223, IC2224, IC2225, IC2227, IC2228, IC2229, IC2230, IC2237, IC2211, IC2212, IC2213, IC2214, IC2215, IC2216, IC2217, IC2218, IC2177, IC2195, IC2196, IC2197, IC2198, IC2200, IC2201, IC2202, IC2203, IC2208, IC2209, IC2210, IC2199, IC2185, IC2186, IC2187, IC2188, IC2189, IC2190, IC2191, IC2192, IC2193, IC2194, IC2168, IC2169, IC2170, IC2180, IC2181, IC2182, IC2183, IC2184, IC2148, IC2149, IC2150, IC2152, IC2153, IC2155, IC2157, IC2158, IC2159, IC2160, IC2161, IC2162, IC2163, IC2164, IC2165, IC2166, IC2167, IC2154, IC1947, IC2138, IC2140, IC2146, IC2147, IC1970, IC2134, IC2135, IC2136, IC2137, IC2139, IC2141, IC2142, IC2103, IC2131, IC2132, IC2133, IC2124, IC2125, IC2126, IC2127, IC2128, IC2129, IC2130, IC2114, IC2115, IC2116, IC2117, IC2118, IC2119, IC2120, IC2121, IC2122, IC2104, IC2105, IC2106, IC2107, IC2108, IC2109, IC2110, IC2111, IC2112, IC2113, IC2089, IC2090, IC2091, IC2092, IC2093, IC2094, IC2087, IC2088, IC2079, IC2080, IC2081, IC2082, IC2083, IC2084, IC2085, IC2086, IC2072, IC2073, IC2074, IC2075, IC2076, IC2077, IC2078, IC2056, IC2057, IC2058, IC2059, IC2060, IC2061, IC2062, IC2063, IC2064, IC2065, IC2066, IC2067, IC2068, IC2070, IC2069, IC2071, IC2048, IC2049, IC2050, IC2051, IC2052, IC2053, IC2054, IC2055, IC2036, IC2041, IC2042, IC2043, IC2044, IC2045, IC2046, IC2047, IC2022, IC2026, IC2027, IC2028, IC2030, IC2031, IC2029, IC2007, IC2008, IC2010, IC2011, IC2012, IC2013, IC2014, IC2015, IC2016, IC2017, IC2018, IC2019, IC2020, IC2004, IC1994, IC1995, IC1996, IC1997, IC1998, IC1999, IC2000, IC2001, IC2002, IC2003, IC1981, IC1988, IC1989, IC1990, IC1991, IC1992, IC1993, IC1979, IC1980, IC1982, IC1983, IC1984, IC1985, IC1986, IC1987, IC1962, IC1963, IC1965, IC1966, IC1967, IC1968, IC1969, IC1964, IC1956, IC1957, IC1958, IC1959, IC1960, IC1961, IC1955, IC1948, IC1949, IC1950, IC1951, IC1952, IC1953, IC1954, IC1946, IC1936, IC1937, IC1938, IC1939, IC1940, IC1941, IC1942, IC1943, IC1945, IC1944, IC1922, IC1923, IC1924, IC1925, IC1926, IC1928, IC1929, IC1930, IC1931, IC1932, IC1933, IC1934, IC1935, IC1851, IC1895, IC1896, IC1897, IC1898, IC1899, IC1900, IC1914, IC1915, IC1916, IC1917, IC1918, IC1919, IC1920, IC1921, IC1885, IC1887, IC1888, IC1889, IC1890, IC1891, IC1892, IC1893, IC1894, IC1843, IC1778, IC1797, IC1802, IC1864, IC1875, IC1877, IC1878, IC1879, IC1880, IC1882, IC1883, IC1884, IC1854, IC1855, IC1857, IC1858, IC1859, IC1860, IC1861, IC1862, IC1863, IC1865, IC1866, IC1867, IC1842, IC1844, IC1845, IC1846, IC1847, IC1848, IC1849, IC1850, IC1852, IC1853, IC1834, IC1835, IC1836, IC1837, IC1838, IC1839, IC1840, IC1841, IC1812, IC1814, IC1815, IC1816, IC1817, IC1818, IC1819, IC1820, IC1827, IC1828, IC1829, IC1830, IC1831, IC1832, IC1833, IC1813, IC1799, IC1801, IC1803, IC1804, IC1805, IC1806, IC1807, IC1808, IC1809, IC1810, IC1811, IC1793, IC1795, IC1800, IC1798, IC1770, IC1772, IC1773, IC1774, IC1776, IC1777, IC1779, IC1781, IC1783, IC1784, IC1785, IC1786, IC1787, IC1788, IC1789, IC1771, IC1748, IC1749, IC1750, IC1751, IC1752, IC1753, IC1754, IC1755, IC1756, IC1757, IC1758, IC1759, IC1760, IC1738, IC1739, IC1740, IC1742, IC1743, IC1744, IC1746, IC1747, IC1741, IC1728, IC1729, IC1730, IC1731, IC1732, IC1733, IC1734, IC1735, IC1736, IC1737, IC1712, IC1713, IC1714, IC1716, IC1717, IC1718, IC1719, IC1711, IC1701, IC1702, IC1703, IC1704, IC1705, IC1707, IC1709, IC1710, IC1708, IC1687, IC1688, IC1689, IC1690, IC1691, IC1692, IC1693, IC1694, IC1695, IC1696, IC1697, IC1698, IC1679, IC1680, IC1681, IC1682, IC1683, IC1686, IC1666, IC1668, IC1669, IC1670, IC1671, IC1672, IC1673, IC1674, IC1675, IC1654, IC1655, IC1656, IC1657, IC1658, IC1659, IC1660, IC1661, IC1662, IC1663, IC1641, IC1642, IC1643, IC1644, IC1645, IC1646, IC1648, IC1650, IC1651, IC1652, IC1629, IC1630, IC1631, IC1632, IC1633, IC1634, IC1635, IC1636, IC1637, IC1638, IC1639, IC1640, IC1621, IC1622, IC1623, IC1624, IC1625, IC1626, IC1627, IC1628, IC1596, IC1599, IC1600, IC1601, IC1602, IC1603, IC1604, IC1605, IC1606, IC1607, IC1608, IC1609, IC1610, IC1612, IC1613, IC1614, IC1615, IC1616, IC1617, IC1618, IC1558, IC1559, IC1560, IC1561, IC1574, IC1575, IC1562, IC1563, IC1564, IC1565, IC1566, IC1567, IC1568, IC1569, IC1570, IC1571, IC1550, IC1551, IC1552, IC1553, IC1554, IC1555, IC1556, IC1557, IC1526, IC1527, IC1528, IC1529, IC1530, IC1531, IC1532, IC1533, IC1534, IC1535, IC1536, IC1537, IC1538, IC1539, IC1540, IC1541, IC1497, IC1498, IC1499, IC1500, IC1501, IC1502, IC1503, IC1504, IC1505, IC1506, IC1507, IC1508, IC1509, IC1510, IC1512, IC1513, IC1514, IC1515, IC1516, IC1517, IC1518, IC1511, IC1486, IC1495, IC1496, IC1489, IC1491, IC1492, IC1476, IC1477, IC1478, IC1479, IC1483, IC1484, IC1485, IC1451, IC1452, IC1453, IC1454, IC1455, IC1456, IC1457, IC1458, IC1459, IC1460, IC1461, IC1462, IC1463, IC1464, IC1465, IC1466, IC1467, IC1469, IC1470, IC1471, IC1472, IC1473, IC1474, IC1475, IC1368, IC1444, IC1445, IC1446, IC1448, IC1449, IC1450, IC1432, IC1433, IC1434, IC1435, IC1436, IC1437, IC1438, IC1439, IC1440, IC1441, IC1442, IC1443, IC1424, IC1425, IC1426, IC1427, IC1428, IC1429, IC1430, IC1431, IC1418, IC1419, IC1420, IC1421, IC1422, IC1423, IC1412, IC1414, IC1417, IC1394, IC1400, IC1401, IC1402, IC1403, IC1404, IC1405, IC1406, IC1407, IC1408, IC1409, IC1410, IC1390, IC1382, IC1372, IC1373, IC1374, IC1375, IC1376, IC1377, IC1378, IC1379, IC1380, IC1359, IC1360, IC1361, IC1362, IC1363, IC1364, IC1365, IC1351, IC1352, IC1353, IC1354, IC1355, IC1356, IC1357, IC1358, IC1334, IC1337, IC1338, IC1339, IC1342, IC1329, IC1333, IC1335, IC1336, IC1326, IC1327, IC1328, IC1330, IC1331, IC1332, IC1312, IC1313, IC1314, IC1315, IC1316, IC1317, IC1319, IC1320, IC1321, IC1322, IC1325, IC1306, IC1307, IC1309, IC1310, IC1311, IC1289, IC1290, IC1291, IC1292, IC1293, IC1294, IC1295, IC1296, IC1297, IC1298, IC1299, IC1300, IC1270, IC1271, IC1272, IC1273, IC1274, IC1269, IC1259, IC1261, IC1262, IC1263, IC1264, IC1246, IC1247, IC1250, IC1251, IC1252, IC1253, IC1254, IC1255, IC1256, IC1257, IC1258, IC1242, IC1243, IC1244, IC1245, IC1239, IC1240, IC1237, IC1238, IC1228, IC1229, IC1230, IC1233, IC1234, IC1235, IC1236, IC1231, IC1204, IC1205, IC1206, IC1207, IC1203, IC1200, IC1201, IC1202, IC1208, IC1209, IC1210, IC1211, IC1212, IC1213, IC1214, IC1215, IC1220, IC1221, IC1222, IC1223, IC1225, IC1226, IC1227, IC1224, IC1198, IC1199, IC1194, IC1196, IC1197, IC1218, IC1219, IC1192, IC1193, IC1161, IC1187, IC1188, IC1189, IC1190, IC1191, IC1155, IC1183, IC1184, IC1185, IC1186, IC1179, IC1180, IC1181, IC1182, IC1160, IC1176, IC1177, IC1178, IC1168, IC1169, IC1170, IC1171, IC1172, IC1173, IC1174, IC1175, IC1162, IC1163, IC1164, IC1165, IC1166, IC1167, IC1156, IC1159, IC1157, IC1158, IC1153, IC1154, IC1151, IC1152, IC1150, IC1145, IC1146, IC1147, IC1148, IC1149, IC1121, IC1123, IC1122, IC1143, IC1144, IC1139, IC1140, IC1141, IC1142, IC1134, IC1135, IC1136, IC1137, IC1138, IC1124, IC1132, IC1133, IC1110, IC1111, IC1112, IC1114, IC1115, IC1116, IC1117, IC1119, IC1120, IC1102, IC1103, IC1104, IC1105, IC1106, IC1108, IC1089, IC1093, IC1094, IC1095, IC1097, IC1099, IC1100, IC1096, IC1098, IC1081, IC1082, IC1083, IC1084, IC1085, IC1086, IC1087, IC1088, IC1080, IC1077, IC1078, IC1079, IC1076, IC1058, IC1075, IC1068, IC1069, IC1070, IC1071, IC1072, IC1073, IC1074, IC1064, IC1065, IC1066, IC1067, IC1050, IC1060, IC1061, IC1062, IC1047, IC1051, IC1030, IC1056, IC1057, IC1040, IC1052, IC1053, IC1055, IC1037, IC1038, IC1039, IC1041, IC1042, IC1043, IC1044, IC1046, IC1049, IC1036, IC1031, IC1032, IC1033, IC1035, IC1026, IC1027, IC1028, IC1029, IC1024, IC1025, IC1019, IC1017, IC1015, IC1014, IC1009, IC1007, IC1008, IC1005, IC1006, IC1001, IC1002, IC1004.
11. Product Code: 0042-0000-US-L; GTIN Number: 813502010022; Serial Numbers: IC1370, IC5021, IC6902, IC6903, IC6890, IC6891, IC6892, IC6821, IC6113, IC6114, IC6126, IC6127, IC6108, IC6109, IC6110, IC6111, IC6112, IC5913, IC5834, IC5835, IC5837, IC5782, IC5783, IC5784, IC5785, IC5786, IC5618, IC5744, IC5426, IC5427, IC5428, IC5429, IC5430, IC5431, IC5432, IC5065, IC5248, IC5249, IC5250, IC5121, IC5186, IC5187, IC5016, IC5010, IC5011, IC5012, IC5013, IC5014, IC4742, IC4743, IC4745, IC4746, IC4747, IC4592, IC4586, IC4587, IC4589, IC4590, IC4553, IC4338, IC4296, IC3905, IC3757, IC3758, IC3759, IC3760, IC3761, IC3762, IC3763, IC3764, IC3765, IC3767, IC3692, IC3678, IC3655, IC3676, IC3677, IC3639, IC3593, IC3613, IC3614, IC3615, IC3611, IC3612, IC3619, IC3620, IC3621, IC3606, IC3607, IC3608, IC3616, IC3617, IC3618, IC3598, IC3605, IC3597, IC3584, IC3585, IC3586, IC3587, IC3588, IC3392, IC3559, IC3560, IC3442, IC3430, IC3431, IC3344, IC3402, IC3395, IC3389, IC3387, IC3391, IC3256, IC2967, IC2968, IC2911, IC2912, IC2903, IC2904, IC2905, IC2906, IC2894, IC2900, IC2902, IC2750, IC2683, IC2685, IC2688, IC2673, IC2455, IC2346, IC2156, IC2143, IC2144, IC2145, IC2123, IC1856, IC1796, IC1745, IC1706, IC1667, IC1647, IC1649, IC1611, IC1468, IC1447, IC1318, IC1323, IC1281, IC1282, IC1266, IC1267, IC1268, IC1260, IC1248, IC1195, IC1125, IC1118, IC1090, IC1091, IC1063, IC1048, IC1045, IC1016, IC1010, IC1012, IC1013.
12. Product Code: 0042-0010-AU; GTIN Number: 813502010930; Serial Numbers: IC5130, IC5131.
13. Product Code: 0042-0010-EU; GTIN Number: 813502010961; Serial Numbers: IC12346, IC11742, IC11747, IC11741, IC11745, IC11748, IC11749, IC11162, IC11485, IC11498, IC11451, IC11491, IC11497, IC11500, IC9996, IC9998, IC10023, IC10024, IC9961, IC9972, IC9973, IC9974, IC9801, IC9865, IC9866, IC9867, IC9834, IC9835, IC9832, IC9833, IC9804, IC9805, IC9783, IC9784, IC9796, IC9799, IC9501, IC8476A, IC9181A, IC8875, IC9299, IC9320, IC9349, IC9356, IC9372, IC9383, IC9230, IC9363, IC9364, IC9365, IC9366, IC9367, IC9368, IC9369, IC9370, IC9371, IC9005A, IC9006A, IC9031, IC9085, IC9090, IC9091, IC9092, IC9093, IC9094, IC9095, IC9096, IC9025, IC9027, IC9028, IC9008, IC8861, IC8863, IC8864, IC8854, IC8855, IC8856, IC8857, IC8858, IC8859, IC8860, IC8655, IC8714, IC8775, IC8812, IC8813, IC8755, IC8751, IC8752, IC8753, IC8754, IC8761, IC8762, IC8763, IC8764, IC8765, IC8644, IC8696, IC8723, IC8725, IC8726, IC8727, IC8729, IC8737, IC8739, IC8748, IC8750, IC8703, IC8704, IC8708, IC8683, IC8690, IC8691, IC8692, IC8693, IC8661, IC8662, IC8663, IC8664, IC8665, IC8639, IC8640, IC8641, IC8642, IC8643, IC8645, IC8646, IC8647, IC8648, IC8649, IC8629, IC8630, IC8560, IC8591, IC8596, IC8600, IC8608, IC8610, IC8621, IC8622, IC8623, IC8625, IC8584, IC8586, IC8588, IC8589, IC8590, IC8592, IC8488, IC8494, IC8495, IC8499, IC8503, IC8511, IC8512, IC8513, IC8514, IC8486, IC8439, IC8457, IC8482, IC8487, IC8441, IC8442, IC8443, IC8444, IC8445, IC8447, IC8448, IC8449, IC8450, IC8451, IC8419, IC8420, IC8421, IC8422, IC8423, IC8424, IC8425, IC8426, IC8427, IC8428, IC8387, IC8388, IC8395, IC8396, IC8397, IC8400, IC8401, IC8402, IC8403, IC8404, IC8405, IC6277, IC8336, IC8337, IC8338, IC8339, IC8340, IC8341, IC8342, IC8343, IC8344, IC8345, IC8252, IC8253, IC8254, IC8255, IC8256, IC8257, IC8258, IC8259, IC8260, IC8261, IC8262, IC8263, IC8217, IC8218, IC8219, IC8220, IC8221, IC8222, IC8225, IC8226, IC8201, IC8202, IC8203, IC8204, IC8205, IC8206, IC8157, IC8158, IC8159, IC8160, IC8161, IC8164, IC8165, IC8162, IC8145, IC8146, IC8147, IC8032, IC8128, IC8130, IC8131, IC8132, IC8133, IC8134, IC8126, IC8117, IC8118, IC8119, IC8120, IC8121, IC8122, IC8123, IC8124, IC8125, IC8067, IC8068, IC8069, IC8070, IC8071, IC8072, IC8073, IC8074, IC8075, IC8056, IC8057, IC8058, IC8059, IC8060, IC8062, IC8063, IC8064, IC8065, IC8066, IC8011, IC8012, IC8013, IC7920, IC7964, IC7965, IC8061, IC7953, IC7954, IC7955, IC7956, IC7957, IC7958, IC7937, IC7939, IC7940, IC7941, IC7942, IC7943, IC7944, IC7947, IC7860, IC7861, IC7862, IC7863, IC7864, IC7865, IC7866, IC7867, IC7868, IC7869, IC7821, IC7822, IC7823, IC7824, IC7825, IC7826, IC7827, IC7828, IC7829, IC7830, IC7792, IC7793, IC7794, IC7795, IC7796, IC7797, IC7798, IC7799, IC7800, IC7801, IC7762, IC7763, IC7764, IC7765, IC7738, IC7739, IC7740, IC7741, IC7710, IC7711, IC7636, IC7637, IC7638, IC7639, IC7640, IC7620, IC7621, IC7622, IC7623, IC7624, IC7625, IC7626, IC7627, IC7628, IC7629, IC7523, IC7524, IC7525, IC7526, IC7528, IC7529, IC7589, IC7590, IC7588, IC7460, IC7586, IC7587, IC7464, IC7465, IC7458, IC7459, IC7461, IC7462, IC7463, IC7358, IC7359, IC7360, IC7361, IC7362, IC7363, IC7364, IC7456, IC7457, IC7268, IC7269, IC7270, IC7271, IC7272, IC7273, IC7274, IC7275, IC7356, IC7357, IC7234, IC7235, IC7266, IC7267, IC7232, IC7226, IC7227, IC7233, IC7156, IC7157, IC7158, IC7159, IC7160, IC7161, IC7054, IC7055, IC6994, IC7047, IC7048, IC7049, IC7050, IC7051, IC7052, IC7053, IC6946, IC6950, IC6951, IC6952, IC6954, IC6955, IC6986, IC6987, IC6988, IC6944, IC6945, IC6947, IC6948, IC6949, IC6703, IC6872, IC6873, IC6874, IC6875, IC6936, IC6937, IC6938, IC6939, IC6940, IC6941, IC6942, IC6758, IC6868, IC6869, IC6870, IC6871, IC6788, IC6789, IC6790, IC6791, IC6792, IC6793, IC6794, IC6795, IC6866, IC6867, IC6776, IC6777, IC6779, IC6780, IC6781, IC6784, IC6785, IC6459, IC6460, IC6461, IC6462, IC6463, IC6465, IC6698, IC6383, IC6384, IC6416, IC6417, IC6418, IC6424, IC6421, IC6422, IC6344, IC6345, IC6376, IC6377, IC6378, IC6379, IC6304, IC6305, IC6336, IC6337, IC6338, IC6339, IC6340, IC6341, IC6342, IC6343, IC6301, IC6299, IC6302, IC6303, IC6297, IC6298, IC6265, IC6276, IC6278, IC6279, IC6280, IC6281, IC6282, IC6283, IC6070, IC6235, IC6256, IC6257, IC6258, IC6259, IC6260, IC6261, IC6263, IC6229, IC6230, IC6231, IC6232, IC6234, IC6121, IC6122, IC6123, IC6166, IC6167, IC6168, IC6169, IC6175, IC6226, IC6171, IC6172, IC6173, IC6174, IC6120, IC6124, IC6075, IC6116, IC6117, IC6118, IC6119, IC6073, IC6074, IC6019, IC6020, IC6021, IC6022, IC6023, IC6024, IC6025, IC6067, IC6068, IC6069, IC6012, IC6014, IC6015, IC6016, IC6017, IC6013, IC6018, IC5923, IC5924, IC5963, IC5964, IC5965, IC5966, IC5967, IC5968, IC5902, IC5903, IC5905, IC5916, IC5918, IC5919, IC5920, IC5921, IC5922, IC5925, IC5801, IC5904, IC5917, IC5814, IC5896, IC5897, IC5899, IC5900, IC5898, IC5811, IC5812, IC5813, IC5808, IC5809, IC5810, IC5802, IC5803, IC5804, IC5805, IC5806, IC5807, IC5671, IC5672, IC5644, IC5645, IC5668, IC5669, IC5666, IC5667, IC5638, IC5639, IC5640, IC5641, IC5642, IC5643, IC5595, IC5636, IC5635, IC5533, IC5567, IC5628, IC5629, IC5630, IC5631, IC5632, IC5633, IC5634, IC5588, IC5589, IC5590, IC5591, IC5594, IC5627, IC5578, IC5579, IC5580, IC5581, IC5582, IC5583, IC5584, IC5585, IC5586, IC5587, IC5626, IC5569, IC5570, IC5571, IC5572, IC5573, IC5574, IC5575, IC5577, IC5566, IC5568, IC5526, IC5532, IC5534, IC5535, IC5576, IC5350, IC5414, IC5415, IC5527, IC5528, IC5529, IC5530, IC5531, IC5410, IC5411, IC5412, IC5413, IC5383, IC5407, IC5408, IC5401, IC5402, IC5403, IC5404, IC5396, IC5397, IC5398, IC5399, IC5400, IC5355, IC5376, IC5377, IC5378, IC5283, IC5284, IC5285, IC5282, IC5235, IC5276, IC5277, IC5278, IC5279, IC5280, IC5281, IC5233, IC5234, IC5082, IC5157, IC5161, IC5162, IC5163, IC5164, IC5140, IC5141, IC5142, IC5143, IC5144, IC5145, IC5156, IC5158, IC5159, IC5160, IC5132, IC5133, IC5134, IC5135, IC5136, IC5137, IC5138, IC5139, IC5114, IC5115, IC5126, IC5127, IC5128, IC5129, IC5098, IC5101, IC5102, IC5103, IC5104, IC5105, IC5106, IC5107, IC5108, IC5109, IC5110, IC5111, IC5112, IC5113, IC5084, IC5085, IC5096, IC5097, IC5099, IC5100, IC5076, IC5077, IC5078, IC5079, IC5080, IC5081, IC5039, IC5040, IC5041, IC5042, IC5044, IC5045, IC5037, IC5038, IC4957, IC4958, IC4959, IC4960, IC4961, IC4946, IC4947, IC4948, IC4949, IC4950, IC4951, IC4952, IC4953, IC4954, IC4955, IC4651, IC4652, IC4653, IC9000, IC9001, IC9002, IC9003, IC9004, IC9005, IC9385, IC9183.
14. Product Code: 0042-0010-EU-L; GTIN Number: 813502010961; Serial Numbers: IC12317, IC12333, IC12335, IC12336, IC12337, IC12338, IC12339, IC12340, IC12341, IC12342, IC12343, IC12345, IC12321, IC12322, IC12323, IC12324, IC12325, IC11740, IC11744, IC11752, IC11728, IC11729, IC11730, IC11731, IC11732, IC11733, IC11734, IC11735, IC11737, IC11739, IC11655, IC11663, IC11664, IC11667, IC11668, IC11669, IC11671, IC11681, IC11691, IC11698, IC11067, IC11317, IC11336, IC11341, IC11342, IC11344, IC11345, IC11346, IC10177, IC10178, IC10179, IC10182, IC10145, IC10032, IC10033, IC10034, IC10035, IC10037, IC9798, IC9800, IC9802, IC9803, IC9400, IC9432, IC9433, IC9434, IC9435, IC9436, IC9437, IC9438, IC9439, IC9440, IC9358, IC9386, IC9387, IC9388, IC9390, IC9391, IC9392, IC9393, IC9394, IC9395, IC8920, IC8929, IC8930, IC8931, IC8932, IC8933, IC8814, IC8815, IC8816, IC8817, IC8504, IC8554, IC8575, IC8578, IC8579, IC8580, IC8581, IC8582, IC8576, IC8577, IC7945, IC7946, IC7365, IC7228, IC7229, IC7230, IC7231, IC7162, IC7163, IC7164, IC7165, IC6995, IC7046, IC6953, IC6943, IC6778, IC6782, IC6783, IC6464, IC6696, IC6697, IC6419, IC6420, IC6300, IC6264, IC6262, IC6170, IC5673, IC5675, IC5796, IC5674, IC5592, IC5593, IC5409, IC5352.
15. Product Code: 0042-0010-IN; GTIN Number: 813502011425; Serial Numbers: IC12407, IC12408, IC12409, IC10366, IC10367, IC10341, IC10238, IC10239, IC10149, IC10150, IC10063, IC10064, IC9969, IC9970, IC9908, IC9921, IC9744, IC9745, IC9746, IC9557, IC9558, IC9324, IC9328, IC9177, IC9178, IC8609, IC8980.
16. Product Code: 0042-0010-UK; GTIN Number: 813502010978; Serial Numbers: IC12505, IC12506, IC12507, IC11440, IC11441, IC11442, IC11444, IC11445, IC11448, IC10808, IC11048, IC11071, IC11183, IC11186, IC11190, IC11201, IC11202, IC11211, IC11086, IC11019, IC11023, IC11026, IC11027, IC11028, IC10646, IC10648, IC10649, IC10640, IC10641, IC10645, IC10319, IC10320, IC10039, IC10040, IC10043, IC10044, IC10012, IC10013, IC10014, IC10015, IC9599, IC9441, IC9442, IC9443, IC9444, IC9445, IC9446, IC9447, IC9448, IC9449, IC9450, IC9133, IC9163, IC8895, IC8767, IC8768, IC8774, IC8561, IC8563, IC8593, IC8594, IC8541, IC8542, IC8543, IC8544, IC8308, IC8315, IC8316, IC8335, IC8331, IC8332, IC8333, IC8334, IC8297, IC8295, IC8127, IC8129, IC8113, IC8114, IC8115, IC8116, IC8101, IC8102, IC8103, IC8104, IC8105, IC8106.
17. Product Code: 0042-0010-UK-L; GTIN Number: 813502010978; Serial Numbers: IC9018, IC8294, IC8296.
18. Product Code: 0042-0010-US; GTIN Number: 813502010985; Serial Numbers: IC12758, IC12414, IC12415, IC12417, IC12418, IC12419, IC12316, IC12214, IC12279, IC11656, IC11670, IC11674, IC11677, IC11685, IC11174, IC11299, IC11410, IC10979, IC10983, IC10984, IC10985, IC10463, IC10587, IC10589, IC10318, IC10278, IC10302, IC10102, IC10103, IC10049, IC10066, IC9989, IC9797, IC9915, IC9919, IC9922, IC9777, IC9793, IC9773, IC9780, IC9781, IC9782, IC9610, IC9622, IC9648, IC9581, IC9500, IC9348, IC9228, IC9229, IC8037, IC8040, IC8041, IC8026, IC8029, IC8034, IC8035, IC8002, IC8003, IC8004, IC8006, IC8008, IC7974, IC7890, IC7893, IC7894, IC7906, IC7908, IC7877, IC7878, IC7879, IC7882, IC7883, IC7884, IC7880, IC7775, IC7802, IC7803, IC7804, IC7805, IC7816, IC7817, IC7818, IC7819, IC7820, IC7771, IC7772, IC7773, IC7774, IC7786, IC7788, IC7789, IC7790, IC7791, IC7787, IC7742, IC7743, IC7744, IC7745, IC7756, IC7757, IC7758, IC7759, IC7760, IC7761, IC7728, IC7729, IC7730, IC7731, IC7732, IC7733, IC7734, IC7735, IC7736, IC7737, IC7712, IC7713, IC7714, IC7715, IC7726, IC7727, IC7604, IC7699, IC7700, IC7701, IC7702, IC7703, IC7704, IC7705, IC7706, IC7707, IC7658, IC7659, IC7660, IC7661, IC7662, IC7663, IC7664, IC7665, IC7676, IC7678, IC7679, IC7677, IC7599, IC7600, IC7601, IC7602, IC7603, IC7605, IC7616, IC7617, IC7618, IC7619, IC7544, IC7547, IC7548, IC7549, IC7558, IC7561, IC7564, IC7570, IC7573, IC7574, IC7545, IC7550, IC7551, IC7552, IC7553, IC7554, IC7555, IC7556, IC7557, IC7559, IC7534, IC7535, IC7536, IC7537, IC7538, IC7539, IC7540, IC7541, IC7542, IC7543, IC7509, IC7510, IC7511, IC7512, IC7513, IC7514, IC7515, IC7516, IC7517, IC7518, IC7501, IC7502, IC7503, IC7504, IC7505, IC7506, IC7507, IC7508, IC7491, IC7492, IC7493, IC7494, IC7495, IC7496, IC7497, IC7498, IC7499, IC7500, IC7484, IC7485, IC7486, IC7487, IC7488, IC7489, IC7490, IC7474, IC7476, IC7477, IC7478, IC7479, IC7480, IC7481, IC7482, IC7483, IC7404, IC7405, IC7466, IC7467, IC7468, IC7469, IC7470, IC7471, IC7472, IC7473, IC7374, IC7375, IC7396, IC7397, IC7398, IC7399, IC7400, IC7401, IC7402, IC7403, IC7344, IC7345, IC7346, IC7347, IC7348, IC7349, IC7350, IC7351, IC7352, IC7353, IC7354, IC7355, IC7366, IC7367, IC7368, IC7369, IC7370, IC7371, IC7372, IC7373, IC7329, IC7330, IC7331, IC7332, IC7333, IC7334, IC7335, IC7336, IC7337, IC7338, IC7243, IC7244, IC7245, IC7246, IC7305, IC7324, IC7325, IC7326, IC7327, IC7328, IC7313, IC7315, IC7316, IC7317, IC7318, IC7319, IC7320, IC7321, IC7322, IC7298, IC7299, IC7300, IC7301, IC7302, IC7303, IC7304, IC7306, IC7307, IC7308, IC7247, IC7249, IC7250, IC7251, IC7252, IC7253, IC7254, IC7255, IC7276, IC7277, IC7278, IC7279, IC7280, IC7281, IC7282, IC7283, IC7284, IC7285, IC7296, IC7297, IC7222, IC7223, IC7224, IC7225, IC7236, IC7237, IC7238, IC7240, IC7239, IC7241, IC7208, IC7209, IC7210, IC7211, IC7212, IC7215, IC7216, IC7217, IC7213, IC7214, IC7192, IC7193, IC7194, IC7195, IC7196, IC7197, IC7198, IC7199, IC7200, IC7201, IC7202, IC7203, IC7204, IC7205, IC7206, IC7170, IC7171, IC7172, IC7173, IC7174, IC7175, IC7176, IC7177, IC7178, IC7179, IC7150, IC7151, IC7152, IC7153, IC7154, IC7155, IC7166, IC7167, IC7168, IC7169, IC7135, IC7136, IC7137, IC7138, IC7139, IC7038, IC7114, IC7117, IC7125, IC7127, IC7128, IC7129, IC7130, IC7131, IC7132, IC7123, IC7109, IC7110, IC7111, IC7112, IC7113, IC7115, IC7116, IC7118, IC7119, IC7120, IC6996, IC7012, IC7016, IC7021, IC7022, IC7023, IC7024, IC7025, IC7037, IC7036, IC7013, IC7014, IC7015, IC7017, IC7018, IC7019, IC7020, IC7009, IC7010, IC7011, IC6998, IC6999, IC7000, IC7001, IC7004, IC7005, IC7006, IC7007, IC7002, IC7003, IC6972, IC6973, IC6974, IC6975, IC6997, IC6956, IC6958, IC6959, IC6960, IC6961, IC6963, IC6964, IC6965, IC6966, IC6967, IC6968, IC6969, IC6970, IC6971, IC6962, IC6957, IC5971.
19. Product Code: 0042-0010-US-L; GTIN Number: 813502010985; Serial Numbers: IC7530, IC6699, IC6702, IC5959, IC8031, IC7996, IC7966, IC7967, IC7968, IC7969, IC7970, IC7971, IC7972, IC7973, IC7975, IC7910, IC7885, IC7886, IC7887, IC7888, IC7889, IC7475, IC7314, IC7248, IC7309, IC7310, IC7311, IC7312, IC7207, IC7218, IC7219, IC7220, IC7221, IC7140, IC7141, IC7142, IC7143, IC7144, IC7145, IC7146, IC7147, IC7148, IC7149, IC7121, IC7122, IC7124, IC7126, IC7039, IC7040, IC7041, IC7042, IC7043, IC7044, IC7045, IC7106, IC7107, IC7108, IC7008, IC6705, IC6756, IC6757, IC6759, IC6760, IC5969, IC5970, IC5972, IC5973, IC5974, IC5975, IC6006, IC6007, IC6008.
20. Product Code: 0042-0040-AU; GTIN Number: 813502012972; Serial Numbers: IC12583, IC12584, IC12008, IC12054, IC11472, IC11487, IC11489, IC11490, IC11544, IC11546, IC11566, IC10747, IC10748, IC10387, IC10388, IC10162, IC10163, IC12674, IC12675, IC12151.
21. Product Code: 0042-0040-CA; GTIN Number: 813502012958; Serial Numbers: IC12604, IC11171, IC11172, IC10306, IC10312.
22. Product Code: 0042-0040-CT-L; GTIN Number: 813502012958; Serial Numbers: IC8505, IC8545, IC8186, IC8187, IC8188, IC7645A, IC7657A.
23. Product Code: 0042-0040-EU; GTIN Number: 813502012217; Serial Numbers: IC12541, IC12574, IC12575, IC12586, IC12593, IC12594, IC12595, IC12596, IC12585, IC12590, IC12591, IC12597, IC12523, IC12550, IC12558, IC12567, IC12568, IC12556, IC12555, IC12565, IC12566, IC12554, IC12524, IC12542, IC12543, IC12544, IC12545, IC12546, IC12548, IC12551, IC12552, IC12553, IC12470, IC12486, IC12487, IC12488, IC12490, IC12491, IC12493, IC12494, IC12495, IC12497, IC12480, IC12481, IC12483, IC12484, IC12485, IC12455, IC12456, IC12457, IC12458, IC12459, IC12463, IC12464, IC12465, IC12466, IC12467, IC12442, IC12443, IC12444, IC12445, IC12446, IC12447, IC12451, IC12452, IC12453, IC12454, IC12307, IC12308, IC12309, IC12310, IC12311, IC12312, IC12313, IC12314, IC12320, IC12278, IC12280, IC12281, IC12282, IC12284, IC12287, IC12288, IC12289, IC12290, IC12291, IC12267, IC12269, IC12270, IC12271, IC12272, IC12273, IC12274, IC12275, IC12276, IC12277, IC12257, IC12258, IC12259, IC12260, IC12261, IC12262, IC12263, IC12264, IC12265, IC12266, IC12012, IC12013, IC12014, IC12015, IC12219, IC12220, IC12221, IC12222, IC12223, IC12224, IC12195, IC12206, IC12208, IC12210, IC12212, IC12213, IC12215, IC12216, IC12217, IC12218, IC11991, IC11992, IC11993, IC11995, IC11997, IC11998, IC11999, IC12000, IC12001, IC12002, IC11977, IC11978, IC11980, IC11981, IC11983, IC11984, IC11985, IC11986, IC11987, IC11988, IC11967, IC11968, IC11969, IC11970, IC11971, IC11973, IC11974, IC11975, IC11976, IC11989, IC11934, IC11952, IC11959, IC11960, IC11961, IC11962, IC11963, IC11964, IC11965, IC11966, IC11948, IC11949, IC11950, IC11951, IC11953, IC11954, IC11955, IC11956, IC11957, IC11958, IC11928, IC11929, IC11930, IC11931, IC11937, IC11942, IC11943, IC11945, IC11946, IC11947, IC11878, IC11879, IC11900, IC11904, IC11909, IC11910, IC11911, IC11913, IC11914, IC11915, IC11713, IC11714, IC11715, IC11716, IC11717, IC11718, IC11719, IC11721, IC11722, IC11726, IC11679, IC11699, IC11705, IC11706, IC11707, IC11708, IC11709, IC11710, IC11711, IC11712, IC11149, IC11283, IC11358, IC11360, IC11361, IC11363, IC11364, IC11367, IC11368, IC11366, IC11257, IC11258, IC11262, IC11264, IC11265, IC11266, IC11277, IC11278, IC11279, IC11280, IC11281, IC11282, IC11286, IC11287, IC11288, IC11289, IC11321, IC11322, IC11323, IC11324, IC11326, IC11328, IC11330, IC11331, IC11332, IC11334, IC11302, IC11223, IC11224, IC11225, IC11226, IC11228, IC11233, IC11234, IC11235, IC11229, IC11240, IC11090, IC11193, IC11209, IC11214, IC11215, IC11217, IC11219, IC11220, IC11221, IC10799, IC11072, IC11132, IC11136, IC11137, IC11138, IC11140, IC11141, IC11142, IC11146, IC11095, IC11116, IC11119, IC11120, IC11122, IC11123, IC11124, IC11125, IC11074, IC11121, IC10809, IC10909, IC11003, IC11004, IC11005, IC11006, IC11007, IC11009, IC11010, IC11011, IC10858, IC10870, IC10946, IC10947, IC10949, IC10950, IC10951, IC10953, IC10954, IC10955, IC10898, IC10943, IC10944, IC10945, IC10952, IC10938, IC10939, IC10940, IC10942, IC10948, IC10916, IC10917, IC10919, IC10920, IC10921, IC10887, IC10889, IC10890, IC10891, IC10892, IC10883, IC10884, IC10885, IC10886, IC10882, IC10828, IC10835, IC10798, IC10767, IC10768, IC10819, IC10820, IC10580, IC10629, IC10720, IC10738, IC10778, IC10779, IC10781, IC10782, IC10721, IC10772, IC10773, IC10775, IC10698, IC10706, IC10707, IC10710, IC10711, IC10713, IC10665, IC10670, IC10679, IC10681, IC10682, IC10683, IC10671, IC10673, IC10674, IC10675, IC10676, IC10677, IC10678, IC10680, IC10621, IC10635, IC10636, IC10638, IC10639, IC10627, IC10628, IC10632, IC10633, IC10634, IC10544, IC10551, IC10555, IC10559, IC10416, IC10468, IC10590, IC10591, IC10556, IC10558, IC10560, IC10552, IC10540, IC10541, IC10542, IC10543, IC10402, IC10505, IC10506, IC10507, IC10508, IC10510, IC10415, IC10424, IC10408, IC10425, IC10329, IC10423, IC10433, IC10438, IC10439, IC10440, IC10441, IC10442, IC10382, IC10419, IC10434, IC10437, IC10401, IC10403, IC10404, IC10389, IC10392, IC10393, IC10394, IC10395, IC10396, IC10397, IC10398, IC10400, IC10371, IC10372, IC10326, IC10347, IC10360, IC10369, IC10370, IC10256, IC10257, IC10258, IC10259, IC10260, IC10261, IC10262, IC10263, IC10264, IC10265, IC10180, IC10181, IC10183, IC10184, IC10185, IC10186, IC10166, IC10167, IC10168, IC10169, IC10171, IC10172, IC10173, IC10174, IC10175, IC10176, IC10152, IC10153, IC10154, IC10155, IC10156, IC10157, IC10158, IC10159, IC10160, IC10161, IC10138, IC10139, IC10140, IC10141, IC10142, IC10143, IC10144, IC10146, IC10148, IC10123, IC10124, IC10125, IC10126, IC10127, IC9949, IC9950, IC9951, IC9955, IC9957, IC9958, IC9954, IC9956, IC9959, IC9960, IC9962, IC9964, IC9965, IC9966, IC9967, IC9968, IC9942, IC9944, IC9945, IC9946, IC9948, IC9868, IC9896, IC9761, IC9762, IC9763, IC9764, IC9765, IC9766, IC9824, IC9827, IC9828, IC9830, IC9757, IC9758, IC9759, IC9760, IC9726, IC9727, IC9728, IC9729, IC9730, IC9687, IC9689, IC9690, IC9692, IC9693, IC9555, IC9554, IC9556, IC9623, IC9624, IC9626, IC9644, IC9645, IC9646, IC9647, IC9643, IC9281, IC9420, IC9424, IC9425, IC9426, IC9427, IC9428, IC9430, IC9431, IC9552, IC9553, IC9429, IC9352, IC9389, IC9396, IC9398, IC9399, IC9401, IC9402, IC9403, IC9404, IC9330, IC9342, IC9343, IC9345, IC9251, IC9252, IC9253, IC9254, IC9255, IC9256, IC9257, IC9258, IC9259, IC9260, IC9215, IC9216, IC9217, IC9189, IC9190, IC9191, IC9192, IC9193, IC9194, IC9167, IC9168, IC9169, IC9170, IC9171, IC9172, IC9173, IC9174, IC9175, IC9176, IC9179, IC9180, IC9106, IC9107, IC9108, IC9109, IC9110, IC9111, IC9112, IC9113, IC9114, IC9115, IC9071, IC9072, IC9073, IC9075, IC9076, IC9077, IC9078, IC9039, IC9070, IC9050, IC9051, IC9052, IC9069, IC9000A, IC9001A, IC9002A, IC9003A, IC9004A, IC8840, IC9015, IC9016, IC9017, IC9019, IC9020, IC9021, IC9023, IC9024, IC9014, IC9022, IC9009, IC9010, IC9011, IC9012, IC9013, IC8910, IC8911, IC8912, IC8913, IC8914, IC8898, IC8899, IC8900, IC8901, IC8902, IC8865, IC8866, IC8868, IC8869, IC8870, IC8871, IC8872, IC8873, IC8874, IC8876, IC8837, IC8838, IC8841, IC8842, IC8836, IC8839, IC8818, IC8819, IC8820, IC8821, IC8822, IC8823, IC8824, IC8825, IC8826, IC8827, IC12319.
24. Product Code: 0042-0040-EU-L; GTIN Number: 813502012217; Serial Numbers: IC11259, IC11260, IC11261, IC11263, IC11300, IC11301, IC11303, IC11304, IC11305, IC11306, IC11307, IC11308, IC11309, IC10630, IC10829, IC11126, IC11127, IC11128, IC11130, IC11131, IC11133, IC11134, IC11135, IC9397, IC9105.
25. Product Code: 0042-0040-JP; GTIN Number: 813502011814; Serial Numbers: IC12612, IC12609, IC12610, IC12613, IC12614, IC12515, IC12516, IC12517, IC12518, IC12434, IC12437, IC12438, IC11499, IC12026, IC12028, IC12029, IC12030, IC12031, IC11777, IC11778, IC11672, IC11720, IC11725, IC11727, IC11736, IC11738, IC11743, IC11750, IC11751, IC10853, IC11015, IC11088, IC11562, IC11570, IC11623, IC11636, IC11666, IC11678, IC11687, IC11688, IC11689, IC11690, IC11693, IC11694, IC11453, IC11454, IC11459, IC11405, IC11413, IC11414, IC11416, IC11417, IC11329, IC11343, IC11369, IC11370, IC11376, IC11377, IC11378, IC11379, IC11267, IC11325, IC11337, IC11338, IC11339, IC11340, IC11252, IC11253, IC11254, IC11255, IC11256, IC11192, IC11195, IC11213, IC11216, IC11218, IC10928, IC10930, IC10931, IC10932, IC10933, IC10869, IC10903, IC10904, IC10906, IC10907, IC10806, IC10807, IC10810, IC10290, IC10803, IC10804, IC10805, IC10686, IC10699, IC10700, IC10701, IC10702, IC10705, IC10473, IC10563, IC10583, IC10536, IC10537, IC10538, IC10539, IC10411, IC10499, IC10503, IC10504, IC10409, IC10413, IC10426, IC10427, IC10405, IC10406, IC10414, IC10417, IC10363, IC10368, IC10354, IC10355, IC10356, IC10357, IC10359, IC10267, IC10268, IC10270, IC10272, IC10273, IC10274, IC10275, IC10276, IC10280, IC10281, IC10068, IC10105, IC10106, IC10107, IC10108, IC10109, IC10119, IC10120, IC10121, IC10122, IC10016, IC10017, IC10018, IC10019, IC10020, IC10021, IC10022, IC9892, IC9894, IC9923, IC9925, IC9926, IC9927, IC9928, IC9929, IC9930, IC9931, IC9811, IC9838, IC9785, IC9786, IC9787, IC9788, IC9789, IC9790, IC9791, IC9694, IC9702, IC9703, IC9704, IC9705, IC9706, IC9707, IC9708, IC9709, IC9494, IC9462, IC9463, IC9485, IC9540, IC9559, IC9560, IC9561, IC9562, IC9563, IC9580, IC9411, IC9489, IC9490, IC9491, IC9492, IC9495, IC9496, IC9497, IC9498, IC9499, IC9292, IC9321, IC9322, IC9323, IC9325, IC9326, IC9329, IC9331, IC9332, IC8937, IC9288, IC9289, IC9290, IC9291, IC9294, IC9295, IC9296, IC9297, IC9298, IC9125, IC9126, IC9127, IC9128, IC9129, IC9130, IC9131, IC9132, IC9134, IC9135, IC9136, IC9137, IC9138, IC9139, IC9140, IC8915, IC8916, IC8917, IC8918, IC8919, IC8921, IC8922, IC8923, IC8924, IC8927, IC8886, IC8887, IC8893, IC8894, IC8896, IC8897, IC8779, IC8780, IC8781, IC8782, IC8784, IC8785, IC8736, IC8731, IC8732, IC8733, IC8734, IC8735, IC8666, IC8670, IC8679, IC8680, IC8681, IC8682, IC8684, IC8685, IC8686, IC8687, IC8369, IC8583, IC8595, IC8612, IC8614, IC8619, IC8626, IC8531, IC8559, IC8558, IC8571, IC8573, IC8585, IC8597, IC8599, IC8601, IC8523, IC8524, IC8525, IC8526, IC8527, IC8489, IC8490, IC8491, IC8492, IC8493, IC8370, IC8371, IC8372, IC8374, IC8375, IC8376, IC8377, IC8364, IC8311, IC8298, IC8299, IC8300, IC8301, IC8302, IC8303, IC8304, IC8306, IC8309, IC8310, IC8264, IC8245, IC8246, IC8237, IC8238, IC8239, IC8240, IC8242, IC8243, IC8244, IC8196, IC8197, IC8889, IC8890, IC8892, IC8888.
26. Product Code: 0042-0040-JP-L; GTIN Number: 813502011814; Serial Numbers: IC8241.
27. Product Code: 0042-0040-UK; GTIN Number: 813502012484; Serial Numbers: IC12561, IC12563, IC12315, IC12318, IC12255, IC12256, IC11723, IC11724, IC11682, IC11683, IC11640, IC11642, IC11196, IC11197, IC11198, IC10788, IC10789, IC10790, IC10792, IC10793, IC10592, IC10593, IC10598, IC10601, IC10240, IC10241, IC10224, IC10225, IC9990, IC10008, IC10011, IC10025, IC10164, IC10165.
28. Product Code: 0042-0040-US; GTIN Number: 813502011401; Serial Numbers: IC12634, IC12641, IC12638, IC12642, IC12510, IC12511, IC12540, IC12440, IC12448, IC12449, IC12450, IC12462, IC12441, IC12461, IC12435, IC12436, IC12433, IC12475, IC12477, IC12460, IC12472, IC12473, IC12474, IC12478, IC12476, IC12469, IC12468, IC12430, IC12406, IC12426, IC12428, IC12431, IC12432, IC12396, IC12399, IC12400, IC12381, IC12382, IC12383, IC12384, IC12385, IC12386, IC12387, IC12389, IC12390, IC12391, IC12368, IC12369, IC12370, IC12371, IC12372, IC12373, IC12374, IC12375, IC12376, IC12377, IC12350, IC12358, IC12360, IC12361, IC12362, IC12363, IC12364, IC12365, IC12366, IC12367, IC12347, IC12348, IC12349, IC12351, IC12352, IC12353, IC12354, IC12355, IC12356, IC12357, IC12297, IC12298, IC12299, IC12300, IC12301, IC12302, IC12303, IC12304, IC12305, IC12306, IC12160, IC12268, IC12283, IC12285, IC12286, IC12292, IC12293, IC12294, IC12295, IC12296, IC12245, IC12246, IC12247, IC12248, IC12249, IC12250, IC12251, IC12252, IC12253, IC12254, IC12235, IC12236, IC12237, IC12238, IC12239, IC12240, IC12241, IC12242, IC12243, IC12244, IC12225, IC12226, IC12227, IC12228, IC12229, IC12230, IC12231, IC12232, IC12233, IC12234, IC12122, IC12125, IC12126, IC12128, IC12129, IC12134, IC12135, IC12140, IC12211, IC12141, IC12142, IC12143, IC12144, IC12145, IC12146, IC12150, IC12183, IC12184, IC12189, IC12104, IC12113, IC12114, IC12115, IC12116, IC12117, IC12119, IC12120, IC12121, IC12137, IC12177, IC12185, IC12188, IC12190, IC12191, IC12192, IC12193, IC12194, IC12196, IC12197, IC12198, IC12199, IC12200, IC12201, IC12202, IC12203, IC12204, IC12205, IC12207, IC12209, IC12097, IC12100, IC12130, IC12131, IC12132, IC12133, IC12136, IC12138, IC12139, IC11488, IC11932, IC11944, IC11979, IC12036, IC12080, IC12081, IC12082, IC12083, IC12085, IC12086, IC12088, IC12089, IC12091, IC12092, IC12166, IC12167, IC12172, IC12173, IC12174, IC12075, IC12078, IC12079, IC12093, IC12095, IC12096, IC12099, IC12110, IC12111, IC12152, IC12118, IC12124, IC12147, IC12148, IC12149, IC12153, IC12154, IC12155, IC12156, IC12164, IC12046, IC12058, IC12084, IC12087, IC12090, IC12157, IC12161, IC12162, IC12165, IC12175, IC12176, IC12178, IC12179, IC12180, IC12181, IC12182, IC12186, IC11563, IC11564, IC11644, IC11982, IC12038, IC12052, IC12055, IC12060, IC12061, IC12062, IC12064, IC12065, IC12066, IC12070, IC12072, IC12073, IC12074, IC12076, IC12077, IC12158, IC12159, IC12168, IC12169, IC12170, IC12171, IC12094, IC12102, IC12103, IC12105, IC12106, IC12107, IC12108, IC12109, IC12112, IC12045, IC12047, IC12048, IC12049, IC12050, IC12051, IC12053, IC12056, IC12057, IC12059, IC12033, IC12034, IC12035, IC12037, IC12039, IC12040, IC12041, IC12042, IC12044, IC12043, IC12098, IC12016, IC12017, IC12018, IC12019, IC12020, IC12021, IC12022, IC12023, IC12024, IC12025, IC11990, IC11994, IC12003, IC12004, IC12005, IC12006, IC12007, IC12009, IC12010, IC12011, IC11924, IC11925, IC11927, IC11933, IC11935, IC11936, IC11938, IC11939, IC11940, IC11941, IC11872, IC11876, IC11877, IC11917, IC11918, IC11919, IC11920, IC11921, IC11923, IC11926, IC11764, IC11776, IC11781, IC11821, IC11826, IC11862, IC11863, IC11869, IC11870, IC11811, IC11812, IC11813, IC11814, IC11825, IC11834, IC11838, IC11839, IC11840, IC11820, IC11106, IC11335, IC11560, IC11633, IC11792, IC11809, IC11810, IC11823, IC11824, IC11887, IC11889, IC11891, IC11892, IC11905, IC11908, IC11912, IC11916, IC11819, IC11516, IC11894, IC11896, IC11897, IC11898, IC11899, IC11901, IC11902, IC11903, IC11907, IC11895, IC11856, IC11857, IC11881, IC11882, IC11883, IC11884, IC11885, IC11886, IC11855, IC11880, IC11355, IC11388, IC11767, IC11773, IC11789, IC11799, IC11822, IC11828, IC11843, IC11853, IC11860, IC11861, IC11864, IC11866, IC11868, IC11871, IC11873, IC11874, IC11875, IC11795, IC11796, IC11797, IC11798, IC11816, IC11827, IC11842, IC11845, IC11888, IC11890, IC11830, IC11844, IC11846, IC11847, IC11848, IC11849, IC11850, IC11851, IC11852, IC11854, IC11817, IC11818, IC11829, IC11831, IC11832, IC11833, IC11835, IC11836, IC11837, IC11841, IC11629, IC11753, IC11762, IC11763, IC11765, IC11769, IC11770, IC11774, IC11779, IC11783, IC11784, IC11785, IC11786, IC11787, IC11788, IC11790, IC11791, IC11794, IC11793, IC11772, IC11800, IC11801, IC11802, IC11803, IC11804, IC11805, IC11806, IC11807, IC11808, IC11815, IC11754, IC11755, IC11756, IC11757, IC11758, IC11759, IC11760, IC11761, IC11766, IC11768, IC11700, IC11702, IC11704, IC11641, IC11654, IC11662, IC11680, IC11684, IC11686, IC11692, IC11695, IC11696, IC11697, IC11535, IC11610, IC11625, IC11626, IC11627, IC11628, IC11631, IC11632, IC11634, IC11518, IC11539, IC11606, IC11609, IC11621, IC11622, IC11635, IC11637, IC11638, IC11639, IC11569, IC11572, IC11581, IC11597, IC11613, IC11614, IC11615, IC11617, IC11618, IC11624, IC11333, IC11359, IC11373, IC11374, IC11375, IC11380, IC11382, IC11384, IC11386, IC11389, IC11390, IC11391, IC11396, IC11398, IC11402, IC11406, IC11409, IC11411, IC11552, IC11565, IC11540, IC11549, IC11550, IC11551, IC11555, IC11556, IC11561, IC11584, IC11585, IC11586, IC11422, IC11435, IC11438, IC11447, IC11452, IC11503, IC11522, IC11523, IC11524, IC11525, IC11526, IC11529, IC11531, IC11534, IC11536, IC11537, IC11557, IC11558, IC11559, IC11567, IC11568, IC11573, IC11574, IC11582, IC11583, IC11579, IC10757, IC11055, IC11425, IC11469, IC11473, IC11474, IC11477, IC11479, IC11492, IC11494, IC11495, IC11510, IC11511, IC11512, IC11513, IC11514, IC11519, IC11521, IC11530, IC11505, IC11420, IC11421, IC11423, IC11426, IC11427, IC11428, IC11429, IC11430, IC11431, IC11432, IC11449, IC11476, IC10832, IC11480, IC11484, IC11486, IC11493, IC11528, IC11532, IC11533, IC11538, IC11462, IC11268, IC11269, IC11270, IC11271, IC11272, IC11273, IC11274, IC11275, IC11276, IC11294, IC11241, IC11242, IC11243, IC11244, IC11245, IC11246, IC11247, IC11249, IC11251, IC11285, IC11362, IC11371, IC11381, IC11383, IC11385, IC11387, IC11392, IC11394, IC11395, IC11397, IC11408, IC11443, IC11457, IC11466, IC11471, IC11478, IC11481, IC11482, IC11284, IC11290, IC11291, IC11292, IC11293, IC11295, IC11296, IC11297, IC11310, IC11311, IC11312, IC11313, IC11314, IC11315, IC11316, IC11318, IC11319, IC11320, IC11460, IC11461, IC11464, IC11465, IC11467, IC11327, IC11419, IC11424, IC11446, IC11458, IC11347, IC11348, IC11349, IC11350, IC11351, IC11352, IC11353, IC11354, IC11356, IC11357, IC11016, IC11298, IC11403, IC11412, IC11415, IC11418, IC11433, IC11434, IC11437, IC11439, IC11024, IC11075, IC11180, IC11194, IC11200, IC11203, IC11205, IC11206, IC11207, IC11208, IC11147, IC11173, IC11175, IC11176, IC11178, IC11179, IC11184, IC11185, IC11187, IC11191, IC11139, IC11156, IC11168, IC11188, IC11189, IC11155, IC11157, IC11161, IC11164, IC11165, IC11166, IC11167, IC11169, IC11170, IC11181, IC10897, IC11021, IC11150, IC11151, IC11152, IC11154, IC11158, IC11159, IC11160, IC11163, IC10787, IC11008, IC11082, IC11117, IC11129, IC10999, IC11092, IC11093, IC11094, IC11096, IC11097, IC11103, IC11107, IC11108, IC11109, IC11001, IC11085, IC11087, IC11089, IC11091, IC11065, IC11069, IC11073, IC11076, IC11077, IC11078, IC11079, IC11080, IC11081, IC11068, IC10825, IC11022, IC11051, IC11052, IC11053, IC11054, IC11056, IC11058, IC11059, IC11060, IC11062, IC11063, IC11064, IC11066, IC11070, IC10941, IC10978, IC11041, IC11042, IC11043, IC11044, IC11045, IC11046, IC11047, IC10844, IC10895, IC10918, IC11031, IC11032, IC11033, IC11034, IC11038, IC11039, IC11040, IC10800, IC11030, IC11035, IC11036, IC11037, IC11012, IC11013, IC11014, IC11017, IC11018, IC10818, IC10830, IC10929, IC10998, IC11002, IC10971, IC10972, IC10974, IC10975, IC10957, IC10958, IC10959, IC10962, IC10961, IC10845, IC10850, IC10980, IC10981, IC10982, IC10888, IC10922, IC10935, IC10936, IC10937, IC10911, IC10912, IC10913, IC10915, IC10934, IC10923, IC10924, IC10925, IC10926, IC10927, IC10899, IC10900, IC10901, IC10908, IC10910, IC10877, IC10878, IC10879, IC10880, IC10881, IC10719, IC10875, IC10876, IC10847, IC10848, IC10852, IC10855, IC10856, IC10857, IC10859, IC10860, IC10861, IC10873, IC10774, IC10783, IC10791, IC10795, IC10796, IC10831, IC10833, IC10838, IC10839, IC10842, IC10776, IC10822, IC10823, IC10824, IC10826, IC10758, IC10770, IC10821, IC10769, IC10777, IC10811, IC10812, IC10813, IC10814, IC10815, IC10816, IC10817, IC10744, IC10754, IC10756, IC10759, IC10760, IC10763, IC10764, IC10669, IC10697, IC10750, IC10752, IC10753, IC10741, IC10742, IC10743, IC10745, IC10746, IC10749, IC10751, IC10730, IC10731, IC10732, IC10733, IC10734, IC10736, IC10737, IC10739, IC10740, IC10722, IC10723, IC10724, IC10725, IC10726, IC10728, IC10729, IC10714, IC10715, IC10716, IC10717, IC10718, IC10704, IC10708, IC10709, IC10712, IC10684, IC10685, IC10687, IC10688, IC10690, IC10691, IC10692, IC10693, IC10694, IC10695, IC10696, IC10658, IC10659, IC10660, IC10661, IC10662, IC10637, IC10647, IC10655, IC10656, IC10626, IC10653, IC10654, IC10657, IC10642, IC10643, IC10644, IC10615, IC10616, IC10617, IC10618, IC10619, IC10620, IC10622, IC10623, IC10624, IC10625, IC10610, IC10611, IC10612, IC10613, IC10614, IC10407, IC10516, IC10524, IC10548, IC10549, IC10557, IC10599, IC10603, IC10604, IC10608, IC10578, IC10605, IC10606, IC10607, IC10609, IC10321, IC10336, IC10361, IC10383, IC10436, IC10448, IC10466, IC10482, IC10533, IC10546, IC10509, IC10566, IC10596, IC10597, IC10600, IC10602, IC10567, IC10568, IC10569, IC10570, IC10571, IC10572, IC10581, IC10582, IC10561, IC10562, IC10564, IC10565, IC10545, IC10547, IC10550, IC10553, IC10554, IC10316, IC10527, IC10532, IC10534, IC10535, IC10526, IC10528, IC10529, IC10530, IC10531, IC10519, IC10521, IC10522, IC10523, IC10525, IC10461, IC10462, IC10520, IC10511, IC10512, IC10513, IC10514, IC10515, IC10480, IC10481, IC10487, IC10478, IC10479, IC10484, IC10485, IC10486, IC10380, IC10435, IC10470, IC10471, IC10472, IC10474, IC10475, IC10476, IC10477, IC10454, IC10455, IC10459, IC10449, IC10450, IC10451, IC10452, IC10453, IC10444, IC10445, IC10446, IC10447, IC10412, IC10430, IC10432, IC10443, IC10428, IC10429, IC10431, IC10410, IC10418, IC10420, IC10421, IC10422, IC10358, IC10384, IC10390, IC10391, IC10381, IC10385, IC10386, IC10377, IC10378, IC10379, IC10375, IC10376, IC10373, IC10374, IC10348, IC10349, IC10350, IC10351, IC10352, IC10342, IC10343, IC10344, IC10345, IC10346, IC10323, IC10324, IC10325, IC10327, IC10339, IC10266, IC10322, IC10328, IC10330, IC10331, IC10332, IC10333, IC10334, IC10335, IC10337, IC10340, IC10313, IC10314, IC10315, IC10317, IC10287, IC10307, IC10308, IC10309, IC10310, IC10311, IC10271, IC10279, IC10303, IC10304, IC10277, IC10282, IC10298, IC10299, IC10301, IC10254, IC10269, IC10294, IC10295, IC10296, IC10297, IC10251, IC10283, IC10284, IC10285, IC10286, IC10288, IC10289, IC10291, IC10292, IC10293, IC10228, IC10242, IC10246, IC10247, IC10248, IC10249, IC10250, IC10252, IC10253, IC10255, IC10219, IC10235, IC10232, IC10233, IC10237, IC10197, IC10208, IC10216, IC10217, IC10218, IC10223, IC10226, IC10227, IC10229, IC10230, IC10231, IC10147, IC10210, IC10220, IC10221, IC10222, IC10212, IC10213, IC10214, IC10215, IC10151, IC10203, IC10204, IC10205, IC10206, IC10207, IC10209, IC10211, IC10187, IC10188, IC10189, IC10190, IC10191, IC10192, IC10193, IC10194, IC10195, IC10196, IC10198, IC10199, IC10200, IC10201, IC10202, IC10128, IC10129, IC10130, IC10131, IC10132, IC10133, IC10134, IC10135, IC10136, IC10137, IC10104, IC10110, IC10111, IC10112, IC10113, IC10114, IC10115, IC10116, IC10117, IC10118, IC9963, IC10081, IC10099, IC10100, IC10101, IC10087, IC10088, IC10089, IC10090, IC10091, IC10092, IC10093, IC10094, IC10095, IC10096, IC10097, IC10098, IC10041, IC10077, IC10078, IC10079, IC10080, IC10082, IC10083, IC10084, IC10085, IC10086, IC9924, IC10000, IC10051, IC10059, IC10071, IC10072, IC10073, IC10074, IC10075, IC10076, IC10042, IC10065, IC10067, IC10069, IC10070, IC10036, IC10053, IC10054, IC10055, IC10056, IC10057, IC10058, IC10060, IC10061, IC10062, IC9842, IC9982, IC10031, IC10038, IC10045, IC10046, IC10047, IC10048, IC10050, IC10052, IC10026, IC10027, IC10028, IC10029, IC10030, IC9999, IC10001, IC10002, IC10003, IC10004, IC10005, IC10006, IC10007, IC10009, IC10010, IC9903, IC9976, IC9987, IC9991, IC9992, IC9993, IC9994, IC9995, IC9997, IC9493, IC9975, IC9977, IC9978, IC9979, IC9980, IC9983, IC9984, IC9985, IC9986, IC9988, IC9932, IC9933, IC9934, IC9935, IC9936, IC9937, IC9938, IC9939, IC9940, IC9941, IC9567, IC9899, IC9910, IC9912, IC9913, IC9914, IC9916, IC9917, IC9918, IC9920, IC9829, IC9885, IC9900, IC9904, IC9906, IC9909, IC9907, IC9911, IC9854, IC9870, IC9876, IC9881, IC9882, IC9883, IC9884, IC9886, IC9887, IC9888, IC9869, IC9871, IC9872, IC9873, IC9874, IC9875, IC9877, IC9878, IC9879, IC9880, IC9850, IC9851, IC9852, IC9853, IC9864, IC9840, IC9846, IC9847, IC9848, IC9849, IC9855, IC9856, IC9857, IC9858, IC9859, IC9860, IC9861, IC9862, IC9863, IC9806, IC9807, IC9808, IC9809, IC9810, IC9812, IC9823, IC9826, IC9792, IC9794, IC9795, IC9825, IC9831, IC9836, IC9837, IC9839, IC9841, IC9843, IC9844, IC9845, IC9657, IC9768, IC9769, IC9770, IC9771, IC9772, IC9774, IC9775, IC9776, IC9778, IC9779, IC9750, IC9751, IC9752, IC9753, IC9754, IC9755, IC9756, IC9741, IC9742, IC9743, IC9747, IC9748, IC9749, IC9653, IC9655, IC9739, IC9740, IC9614, IC9649, IC9731, IC9733, IC9734, IC9735, IC9736, IC9737, IC9738, IC9652, IC9685, IC9691, IC9719, IC9720, IC9721, IC9722, IC9723, IC9724, IC9725, IC9663, IC9686, IC9710, IC9711, IC9712, IC9713, IC9714, IC9715, IC9716, IC9717, IC9718, IC9697, IC9698, IC9699, IC9700, IC9701, IC9656, IC9661, IC9669, IC9673, IC9678, IC9680, IC9681, IC9682, IC9683, IC9684, IC9688, IC9668, IC9672, IC9675, IC9676, IC9677, IC9679, IC9650, IC9658, IC9659, IC9660, IC9662, IC9665, IC9666, IC9667, IC9670, IC9671, IC9617, IC9618, IC9619, IC9620, IC9621, IC9625, IC9651, IC9654, IC9596, IC9597, IC9598, IC9606, IC9607, IC9608, IC9609, IC9611, IC9612, IC9613, IC9510, IC9582, IC9583, IC9585, IC9586, IC9587, IC9588, IC9589, IC9594, IC9576, IC9577, IC9578, IC9579, IC9575, IC9591A, IC9592A, IC9590A, IC9451, IC9571, IC9572, IC9573, IC9574, IC9536, IC9537, IC9538, IC9539, IC9541, IC9542, IC9543, IC9544, IC9545, IC9546, IC9547, IC9548, IC9549, IC9550, IC9551, IC9525, IC9526, IC9527, IC9528, IC9530, IC9532, IC9533, IC9534, IC9535, IC9516, IC9517, IC9518, IC9519, IC9520, IC9521, IC9522, IC9523, IC9524, IC9511, IC9512, IC9513, IC9514, IC9515, IC9478, IC9479, IC9480, IC9481, IC9482, IC9483, IC9484, IC9486, IC9488, IC9473, IC9474, IC9475, IC9476, IC9477, IC9461, IC9465, IC9466, IC9467, IC9468, IC9469, IC9470, IC9471, IC9472, IC9327, IC9452, IC9453, IC9454, IC9455, IC9456, IC9457, IC9458, IC9460, IC9412, IC9418, IC9421, IC9422, IC9423, IC9373, IC9374, IC9375, IC9376, IC9377, IC9378, IC9379, IC9380, IC9381, IC9382, IC9405, IC9406, IC9408, IC9409, IC9410, IC9333, IC9334, IC9335, IC9337, IC9338, IC9339, IC9350, IC9351, IC9353, IC9355, IC9357, IC9359, IC9360, IC9361, IC9362, IC9308, IC9311, IC9312, IC9313, IC9314, IC9315, IC9316, IC9317, IC9318, IC9319, IC9300, IC9301, IC9302, IC9303, IC9304, IC9305, IC9306, IC9307, IC9309, IC9310, IC9282, IC9283, IC9284, IC9285, IC9286, IC9287, IC9271, IC9272, IC9273, IC9274, IC9275, IC9276, IC9277, IC9278, IC9279, IC9280, IC9266, IC9267, IC9268, IC9269, IC9270, IC9231, IC9232, IC9243, IC9244, IC9245, IC9246, IC9248, IC9249, IC9250, IC9247, IC9218, IC9219, IC9220, IC9221, IC9222, IC9223, IC9224, IC9225, IC9226, IC9227, IC9204, IC9205, IC9206, IC9207, IC9208, IC9209, IC9210, IC9211, IC9212, IC9213, IC9214, IC9184, IC9185, IC9186, IC9187, IC9188, IC9195, IC9196, IC9197, IC9199, IC9198, IC9151, IC9152, IC9153, IC9154, IC9155, IC9156, IC9157, IC9158, IC9159, IC9160, IC9161, IC9162, IC9164, IC9165, IC9166, IC9141, IC9142, IC9143, IC9144, IC9145, IC9146, IC9147, IC9148, IC9149, IC9150, IC8766, IC9116, IC9117, IC9118, IC9119, IC9120, IC9121, IC9122, IC9123, IC9124, IC8953, IC8961, IC8962, IC8963, IC8964, IC9044, IC9068, IC9097, IC9098, IC9099, IC9100, IC9101, IC9102, IC9103, IC9104, IC9026, IC9033, IC9038, IC9040, IC9074, IC9084, IC9086, IC9087, IC9088, IC9089, IC8884, IC9079, IC9080, IC9083, IC9081, IC9082, IC8881, IC8882, IC8883, IC8885, IC8365, IC8633, IC8656, IC8783, IC8995, IC8996, IC8997, IC8998, IC8999, IC9045, IC9046, IC9047, IC9048, IC9049, IC9029, IC9030, IC9032, IC9034, IC9035, IC9037, IC9041, IC9042, IC9043, IC8925, IC8981, IC8992, IC8993, IC8994, IC8970, IC8971, IC8972, IC8973, IC8974, IC8975, IC8976, IC8977, IC8978, IC8979, IC8965, IC8966, IC8967, IC8968, IC8969, IC8943, IC8944, IC8946, IC8947, IC8949, IC8945, IC8948, IC8950, IC8951, IC8952, IC8954, IC8955, IC8956, IC8957, IC8958, IC8959, IC8960, IC8926, IC8928, IC8934, IC8935, IC8936, IC8938, IC8939, IC8940, IC8941, IC8942, IC8847, IC8862, IC8903, IC8904, IC8905, IC8907, IC8908, IC8909, IC8906, IC8848, IC8849, IC8850, IC8851, IC8852, IC8853, IC8867, IC8878, IC8879, IC8807, IC8809, IC8810, IC8808, IC8500, IC8794, IC8828, IC8829, IC8830, IC8831, IC8832, IC8833, IC8834, IC8835, IC8791, IC8793, IC8795, IC8796, IC8797, IC8798, IC8799, IC8800, IC8801, IC8802, IC8803, IC8804, IC8805, IC8806, IC8719, IC8742, IC8757, IC8759, IC8769, IC8778, IC8786, IC8787, IC8789, IC8790, IC8756, IC8758, IC8760, IC8770, IC8771, IC8772, IC8773, IC8776, IC8777, IC8738, IC8743, IC8744, IC8745, IC8746, IC8747, IC8702, IC8709, IC8711, IC8712, IC8713, IC8715, IC8716, IC8717, IC8718, IC8720, IC8721, IC8722, IC8724, IC8730, IC8710, IC8688, IC8689, IC8695, IC8667, IC8668, IC8669, IC8671, IC8672, IC8673, IC8674, IC8675, IC8676, IC8678, IC8598, IC8620, IC8624, IC8627, IC8631, IC8632, IC8634, IC8635, IC8636, IC8637, IC8638, IC8602, IC8603, IC8604, IC8611, IC8613, IC8615, IC8618, IC8617, IC8546, IC8552, IC8553, IC8562, IC8564, IC8566, IC8533, IC8537, IC8538, IC8508, IC8528, IC8530, IC8532, IC8534, IC8535, IC8536, IC8539, IC8529, IC8478, IC8497, IC8507, IC8510, IC8517, IC8518, IC8519, IC8520, IC8521, IC8522, IC8480, IC8481, IC8483, IC8484, IC8485, IC8458, IC8459, IC8460, IC8470, IC8472, IC8473, IC8474, IC8475, IC8477, IC8479, IC8464, IC8466, IC8468, IC8469, IC8471, IC8318, IC8353, IC8361, IC8437, IC8452, IC8453, IC8454, IC8455, IC8456, IC8383, IC8407, IC8429, IC8430, IC8431, IC8432, IC8433, IC8434, IC8435, IC8436, IC8438, IC8440, IC8380, IC8391, IC8394, IC8398, IC8413, IC8414, IC8416, IC8417, IC8415, IC8418, IC8382, IC8384, IC8385, IC8386, IC8293, IC8390, IC8392, IC8393, IC8406, IC8408, IC8411, IC8412, IC8355, IC8366, IC8367, IC8163, IC8215, IC8312, IC8358, IC8359, IC8360, IC8362, IC8363, IC8346, IC8347, IC8348, IC8350, IC8351, IC8352, IC8354, IC8356, IC8357, IC8349, IC8270, IC8271, IC8272, IC8273, IC8322, IC8325, IC8328, IC8329, IC8330, IC8313, IC8314, IC8317, IC8319, IC8320, IC8321, IC8323, IC8324, IC8326, IC8327, IC8247, IC8265, IC8305, IC8307, IC8399, IC8274, IC8275, IC8286, IC8287, IC8288, IC8289, IC8290, IC8291, IC8292, IC8231, IC8232, IC8233, IC8234, IC8235, IC8236, IC8230, IC8198, IC8199, IC8200, IC8211, IC8212, IC8213, IC8214, IC8216, IC8191, IC8192, IC8193, IC8194, IC8195, IC8176, IC8177, IC8178, IC8180, IC8181, IC8182, IC8183, IC8184, IC8185, IC8179, IC8148, IC8149, IC8150, IC8151, IC8152, IC8154, IC8155, IC8156, IC8088, IC8093, IC8141, IC8142, IC8143, IC8144, IC8135, IC8136, IC8137, IC8138, IC8139, IC8140, IC8086, IC8087, IC8089, IC8090, IC8091, IC8092, IC8094, IC8095, IC8096, IC8097, IC8098, IC8099, IC8100, IC8108, IC8109, IC8110, IC8111, IC8112, IC8020, IC8021, IC8022, IC8023, IC8038, IC8039, IC8042, IC8043, IC8044, IC8045, IC8024, IC8036, IC8025, IC8027, IC8028, IC8030, IC8033, IC8014, IC8017, IC8018, IC8019, IC7997, IC7998, IC7999, IC8000, IC8001, IC8005, IC8007, IC8009, IC8010, IC8015, IC7961, IC7962, IC7963, IC7986, IC7987, IC7988, IC7989, IC7990, IC7991, IC7992, IC7993, IC7994, IC7995, IC7960, IC7959, IC7948, IC7949, IC7950, IC7951, IC7952, IC7923, IC7924, IC7925, IC7936, IC7938, IC7913, IC7914, IC7915, IC7916, IC7917, IC7918, IC7921, IC7912, IC7891, IC7892, IC7895, IC7907, IC7909, IC7870, IC7871, IC7872, IC7873, IC7874, IC7875, IC7876, IC7881, IC7845, IC7856, IC7857, IC7858, IC7859, IC7767, IC7840, IC7841, IC7842, IC7843, IC7844, IC7831, IC7832, IC7833, IC7834, IC7835, IC7766, IC7768, IC7769, IC7770, IC7684, IC7685, IC7696, IC7697, IC7698, IC7641, IC7643, IC7656, IC7631, IC7632, IC7633, IC7634, IC7635, IC7560, IC7562, IC7563, IC7565, IC7566, IC7567, IC7568, IC7569, IC7571, IC7572, IC7521, IC7519, IC7520, IC7339, IC7340, IC7341, IC7342, IC7343, IC7190, IC7191, IC7187, IC7188, IC7189, IC7182, IC7183, IC7184, IC7185, IC7186.
29. Product Code: 0042-0040-US-L; GTIN Number: 813502011401; Serial Numbers: IC12127, IC12588, IC12589, IC12598, IC12599, IC12600, IC12603, IC12606, IC12608, IC12537, IC12547, IC12562, IC12564, IC12569, IC12570, IC12571, IC12572, IC12573, IC12607, IC12471, IC12482, IC12489, IC12498, IC12499, IC12500, IC12502, IC12525, IC12526, IC12527, IC12528, IC12529, IC12531, IC12532, IC12533, IC12534, IC12539, IC11083, IC11527, IC11612, IC11643, IC11647, IC11648, IC11649, IC11650, IC11651, IC11652, IC11653, IC11554, IC11580, IC11590, IC11601, IC11602, IC11603, IC11605, IC11607, IC11608, IC11620, IC11587, IC11588, IC11589, IC11591, IC11592, IC11593, IC11594, IC11595, IC11596, IC11599, IC11508, IC11515, IC11517, IC11541, IC11542, IC11543, IC11545, IC11547, IC11548, IC11553, IC10905, IC10914, IC11025, IC11050, IC11057, IC11061, IC11204, IC11212, IC11436, IC11450, IC11455, IC11463, IC11468, IC11470, IC11475, IC11483, IC11501, IC11502, IC11084, IC11104, IC11105, IC11110, IC11111, IC11112, IC11113, IC11114, IC11115, IC11118, IC11098, IC11099, IC11100, IC11101, IC11102, IC10976, IC10977, IC10988, IC10989, IC10990, IC10991, IC10992, IC10993, IC10996, IC10997, IC10836, IC10902, IC10963, IC10964, IC10965, IC10966, IC10967, IC10968, IC10969, IC10970, IC10956, IC10986, IC10987, IC10994, IC10995, IC10893, IC10894, IC10896, IC10867, IC10868, IC10871, IC10872, IC10874, IC10862, IC10863, IC10864, IC10865, IC10866, IC10840, IC10780, IC10834, IC10841, IC10846, IC10794, IC10797, IC10802, IC10761, IC10765, IC10766, IC10771, IC10663, IC10664, IC10666, IC10667, IC10668, IC10573, IC10574, IC10575, IC10576, IC10577, IC10579, IC10498, IC10500, IC10501, IC10502, IC10495, IC10488, IC10489, IC10490, IC10491, IC10399, IC10456, IC10457, IC10458, IC10460, IC10464, IC10465, IC10467, IC10469, IC10305, IC10234, IC9981, IC9627, IC9664, IC9695, IC9696, IC9674, IC9584, IC9593, IC9615, IC9616, IC9600, IC9601, IC9602, IC9603, IC9604, IC9605, IC9595, IC9529A, IC9564, IC9565, IC9566, IC9568, IC9569, IC9570, IC9407, IC9419, IC9502, IC9503, IC9504, IC9505, IC9506, IC9507, IC9508, IC9509, IC9487, IC9459, IC9354, IC9261, IC9262, IC9263, IC9264, IC9265, IC9200, IC9201, IC9202, IC9203, IC8877, IC8788, IC8811, IC8843, IC8844, IC8845, IC8846, IC8792, IC8677, IC8728, IC8740, IC8741, IC8749, IC8697, IC8698, IC8699, IC8700, IC8701, IC8628, IC8650, IC8651, IC8652, IC8653, IC8654, IC8657, IC8658, IC8659, IC8660, IC8574, IC8570, IC8572, IC8587, IC8605, IC8606, IC8607, IC8616, IC8389, IC8496, IC8555, IC8565, IC8498, IC8506, IC8509, IC8515, IC8516, IC8446, IC8461, IC8462, IC8463, IC8465, IC8373, IC8368, IC8378, IC8379, IC8381, IC8248, IC8249, IC8250, IC8251, IC8266, IC8267, IC8268, IC8269, IC8189, IC8190, IC8175, IC8107, IC8167, IC8168, IC8169, IC8170, IC8171, IC8172, IC8173, IC8174, IC8153, IC7522, IC10801.
30. Product Code: 1000201; GTIN Number: 813502010442; Serial Numbers: IC9813, IC9814, IC9815, IC9816, IC9817, IC9818, IC9819, IC9820, IC9821, IC9632, IC9633, IC9636, IC9637, IC9641, IC9629, IC9635.
31. Product Code: 1000432; GTIN Number: 813502013030; Serial Numbers: IC11600, IC11611, IC11616, IC11619, IC10827, IC11399, IC11400, IC11401, IC11404, IC11210, IC11227, IC11230, IC11231, IC11232, IC11236, IC11237, IC11238, IC11239, IC11143, IC11145, IC11153, IC11182, IC10837, IC10849, IC10851, IC10854, IC10300, IC10703, IC10762, IC10784, IC10785, IC10786, IC10672, IC10689, IC10727, IC10755, IC10631, IC10650, IC10651, IC10652, IC10584, IC10585, IC10586, IC10588, IC10517, IC10518, IC10483, IC10492, IC10494, IC10496, IC10493, IC10497, IC10338, IC10362, IC10364, IC10365, IC10170, IC10353. |
| FEI Number |
1220648
|
Recalling Firm/ Manufacturer |
Abiomed, Inc. 24 Cherry Hill Dr Danvers MA 01923-2575
|
| For Additional Information Contact | 978-646-1400 |
Manufacturer Reason for Recall | A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected |
FDA Determined Cause 2 | Device Design |
| Action | Initially Abiomed posted their recall notification on their website, HeartRecover.com on 06/23/2025. Consignees were also notified on 06/25/2025 via letter. Consignees were advised that devices were not being removed. They were advised to have a back-up Automated Impella Controller (AIC) available in the event of a device failure and were provided work around instruction for console-to-console transfer and case starts. Additionally, consignees were instructed to forward the notification to all pertinent personnel and customers if further distributed and complete and return the business response form provided. |
| Quantity in Commerce | 11,031 units |
| Distribution | Worldwide distribution. US Nationwide, Australia, Austria, Belgium, Brunei Darussalam, Canada, Colombia, Croatia, Czechia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, and United Kingdom.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| PMA Database | PMAs with Product Code = OZD
|
|
|
|