| Date Initiated by Firm | June 03, 2025 |
| Date Posted | August 19, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2328-2025 |
| Recall Event ID |
97079 |
| 510(K)Number | K201658 |
| Product Classification |
Ventilator, continuous, facility use - Product Code CBK
|
| Product | Hamilton C6 REF 160021 (ESM Board PN 10144626
UDI-DI code: 76300028PN160021ZW
Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates. |
| Code Information |
Model/Catalog Number: 160021
ESM board Part Number: 10144626
GTIN: 07630002808590
UDI-DI: 76300028PN160021ZW
Affected SN range: 13636 till 13950 and 13966 till 14753
13797
13801
13806
13855
13865
13981
13988
13989
14012
14015
14019
14034
14035
14042
14052
14054
14102
14109
14112
14113
14114
14115
14116
14119
14120
14121
14122
14123
14124
14126
14127
14128
14163
14168
14169
14170
14172
14173
14175
14177
14187
14188
14189
14195
14197
|
| FEI Number |
3014766741
|
Recalling Firm/ Manufacturer |
Hamilton Medical AG Parc Industrial Vial 10 Domat/Ems Switzerland
|
| For Additional Information Contact | Sachin Thanawala 800-426-6331 Ext. 6004 |
Manufacturer Reason for Recall | A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation. |
FDA Determined Cause 2 | Equipment maintenance |
| Action | URGENT: Medical Device Correction letters were emailed to customers on 06/06/2025, 06/20/2025, and 07/07/2025, informing them that due to damage to the capacitor in the ESM-Board within the Hamilton-C6 ventilator the capacitor could, over time, lead to a loss of its function due to oxidation of similar degradation processes.
Customers are instructed to:
1. Review their Hamilton-C6 ventilators and check for the serial number within the scope mentioned in the letter.
2. for affected ventilators must have the ESM-Boards replaced. The replacement will be organized by the local partner as per servicing protocol.
Note - the Hamilton-C6 ventilators do not need to be removed from use/service. These devices were tested prior to release and no test will be able to detect if the ESM board will fail in the future.
3. Keep the recall notice within the respective ventilator's data records.
For questions or assistance, contact Technical Support at 1-800-426-6331 option 2 or email reno.techsupport@hamiltonmedical.com
|
| Quantity in Commerce | 1040 systems |
| Distribution | US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = CBK
|