| Date Initiated by Firm | September 18, 2025 |
| Date Posted | October 15, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0157-2026 |
| Recall Event ID |
97726 |
| 510(K)Number | K214039 |
| Product Classification |
System, imaging, pulsed doppler, ultrasonic - Product Code IYN
|
| Product | LOGIQ P9 R4.5 ultrasound system with system software version R4.5.7., Model Number 5877533 |
| Code Information |
Model 5877533 UDI-DI: 00195278464637,
Serial Numbers: LP9490416
LP9490175
LP9490393
LP9490399
LP9490376
LP9490463
LP9443017
LP9442472
LP9443231
LP9442094
LP9443291
LP9443230
LP9440403
LP9442483
LP9440972
LP9440166
LP9440237
LP9440933
LP9440348
LP9442196
LP9440631
LP9440926
LP9441099
LP9441225
LP9441097
LP9440929
LP9441030
LP9440921
LP9441035
LP9440507
LP9440920
LP9440459
LP9440460
LP9441332
LP9442656
LP9443022
LP9440986 |
Recalling Firm/ Manufacturer |
GE Medical Systems, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
| For Additional Information Contact | GE HealthCare Service 800-437-1171 |
Manufacturer Reason for Recall | The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care. |
FDA Determined Cause 2 | Software design |
| Action | An URGENT MEDICAL DEVICE CORRECTION notification letter dated 9/18/25 was sent to customers.
Actions to be taken by Customer/User
For LOGIQ P9 and LOGIQ P10 series ultrasound systems with system software version R4.5.7, discontinue using the UGAP feature (please see Appendix for product identification). You may continue using other imaging features.
After the correction has been implemented, please destroy the installation media for software R4.5.7 at your site.
Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions.
Please retain this document for your records.
Please complete and return the attached acknowledgement form to recall.74084@gehealthcare.com.
GE HealthCare will correct all affected products at no cost to you.
A GE HealthCare representative will contact you to arrange for the correction.
If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity in Commerce | 37 units |
| Distribution | US Nationwide. Global Distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = IYN
|