• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall HeartMate

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall HeartMatesee related information
Date Initiated by FirmOctober 09, 2025
Date PostedNovember 14, 2025
Recall Status1 Open3, Classified
Recall NumberZ-0508-2026
Recall Event ID 97782
PMA NumberP060040 
Product Classification Ventricular (assist) bypass - Product Code DSQ
ProductBrand Name: HeartMate Product Name: HeartMate II Controller (standalone) Model/Catalog Number: 106017, HeartMate II Pocket Controller, US; 106762, HeartMate II Pocket Controller, US; 107801, HeartMate II Pocket Controller Software Version: N/A Product Description: The HeartMate II System Controller is a small computer that controls and monitors system operation. It sends power and operating signals to the Left Ventricular Assist Device and adjusts device operation to maintain programmed levels of cardiac support. The System Controller identifies alarm conditions and initiates Hazard and Advisory alarms. It records device performance and alarm data, and transfers the information to the HeartMate Touch Communication System. Component: Yes, HeartMate II LVAS Implant Kit
Code Information Lot Code: All units of the following models are impacted: Model 106017 UDI: 00813024011248 Model 106762 UDI: 00813024011286 Model 107801 UDI: 00813024011996 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used.
FEI Number 2916596
Recalling Firm/
Manufacturer
Thoratec LLC
6035 Stoneridge Dr
Pleasanton CA 94588-3270
For Additional Information ContactMs. Shelley Lange
1-612-3463514
Manufacturer Reason
for Recall
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
FDA Determined
Cause 2
Device Design
ActionOn 10/09/2025, the firm sent via email an "URGENT MEDICAL DEVICE RECALL (CORRECTION)" Letter to customers informing them that Abbott has observed an increase in customer complaints about the controller Backup Battery Fault Alarm. The issue is potentially due to excessive handling, movement, stress and pulling at the interface between the ribbon cable and controller which results in corrosion of the ribbon cable connectors. What Customers should know: When the Backup Battery Fault Alarm is active, this Advisory alarm will appear as a flashing yellow wrench on the user interface and Call Hospital Contact; Backup Battery Fault will appear on the screen, accompanied by a slow beep tone. When in a clinical setting, the HeartMate Touch App will have an active alarm that reads Replace Backup Battery. As a reminder, when this alarm is active and the pump is running and flow is maintained (indicated by the green arrow symbol ), this alarm does not affect the controller or pump functionality. Customers are instructed: -When performing Backup Battery installation or replacement, clinicians should minimize excessive handling, movement, stress and pulling at the interface between the ribbon cable and controller. An example of excessive handling would be using the ribbon cable to hold the weight of the controller. -Remind patients and caregivers to follow the Patient Handbook: If System Controller Backup Battery Fault Alarm is active, patients should first ensure that no other alarms are active, and then they should call their hospital contact as soon as possible for diagnosis and instructions. Patients and caregivers should not immediately attempt to exchange the System Controller without clinical guidance. For questions or further assistance is needed regarding the Backup Battery Alarm after following Instructions for Use, contact Abbott Technical Service at 1-800-456-1477 (Business hours: 8 AM EST to 7 PM EST).
Quantity in Commerce2424
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
PMA DatabasePMAs with Product Code = DSQ
-
-