| | Class 3 Device Recall CONSTRUX" Mini Ti Spacer System |  |
| Date Initiated by Firm | November 03, 2025 |
| Date Posted | December 11, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0896-2026 |
| Recall Event ID |
97862 |
| 510(K)Number | K203342 |
| Product Classification |
Intervertebral fusion device with bone graft, cervical - Product Code ODP
|
| Product | Brand Name: CONSTRUX" Mini Ti Spacer System
Product Name: CONSTRUX" Mini Ti Spacer System
Model/Catalog Number: 37-6005SP 12mm W X 12mm L, 5mm H, 5 Lordotic
37-6005SP 12mm W X 12mm L, 5mm H, 5 Lordotic
37-6005SP 12mm W X 12mm L, 5mm H, 5 Lordotic
37-6006SP 12mm W X 12mm L, 6mm H, 5 Lordotic
37-6006SP 12mm W X 12mm L, 6mm H, 5 Lordotic
37-6006SP 12mm W X 12mm L, 6mm H, 5 Lordotic
37-6007SP 12mm W X 12mm L, 7mm H, 5 Lordotic
37-6007SP 12mm W X 12mm L, 7mm H, 5 Lordotic
37-6007SP 12mm W X 12mm L, 7mm H, 5 Lordotic
37-6008SP 12mm W X 12mm L, 8mm H, 5 Lordotic
37-6008SP 12mm W X 12mm L, 8mm H, 5 Lordotic
37-6008SP 12mm W X 12mm L, 8mm H, 5 Lordotic
37-6009SP 12mm W X 12mm L, 9mm H, 5 Lordotic
37-6009SP 12mm W X 12mm L, 9mm H, 5 Lordotic
37-6009SP 12mm W X 12mm L, 9mm H, 5 Lordotic
37-6010SP 12mm W X 12mm L, 10mm H, 5 Lordotic
37-6010SP 12mm W X 12mm L, 10mm H, 5 Lordotic
37-6010SP 12mm W X 12mm L, 10mm H, 5 Lordotic
37-6011SP 12mm W X 12mm L, 11mm H, 5 Lordotic
37-6011SP 12mm W X 12mm L, 11mm H, 5 Lordotic
37-6011SP 12mm W X 12mm L, 11mm H, 5 Lordotic
37-6012SP 12mm W X 12mm L, 12mm H, 5 Lordotic
37-6012SP 12mm W X 12mm L, 12mm H, 5 Lordotic
37-6012SP 12mm W X 12mm L, 12mm H, 5 Lordotic
37-6105SP 12mm W X 12mm L, 5mm H, Parallel
37-6105SP 12mm W X 12mm L, 5mm H, Parallel
37-6105SP 12mm W X 12mm L, 5mm H, Parallel
37-6106SP 12mm W X 12mm L 6mm H, Parallel
37-6106SP 12mm W X 12mm L 6mm H, Parallel
37-6106SP 12mm W X 12mm L 6mm H, Parallel
37-6107SP 12mm W X 12mm L, 7mm H, Parallel
37-6107SP 12mm W X 12mm L, 7mm H, Parallel
37-6107SP 12mm W X 12mm L, 7mm H, Parallel
37-6108SP 12mm W X 12mm L, 8mm H, Parallel
37-6108SP 12mm W X 12mm L, 8mm H, Parallel
37-6108SP 12mm W X 12mm L, 8mm H, Parallel
37-6109SP 12mm W X 12mm L, 9mm H, Parallel
37-6109SP 12mm W X 12mm L, 9mm H, Parallel
37-6109SP 12mm W X 12mm L, 9mm H, Parallel
37-6110SP 12mm W X 12mm L, 10mm H, Parallel
37-6110SP 12mm W X 12mm L, 10mm H, Parallel
37-6110SP 12mm W X 12mm L, 10mm H, Parallel
37-6111SP 12mm W X 12mm L, 11mm H, Parallel
37-6111SP 12mm W X 12mm L, 11mm H, Parallel
37-6111SP 12mm W X 12mm L, 11mm H, Parallel
37-6112SP 12mm W X 12mm L, 12mm H, Parallel
37-6112SP 12mm W X 12mm L, 12mm H, Parallel
37-6112SP 12mm W X 12mm L, 12mm H, Parallel
37-6206SP 12mm W X 12mm L, 6mm H, 10 Lordotic
37-6206SP 12mm W X 12mm L, 6mm H, 10 Lordotic
37-6206SP 12mm W X 12mm L, 6mm H, 10 Lordotic
37-6207SP 12mm W X 12mm L, 7mm H, 10 Lordotic
37-6207SP 12mm W X 12mm L, 7mm H, 10 Lordotic
37-6207SP 12mm W X 12mm L, 7mm H, 10 Lordotic
37-6208SP 12mm W X 12mm L, 8mm H, 10 Lordotic
37-6208SP 12mm W X 12mm L, 8mm H, 10 Lordotic
37-6208SP 12mm W X 12mm L, 8mm H, 10 Lordotic
37-6209SP 12mm W X 12mm L, 9mm H, 10 Lordotic
37-6209SP 12mm W X 12mm L, 9mm H, 10 Lordotic
37-6209SP 12mm W X 12mm L, 9mm H, 10 Lordotic
37-6210SP 12mm W X 12mm L, 10mm H, 10 Lordotic
37-6210SP 12mm W X 12mm L, 10mm H, 10 Lordotic
37-6210SP 12mm W X 12mm L, 10mm H, 10 Lordotic
37-6211SP 12mm W X 12mm L, 11mm H, 10 Lordotic
37-6211SP 12mm W X 12mm L, 11mm H, 10 Lordotic
37-6211SP 12mm W X 12mm L, 11mm H, 10 Lordotic |
| Code Information |
Lot Code: All Lot Codes/ UDI: 37-6005SP 18257200150151
37-6005SP 18257200150151
37-6005SP 18257200150151
37-6006SP 18257200150168
37-6006SP 18257200150168
37-6006SP 18257200150168
37-6007SP 18257200150175
37-6007SP 18257200150175
37-6007SP 18257200150175
37-6008SP 18257200150182
37-6008SP 18257200150182
37-6008SP 18257200150182
37-6009SP 18257200150199
37-6009SP 18257200150199
37-6009SP 18257200150199
37-6010SP 18257200150205
37-6010SP 18257200150205
37-6010SP 18257200150205
37-6011SP 18257200150212
37-6011SP 18257200150212
37-6011SP 18257200150212
37-6012SP 18257200150229
37-6012SP 18257200150229
37-6012SP 18257200150229
37-6105SP 18257200149759
37-6105SP 18257200149759
37-6105SP 18257200149759
37-6106SP 18257200149766
37-6106SP 18257200149766
37-6106SP 18257200149766
37-6107SP 18257200149773
37-6107SP 18257200149773
37-6107SP 18257200149773
37-6108SP 18257200149780
37-6108SP 18257200149780
37-6108SP 18257200149780
37-6109SP 18257200149797
37-6109SP 18257200149797
37-6109SP 18257200149797
37-6110SP 18257200149803
37-6110SP 18257200149803
37-6110SP 18257200149803
37-6111SP 18257200149810
37-6111SP 18257200149810
37-6111SP 18257200149810
37-6112SP 18257200149827
37-6112SP 18257200149827
37-6112SP 18257200149827
37-6206SP 18257200150557
37-6206SP 18257200150557
37-6206SP 18257200150557
37-6207SP 18257200150564
37-6207SP 18257200150564
37-6207SP 18257200150564
37-6208SP 18257200150571
37-6208SP 18257200150571
37-6208SP 18257200150571
37-6209SP 18257200150588
37-6209SP 18257200150588
37-6209SP 18257200150588
37-6210SP 18257200150595
37-6210SP 18257200150595
37-6210SP 18257200150595
37-6211SP 18257200150601
37-6211SP 18257200150601
37-6211SP 18257200150601
37-6212SP 18257200150618
37-6212SP 18257200150618
37-6212SP 18257200150618
37-6306SP 18257200150915
37-6306SP 18257200150915
37-6306SP 18257200150915
37-6307SP 18257200150922
37-6307SP 18257200150922
37-6307SP 18257200150922
37-6308SP 18257200150939
37-6308SP 18257200150939
37-6308SP 18257200150939
37-6309SP 18257200150946
37-6309SP 18257200150946
37-6309SP 18257200150946
37-6310SP 18257200150953
37-6310SP 18257200150953
37-6310SP 18257200150953
37-6311SP 18257200150960
37-6311SP 18257200150960
37-6311SP 18257200150960
37-6312SP 18257200150977
37-6312SP 18257200150977
37-6312SP 18257200150977
37-7005SP 18257200150250
37-7005SP 18257200150250
37-7005SP 18257200150250
37-7006SP 18257200150267
37-7006SP 18257200150267
37-7006SP 18257200150267
37-7007SP 18257200150274
37-7007SP 18257200150274
37-7007SP 18257200150274
37-7008SP 18257200150281
37-7008SP 18257200150281
37-7008SP 18257200150281
37-7009SP 18257200150298
37-7009SP 18257200150298
37-7009SP 18257200150298
37-7010SP 18257200150304
37-7010SP 18257200150304
37-7010SP 18257200150304
37-7011SP 18257200150311
37-7011SP 18257200150311
37-7011SP 18257200150311
37-7012SP 18257200150328
37-7012SP 18257200150328
37-7012SP 18257200150328
37-7105SP 18257200149858
37-7105SP 18257200149858
37-7105SP 18257200149858
37-7106SP 18257200149865
37-7106SP 18257200149865
37-7106SP 18257200149865
37-7107SP 18257200149872
37-7107SP 18257200149872
37-7107SP 18257200149872
37-7108SP 18257200149889
37-7108SP 18257200149889
37-7108SP 18257200149889
37-7109SP 18257200149896
37-7109SP 18257200149896
37-7109SP 18257200149896
37-7110SP 18257200149902
37-7110SP 18257200149902
37-7110SP 18257200149902
37-7111SP 18257200149919
37-7111SP 18257200149919
37-7111SP 18257200149919
37-7112SP 18257200149926
37-7112SP 18257200149926
37-7112SP 18257200149926
37-7206SP 18257200150649
37-7206SP 18257200150649
37-7206SP 18257200150649
37-7207SP 18257200150656
37-7207SP 18257200150656
37-7207SP 18257200150656
37-7208SP 18257200150663
37-7208SP 18257200150663
37-7208SP 18257200150663
37-7209SP 18257200150670
37-7209SP 18257200150670
37-7209SP 18257200150670
37-7210SP 18257200150687
37-7210SP 18257200150687
37-7210SP 18257200150687
37-7211SP 18257200150694
37-7211SP 18257200150694
37-7211SP 18257200150694
37-7212SP 18257200150700
37-7212SP 18257200150700
37-7212SP 18257200150700
37-7306SP 18257200151004
37-7306SP 18257200151004
37-7306SP 18257200151004
37-7307SP 18257200151011
37-7307SP 18257200151011
37-7307SP 18257200151011
37-7308SP 18257200151028
37-7308SP 18257200151028
37-7308SP 18257200151028
37-7309SP 18257200151035
37-7309SP 18257200151035
37-7309SP 18257200151035
37-7310SP 18257200151042
37-7310SP 18257200151042
37-7310SP 18257200151042
37-7311SP 18257200151059
37-7311SP 18257200151059
37-7311SP 18257200151059
37-7312SP 18257200151066
37-7312SP 18257200151066
37-7312SP 18257200151066
37-8005SP 18257200150359
37-8005SP 18257200150359
37-8005SP 18257200150359
37-8006SP 18257200150366
37-8006SP 18257200150366
37-8006SP 18257200150366
37-8007SP 18257200150373
37-8007SP 18257200150373
37-8007SP 18257200150373
37-8008SP 18257200150380
37-8008SP 18257200150380
37-8008SP 18257200150380
37-8009SP 18257200150397
37-8009SP 18257200150397
37-8009SP 18257200150397
37-8011SP 18257200150410
37-8011SP 18257200150410
37-8011SP 18257200150410
37-8012SP 18257200150427
37-8012SP 18257200150427
37-8012SP 18257200150427
37-8015SP 18257200150403
37-8015SP 18257200150403
37-8015SP 18257200150403
37-8105SP 18257200149957
37-8105SP 18257200149957
37-8105SP 18257200149957
37-8106SP 18257200149964
37-8106SP 18257200149964
37-8106SP 18257200149964
37-8107SP 18257200149971
37-8107SP 18257200149971
37-8107SP 18257200149971
37-8108SP 18257200149988
37-8108SP 18257200149988
37-8108SP 18257200149988
37-8109SP 18257200149995
37-8109SP 18257200149995
37-8109SP 18257200149995
37-8110SP 18257200150007
37-8110SP 18257200150007
37-8110SP 18257200150007
37-8111SP 18257200150014
37-8111SP 18257200150014
37-8111SP 18257200150014
37-8112SP 18257200150021
37-8112SP 18257200150021
37-8112SP 18257200150021
37-8206SP 18257200150731
37-8206SP 18257200150731
37-8206SP 18257200150731
37-8207SP 18257200150748
37-8207SP 18257200150748
37-8207SP 18257200150748
37-8208SP 18257200150755
37-8208SP 18257200150755
37-8208SP 18257200150755
37-8209SP 18257200150762
37-8209SP 18257200150762
37-8209SP 18257200150762
37-8210SP 18257200150779
37-8210SP 18257200150779
37-8210SP 18257200150779
37-8211SP 18257200150786
37-8211SP 18257200150786
37-8211SP 18257200150786
37-8212SP 18257200150793
37-8212SP 18257200150793
37-8212SP 18257200150793
37-8306SP 18257200151097
37-8306SP 18257200151097
37-8306SP 18257200151097
37-8307SP 18257200151103
37-8307SP 18257200151103
37-8307SP 18257200151103
37-8308SP 18257200151110
37-8308SP 18257200151110
37-8308SP 18257200151110
37-8309SP 18257200151127
37-8309SP 18257200151127
37-8309SP 18257200151127
37-8310SP 18257200151134
37-8310SP 18257200151134
37-8310SP 18257200151134
37-8311SP 18257200151141
37-8311SP 18257200151141
37-8311SP 18257200151141
37-8312SP 18257200151158
37-8312SP 18257200151158
37-8312SP 18257200151158
37-9005SP 18257200150458
37-9005SP 18257200150458
37-9005SP 18257200150458
37-9006SP 18257200150465
37-9006SP 18257200150465
37-9006SP 18257200150465
37-9007SP 18257200150472
37-9007SP 18257200150472
37-9007SP 18257200150472
37-9008SP 18257200150489
37-9008SP 18257200150489
37-9008SP 18257200150489
37-9009SP 18257200150496
37-9009SP 18257200150496
37-9009SP 18257200150496
37-9011SP 18257200150519
37-9011SP 18257200150519
37-9011SP 18257200150519
37-9012SP 18257200150526
37-9012SP 18257200150526
37-9105SP 18257200150052
37-9105SP 18257200150052
37-9106SP 18257200150069
37-9106SP 18257200150069
37-9107SP 18257200150076
37-9107SP 18257200150076
37-9108SP 18257200150083
37-9108SP 18257200150083
37-9109SP 18257200150090
37-9109SP 18257200150090
37-9110SP 18257200150106
37-9110SP 18257200150106
37-9111SP 18257200150113
37-9111SP 18257200150113
37-9112SP 18257200150120
37-9112SP 18257200150120
37-9206SP 18257200150823
37-9206SP 18257200150823
37-9207SP 18257200150830
37-9207SP 18257200150830
37-9208SP 18257200150847
37-9208SP 18257200150847
37-9209SP 18257200150854
37-9209SP 18257200150854
37-9210SP 18257200150861
37-9210SP 18257200150861
37-9211SP 18257200150878
37-9211SP 18257200150878
37-9212SP 18257200150885
37-9212SP 18257200150885
37-9306SP 18257200151189
37-9306SP 18257200151189
37-9307SP 18257200151196
37-9307SP 18257200151196
37-9308SP 18257200151202
37-9308SP 18257200151202
37-9309SP 18257200151219
37-9309SP 18257200151219
37-9310SP 18257200151226
37-9310SP 18257200151226
37-9311SP 18257200151233
37-9311SP 18257200151233
37-9312SP 18257200151240
37-9312SP 18257200151240
|
Recalling Firm/ Manufacturer |
Orthofix U.S. LLC 3451 Plano Pkwy Lewisville TX 75056-9453
|
| For Additional Information Contact | Orthofix Field Actions 214-937-2000 |
Manufacturer Reason for Recall | Labeling contains claims that are not consistently present. |
FDA Determined Cause 2 | Device Design |
| Action | On November 3, 2025 Orthofix issued a recall notification via Email to affected consignees. In addition to informing consignees about the recall , Orthofix ask consignees to take the following actions:
1. Orthofix is removing claims regarding Nanovate Technology in materials, including but not limited to Operative Techniques, Product Brochures, Medical Education Documents, and the Orthofix website.
2. No product return is being requested. You may continue to use the product.
3. Distribute this Notice to all relevant personnel within your organization.
4. Forward this Notice to other organization or facilities where affected products may have been transferred.
5. Acknowledge this Notice and return the completed receipt form to fieldactions@orthofix.com
|
| Quantity in Commerce | 40956 |
| Distribution | Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WY and the countries of AU, DE, PR, GB, HU, CZ, CA, CL, AE, ZA, BR.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = ODP
|
|
|
|