| | Class 2 Device Recall Codman Microsensor Basic Kit |  |
| Date Initiated by Firm | December 12, 2025 |
| Date Posted | January 08, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1036-2026 |
| Recall Event ID |
98091 |
| 510(K)Number | K153347 |
| Product Classification |
Device, monitoring, intracranial pressure - Product Code GWM
|
| Product | Codman Microsensor Basic Kit. Catalog Number: 626631US.
Use of the CODMAN MICROSENSOR Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only. |
| Code Information |
Catalog Number: 626631US. UDI-DI: 10381780514473. Lot Numbers: 7288007, 7288040, 7294017, 7305988, 7305989, 7305990, 7305991, 7307683, 7325489, 7336088, 7336089, 7336090, 7305992, 7346223, 7391752, 7391751, 7406650, 7406651, 7406652, 7408825, 7408826, 7408827, 7456174, 7456175, 7456176, 7456179, 7478039, 7478040, 7480073, 7480072, 7482162, 7503704, 7507495, 7503608, 7510153, 7510155, 7510154, 7534856, 7535008, 7535009, 7411809.
|
Recalling Firm/ Manufacturer |
Integra LifeSciences Corp. (NeuroSciences) 1100 Campus Rd Princeton NJ 08540-6650
|
| For Additional Information Contact | Mary O'Neill 609-275-0500 |
Manufacturer Reason for Recall | Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits. |
FDA Determined Cause 2 | Process control |
| Action | Integra LifeSciences Corporation notified consignees on about 12/12/2025 via FedEx overnight mailing. Consignees were instructed to identify impacted part numbers and lot numbers on hand, remove them immediately from service and quarantine them, complete and return the Acknowledgement Form, and maintain the notice for records. Consignees were also instructed to forward the notification to those who utilize the product for awareness of the recall. Once the Acknowledgement Form is received, and noted that affected product are on hand, Customer Service will contact consignees and provide a Return Material Authorization (RMA) number and directions to return the affected product. |
| Quantity in Commerce | 1,301 units |
| Distribution | Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the). |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|