| Date Initiated by Firm | February 27, 2026 |
| Date Posted | April 17, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1879-2026 |
| Recall Event ID |
98601 |
| Product Classification |
CT biopsy tray - Product Code OFG
|
| Product | Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes
BIOPSY PACK DYNJ38587A, DYNJ38587B
BIOPSY/DRAINAGE TRAY-LF DYNJ30078D
W LINQ PACK DYNJ59108B |
| Code Information |
DYNJ38587A
UDI-DI 10193489754070 (EA) 40193489754071 (CS)
LOTS 20BBF696
20EBH917
20FBW435
20HBM667
20JBF107
20KBW292
20LBS930
21EBB195
21FBR610
21GBS936
21GBS940
21JBU912
21KBP999
22ABX182
22EBI922
22GBZ788
22JBL031
23BBN160
DYNJ38587B
UDI-DI 10195327397326 (EA) 40195327397327 (CS)
LOTS 23DBP789
23FBM124
23HBW639
23JBN841
24ABG236
24CBM867
24FME120
24JMD066
24LME187
DYNJ30078D
UDI-DI 10889942138647 (EA) 40889942138648 (CS)
LOTS 23CBK546
23FBQ481
23IBQ435
23JBR634
23LBG856
24ABE287
24BBS264
24EBQ336
24GBJ987
24JBO830
DYNJ59108B
UDI-DI 10198459138461 (EA) 40198459138462 (CS)
LOT 24JBP861
|
| FEI Number |
1417592
|
Recalling Firm/ Manufacturer |
Medline Industries, LP 3 Lakes Dr Northfield IL 60093-2753
|
| For Additional Information Contact | Haley Barclay 800-633-5463 |
Manufacturer Reason for Recall | Unapproved design changes to the products outside of the 510(k) clearance. |
FDA Determined Cause 2 | Component change control |
| Action | On February 27, 2026, the firm began notifying customers via email and first-class mail.
Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply. |
| Quantity in Commerce | 270,311 total |
| Distribution | Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|