Date Initiated by FirmJanuary 07, 2026
Date PostedMay 07, 2026
Recall Status1 Open3, Classified
Recall NumberZ-2095-2026
Recall Event ID 98642
Product Classification CT biopsy tray - Product Code OFG
ProductMedline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number:DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number:DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number:DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number:DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number:DYNDH1934; 6) BREAST BIOPSY PACK, Model Number:DYNJ17489I; 7) C T BIOPSY TRAY-LF, Model Number:DYNJ17558D; 8) RIGHT HEART BIOPSY TRAY, Model Number:DYNJ19063K; 9) VOR PACK-LF, Model Number:DYNJ48539; 10) VOR PACK-LF, Model Number:DYNJ48539A; 11) IR BIOPSY PACK, Model Number:DYNJ59897B; 12) BREAST PACK, Model Number:DYNJ82998; 13) LIVER BIOPSY KIT, Model Number:DYNJ910654
Code Information 1) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBQ267; 2) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBU924; 3) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 23ABK216; 4) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBK913; 5) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBM302; 6) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23ABU170; 7) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBC364; 8) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBL439; 9) DYNDH1143C, UDI-DI: 10193489193336(each), 40193489193337(case), Lot Number: 22GBZ268; 10) DYNDH1873, UDI-DI: 10195327486464(each), 40195327486465(case), Lot Number: 23KBK476; 11) DYNDH1934, UDI-DI: 10195327519728(each), 40195327519729(case), Lot Number: 24FMF173; 12) DYNJ17489I, UDI-DI: 10195327224783(each), 40195327224784(case), Lot Number: 23CBL337; 13) DYNJ17558D, UDI-DI: 10195327504205(each), 40195327504206(case), Lot Number: 23JLB107; 14) DYNJ19063K, UDI-DI: 10888277828025(each), 40888277828026(case), Lot Number: 24DBL079; 15) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 20LME482; 16) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 20LMF497; 17) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 21JMC430; 18) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 22BMH767; 19) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 22EMD675; 20) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23AMC919; 21) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23DMD933; 22) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23GBI270; 23) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23HBR688; 24) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23KBD033; 25) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24ABL751; 26) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24DBK656; 27) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24EBQ366; 28) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24GBD309; 29) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 24IBT900; 30) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25EBA588; 31) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25HBK043; 32) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25IBA545; 33) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25JBA908; 34) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 22LBH614; 35) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23EBQ760; 36) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23FBT785; 37) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23LBA546; 38) DYNJ82998, UDI-DI: 10195327222987(each), 40195327222988(case), Lot Number: 22JMG410; 39) DYNJ82998, UDI-DI: 10195327222987(each), 40195327222988(case), Lot Number: 22LMF004; 40) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25CBM821; 41) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25DBP029; 42) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25EBU142; 43) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25GBO011; 44) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25HBG682
FEI Number 1417592
Recalling Firm/
Manufacturer
Medline Industries, LP
3 Lakes Dr
Northfield IL 60093-2753
For Additional Information Contact Haley Barclay
866-359-1704
Manufacturer Reason
for Recall
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
FDA Determined
Cause 2
Process control
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 1/7/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/14/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-003-
Quantity in Commerce4064 units
DistributionWorldwide - US Nationwide distribution.