| | Class 2 Device Recall DEX Ophthalmic Stiff Super Grip Forceps |  |
| Date Initiated by Firm | March 25, 2026 |
| Date Posted | May 06, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2080-2026 |
| Recall Event ID |
98660 |
| Product Classification |
Forceps, ophthalmic - Product Code HNR
|
| Product | DEX Ophthalmic Tissue Forceps, 27ga DEX Stiff Super Grip Forceps, Model/Catalog Number: DVF4034-27 |
| Code Information |
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250429(17)280429(10)M47331
Lot Number: M47331
Expiration Date: 04/29/2028
Model No: DVF4034-27
UDI-DI: (01)1081748902557(11)250527(17)280527(10)M48044
Lot Number: M48044
Expiration Date: 05/27/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250610(17)280610(10)M48163
Lot Number: M48163
Expiration Date: 06/10/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250527(17)280527(10)M47612
Lot Number: M47612
Expiration Date: 05/27/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250617(17)280617(10)M48071
Lot Number: M48071
Expiration Date: 06/17/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250610(17)280610(10)M48164
Lot Number: M48164
Expiration Date: 06/10/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250626(17)280626(10)M48470
Lot Number: M48470
Expiration Date: 06/26/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250626(17)280626(10)M48471
Lot Number: M48471
Expiration Date: 06/26/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250527(17)280527(10)M48043
Lot Number: M48043
Expiration Date: 05/27/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250527(17)280527(10)M47611
Lot Number: M47611
Expiration Date: 05/27/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250626(17)280626(10)M48472
Lot Number: M48472
Expiration Date: 06/26/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250709(17)280709(10)M48658
Lot Number: M48658
Expiration Date: 07/09/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250709(17)280709(10)M48657
Lot Number: M48657
Expiration Date: 07/09/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250807(17)280807(10)M49169
Lot Number: M49169
Expiration Date: 08/07/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250806(17)280806(10)M49170
Lot Number: M49170
Expiration Date: 08/06/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250819(17)280819(10)M49287
Lot Number: M49287
Expiration Date: 08/19/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250825(17)280825(10)M49777
Lot Number: M49777
Expiration Date: 08/25/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250825(17)280825(10)M49776
Lot Number: M49776
Expiration Date: 08/25/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250826(17)280826(10)M49830
Lot Number: M49830
Expiration Date: 08/26/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250826(17)280826(10)M49831
Lot Number: M49831
Expiration Date: 08/26/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250826(17)280826(10)M49907
Lot Number: M49907
Expiration Date: 08/26/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250904(17)280904(10)M49906
Lot Number: M49906
Expiration Date: 09/04/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250909(17)280909(10)M49925
Lot Number: M49925
Expiration Date: 09/09/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)250909(17)280909(10)M49926
Lot Number: M49926
Expiration Date: 09/09/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251003(17)281003(10)M50538
Lot Number: M50538
Expiration Date: 10/03/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251003(17)281003(10)M50628
Lot Number: M50628
Expiration Date: 10/03/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251003(17)281003(10)M50629
Lot Number: M50629
Expiration Date: 10/03/2028
Model No:DVF4034-27
UDI-DI: (01)10817489025557(11)251008(17)281008(10)M50897
Lot Number: M50897
Expiration Date: 10/08/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251008(17)281008(10)M50898
Lot Number: M50898
Expiration Date: 10/08/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251007(17)281007(10)M50678
Lot Number: M50678
Expiration Date: 10/07/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251015(17)281015(10)M50917
Lot Number: M50917
Expiration Date: 10/15/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251017(17)281017(10)M50932
Lot Number: M50932
Expiration Date: 10/17/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251107(17)281107(10)M50993
Lot Number: M50993
Expiration Date: 11/07/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251113(17)281113(10)M51034
Lot Number: M51034
Expiration Date: 11/13/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251209(17)281209(10)M51076
Lot Number: M51076
Expiration Date: 12/09/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251209(17)281209(10)M51077
Lot Number: M51077
Expiration Date: 12/09/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251209(17)281209(10)M51131
Lot Number: M51131
Expiration Date: 12/09/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251212(17)281212(10)M51235
Lot Number: M51235
Expiration Date: 12/12/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251212(17)281212(10)M51236
Lot Number: M51236
Expiration Date: 12/12/2028
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)251218(17)281218(10)M51292
Lot Number: M51292
Expiration Date: 12/18/2028
Model No: DVF4034-27
UDI-DI: (01)1081748902557(11)251218(17)281218(10)M51293
Lot Number: M51293
Expiration Date: 12/18/2028
Model No: DVF4034-27
UDI-DI: (01)10718489025557(11)260107(17)290107(10)M51329
Lot Number: 51329
Expiration Date: 01/07/2029
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)260115(17)290115(10)M51630
Lot Number: M51630
Expiration Date: 01/15/2029
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)260115(17)290115(10)M51629
Lot Number: M51629
Expiration Date: 01/15/2029
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)260120(17)290120(10)M51699
Lot Number: M51699
Expiration Date: 01/20/2029
Model No: DVF4034-27
UDI-DI: (01)10817489025557(11)260120(17)290120(10)M51700
Lot Number: M51700
Expiration Date: 01/20/2029 |
| FEI Number |
3007589150
|
Recalling Firm/ Manufacturer |
Katalyst Surgical, LLC
722 Goddard Ave
Chesterfield
MO
63005-1100
|
| For Additional Information Contact | Alex Bergoudian +1-636-5365950 |
Manufacturer Reason for Recall | Ophthalmic Tissue Forceps (DEX") are sterile, hand-held ophthalmic surgical instruments designed for grasping, manipulating, compressing, pulling, or joining eye and/or surrounding tissues during surgical procedures. These devices fall under the GMDN code 62478 entitled Ophthalmic soft-tissue surgical forceps, probe-like, single use . Double pouched in tyvek pouches, sold in boxes of 5 with an inner and outer box. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | A Field Safety Notification: Field Safety Corrective Action for IFU in DEX Forceps and Scissors dated 3/19/26 was sent to customers.
FIELD SAFETY ACTIONS REQUIRED
As a customer identified in the receipt of one or more of the affected lots, you are requested/required to take the following steps:
1. Please confirm receipt of this Field Safety Notification by email, either by replying to the sender or using the email addresses listed in the signature at the bottom of this notice.
2. Review the Individualized List of Affected Lot Numbers by Customer provided in the email for lots that were distributed to your facility. Identify all of these lots that currently remain in your stock and place them on hold.
3. Provide information to Katalyst Surgical team identifying which lots are currently in stock and on hold at your facility. Katalyst Surgical will send a replacement correct IFU for each affected box.
4. Upon receipt of the correct IFU, remove the incorrect IFU, replace the correct IFU in the box, and destroy the incorrect IFU. Approval for Modification including instructions will accompany the replacement IFUs.
5. We recommend you reach out to your customers/users of these DEX products with this information. Katalyst Surgical will gladly provide correct replacement IFU documents to you for further distribution.
Katalyst Surgical deeply regrets the inconvenience this causes. We assure you of our continuous support and commitment to safety and quality. Please do not hesitate to contact our team if you have any further questions. |
| Quantity in Commerce | 1330 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Peru, Argentina, South Korea, Japan, France. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|