Class 2 Device Recall ContinuFlo |
|
Date Initiated by Firm |
July 16, 2003 |
Date Posted |
August 21, 2003 |
Recall Status1 |
Terminated 3 on June 29, 2004 |
Recall Number |
Z-1169-03 |
Recall Event ID |
26772 |
510(K)Number |
K003225
|
Product Classification |
Set, Administration, Intravascular - Product Code FPA
|
Product |
Continu-Flo Solution Set, 10 drops/mL, 110'', 3 Luer Activated Valves, Male Luer Lock Adapter, catalog 2C9905; Baxter Healthcare Corporation, Deerfield, IL 60015 USA |
Code Information |
catalog 2C9905, lot numbers U556902R through UR126946. Please note: Baxter lot numbers are represented by an internal reference code ''UR'' followed by a six digit sequential number, or by the letter ''U'' followed by a six digit sequential number with and ''R'' suffix. Any lot number appearing within the identified six digit sequential number range is subject to the recall. |
Recalling Firm/ Manufacturer |
Baxter Healthcare Corp. Rt 120 & Wilson Rd Round Lake IL 60073
|
For Additional Information Contact |
Center for One Baxter 800-422-9837
|
Manufacturer Reason for Recall |
There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.
|
FDA Determined Cause 2 |
Other |
Action |
Recalled by letter dated 7/16/03. The accounts were informed of the potential for leaks in the set tubing, and asked to inspect their inventory for the affected product codes and lots, segregate any affected products found and discard them according to local procedures. The customers will receive credit for the discarded affected product. |
Quantity in Commerce |
299,839 units |
Distribution |
Nationwide, Mexico, Canada, Columbia, Australia, New Zealand and Taiwan |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = FPA and Original Applicant = BAXTER HEALTHCARE CORP.
|
|
|
|