| | Class 2 Device Recall P.F.C. |  |
| Date Initiated by Firm | May 26, 2005 |
| Date Posted | August 12, 2005 |
| Recall Status1 |
Terminated 3 on December 02, 2005 |
| Recall Number | Z-1183-05 |
| Recall Event ID |
32192 |
| 510(K)Number | K033272 |
| Product Classification |
Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer - Product Code JWH
|
| Product | Depuy P.F.C. SIGMA stabilized insert; polyethylene tibial insert, Sz 3, 17.5 mm; Ref 1581-23-017. |
| Code Information |
Lots Y45KN4 and Y5CVW4. |
| FEI Number |
1818910
|
Recalling Firm/ Manufacturer |
Depuy Orthopaedics, Inc. 700 Orthopaedic Dr Warsaw IN 46582-3994
|
| For Additional Information Contact | 800-366-8143 |
Manufacturer Reason for Recall | An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant. |
FDA Determined Cause 2 | Other |
| Quantity in Commerce | 12 |
| Distribution | Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|