• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall LIFEPAK 12

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall LIFEPAK 12see related information
Date Initiated by FirmMarch 22, 2006
Date PostedApril 04, 2006
Recall Status1 Terminated 3 on September 06, 2011
Recall NumberZ-0710-06
Recall Event ID 34947
510(K)NumberK010918 
Product Classification Automated External Defibrillators (Non-Wearable) - Product Code MKJ
ProductLIFEPAK 12 defibrillator/monitor
Code Information Domestic Serial Numbers: 34106383, 33996555, 33992033, 33992030, 33992029, 33992031, 33992027, 33990244, 33990241, 33990245, 33990242, 33990240, 34096406, 34086957, 34039438, 34100891, 33917921, 34100498, 34019560, 34011939, 34011940, 33985062, 33985063, 33985064, 33985065, 33985066, 33985067, 33985068, 33985070, 33985069, 33985074, 33985061, 33985071, 33985073, 33915256, 34016156, 34016153, 34016154, 34016155, 34016157, 34016158, 33989218, 33989220, 33989216, 33989219, 33989217, 34039448, 34048193, 34100900, 34020194, 34020192, 34060159, 33988136, 33988137, 34020193, 33995357, 33949669, 34074066, 33913884, 34014205, 11201950, 34091725, 34064561, 33913893, 33913894, 33913895, 34004873, 33995317, 33995318, 33995316, 33995315, 33995319, 34064439, 34064440, 34064441, 34020190, 34020507, 34020187, 34020189, 34020188, 34019563, 34019572, 34019573, 33904159, 33905092, 34095289, 34095297, 34095301, 33991999, 33992000, 33992003, 33992023, 33992028, 33992016, 33991998, 34087619, 34087620, 34087605, 34087610, 34087612, 34087035, 34087892, 34087602, 34087615, 34087616, 34087617, 34087618, 34086700, 34086709, 34086715, 34087613, 34086714, 34087609, 34087607, 34087614, 34087608, 34086718, 34086706, 34086719, 34086711, 34086713, 34086705, 34086717, 34086702, 34087606, 34086703, 34087033, 34086708, 34086712, 34087603, 34086707, 34087604, 34086716, 34087601, 34087034, 34091720, 33949930, 34091120, 34039430, 34039429, 34015786, 34011974, 33982989, 33899585, 33995356, 33995351, 33995350, 33995346, 33995360, 33995355, 33995369, 33995349, 33995348, 33995352, 33995347, 33995359, 33995353, 33995345, 33995358, 33995354, 12737672, 33913878, 34094631, 34049603, 34049601, 34049604, 34049611, 34049606, 34049607, 34049609, 34049612, 34049608, 34049610, 34049605, 34049602, 30859396, 33978921, 33978922, 33978923, 33978924, 34039468, 34039433, 34039470, 34083705, 34096379, 33915357, 33988150, 33992032, 33931367, 33931365, 33931364, 33931363, 33931362, 33931366, 34052702, 34052703, 33795379, 34095303, 34095298, 34095299, 34095304, 34095300, 34096375, 34095314, 34095313, 34095312, 34095307, 34095306, 34095305, 33862271, 34019587, 34020177, 34019586, 34019585, 34090338, 34090337, 34049130, 34043663, 34086675, 34086676, 34086671, 34086672, 34086674, 34086673, 33832318, 33888386, 33888372, 33888377, 33888376, 33888378, 33888379, 33888383, 33888385, 33888384, 33888373, 34044798, 34064654, 34019555, 34019554, 33802295, 33990042, 33982418, 33982417, 34074060, 34074061, 34095296, 34039463, 33950592, 33983052, 34099021, 34095323, 34082638, 34078647, 34078648, 34078646, 34082590, 34082591, 34048199, 34048200, 34049140, 34049141, 34048196, 34048198, 34045124, 34048194, 34010325, 33899587, 34096386, 34074089, 33974887, 33974888, 34078076, 34100491, 34100492, 34099179, 34099180, 34099178, 34099177, 34099176, 34090571, 34067899, 33995913, 33995914, 33974374, 33974373, 34013030, 34083205, 34019584, 10732394, 10732404, 34079488, 34084291, 34084290, 34004900, 34019558, 34019561, 33913867, 33913879, 33913888, 33995281, 33995280, 34090554, 34090556, 34090555, 34052280, 33991462, 34065273, 34048197, 34048195, 34034132, 34004904, 34082921, 34060126, 34060127, 34060128, 34086704, 34086701, 34086710, 33992506, 34074081, 34074073, 33855207, 33995302, 33995301, 34054626, 34054625, 33904163, 12776983, 33997196, 33987972, 33987973, 33928602, 34010320, 33974366, 34095302, 33938189, 34040207, 34039481, 34060174, 34060158, 34060161, 34060160, 34064667, 34003616, 34039469, 34060746, 34034134, 34106173, 33995035, 34074053, 34010323, 34090366, 34087611, 34090303, 34090302, 34091128, 34091129, 34095293, 33897593, 34078783, 34078784, 34078785, 34074072, 33992007, 33992020, 33993956, 33992022, 33992002, 33992021, 33992024, 33992025, 33992026, 33992004, 33992009, 33992015, 33992001, 33990248, 33992006, 33992005, 33992008, 33992014, 33992018, 33992017, 33990243, 33990246, 33990247, 33992011, 33992012, 33992013, 33992019, 33992010, 34019904, 33899702, 33960797, 33960796, 33960799, 33960798, 33960795, 33995049, 34061036, 34090589, 33931359, 33931358, 33931355, 33931356, 33931357, 34079490, 34079489, 34079833, 34079492, 34079491, 34079487, 34079486, 34079485, 33936114, 34068040, 34068041, 34078675, 33925842, 33925844, 33925846, 33925841, 33925843, 33925845, 33935283, 33935282, 33935281, 33935280, 33928109, 34090546, 34039464, 34082588, 33911101, 33911100, 33965863, 34074054, 34070493, 33997671, 33997672, 34078070, 34064659, 33996557, 33996553, 33996551, 33996550, 33996556, 33996549, 33996849, 33996558, 33996554, 33996552, 33993444, 34095321, 34095327, 34095331, 34095330, 34095326, 34095329, 34095320, 34095324, 34095328, 34095322, 33983339, 33976090, 33983346, 33983345, 33983338, 33983337, 33983336, 33983335, 32525390, 33995263, 34020191, 34019826, 34019557, 34019827, 34019556, 13583628, 34005592, 34005591, 34005587, 34005588, 34005589, 34005590, 33969580, 33969582, 33969581, 33969583, 33969584, 33995222, 34045762, 34099186, 34099185, 33991446, 33903858, 33997197, 34084471, 34075613, 34075614.    International Serial Numbers: 34098600, 34098601, 34090290, 34049509, 34049514, 34049515, 34083701, 34083702, 34044802, 33982442, 33982438, 33982439, 33934527, 33903205, 33903206, 33903208, 33903204, 33903198, 33903203, 33903207, 33903201, 33903210, 33903202, 33903199, 33903200, 33934528, 33934531, 33934529, 33934526, 33934530, 33928168, 33903217, 33928174, 33928165, 33982454, 33982437, 33982443, 33982440, 33903216, 33903215, 33982453, 33982452, 33982451, 33982450, 33982441, 33982448, 33982447, 33982446, 33982445, 33982444, 33982449, 33903209, 33928169, 33928160, 33928161, 33928162, 33928164, 33928170, 33928171, 33928172, 33928173, 33928175, 33928176, 33928177, 33928167, 33903211, 33903212, 33903213, 33903214, 33928159, 33950039, 33950038, 33950048, 33952773, 33952774, 33952775, 33952771, 34105152, 33950032, 33876945, 33876930, 33821315, 33876931, 34078216, 33876941, 33876943, 33876942, 33876938, 33876944, 33876952, 33876928, 33876946, 33876951, 33876950, 33876948, 33876940, 33876933, 33876935, 33876936, 33876937, 33950037, 33929937, 33950034, 33950043, 33950042, 33950041, 33950040, 33950053, 33950052, 33950051, 33950049, 33950045, 33950046, 33950047, 33950035, 33950050, 33950054, 33876949, 33876947, 33876939, 33876934, 33876932, 34079098, 33876929, 33950044, 33950033, 33950031, 33950030, 33950036, 34078082, 34078078, 34078081, 34078080, 34079097, 34083393, 34078079, 33985301, 33985300, 33985299, 33985307, 33986568, 33952772, 34079096, 34078215, 34078217, 34078225, 33982471, 33985298, 33985303, 33985304, 33986537, 33985302, 34009595, 34009594, 34009593, 34083729, 34083732, 34077752, 34106204, 34091376, 13669402, 33928605, 33985022, 33983183, 33979582, 33979583, 33974289, 33974293, 33952829, 33952830, 33798619, 33913915, 33798616, 33913909, 33918168, 33918735, 33918736, 33872713, 33918165, 33918166, 33918167, 33918181, 33795390, 33798614, 33896029, 33887441, 33798612, 33795399, 33795397, 33932561, 34094348, 34094349, 34091746, 33986283, 34039425, 33986294, 33983009, 33795389, 33983184, 33983178, 33983010, 33961531, 34074827, 34087714, 33974286, 34049521, 34049526, 34049527, 34049528, 34049529, 34049530, 34053704, 34049523, 33913897, 33913904, 34008743, 34008737, 34003664, 34003634, 34020680, 34020681, 34020164, 34020510, 34034118, 34034725, 33983556, 33974288, 33974287, 33979581, 33974290, 33974291, 34049522, 33983555, 33983554, 33979580, 33974229, 33974292, 33986320, 33913913, 34049520, 33986261, 33974295, 34053733, 34049525, 34053699, 34049524, 33913899, 33913900, 33913901, 33913902, 33985306, 33913905, 34069183, 33985305, 33913898, 33913903, 33913906, 33986313, 33986285, 34003817, 34049519, 34003650, 34003796, 34004911, 34008686, 34008744, 34008742, 34008740, 34008739, 34008738, 34008687, 34008745, 34008741, 34003658, 34020516, 34084458, 34084468, 34074427, 34084092, 34087711, 34087496, 33937399, 33937400, 33937398, 34090742, 33937401, 33937404, 34074835, 33937402, 34083715, 33937403, 33937397, 34090743, 33937405, 33910410, 33910411, 34090741, 34083714, 34044811, 34044812, 34039416, 34074102, 33979590, 34074107, 33983550, 34039419, 34073371, 34039413, 33958404, 33958774, 34074106, 33855054, 33921591, 33921592, 33880013, 33880012, 33880011, 33895654, 33900603, 33899646, 33895655, 33961098, 33961097, 33978886, 34074101, 34106223, 33973952, 33978885, 33979004, 33978888, 33978887, 33983551, 34039415, 34039409, 34039408, 33979564, 33979567, 34079036, 33895653, 33895652, 33895651, 33895650, 33895649, 33893664, 33892562, 33945926, 34074108, 34064246, 34039418, 33895656, 34064247, 34074105, 33968953, 34074099, 34074595, 34074596, 34074104, 34074098, 33949217, 33968972, 33969016, 34074103, 34008392, 34039421, 34012451, 34039422, 34040777, 34040525, 33969015, 34039406, 33968973, 34039405, 33968970, 33968962, 33968966, 33968964, 33969014, 34039407, 33968968, 34034122, 34034124, 34034127, 34034129, 34073370, 33983559, 34073365, 34073366, 34073376, 34065724, 34073364, 33905086, 33905087, 33905088, 33905089, 34003929, 34073369, 34073372, 34073373, 34073374, 34073375, 33968961, 33968963, 33983561, 34079035, 34043644, 34079040, 34079039, 34079037, 34073368, 34073367, 34073363, 34079038, 33928097, 34039410, 34043602, 33983560, 34039411, 34039412, 34043601, 34039414, 33928392, 33968957, 33968956, 33968955, 33968954, 34004876, 34043599, 34044807, 34004928, 34004927, 33968971, 34039417, 34003928, 34004922, 34004924, 34039423, 33968959, 34106425, 34106224, 34106225, 34086724, 34091392, 34086725.  
FEI Number 3015876
Recalling Firm/
Manufacturer
Medtronic Emergency Response Systems, Inc.
11811 Willows Rd Ne
Redmond WA 98052-2003
For Additional Information ContactTodd Bandy
425-867-4514
Manufacturer Reason
for Recall
Therapy connectors may have connector sockets that do not meet design specs - short drill depth. Interruption of the connections may cause unexpected, intermittent or complete loss of pacing and defibrillation functions.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionOn 3/22/06, the firm began sending out letters dated March 2006, and flagged URGENT MEDICAL DEVICE CORRECTION. The letters advised customers of the issue and that a service representative would visit and inspect the therapy connector sockets, and replace the therapy connector and the mating cable if short drill sockets are found.
Quantity in Commerce942 devices
DistributionNationwide, and Canada, Netherlands, Germany, Australia, Guam, Hong Kong, New Zealand, New Caledonia, Argentina, Brazil, Chile, Costa Rica, Mexico, Uruguay, Venezuela. Domestic distribution includes hospitals, clinics,
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MKJ
-
-