|
Class 2 Device Recall Allura Xper FD 10/10 |
|
Date Initiated by Firm |
August 24, 2006 |
Date Posted |
November 08, 2006 |
Recall Status1 |
Terminated 3 on May 29, 2008 |
Recall Number |
Z-0161-2007 |
Recall Event ID |
36196 |
510(K)Number |
K041949
|
Product Classification |
x-ray, angiographic system - Product Code IZI
|
Product |
Allura Xper FD 10/10 x-ray, angiographic system |
Code Information |
Site numbers for both FD 10/10 and FD 10: 532800, 544506, 543702, 543289, 543209, 543200, 543194, 542714, 542557, 542413, 542117, 542113, 541822, 541592, 541577, 541409, 541029, 541026, 540841, 540488, 540386, 540311, 540301, 540269, 540237, 540235, 540234, 540184, 540070, 540016, 539641, 539628, 539595, 539480, 539455, 539326, 539107, 538903, 538744, 538516, 538514, 538486, 538485, 538476, 538409, 538368, 538157, 538084, 538028, 537855, 537804, 537745, 537698, 537631, 537630, 536966, 536721, 536379, 536138, 536042, 536037, 536035, 536034, 536031, 536028, 535769, 535667, 535416, 535415, 534941, 534731, 534678, 534638, 534634, 533762, 533397, 533289, 533283, 533282, 533278, 532799, 532611, 531942, 530667, 530666, 104497, 103283 |
Recalling Firm/ Manufacturer |
Philips Medical Systems North America Co. Phillips 22100 Bothell Everett Hwy Bothell WA 98021-8431
|
For Additional Information Contact |
Sarah Baxter 425-487-7665
|
Manufacturer Reason for Recall |
Potential for x-ray system to freeze during an examination, and inability of table to be re-positioned.
|
FDA Determined Cause 2 |
Other |
Action |
A letter dated September 1, 2006 was sent to customers on September 5, 2006. The letter advised customers of the problem and advised a service representative will visit the consignee to install a new software upgrade. |
Quantity in Commerce |
90 total devices - FD 10/10 & FD 10 |
Distribution |
Nationwide. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = IZI and Original Applicant = PHILIPS MEDICAL SYSTEMS NORTH AMERICA CO.
|
|
|
|