| Date Initiated by Firm | August 24, 2006 |
| Date Posted | November 08, 2006 |
| Recall Status1 |
Terminated 3 on May 29, 2008 |
| Recall Number | Z-0162-2007 |
| Recall Event ID |
36196 |
| 510(K)Number | K041949 |
| Product Classification |
x-ray, angiographic system - Product Code IZI
|
| Product | Allura Xper FD 10 x-ray, angiographic system |
| Code Information |
Site codes listed in Z-0161-2007 for FD 10/10. |
| FEI Number |
1000524572
|
Recalling Firm/ Manufacturer |
Philips Medical Systems North America Co. Phillips 22100 Bothell Everett Hwy Bothell WA 98021-8431
|
| For Additional Information Contact | Sarah Baxter 425-487-7665 |
Manufacturer Reason for Recall | Potential for x-ray system to freeze during an examination, and inability of table to be re-positioned. |
FDA Determined Cause 2 | Other |
| Action | A letter dated September 1, 2006 was sent to customers on September 5, 2006. The letter advised customers of the problem and advised a service representative will visit the consignee to install a new software upgrade. |
| Quantity in Commerce | 90 total devices - FD 10/10 & FD 10 |
| Distribution | Nationwide. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IZI
|