| | Class 2 Device Recall Pulmonetic Systems LTV 950 Ventilator |  |
| Date Initiated by Firm | February 22, 2008 |
| Date Posted | August 13, 2008 |
| Recall Status1 |
Terminated 3 on August 16, 2012 |
| Recall Number | Z-1464-2008 |
| Recall Event ID |
46850 |
| 510(K)Number | K002881 K010508 K040540 K981371 K984056 |
| Product Classification |
Ventilator - Product Code CBK
|
| Product | Viasys healthcare, Pulmonetic Systems LTV 950 Ventilator
The ventilator is suitable for use in institutional, home or transport settings. Fully self-contained ventilator in a compact and lightweight design. It can operate from an internal battery, external battery or AC power. Applications include long term acute care or short term / recovery room.
Pulmonetic Systems, 17400 Medina Rd., Suite 100, Minneapolis, MN 55447. |
| Code Information |
Serial numbers: C01001, C01002, C01003, C01022, C01026, C01027, C01029, C01030, C01031, C01032, C01033, C01034, C01039, C01045, C01054, C01055, C01060, C01061, C01062, C01064, C01066, C01067, C01068, C01069, C01070, C01072, C01074, C01076, C01077, C01078, C01079, C01080, C01081, C01082, C01084, C01085, C01086, C01087, C01089, C01090, C01092, C01093, C01094, C01095, C01096, C01097, C01098, C01099, C01100, C01101, C01102, C01103, C01104, C01105, C01106, C01107, C01108, C01110, C01111, C01113, C01114, C01115, C01116, C01117, C01118, C01119, C01122, C01123, C01124, C01126, C01130, C01131, C01132, C01133, C01134, C01138, C01139, C01140, C01141, C01142, C01144, C01147, C01148, C01149, C01151, C01153, C01154, C01163, C01166, C01167, C01173, C01177, C01179, C01180, C01182, C01183, C01185, C01186, C01187, C01188, C01189, C01192, C01193, C01194, C01195, C01196, C01198, C01199, C01200, C01201, C01203, C01204, C01205, C01206, C01207, C01208, C01209, C01210, C01211, C01212, C01213, C01214, C01228, C01231, C01232, C01237, C01240, C01250, C01251, C01252, C01255, C01256, C01257, C01259, C01261, C01262, C01264, C01265, C01267, C01268, C01269, C01270, C01271, C01273, C01274, C01283, C01284, C01286, C01292, C01302, C01303, C01304, C01305, C01307, C01309, C01310, C01311, C01312, C01313, C01314, C01315, C01319, C01320, C01321, C01322, C01323, C01326, C01327, C01328, C01329, C01333, C01339, C01340, C01343, C01344, C01345, C01346, C01348, C01350, C01351, C01353, C01354, C01355, C01356, C01357, C01358, C01359, C01363, C01364, C01374, C01385, C01386, C01388, C01389, C01390, C01391, C01392, C01393, C01394, C01395, C01398, C01401, C01402, C01403, C01405, C01406, C01407, C01408, C01413, C01420, C01429, C01430, C01431, C01432, C01433, C01435, C01437, C01438, C01439, C01441, C01443, C01444, C01445, C01446, C01448, C01452, C01453, C01454, C01461, C01475, C01476, C01477, C01478, C01479, C01480, C01481, C01482, C01483, C01484, C01485, C01486, C01488, C01489, C01492, C01494, C01496, C01497, C01499, C01505, C01506, C01514, C01519, C01520, C01521, C01522, C01523, C01524, C01525, C01526, C01527, C01529, C01530, C01531, C01532, C01533, C01534, C01536, C01537, C01539, C01540, C01541, C01553, C01557, C01559, C01560, C01561, C01562, C01563, C01564, C01565, C01567, C01570, C01571, C01572, C01573, C01574, C01576, C01577, C01578, C01579, C01580, C01581, C01582, C01583, C01585, C01586, C01589, C01591, C01592, C01594, C01595, C01596, C01597, C01598, C01600, C01601, C01603, C01604, C01611, C01612, C01613, C01615, C01616, C01624, C01625, C01632, C01635, C01638, C01639, C01640, C01641, C01642, C01644, C01645, C01646, C01647, C01648, C01650, C01651, C01652, C01654, C01656, C01657, C01658, C01659, C01660, C01662, C01670, C01681, C01694, C01695, C01697, C01699, C01700, C01701, C01703, C01705, C01707, C01708, C01710, C01711, C01712, C01713, C01714, C01716, C01717, C01725, C01743, C01744, C01746, C01747, C01749, C01750, C01752, C01753, C01754, C01757, C01762, C01764, C01768, C01769, C01770, C01771, C01772, C01777, C01778, C01779, C01780, C01781, C01783, C01784, C01785, C01786, C01789, C01790, C01791, C01792, C01794, C01795, C01796, C01797, C01824, C01829, C01831, C01832, C01834, C01836, C01837, C01838, C01840, C01841, C01842, C01843, C01844, C01845, C01845, C01848, C01849, C01850, C01852, C01853, C01854, C01855, C01858, C01861, C01862, C01863, C01864, C01865, C01866, C01867, C01868, C01869, C01871, C01873, C01874, C01875, C01876, C01877, C01878, C01880, C01881, C01883, C01884, C01885, C01895, C01896, C01901, C01903, C01904, C01907, C01908, C01909, C01910, C01913, C01914, C01916, C01917, C01919, C01920, C01921, C01922, C01923, C01929, C01930, C01933, C01935, C01939, C01941, C01942, C01943, C01945, C01946, C01948, C01949, C01950, C01966, C01980, C01981, C01982, C01983, C01984, C01985, C01986, C01987, C01988, C01989, C01991, C01993, C01994, C01995, C01996, C01998, C01999, C02000, C02001, C02005, C02006, C02007, C02010, C02021, C02023, C02025, C02026, C02027, C02028, C02029, C02030, C02031, C02032, C02033, C02036, C02037, C02038, C02039, C02041, C02042, C02043, C02044, C02047, C02048, C02049, C02050, C02071, C02072, C02073, C02074, C02076, C02077, C02078, C02080, C02083, C02084, C02085, C02087, C02089, C02090, C02091, C02092, C02094, C02095, C02097, C02102, C02103, C02104, C02105, C02108, C02110, C02111, C02114, C02115, C02123, C02125, C02127, C02128, C02130, C02132, C02133, C02134, C02135, C02136, C02137, C02139, C02140, C02141, C02143, C02144, C02145, C02146, C02147, C02148, C02149, C02150, C02151, C02152, C02156, C02158, C02159, C02160, C02163, C02164, C02186, C02203, C02204, C02207, C02208, C02209, C02210, C02211, C02212, C02214, C02215, C02216, C02217, C02219, C02220, C02221, C02222, C02223, C02224, C02225, C02226, C02227, C02228, C02234, C02238, C02239, C02240, C02241, C02242, C02245, C02246, C02248, C02249, C02250, C02251, C02252, C02253, C02254, C02255, C02260, C02261, C02262, C02263, C02265, C02266, C02268, C02269, C02270, C02275, C02276, C02277, C02280, C02284, C02285, C02286, C02287, C02288, C02289, C02290, C02291, C02292, C02293, C02295, C02297, C02302, C02303, C02304, C02305, C02306, C02307, C02308, C02309, C02311, C02313, C02314, C02315, C02316, C02317, C02319, C02320, C02324, C02326, C02327, C02328, C02358, C02359, C02360, C02363, C02367, C02371, C02375, C02376, C02379, C02381, C02382, C02384, C02385, C02386, C02388, C02389, C02390, C02391, C02392, C02393, C02394, C02396, C02397, C02398, C02399, C02400, C02401, C02402, C02403, C02404, C02405, C02406, C02408, C02409, C02410, C02411, C02412, C02413, C02414, C02415, C02416, C02417, C02418, C02420, C02421, C02430, C02434, C02444, C02445, C02446, C02448, C02450, C02451, C02452, C02454, C02455, C02460, C02461, C02462, C02463, C02466, C02468, C02470, C02472, C02480, C02482, C02483, C02486, C02487, C02488, C02490, C02492, C02493, C02495, C02496, C02498, C02499, C02500, C02501, C02502, C02503, C02506, C02507, C02508, C02509, C02511, C02515, C02516, C02519, C02528, C02530, C02531, C02536, C02537, C02544, C02546, C02547, C02548, C02549, C02550, C02552, C02554, C02556, C02558, C02560, C02561, C02562, C02567, C02568, C02569, C02570, C02571, C02572, C02573, C02574, C02575, C02576, C02578, C02580, C02581, C02582, C02583, C02584, C02585, C02586, C02587, C02588, C02589, C02590, C02597, C02600, C02616, C02617, C02620, C02621, C02624, C02626, C02633, C02634, C02638, C02640, C02642, C02643, C02645, C02647, C02648, C02650, C02651, C02653, C02655, C02656, C02657, C02658, C02659, C02660, C02662, C02664, C02665, C02666, C02667, C02669, C02671, C02672, C02673, C02674, C02677, C02678, C02680, C02681, C02682, C02683, C02685, C02686, C02687, C02688, C02689, C02690, C02691, C02692, C02693, C02696, C02697, C02698, C02699, C02700, C02703, C02704, C02705, C02706, C02711, C02712, C02713, C02715, C02716, C02719, C02720, C02721, C02722, C02723, C02724, C02726, C02727, C02728, C02729, C02730, C02731, C02732, C02733, C02735, C02736, C02737, C02738, C02739, C02740, C02742, C02743, C02744, C02746, C02747, C02749, C02750, C02751, C02752, C02753, C02754, C02755, C02756, C02757, C02758, C02759, C02760, C02761, C02767, C02768, C02769, C02770, C02771, C02772, C02773, C02774, C02776, C02779, C02780, C02810, C02812, C02813, C02814, C02815, C02816, C02817, C02818, C02819, C02821, C02822, C02823, C02824, C02825, C02826, C02827, C02828, C02829, C02830, C02831, C02832, C02833, C02835, C02836, C02838, C02839, C02840, C02841, C02845, C02846, C02848, C02849, C02851, C02852, C02853, C02854, C02855, C02856, C02858, C02859, C02860, C02861, C02862, C02863, C02864, C02865, C02866, C02867, C02868, C02869, C02870, C02873, C02874, C02876, C02877, C02878, C02879, C02880, C02881, C02882, C02883, C02884, C02885, C02886, C02887, C02888, C02889, C02890, C02891, C02893, C02894, C02895, C02896, C02898, C02899, C02900, C02901, C02904, C02905, C02906, C02907, C02908, C02910, C02911, C02912, C02913, C02915, C02916, C02918, C02919, C02920, C02921, C02922, C02923, C02924, C02925, C02926, C02927, C02928, C02929, C02931, C02932, C02933, C02934, C02935, C02936, C02938, C02939, C02940, C02941, C02942, C02944, C02947, C02948, C02949, C02950, C02951, C02953, C02954, C02957, C02958, C02959, C02960, C02961, C02962, C02963, C02964, C02965, C02967, C02968, C02969, C02970, C02971, C02972, C02973, C02974, C02975, C02976, C02977, C02978, C02981, C02982, C02983, C02984, C02987, C02988, C02989, C02990, C02991, C02992, C02993, C02994, C02995, C02996, C02997, C02998, C02999, C03000, C03001, C03003, C03005, C03006, C03007, C03008, C03009, C03010, C03011, C03013, C03014, C03015, C03016, C03018, C03021, C03023, C03026, C03027, C03028, C03030, C03031, C03033, C03034, C03035, C03036, C03037, C03039, C03041, C03042, C03043, C03044, C03045, C03046, C03047, C03048, C03049, C03050, C03051, C03053, C03056, C03057, C03059, C03060, C03061, C03063, C03064, C03067, C03068, C03070, C03071, C03072, C03073, C03075, C03076, C03077, C03078, C03079, C03080, C03084, C03085, C03087, C03088, C03089, C03093, C03094, C03095, C03097, C03098, C03099, C03100, C03102, C03103, C03107, C03109, C03110, C03111, C03114, C03115, C03118, C03119, C03120, C03121, C03124, C03125, C03127, C03128, C03131, C03133, C03134, C03135, C03136, C03139, C03140, C03141, C03147, C03149, C03151, C03152, C03153, C03154, C03156, C03157, C03158, C03159, C03160, C03161, C03162, C03163, C03164, C03165, C03166, C03167, C03180, C03184, C03185, C03189, C03193, C03198, C03200, C03201, C03204, C03205, C03206, C03207, C03209, C03210, C03212, C03215, C03218, C03222, C03223, C03224, C03226, C03228, C03229, C03231, C03232, C03233, C03234, C03235, C03236, C03237, C03239, C03240, C03241, C03242, C03243, C03244, C03245, C03246, C03247, C03248, C03249, C03250, C03251, C03254, C03255, C03257, C03259, C03260, C03261, C03262, C03264, C03265, C03266, C03268, C03269, C03271, C03272, C03275, C03276, C03278, C03279, C03280, C03281, C03282, C03283, C03284, C03285, C03286, C03287, C03289, C03290, C03291, C03292, C03294, C03296, C03297, C03298, C03299, C03300, C03302, C03303, C03304, C03306, C03307, C03308, C03310, C03312, C03313, C03316, C03318, C03319, C03320, C03322, C03324, C03325, C03326, C03328, C03330, C03331, C03332, C03335, C03336, C03338, C03346, C03349, C03352, C03354, C03362, C03364, C03368, C03369, C03370, C03374, C03375, C03377, C03379, C03380, C03381, C03382, C03383, C03384, C03386, C03387, C03388, C03389, C03390, C03391, C03392, C03393, C03394, C03395, C03397, C03398, C03399, C03400, C03401, C03402, C03403, C03404, C03405, C03408, C03412, C03413, C03414, C03416, C03418, C03419, C03420, C03421, C03422, C03424, C03426, C03427, C03428, C03429, C03430, C03431, C03432, C03433, C03434, C03436, C03438, C03439, C03440, C03441, C03443, C03446, C03449, C03451, C03454, C03457, C03458, C03460, C03461, C03462, C03463, C03465, C03467, C03469, C03472, C03473, C03474, C03475, C03476, C03477, C03478, C03479, C03480, C03481, C03482, C03486, C03487, C03488, C03489, C03491, C03493, C03494, C03495, C03496, C03497, C03498, C03499, C03501, C03503, C03504, C03505, C03506, C03508, C03509, C03510, C03513, C03514, C03516, C03517, C03518, C03521, C03523, C03524, C03525, C03526, C03527, C03529, C03531, C03532, C03533, C03534, C03536, C03537, C03538, C03540, C03544, C03546, C03549, C03550, C03551, C03554, C03557, C03558, C03559, C03560, C03561, C03562, C03563, C03564, C03566, C03568, C03569, C03573, C03574, C03575, C03576, C03578, C03580, C03583, C03584, C03589, C03592, C03593, C03594, C03595, C03596, C03598, C03599, C03601, C03602, C03603, C03604, C03606, C03608, C03609, C03614, C03617, C03620, C03621, C03622, C03624, C03629, C03631, C03633, C03634, C03636, C03637, C03639, C03641, C03642, C03644, C03645, C03647, C03648, C03649, C03652, C03656, C03661, C03662, C03666, C03668, C03669, C03670, C03671, C03673, C03680, C03682, C03683, C03684, C03685, C03687, C03688, C03689, C03690, C03691, C03692, C03694, C03695, C03696, C03697, C03699, C03701, C03702, C03703, C03705, C03706, C03710, C03713, C03714, C03716, C03717, C03718, C03719, C03720, C03722, C03724, C03725, C03726, C03730, C03731, C03732, C03734, C03736, C03737, C03738, C03739, C03741, C03742, C03744, C03745, C03746, C03748, C03749, C03750, C03751, C03752, C03753, C03754, C03756, C03757, C03758, C03760, C03761, C03763, C03765, C03766, C03768, C03770, C03771, C03772, C03773, C03774, C03775, C03776, C03777, C03778, C03779, C03780, C03781, C03782, C03783, C03784, C03787, C03788, C03789, C03790, C03791, C03792, C03793, C03794, C03796, C03797, C03798, C03799, C03801, C03803, C03804, C03805, C03808, C03809, C03811, C03813, C03814, C03815, C03817, C03818, C03819, C03820, C03821, C03822, C03823, C03824, C03825, C03827, C03828, C03829, C03830, C03831, C03832, C03833, C03834, C03836, C03837, C03838, C03839, C03840, C03842, C03843, C03844, C03845, C03850, C03851, C03852, C03857, C03858, C03859, C03861, C03862, C03863, C03864, C03866, C03868, C03872, C03873, C03874, C03875, C03877, C03879, C03880, C03881, C03883, C03887, C03888, C03889, C03890, C03891, C03892, C03893, C03894, C03895, C03897, C03898, C03899, C03900, C03901, C03902, C03906, C03908, C03909, C03910, C03911, C03913, C03915, C03916, C03917, C03918, C03919, C03920, C03924, C03926, C03927, C03928, C03929, C03931, C03932, C03934, C03935, C03937, C03938, C03939, C03940, C03941, C03942, C03943, C03944, C03945, C03946, C03947, C03948, C03949, C03950, C03951, C03952, C03953, C03954, C03955, C03956, C03957, C03959, C03960, C03961, C03962, C03963, C03964, C03965, C03966, C03967, C03968, C03969, C03970, C03971, C03972, C03973, C03975, C03976, C03977, C03978, C03980, C03981, C03982, C03983, C03984, C03985, C03987, C03988, C03989, C03990, C03991, C03992, C03993, C03994, C03995, C03996, C03997, C04000, C04001, C04002, C04003, C04004, C04005, C04006, C04007, C04010, C04011, C04012, C04013, C04014, C04016, C04017, C04018, C04019, C04021, C04025, C04026, C04027, C04031, C04032, C04033, C04034, C04037, C04039, C04040, C04042, C04044, C04046, C04047, C04048, C04050, C04051, C04052, C04053, C04055, C04056, C04057, C04058, C04059, C04060, C04061, C04062, C04063, C04064, C04068, C04069, C04070, C04071, C04072, C04073, C04074, C04075, C04076, C04077, C04078, C04080, C04081, C04082, C04083, C04084, C04085, C04086, C04087, C04088, C04089, C04091, C04092, C04093, C04095, C04096, C04097, C04098, C04099, C04102, C04103, C04104, C04105, C04106, C04107, C04108, C04109, C04111, C04112, C04113, C04114, C04115, C04116, C04117, C04118, C04119, C04120, C04121, C04122, C04123, C04126, C04128, C04129, C04130, C04133, C04134, C04136, C04137, C04138, C04139, C04140, C04141, C04143, C04145, C04146, C04148, C04150, C04151, C04152, C04154, C04155, C04158, C04159, C04160, C04161, C04162, C04163, C04165, C04166, C04167, C04168, C04169, C04170, C04172, C04173, C04174, C04175, C04176, C04177, C04178, C04179, C04180, C04181, C04182, C04183, C04184, C04185, C04186, C04187, C04188, C04189, C04190, C04191, C04192, C04193, C04196, C04198, C04199, C04200, C04202, C04203, C04204, C04208, C04209, C04210, C04211, C04213, C04214, C04215, C04217, C04218, C04219, C04220, C04222, C04223, C04225, C04227, C04228, C04229, C04231, C04232, C04233, C04234, C04235, C04237, C04238, C04240, C04242, C04243, C04244, C04245, C04247, C04248, C04249, C04251, C04253, C04257, C04259, C04260, C04262, C04263, C04264, C04265, C04266, C04268, C04269, C04270, C04271, C04272, C04273, C04274, C04275, C04276, C04277, C04278, C04279, C04282, C04283, C04284, C04286, C04287, C04288, C04289, C04290, C04291, C04292, C04294, C04295, C04296, C04297, C04298, C04299, C04300, C04301, C04302, C04303, C04304, C04305, C04306, C04307, C04308, C04309, C04310, C04311, C04312, C04313, C04315, C04316, C04317, C04318, C04319, C04320, C04321, C04322, C04323, C04325, C04327, C04329, C04330, C04332, C04334, C04335, C04336, C04338, C04339, C04340, C04341, C04342, C04344, C04345, C04347, C04349, C04357, C04358, C04359, C04360, C04361, C04362, C04363, C04364, C04365, C04366, C04367, C04368, C04369, C04370, C04371, C04372, C04375, C04376, C04377, C04378, C04379, C04380, C04381, C04382, C04385, C04387, C04390, C04391, C04392, C04393, C04394, C04395, C04397, C04398, C04399, C04400, C04401, C04402, C04403, C04406, C04407, C04408, C04409, C04410, C04415, C04416, C04417, C04418, C04419, C04420, C04421, C04422, C04426, C04427, C04428, C04429, C04431, C04432, C04433, C04436, C04437, C04438, C04439, C04441, C04442, C04443, C04444, C04445, C04446, C04447, C04449, C04450, C04454, C04456, C04457, C04458, C04459, C04460, C04461, C04462, C04463, C04464, C04465, C04466, C04470, C04474, C04477, C04478, C04479, C04481, C04483, C04484, C04485, C04487, C04488, C04489, C04490, C04491, C04493, C04494, C04495, C04496, C04497, C04498, C04501, C04502, C04503, C04504, C04506, C04507, C04509, C04510, C04511, C04516, C04517, C04519, C04520, C04521, C04524, C04525, C04527, C04528, C04530, C04531, C04532, C04533, C04534, C04535, C04537, C04539, C04541, C04544, C04545, C04546, C04547, C04548, C04550, C04551, C04554, C04559, C04561, C04562, C04563, C04564, C04565, C04566, C04569, C04570, C04571, C04572, C04574, C04577, C04581, C04582, C04584, C04586, C04589, C04590, C04592, C04593, C04594, C04595, C04596, C04597, C04598, C04599, C04600, C04601, C04603, C04604, C04605, C04606, C04608, C04609, C04611, C04612, C04615, C04616, C04617, C04619, C04620, C04624, C04625, C04626, C04628, C04630, C04632, C04635, C04636, C04637, C04638, C04639, C04641, C04643, C04645, C04646, C04649, C04650, C04651, C04652, C04653, C04654, C04656, C04657, C04658, C04659, C04660, C04662, C04664, C04674, C04677, C04678, C04680, C04684, C04685, C04691, C04698, C04699, C04701, C04702, C04707, C04708, C04709, C04711, C04712, C04716, C04717, C04719, C04721, C04724, C04729, C04730, C04732, C04733, C04734, C04735, C04737, C04739, C04742, C04748, C04749, C04750, C04752, C04753, C04754, C04755, C04756, C04757, C04759, C04762, C04763, C04765, C04767, C04768, C04769, C04770, C04771, C04772, C04773, C04774, C04775, C04776, C04777, C04778, C04779, C04780, C04781, C04784, C04785, C04786, C04787, C04789, C04790, C04791, C04792, C04793, C04794, C04795, C04796, C04798, C04799, C04800, C04801, C04802, C04803, C04804, C04805, C04806, C04807, C04808, C04809, C04810, C04811, C04812, C04813, C04814, C04815, C04816, C04817, C04818, C04819, C04820, C04823, C04824, C04826, C04828, C04829, C04830, C04831, C04832, C04834, C04836, C04837, C04838, C04840, C04844, C04845, C04850, C04854, C04857, C04858, C04859, C04860, C04861, C04862, C04866, C04875, C04876, C04877, C04878, C04880, C04882, C04883, C04884, C04886, C04887, C04888, C04890, C04894, C04896, C04897, C04898, C04900, C04901, C04903, C04904, C04905, C04906, C04907, C04908, C04909, C04911, C04913, C04914, C04915, C04916, C04917, C04919, C04921, C04922, C04923, C04924, C04925, C04926, C04927, C04930, C04934, C04937, C04938, C04939, C04940, C04941, C04943, C04949, C04950, C04951, C04955, C04956, C04957, C04959, C04960, C04963, C04964, C04966, C04967, C04968, C04970, C04972, C04974, C04975, C04976, C04978, C04981, C04982, C04984, C04985, C04986, C04987, C04991, C04992, C04993, C04995, C04997, C05002, C05003, C05004, C05006, C05007, C05008, C05009, C05010, C05011, C05015, C05016, C05018, C05020, C05022, C05024, C05025, C05026, C05027, C05028, C05029, C05032, C05033, C05034, C05035, C05036, C05039, C05041, C05042, C05043, C05044, C05045, C05046, C05047, C05048, C05049, C05050, C05051, C05052, C05056, C05057, C05059, C05060, C05061, C05063, C05064, C05065, C05066, C05068, C05069, C05070, C05071, C05072, C05073, C05074, C05075, C05076, C05077, C05078, C05080, C05081, C05082, C05083, C05085, C05087, C05088, C05090, C05091, C05092, C05094, C05096, C05097, C05099, C05102, C05103, C05104, C05106, C05107, C05109, C05111, C05112, C05114, C05117, C05118, C05121, C05122, C05124, C05125, C05126, C05127, C05129, C05130, C05131, C05133, C05134, C05135, C05138, C05139, C05140, C05142, C05143, C05144, C05146, C05147, C05148, C05149, C05152, C05153, C05154, C05156, C05157, C05158, C05159, C05160, C05162, C05164, C05165, C05168, C05169, C05170, C05173, C05174, C05175, C05176, C05177, C05178, C05179, C05180, C05181, C05183, C05184, C05185, C05186, C05187, C05188, C05189, C05191, C05192, C05193, C05194, C05195, C05196, C05197, C05199, C05200, C05201, C05202, C05203, C05204, C05207, C05208, C05209, C05210, C05211, C05213, C05214, C05215, C05216, C05217, C05218, C05221, C05223, C05224, C05225, C05226, C05228, C05229, C05230, C05231, C05233, C05236, C05237, C05241, C05242, C05243, C05244, C05245, C05253, C05254, C05255, C05256, C05257, C05258, C05259, C05260, C05263, C05264, C05265, C05266, C05267, C05268, C05269, C05271, C05272, C05273, C05274, C05276, C05277, C05278, C05280, C05281, C05282, C05283, C05284, C05285, C05286, C05287, C05288, C05289, C05290, C05294, C05295, C05296, C05297, C05298, C05299, C05301, C05302, C05305, C05307, C05308, C05309, C05310, C05312, C05313, C05315, C05316, C05318, C05319, C05320, C05321, C05322, C05323, C05326, C05327, C05328, C05329, C05331, C05332, C05333, C05335, C05337, C05338, C05340, C05344, C05345, C05346, C05348, C05350, C05351, C05352, C05353, C05354, C05355, C05357, C05359, C05361, C05362, C05363, C05365, C05366, C05367, C05368, C05369, C05370, C05371, C05373, C05374, C05377, C05378, C05380, C05383, C05385, C05388, C05392, C05393, C05395, C05396, C05397, C05398, C05399, C05400, C05401, C05402, C05403, C05406, C05409, C05412, C05414, C05415, C05416, C05417, C05418, C05419, C05420, C05421, C05422, C05423, C05424, C05426, C05427, C05428, C05430, C05431, C05432, C05433, C05434, C05435, C05437, C05438, C05440, C05441, C05443, C05444, C05448, C05455, C05458, C05459, C05460, C05461, C05463, C05464, C05465, C05467, C05468, C05469, C05470, C05471, C05473, C05474, C05475, C05476, C05478, C05479, C05480, C05483, C05484, C05485, C05486, C05487, C05490, C05491, C05492, C05493, C05494, C05495, C05496, C05497, C05498, C05499, C05500, C05501, C05503, C05504, C05505, C05506, C05507, C05508, C05510, C05512, C05513, C05514, C05515, C05517, C05518, C05519, C05520, C05521, C05522, C05523, C05524, C05526, C05527, C05528, C05529, C05530, C05531, C05532, C05533, C05534, C05535, C05536, C05537, C05538, C05539, C05544, C05545, C05546, C05548, C05549, C05550, C05551, C05552, C05553, C05554, C05557, C05558, C05560, C05562, C05563, C05564, C05565, C05566, C05567, C05569, C05571, C05572, C05573, C05574, C05575, C05576, C05580, C05581, C05582, C05583, C05588, C05589, C05590, C05591, C05592, C05594, C05595, C05597, C05598, C05599, C05601, C05605, C05606, C05610, C05611, C05613, C05614, C05615, C05617, C05618, C05621, C05622, C05623, C05624, C05628, C05629, C05631, C05632, C05633, C05634, C05636, C05637, C05639, C05642, C05645, C05646, C05647, C05651, C05655, C05656, C05660, C05661, C05662, C05663, C05664, C05665, C05666, C05667, C05668, C05671, C05672, C05673, C05674, C05675, C05676, C05677, C05679, C05680, C05681, C05682, C05683, C05684, C05685, C05686, C05689, C05690, C05691, C05692, C05693, C05695, C05696, C05697, C05698, C05699, C05701, C05702, C05703, C05704, C05705, C05706, C05707, C05708, C05709, C05710, C05712, C05714, C05715, C05716, C05717, C05718, C05719, C05721, C05722, C05723, C05725, C05727, C05728, C05729, C05732, C05734, C05735, C05736, C05738, C05739, C05740, C05745, C05748, C05750, C05751, C05752, C05754, C05755, C05757, C05760, C05761, C05764, C05765, C05767, C05768, C05769, C05770, C05771, C05772, C05773, C05774, C05777, C05778, C05779, C05781, C05784, C05785, C05786, C05787, C05791, C05792, C05793, C05794, C05796, C05797, C05802, C05804, C05805, C05806, C05807, C05808, C05809, C05811, C05812, C05813, C05814, C05815, C05816, C06002, C06003, C06008, C06009, C06013, C06017, C06020, C06021, C06022, C06024, C06026, C06031, C06032, C06034, C06035, C06037, C06038, C06039, C06041, C06042, C06043, C06044, C06047, C06048, C06049, C06050, C06055, C06056, C06057, C06058, C06059, C06060, C06061, C06062, C06063, C06064, C06065, C06066, C06067, C06069, C06071, C06072, C06073, C06074, C06075, C06077, C06078, C06079, C06080, C06082, C06083, C06084, C06088, C06089, C06090, C06091, C06092, C06093, C06094, C06095, C06096, C06097, C06098, C06099, C06100, C06101, C06102, C06103, C06104, C06105, C06106, C06107, C06108, C06110, C06111, C06112, C06114, C06115, C06117, C06118, C06119, C06120, C06121, C06122, C06123, C06124, C06125, C06126, C06127, C06130, C06131, C06132, C06133, C06134, C06136, C06139, C06142, C06145, C06150, C06155, C06156, C06157, C06159, C06161, C06162, C06165, C06171, C06174, C06177, C06179, C06180, C06182, C06183, C06188, C06190, C06192, C06193, C06194, C06195, C06197, C06199, C06201, C06202, C06203, C06207, C06208, C06209, C06211, C06212, C06213, C06214, C06216, C06217, C06218, C06219, C06220, C06221, C06224, C06228, C06229, C06230, C06232, C06233, C06234, C06235, C06236, C06237, C06238, C06239, C06240, C06241, C06242, C06243, C06244, C06246, C06247, C06248, C06249, C06250, C06251, C06252, C06253, C06254, C06257, C06258, C06259, C06260, C06261, C06262, C06263, C06264, C06265, C06266, C06267, C06268, C06269, C06271, C06272, C06273, C06284, C06286, C06290, C06293, C06296, C06300, C06304, C06310, C06311, C06313, C06315, C06317, C06319, C06320, C06324, C06326, C06327, C06328, C06329, C06331, C06333, C06334, C06336, C06337, C06341, C06342, C06344, C06345, C06346, C06347, C06349, C06353, C06354, C06356, C06357, C06359, C06360, C06362, C06363, C06364, C06366, C06368, C06369, C06370, C06371, C06372, C06373, C06374, C06376, C06378, C06379, C06381, C06382, C06383, C06384, C06385, C06386, C06387, C06388, C06389, C06390, C06391, C06392, C06393, C06394, C06395, C06396, C06397, C06398, C06399, C06400, C06401, C06403, C06404, C06405, C06407, C06408, C06409, C06411, C06412, C06413, C06414, C06415, C06417, C06418, C06419, C06422, C06423, C06424, C06425, C06427, C06429, C06430, C06431, C06433, C06434, C06435, C06436, C06437, C06438, C06439, C06440, C06443, C06444, C06446, C06448, C06449, C06450, C06452, C06453, C06454, C06455, C06456, C06457, C06458, C06459, C06460, C06461, C06463, C06464, C06465, C06466, C06469, C06470, C06471, C06472, C06473, C06475, C06476, C06478, C06480, C06482, C06483, C06485, C06486, C06489, C06490, C06491, C06494, C06496, C06497, C06498, C06499, C06500, C06501, C06502, C06503, C06504, C06507, C06508, C06510, C06511, C06512, C06513, C06515, C06516, C06518, C06519, C06521, C06522, C06527, C06528, C06529, C06532, C06533, C06534, C06536, C06539, C06540, C06541, C06542, C06543, C06544, C06546, C06547, C06548, C06550, C06554, C06555, C06557, C06558, C06560, C06561, C06562, C06563, C06564, C06566, C06567, C06569, C06571, C06572, C06574, C06575, C06579, C06580, C06581, C06582, C06583, C06584, C06585, C06586, C06587, C06588, C06591, C06592, C06593, C06594, C06595, C06596, C06597, C06598, C06599, C06600, C06601, C06602, C06603, C06604, C06606, C06608, C06609, C06610, C06612, C06614, C06615, C06616, C06617, C06618, C06619, C06621, C06624, C06625, C06626, C06627, C06628, C06629, C06630, C06631, C06632, C06635, C06636, C06637, C06639, C06642, C06643, C06644, C06645, C06646, C06647, C06648, C06649, C06650, C06651, C06652, C06654, C06655, C06656, C06657, C06658, C06659, C06660, C06662, C06663, C06664, C06665, C06666, C06668, C06669, C06672, C06673, C06674, C06675, C06676, C06677, C06678, C06680, C06681, C06682, C06685, C06687, C06690, C06691, C06693, C06694, C06695, C06696, C06697, C06698, C06700, C06702, C06705, C06706, C07001, C07004, C07006, C07007, C07008, C07011, C07012, C07013, C07016, C07017, C07019, C07020, C07021, C07024, C07025, C07026, C07027, C07028, C07029, C07030, C07032, C07033, C07034, C07035, C07037, C07038, C07039, C07040, C07041, C07042, C07044, C07047, C07048, C07049, C07051, C07052, C07053, C07054, C07055, C07056, C07057, C07058, C07060, C07062, C07063, C07064, C07065, C07066, C07067, C07068, C07071, C07074, C07076, C07077, C07079, C07081, C07082, C07084, C07085, C07090, C07095, C07096, C07097, C07098, C07099, C07100, C07103, C07104, C07108, C07112, C07114, C07117, C07119, C07120, C07121, C07127, C07128, C07129, C07131, C07133, C07134, C07135, C07136, C07137, C07138, C07139, C07140, C07146, C07147, C07148, C07153, C07156, C07160, C07163, C07164, C07165, C07166, C07167, C07169, C07171, C07172, C07173, C07174, C07175, C07176, C07177, C07178, C07179, C07180, C07181, C07183, C07184, C07185, C07186, C07187, C07193, C07196, C07197, C07198, C07199, C07200, C07201, C07202, C07203, C07204, C07205, C07206, C07207, C07208, C07209, C07210, C07211, C07212, C07213, C07214, C07215, C07216, C07217, C07218, C07220, C07222, C07223, C07225, C07227, C07228, C07229, C07231, C07232, C07233, C07234, C07239, C07240, C07241, C07242, C07243, C07244, C07245, C07246, C07247, C07248, C07251, C07253, C07254, C07255, C07257, C07258, C07259, C07260, C07261, C07262, C07263, C07264, C07265, C07266, C07267, C07268, C07269, C07270, C07272, C07273, C07274, C07275, C07276, C07277, C07278, C07279, C07280, C07281, C07282, C07283, C07284, C07285, C07288, C07290, C07293, C07295, C07296, C07297, C07299, C07300, C07301, C07303, C07304, C07305, C07306, C07308, C07309, C07310, C07312, C07315, C07316, C07319, C07320, C07321, C07322, C07324, C07325, C07326, C07327, C07328, C07329, C07331, C07332, C07334, C07336, C07337, C07338, C07339, C07340, C07341, C07342, C07343, C07345, C07347, C07348, C07349, C07351, C07353, C07354, C07355, C07357, C07358, C07359, C07360, C07361, C07362, C07363, C07364, C07366, C07367, C07368, C07369, C07371, C07372, C07373, C07374, C07375, C07376, C07377, C07379, C07380, C07381, C07382, C07383, C07384, C07386, C07388, C07389, C07390, C07392, C07393, C07394, C07395, C07396, C07398, C07399, C07400, C07401, C07403, C07404, C07405, C07407, C07408, C07409, C07410, C07411, C07412, C07413, C07415, C07416, C07417, C07418, C07419, C07420, C07421, C07422, C07423, C07424, C07425, C07426, C07427, C07429, C07430, C07431, C07432, C07434, C07436, C07437, C07438, C07440, C07441, C07442, C07444, C07446, C07448, C07449, C07452, C07454, C07455, C07460, C07461, C07462, C07463, C07464, C07465, C07466, C07467, C07468, C07469, C07470, C07472, C07474, C07475, C07476, C07477, C07478, C07479, C07480, C07481, C07482, C07483, C07484, C07485, C07487, C07488, C07489, C07490, C07491, C07492, C07493, C07494, C07495, C07496, C07498, C07499, C07502, C07504, C07505, C07506, C07507, C07508, C07509, C07511, C07512, C07513, C07514, C07515, C07516, C07517, C07519, C07520, C07521, C07522, C07523, C07526, C07527, C07529, C07531, C07532, C07535, C07538, C07539, C07543, C07544, C07545, C07546, C07547, C07549, C07551, C07553, C07555, C07556, C07557, C07558, C07560, C07563, C07565, C07566, C07567, C07569, C07571, C07572, C07573, C07574, C07575, C07578, C07580, C07582, C07583, C07586, C07587, C07588, C07589, C07590, C07591, C07593, C07594, C07595, C07596, C07597, C07598, C07599, C07600, C07601, C07602, C07603, C07604, C07605, C07606, C07607, C07608, C07609, C07610, C07611, C07612, C07613, C07614, C07616, C07617, C07618, C07619, C07620, C07621, C07623, C07624, C07625, C07626, C07627, C07628, C07629, C07630, C07631, C07632, C07633, C07634, C07635, C07636, C07637, C07638, C07639, C07640, C07641, C07642, C07643, C07644, C07646, C07647, C07648, C07649, C07650, C07653, C07654, C07655, C07656, C07657, C07658, C07660, C07661, C07662, C07663, C07664, C07665, C07666, C07667, C07668, C07669, C07670, C07671, C07672, C07673, C07674, C07675, C07676, C07677, C07678, C07679, C07680, C07681, C07682, C07686, C07700, C07710, C07712, C07723, C07726, C07733, C07734, C07735, C07736, C07737, C07738, C07739, C07740, C07741, C07742, C07743, C07744, C07745, C07746, C07747, C07748, C07749, C07750, C07751, C07752, C07753, C07754, C07755, C07756, C07757, C07760, C07762, C07763, C07764, C07765, C07767, C07768, C07769, C07771, C07772, C07774, C07775, C07776, C07777, C07778, C07779, C07780, C07782, C07783, C07784, C07785, C07789, C07790, C07791, C07793, C07794, C07795, C07796, C07797, C07798, C07799, C07802, C07803, C07804, C07805, C07806, C07807, C07808, C07809, C07811, C07812, C07813, C07816, C07818, C07819, C07820, C07821, C07822, C07824, C07825, C07827, C07828, C07829, C07835, C07870, C07880, C07881, C07883, C07884, C07886, C07887, C07889, C07890, C07891, C07892, C07893, C07894, C07895, C07896, C07897, C07898, C07899, C07900, C07901, C07902, C07903, C07904, C07905, C07906, C07907, C07908, C07909, C07910, C07911, C07913, C07914, C07915, C07916, C07917, C07918, C07919, C07921, C07922, C07924, C07925, C07927, C07928, C07929, C07930, C07931, C07932, C07933, C07934, C07936, C07937, C07938, C07939, C07940, C07941, C07942, C07943, C07944, C07945, C07946, C07947, C07949, C07950, C07951, C07952, C07953, C07954, C07957, C07958, C07959, C07960, C07961, C07962, C07963, C07964, C07965, C07966, C07967, C07968, C07970, C07971, C07973, C07974, C07976, C07977, C07978, C07979, C07980, C07981, C07982, C07983, C07984, C07985, C07986, C07988, C07990, C07991, C07992, C07993, C07994, C07995, C07996, C07997, C07998, C07999, C08001, C08002, C08003, C08004, C08005, C08023, C08025, C08035, C08036, C08039, C08044, C08045, C08048, C08055, C08060, C08062, C08063, C08064, C08065, C08066, C08067, C08068, C08069, C08072, C08073, C08074, C08075, C08076, C08077, C08078, C08079, C08080, C08083, C08085, C08086, C08087, C08088, C08089, C08090, C08091, C08092, C08093, C08094, C08095, C08096, C08097, C08098, C08099, C08100, C08101, C08102, C08103, C08104, C08105, C08106, C08107, C08109, C08110, C08111, C08112, C08113, C08114, C08116, C08117, C08118, C08119, C08120, C08121, C08122, C08123, C08124, C08125, C08127, C08128, C08129, C08130, C08131, C08132, C08133, C08134, C08135, C08136, C08137, C08138, C08139, C08140, C08141, C08143, C08144, C08145, C08146, C08147, C08148, C08149, C08150, C08151, C08152, C08153, C08155, C08156, C08157, C08158, C08159, C08160, C08161, C08162, C08163, C08164, C08165, C08166, C08167, C08168, C08169, C08171, C08172, C08173, C08174, C08175, C08176, C08177, C08178, C08179, C08180, C08181, C08182, C08183, C08184, C08185, C08186, C08187, C08188, C08189, C08190, C08191, C08192, C08193, C08194, C08195, C08196, C08197, C08198, C08199, C08202, C08203, C08205, C08206, C08207, C08208, C08209, C08210, C08211, C08212, C08213, C08214, C08215, C08216, C08217, C08218, C08219, C08220, C08221, C08222, C08223, C08224, C08226, C08227, C08228, C08229, C08230, C08231, C08233, C08234, C08235, C08237, C08238, C08240, C08241, C08242, C08244, C08245, C08247, C08248, C08249, C08250, C08251, C08252, C08253, C08254, C08255, C08256, C08257, C08258, C08259, C08260, C08261, C08262, C08263, C08264, C08265, C08266, C08267, C08268, C08269, C08272, C08273, C08274, C08275, C08276, C08277, C08278, C08279, C08280, C08282, C08283, C08284, C08285, C08287, C08289, C08290, C08291, C08292, C08295, C08296, C08297, C08298, C08299, C08300, C08302, C08304, C08305, C08307, C08308, C08309, C08310, C08314, C08315, C08316, C08317, C08319, C08320, C08321, C08323, C08324, C08325, C08326, C08328, C08335, C08338, C08343, C08347, C08352, C08353, C08357, C08359, C08360, C08362, C08363, C08367, C08369, C08370, C08371, C08372, C08373, C08374, C08377, C08378, C08379, C08383, C08384, C08385, C08386, C08387, C08388, C08390, C08391, C08393, C08394, C08395, C08397, C08398, C08399, C08400, C08401, C08402, C08403, C08404, C08406, C08407, C08408, C08409, C08410, C08412, C08413, C08415, C08417, C08418, C08419, C08420, C08421, C08422, C08423, C08424, C08425, C08426, C08427, C08428, C08429, C08430, C08431, C08432, C08433, C08434, C08435, C08436, C08437, C08438, C08439, C08440, C08441, C08442, C08443, C08444, C08445, C08446, C08447, C08448, C08449, C08450, C08451, C08452, C08453, C08454, C08455, C08456, C08457, C08458, C08459, C08460, C08461, C08463, C08464, C08465, C08466, C08467, C08468, C08469, C08470, C08471, C08472, C08473, C08474, C08475, C08476, C08477, C08478, C08479, C08480, C08481, C08482, C08483, C08484, C08485, C08486, C08488, C08489, C08490, C08491, C08492, C08493, C08494, C08496, C08497, C08498, C08499, C08500, C08501, C08502, C08503, C08504, C08505, C08506, C08507, C08508, C08509, C08510, C08511, C08512, C08514, C08515, C08516, C08517, C08518, C08519, C08521, C08522, C08523, C08524, C08525, C08527, C08528, C08529, C08532, C08533, C08535, C08536, C08537, C08546, C08550, C08551, C08552, C08553, C08554, C08555, C08556, C08557, C08558, C08559, C08560, C08561, C08562, C08563, C08564, C08565, C08566, C08567, C08568, C08569, C08570, C08571, C08572, C08573, C08574, C08575, C08576, C08577, C08578, C08579, C08581, C08582, C08583, C08584, C08585, C08587, C08588, C08589, C08590, C08591, C08592, C08593, C08594, C08595, C08596, C08597, C08598, C08599, C08600, C08601, C08602, C08603, C08604, C08605, C08606, C08607, C08610, C08615, C08620, C08648, C08650, C08651, C08652, C08653, C08654, C08655, C08656, C08657, C08658, C08659, C08660, C08661, C08662, C08663, C08664, C08665, C08666, C08667, C08668, C08669, C08670, C08671, C08672, C08673, C08674, C08675, C08676, C08677, C08678, C08679, C08680, C08681, C08684, C08685, C08686, C08687, C08688, C08689, C08690, C08691, C08693, C08694, C08695, C08697, C08704, C08706, C08707, C08714, C08715, C08718, C08719, C08720, C08724, C08728, C08729, C08732, C08733, C08734, C08738, C08739, C08746, C08749, C08750, C08752, C08755, C08765, C08766, C08767, C08768, C08769, C08770, C08771, C08772, C08773, C08774, C08775, C08776, C08777, C08780, C08781, C08783, C08784, C08785, C08787, C08789, C08790, C08791, C08792, C08793, C08794, C08795, C08796, C08797, C08799, C08801, C08802, C08804, C08805, C08806, C08807, C08808, C08809, C08810, C08813, C08814, C08819, C08823, C08825, C08828, C08829, C08830, C08831, C08832, C08833, C08834, C08836, C08838, C08840, C08841, C08842, C08843, C08844, C08846, C08847, C08848, C08849, C08851, C08852, C08853, C08854, C08857, C08858, C08859, C08860, C08861, C08862, C08863, C08864, C08865, C08866, C08867, C08868, C08869, C08870, C08872, C08873, C08874, C08875, C08876, C08877, C08878, C08879, C08880, C08882, C08883, C08884, C08885, C08886, C08887, C08888, C08889, C08890, C08892, C08893, C08894, C08895, C08897, C08898, C08900, C08901, C08902, C08903, C08904, C08906, C08907, C08908, C08909, C08910, C08911, C08912, C08913, C08914, C08915, C08916, C08917, C08918, C08919, C08920, C08922, C08924, C08925, C08926, C08928, C08929, C08930, C08932, C08933, C08934, C08939, C08940, C08941, C08942, C08943, C08944, C08945, C08946, C08947, C08948, C08949, C08951, C08952, C08953, C08956, C08960, C08961, C08962, C08965, C08967, C08968, C08970, C08972, C08973, C08974, C08976, C08979, C08981, C08982, C08984, C08985, C08986, C08987, C08988, C08989, C08990, C08991, C08993, C08995, C08996, C08997, C08998, C08999, C09001, C09002, C09003, C09004, C09005, C09006, C09007, C09008, C09009, C09011, C09013, C09014, C09015, C09016, C09017, C09018, C09020, C09021, C09022, C09023, C09024, C09025, C09026, C09027, C09029, C09030, C09031, C09032, C09033, C09037, C09040, C09042, C09047, C09050, C09052, C09054, C09055, C09057, C09060, C09061, C09062, C09063, C09064, C09065, C09066, C09067, C09069, C09070, C09071, C09072, C09073, C09074, C09075, C09076, C09077, C09078, C09079, C09080, C09082, C09083, C09084, C09085, C09086, C09087, C09088, C09089, C09090, C09091, C09092, C09093, C09094, C09095, C09096, C09097, C09098, C09099, C09100, C09101, C09102, C09103, C09104, C09105, C09106, C09107, C09108, C09109, C09110, C09111, C09112, C09113, C09114, C09115, C09116, C09117, C09118, C09119, C09120, C09121, C09122, C09123, C09124, C09125, C09126, C09127, C09128, C09129, C09130, C09131, C09132, C09133, C09134, C09135, C09136, C09137, C09138, C09139, C09140, C09141, C09142, C09143, C09144, C09146, C09147, C09148, C09149, C09151, C09152, C09153, C09155, C09156, C09157, C09158, C09159, C09160, C09161, C09162, C09163, C09165, C09166, C09167, C09168, C09169, C09170, C09171, C09172, C09173, C09174, C09175, C09176, C09178, C09179, C09180, C09182, C09183, C09184, C09185, C09187, C09188, C09190, C09191, C09192, C09193, C09194, C09195, C09196, C09201, C09204, C09205, C09209, C09210, C09211, C09212, C09214, C09217, C09219, C09221, C09223, C09224, C09225, C09226, C09227, C09229, C09231, C09232, C09233, C09237, C09238, C09239, C09241, C09242, C09243, C09245, C09246, C09247, C09248, C09249, C09250, C09253, C09254, C09255, C09256, C09257, C09258, C09259, C09260, C09261, C09262, C09263, C09264, C09265, C09266, C09267, C09268, C09269, C09270, C09271, C09272, C09273, C09274, C09275, C09276, C09277, C09278, C09279, C09280, C09281, C09282, C09283, C09284, C09285, C09286, C09288, C09289, C09291, C09292, C09297, C09298, C09299, C09303, C09304, C09305, C09310, C09311, C09312, C09313, C09317, C09319, C09320, C09322, C09323, C09324, C09328, C09330, C09331, C09332, C09333, C09334, C09336, C09337, C09338, C09339, C09341, C09345, C09346, C09349, C09350, C09351, C09352, C09353, C09354, C09355, C09356, C09357, C09358, C09359, C09360, C09361, C09362, C09363, C09364, C09365, C09366, C09367, C09368, C09370, C09371, C09373, C09374, C09376, C09378, C09379, C09380, C09381, C09383, C09385, C09386, C09387, C09388, C09389, C09391, C09392, C09393, C09394, C09397, C09398, C09399, C09400, C09402, C09403, C09404, C09405, C09407, C09408, C09409, C09410, C09411, C09412, C09414, C09415, C09416, C09417, C09418, C09419, C09420, C09421, C09422, C09423, C09424, C09425, C09426, C09427, C09428, C09429, C09430, C09431, C09433, C09434, C09435, C09436, C09437, C09438, C09439, C09440, C09441, C09442, C09443, C09444, C09445, C09446, C09447, C09448, C09449, C09450, C09451, C09452, C09453, C09454, C09455, C09456, C09457, C09458, C09459, C09460, C09461, C09462, C09464, C09465, C09466, C09467, C09468, C09469, C09470, C09471, C09472, C09473, C09474, C09475, C09476, C09477, C09478, C09479, C09480, C09481, C09482, C09483, C09484, C09485, C09486, C09488, C09489, C09490, C09491, C09492, C09493, C09494, C09495, C09496, C09497, C09498, C09499, C09500, C09501, C09502, C09503, C09504, C09505, C09506, C09507, C09508, C09510, C09511, C09512, C09513, C09514, C09515, C09516, C09518, C09519, C09521, C09522, C09523, C09525, C09526, C09527, C09528, C09529, C09530, C09532, C09533, C09534, C09535, C09536, C09537, C09538, C09539, C09540, C09541, C09542, C09543, C09544, C09545, C09546, C09547, C09548, C09549, C09550, C09551, C09552, C09553, C09554, C09555, C09556, C09557, C09558, C09559, C09560, C09561, C09562, C09563, C09565, C09566, C09567, C09568, C09569, C09570, C09571, C09572, C09573, C09575, C09576, C09578, C09579, C09581, C09582, C09583, C09585, C09586, C09588, C09589, C09590, C09592, C09594, C09599, C09601, C09602, C09603, C09604, C09605, C09606, C09607, C09608, C09609, C09610, C09611, C09612, C09613, C09614, C09615, C09616, C09617, C09618, C09619, C09620, C09621, C09622, C09623, C09624, C09625, C09626, C09627, C09628, C09629, C09630, C09631, C09632, C09633, C09634, C09635, C09636, C09637, C09638, C09639, C09640, C09641, C09643, C09644, C09645, C09646, C09647, C09648, C09649, C09650, C09651, C09652, C09653, C09655, C09656, C09657, C09658, C09659, C09660, C09661, C09662, C09663, C09664, C09665, C09666, C09667, C09668, C09669, C09671, C09672, C09673, C09674, C09676, C09677, C09678, C09679, C09680, C09681, C09682, C09684, C09685, C09686, C09687, C09688, C09689, C09690, C09691, C09692, C09693, C09694, C09695, C09697, C09698, C09699, C09702, C09703, C09704, C09705, C09706, C09707, C09708, C09709, C09710, C09711, C09712, C09713, C09714, C09715, C09716, C09717, C09718, C09719, C09720, C09721, C09724, C09725, C09726, C09727, C09728, C09729, C09731, C09732, C09733, C09734, C09735, C09736, C09737, C09738, C09739, C09740, C09741, C09742, C09743, C09744, C09745, C09746, C09747, C09748, C09749, C09750, C09751, C09752, C09754, C09755, C09756, C09758, C09759, C09760, C09763, C09765, C09766, C09767, C09768, C09769, C09770, C09771, C09772, C09773, C09774, C09775, C09776, C09777, C09778, C09779, C09780, C09781, C09782, C09783, C09784, C09785, C09786, C09788, C09789, C09790, C09791, C09792, C09793, C09794, C09795, C09796, C09797, C09798, C09799, C09800, C09801, C09802, C09804, C09805, C09807, C09808, C09809, C09810, C09811, C09812, C09813, C09814, C09815, C09816, C09819, C09821, C09822, C09823, C09824, C09825, C09827, C09828, C09830, C09831, C09832, C09833, C09834, C09835, C09836, C09837, C09839, C09841, C09842, C09843, C09844, C09845, C09846, C09847, C09848, C09849, C09850, C09851, C09852, C09853, C09854, C09856, C09857, C09858, C09859, C09860, C09861, C09862, C09863, C09864, C09866, C09867, C09869, C09870, C09871, C09872, C09873, C09874, C09875, C09876, C09877, C09878, C09879, C09880, C09881, C09882, C09883, C09884, C09885, C09886, C09887, C09888, C09889, C09890, C09891, C09892, C09893, C09894, C09895, C09896, C09897, C09898, C09899, C09900, C09901, C09902, C09903, C09904, C09905, C09906, C09907, C09908, C09909, C09910, C09911, C09913, C09914, C09915, C09916, C09917, C09918, C09919, C09920, C09921, C09922, C09923, C09924, C09925, C09927, C09928, C09929, C09930, C09931, C09933, C09934, C09935, C09937, C09938, C09939, C09941, C09943, C09944, C09945, C09946, C09947, C09948, C09949, C09950, C09951, C09952, C09953, C09954, C09955, C09956, C09957, C09958, C09959, C09960, C09961, C09962, C09963, C09964, C09965, C09966, C09967, C09968, C09969, C09970, C09971, C09972, C09974, C09975, C09976, C09977, C09978, C09979, C09980, C09982, C09983, C09984, C09985, C09986, C09987, C09988, C09989, C09990, C09991, C09992, C09993, C09994, C09995, C09997, C09998, C09999, C10000, C10001, C10002, C10003, C10004, C10005, C10006, C10007, C10008, C10009, C10010, C10011, C10012, C10013, C10015, C10016, C10017, C10018, C10019, C10020, C10021, C10022, C10023, C10024, C10025, C10026, C10027, C10028, C10029, C10030, C10031, C10032, C10033, C10034, C10035, C10036, C10037, C10038, C10039, C10040, C10041, C10042, C10043, C10044, C10045, C10046, C10047, C10048, C10049, C10050, C10052, C10053, C10054, C10055, C10056, C10057, C10058, C10059, C10060, C10061, C10062, C10063, C10064, C10065, C10066, C10067, C10068, C10069, C10070, C10071, C10072, C10073, C10074, C10075, C10076, C10077, C10078, C10079, C10080, C10081, C10082, C10083, C10084, C10085, C10086, C10087, C10088, C10089, C10090, C10091, C10092, C10093, C10094, C10095, C10096, C10097, C10098, C10099, C10100, C10101, C10102, C10103, C10104, C10105, C10106, C10107, C10108, C10109, C10110, C10111, C10112, C10113, C10114, C10115, C10116, C10117, C10118, C10119, C10120, C10121, C10122, C10123, C10124, C10125, C10126, C10127, C10128, C10129, C10130, C10131, C10132, C10133, C10134, C10135, C10136, C10137, C10138, C10139, C10140, C10141, C10142, C10143, C10144, C10145, C10146, C10147, C10149, C10150, C10151, C10152, C10153, C10154, C10155, C10156, C10157, C10158, C10159, C10160, C10161, C10162, C10163, C10164, C10165, C10166, C10167, C10168, C10169, C10170, C10171, C10172, C10173, C10174, C10175, C10176, C10177, C10178, C10179, C10180, C10181, C10182, C10183, C10184, C10185, C10186, C10187, C10188, C10189, C10190, C10191, C10192, C10193, C10194, C10195, C10196, C10197, C10198, C10199, C10200, C10201, C10202, C10203, C10204, C10205, C10206, C10207, C10208, C10209, C10210, C10211, C10212, C10213, C10214, C10215, C10216, C10217, C10218, C10219, C10220, C10221, C10222, C10223, C10224, C10225, C10226, C10227, C10228, C10229, C10230, C10231, C10232, C10233, C10234, C10235, C10236, C10237, C10238, C10239, C10240, C10241, C10242, C10243, C10244, C10245, C10246, C10247, C10248, C10249, C10250, C10251, C10252, C10253, C10254, C10255, C10256, C10257, C10258, C10259, C10260, C10261, C10262, C10263, C10264, C10265, C10266, C10267, C10268, C10269, C10270, C10271, C10272, C10273, C10274, C10275, C10276, C10277, C10278, C10279, C10280, C10281, C10282, C10283, C10284, C10285, C10286, C10287, C10288, C10289, C10290, C10291, C10292, C10293, C10294, C10295, C10296, C10297, C10298, C10299, C10300, C10301, C10302, C10303, C10304, C10305, C10306, C10307, C10308, C10309, C10310, C10311, C10312, C10313, C10314, C10315, C10316, C10317, C10318, C10319, C10320, C10321, C10322, C10323, C10324, C10325, C10326, C10327, C10328, C10329, C10330, C10331, C10332, C10333, C10334, C10335, C10336, C10337, C10338, C10339, C10340, C10341, C10342, C10343, C10344, C10345, C10346, C10348, C10349, C10350, C10351, C10352, C10353, C10354, C10355, C10356, C10357, C10358, C10359, C10360, C10361, C10362, C10363, C10364, C10365, C10366, C10367, C10368, C10369, C10371, C10373, C10374, C10375, C10376, C10377, C10378, C10379, C10380, C10381, C10382, C10383, C10385, C10386, C10387, C10388, C10389, C10390, C10391, C10392, C10393, C10394, C10395, C10396, C10397, C10398, C10399, C10400, C10401, C10402, C10403, C10404, C10405, C10406, C10407, C10408, C10409, C10410, C10411, C10412, C10413, C10414, C10415, C10416, C10417, C10418, C10419, C10420, C10421, C10422, C10423, C10424, C10425, C10426, C10427, C10428, C10430, C10431, C10432, C10433, C10434, C10435, C10436, C10437, C10438, C10439, C10440, C10441, C10442, C10443, C10444, C10445, C10446, C10447, C10448, C10449, C10450, C10451, C10452, C10453, C10454, C10455, C10456, C10457, C10458, C10459, C10460, C10461, C10462, C10463, C10464, C10465, C10466, C10467, C10468, C10469, C10470, C10471, C10472, C10473, C10475, C10476, C10477, C10478, C10479, C10480, C10481, C10482, C10483, C10484, C10485, C10486, C10487, C10488, C10489, C10490, C10491, C10492, C10493, C10494, C10495, C10496, C10497, C10498, C10499, C10500, C10501, C10502, C10503, C10504, C10505, C10506, C10507, C10508, C10509, C10510, C10511, C10512, C10513, C10514, C10515, C10516, C10517, C10518, C10519, C10520, C10521, C10522, C10523, C10524, C10525, C10526, C10527, C10528, C10529, C10530, C10531, C10532, C10533, C10534, C10535, C10536, C10537, C10538, C10539, C10540, C10541, C10542, C10543, C10544, C10545, C10546, C10547, C10548, C10549, C10550, C10551, C10552, C10553, C10554, C10555, C10556, C10557, C10558, C10560, C10561, C10562, C10563, C10564, C10565, C10566, C10567, C10568, C10569, C10570, C10571, C10572, C10573, C10575, C10577, C10578, C10580, C10581, C10582, C10583, C10584, C10585, C10586, C10587, C10588, C10589, C10590, C10591, C10592, C10593, C10594, C10595, C10596, C10597, C10598, C10599, C10600, C10601, C10602, C10603, C10604, C10605, C10606, C10607, C10608, C10609, C10610, C10611, C10612, C10613, C10614, C10615, C10616, C10617, C10618, C10619, C10620, C10621, C10622, C10623, C10624, C10625, C10626, C10627, C10628, C10629, C10630, C10631, C10632, C10633, C10634, C10635, C10636, C10637, C10638, C10639, C10640, C10641, C10642, C10643, C10644, C10645, C10646, C10647, C10648, C10649, C10650, C10651, C10653, C10654, C10655, C10656, C10657, C10658, C10659, C10660, C10661, C10662, C10663, C10664, C10665, C10666, C10667, C10668, C10669, C10670, C10671, C10672, C10673, C10674, C10675, C10676, C10677, C10678, C10679, C10681, C10682, C10683, C10684, C10685, C10686, C10687, C10688, C10689, C10690, C10691, C10692, C10693, C10694, C10695, C10696, C10697, C10698, C10699, C10700, C10701, C10702, C10703, C10704, C10705, C10706, C10707, C10708, C10709, C10710, C10711, C10713, C10714, C10715, C10716, C10717, C10718, C10719, C10720, C10721, C10722, C10723, C10724, C10725, C10726, C10727, C10728, C10729, C10730, C10731, C10732, C10733, C10734, C10735, C10736, C10737, C10738, C10739, C10740, C10741, C10742, C10743, C10744, C10745, C10746, C10747, C10748, C10749, C10750, C10751, C10752, C10753, C10754, C10755, C10756, C10757, C10758, C10759, C10760, C10761, C10762, C10763, C10764, C10765, C10766, C10767, C10768, C10769, C10770, C10771, C10772, C10773, C10774, C10775, C10776, C10777, C10778, C10779, C10780, C10781, C10782, C10784, C10785, C10786, C10787, C10788, C10789, C10790, C10791, C10792, C10794, C10796, C10797, C10798, C10799, C10800, C10801, C10802, C10803, C10804, C10805, C10806, C10807, C10808, C10809, C10810, C10811, C10812, C10813, C10814, C10815, C10816, C10817, C10818, C10819, C10820, C10821, C10822, C10823, C10824, C10825, C10826, C10827, C10828, C10829, C10830, C10831, C10832, C10833, C10834, C10835, C10836, C10837, C10838, C10839, C10840, C10841, C10842, C10843, C10844, C10845, C10846, C10847, C10848, C10849, C10850, C10851, C10852, C10853, C10854, C10855, C10856, C10857, C10858, C10859, C10860, C10861, C10862, C10863, C10864, C10865, C10866, C10867, C10868, C10869, C10870, C10871, C10872, C10873, C10874, C10875, C10876, C10877, C10878, C10879, C10880, C10881, C10882, C10883, C10884, C10885, C10886, C10887, C10888, C10889, C10890, C10892, C10893, C10894, C10897, C10898, C10899, C10900, C10901, C10902, C10903, C10904, C10905, C10906, C10907, C10908, C10909, C10910, C10911, C10912, C10913, C10914, C10915, C10916, C10917, C10918, C10919, C10920, C10921, C10922, C10923, C10924, C10925, C10926, C10927, C10928, C10929, C10930, C10931, C10933, C10934, C10935, C10936, C10937, C10938, C10939, C10940, C10941, C10942, C10943, C10944, C10945, C10946, C10947, C10948, C10949, C10950, C10951, C10953, C10954, C10955, C10956, C10957, C10958, C10959, C10960, C10961, C10962, C10963, C10964, C10965, C10966, C10967, C10968, C10969, C10970, C10971, C10972, C10973, C10974, C10975, C10976, C10977, C10978, C10979, C10980, C10981, C10982, C10984, C10985, C10986, C10987, C10988, C10989, C10990, C10991, C10992, C10993, C10994, C10995, C10996, C10997, C10998, C10999, C11000, C11001, C11002, C11003, C11004, C11005, C11006, C11007, C11008, C11009, C11010, C11011, C11012, C11013, C11014, C11015, C11016, C11017, C11018, C11019, C11020, C11021, C11022, C11023, C11024, C11025, C11026, C11027, C11028, C11029, C11030, C11031, C11032, C11033, C11035, C11036, C11037, C11038, C11039, C11040, C11041, C11042, C11043, C11044, C11045, C11046, C11047, C11048, C11049, C11050, C11051, C11052, C11053, C11054, C11055, C11056, C11057, C11058, C11059, C11060, C11061, C11062, C11063, C11064, C11065, C11066, C11067, C11068, C11069, C11070, C11071, C11072, C11073, C11074, C11075, C11076, C11077, C11078, C11079, C11080, C11081, C11082, C11083, C11084, C11085, C11086, C11087, C11088, C11089, C11090, C11091, C11092, C11093, C11095, C11096, C11097, C11098, C11099, C11100, C11101, C11102, C11103, C11104, C11105, C11106, C11107, C11108, C11109, C11110, C11111, C11112, C11113, C11114, C11115, C11116, C11117, C11118, C11119, C11120, C11121, C11122, C11124, C11125, C11126, C11127, C11128, C11129, C11130, C11131, C11132, C11133, C11134, C11135, C11137, C11138, C11139, C11140, C11141, C11142, C11143, C11144, C11145, C11146, C11147, C11148, C11149, C11150, C11151, C11152, C11153, C11154, C11155, C11156, C11157, C11158, C11159, C11160, C11161, C11162, C11163, C11164, C11165, C11166, C11167, C11168, C11169, C11170, C11171, C11172, C11174, C11175, C11176, C11177, C11178, C11179, C11180, C11181, C11182, C11184, C11185, C11186, C11188, C11189, C11190, C11191, C11192, C11193, C11194, C11195, C11196, C11197, C11198, C11199, C11200, C11201, C11202, C11204, C11205, C11206, C11207, C11208, C11209, C11211, C11212, C11213, C11214, C11215, C11216, C11217, C11218, C11219, C11220, C11221, C11222, C11223, C11224, C11225, C11226, C11227, C11228, C11229, C11230, C11231, C11232, C11233, C11234, C11235, C11236, C11237, C11238, C11240, C11241, C11242, C11243, C11244, C11245, C11246, C11247, C11248, C11249, C11250, C11251, C11252, C11253, C11254, C11255, C11256, C11257, C11258, C11259, C11260, C11261, C11263, C11264, C11265, C11266, C11267, C11268, C11269, C11270, C11271, C11272, C11276, C11277, C11278, C11279, C11280, C11281, C11282, C11283, C11284, C11285, C11286, C11287, C11288, C11289, C11290, C11291, C11352, C11353, C11354, C11355, C11356, C11357, C11358, C11359, C11360, C11361, C11362, C11363, C11364, C11365, C11366, C11367, C11368, C11369, C11370, C11371, C11372, C11373, C11374, C11375, C11376, C11377, C11378, C11379, C11380, C11382, C11383, C11384, C11385, C11386, C11387, C11388, C11389, C11390, C11391, C11392, C11393, C11394, C11395, C11396, C11397, C11398, C11400, C11401, C11402, C11403, C11404, C11406, C11407, C11408, C11409, C11410, C11411, C11412, C11432, C11579, C11626, C11743, C11744, C11747, C11842, C12012, C14890, C15138, C15422, C15424, C15505, C15506 |
Recalling Firm/ Manufacturer |
Cardinal Health 203, Inc 17400 Medina Rd Ste 100 Minneapolis MN 55447-1341
|
Manufacturer Reason for Recall | The failure of an electronic component could result in failure of the ventilator to breathe for the patient and/or failure of the ventilator to properly alarm to alert the caregiver. |
FDA Determined Cause 2 | Device Design |
| Action | An Urgent Medical Device Recall letter dated February 22, 2008 was sent to ventilator consignees/owners and distributors. The letter describes the issue, identifies the affected devices, states that the devices will be corrected by Cardinal Health and provides direction to verify ventilator operation and audible alarm function daily. Service Centers were provided information to identify affected printed circuit board assembly replacement kits' lot numbers and asked to return them. An "Update" was mailed 03/04/2008 to identify additional ventilator serial numbers that also need correction. |
| Quantity in Commerce | 6981 (5831 US, 1150 OUS) |
| Distribution | Worldwide Distribution --- USA including Puerto Rico and countries of Algeria, Argentina, Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, China Peoples Republic, Colombia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Israel, Italy, Japan, Malaysia, Mexico, New Zealand, Norway, Panama, Paraguay, Philippines, Portugal, Saudi Arabia, Serbia & Montenegro, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Vietnam, and West Indies. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = CBK 510(K)s with Product Code = CBK
|
|
|
|