Class 2 Device Recall Newport e360 Ventilator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Newport e360 Ventilatorsee related information
Date Initiated by FirmDecember 15, 2008
Date PostedSeptember 04, 2009
Recall Status1 Terminated 3 on November 04, 2011
Recall NumberZ-2029-2009
Recall Event ID 50785
510(K)NumberK053502 
Product Classification Ventilator, continuous, facility use - Product Code CBK
ProductNewport e360 Ventilator, Rx only. The e360 Ventilator System is intended to provide continuous (endotracheal or tracheostomy [ET] tube) or non-continuous (mask) ventilatory support and monitoring for infant, pediatric, and adult patients requiring tidal volumes equal to or greater than 20 milliliters (mL).
Code Information Serial Numbers for the Domestic Units: N07361211863  N06360910375  N06360610268  N09360413995  N07360110859  N07361111725  N07361111620  N06360710310  N06360710311  N06360710312  N07360110797  N08360812873  N06361210693  N07360911465  N36360410225  N07360110864  N07360110870  N09360313832  N07361111713  N08360512383  N08360512384  N08360512385  N08360512386  N07361211793  N06360910351  N08360712658  N08360712659  N08360712660  N08360712661  N08360712662  N08361013165  N08361013166  N08361013167  N08361013168  N08361013169  N06361210693  N07360110855  N08361213295  N09360113451  N06360610258  N07360110853  N06360610271  N06361010529  N06361010532  N07360110825  N07360811393  N08360212053  N07361211869  N07361211870  N07361211871  N07361211872  N09360213550  N06360810339  N06360910350  N06360910352  N06360910354  N06360910388  N06360910389  N07360411127  N07360811396  N07361111626  N07361111627  N07361111711  N07361111712  N08360512396  N08360612528  N08360612561  N08360712687  N08360712688  N08360712689  N08360712690  N08360712691  N08360712692  N08360712693  N08360712694  N08360712695  N08360712696  N08360712697  N08360712698  N08360712699  N08360712700  N08360712701  N08360712702  N08360712703  N08360712704  N08360712705  N08360712706  N08360712707  N08360712708  N08360712709  N08360712710  NO8360712709  N06360710307  N06360710308  N06360710309  N06360710313  N06360710314  N06360610280  N06361010516  N06361010518  N06361010519  N06361010522  N06361010536  N07360110856  N07360110862  N07360110867  N07360210926  N07360911464  N07360911489  N07360911490  N07360210952  N07360210953  N07360210954  N08361013107  N08361013108  N08361013109  N08361213283  N08361213284  N08360512381  N08360512382  N08360512398  N08360512399  N08360512400  N08360512402  N06361010542  N06361010543  N06361010544  N06361010545  N06361010546  N06361010547  N06361010548  N06361010550  N06361110551  N06361110552  N06361110553  N06361110554  N06361110555  N06361110556  N06361110557  N06361110558  N06361110559  N06361110560  N06361110561  N06361110562  N06361110563  N06361110564  N06361110572  N06360910373  N07306011235  N07360611236  N07360711360  N07360711361  N07360711363  N07360711364  N07360711368  N07360711372  N07360911491  N07360911492  N07361211769  N07361211783  N07361211784  N07361211785  N07361211786  N07361211787  N07361211788  N07361211789  N07361211790  N07361211791  N07361211792  N08360111874  N08360111875  N08360512395  N08361013110  N08361013111  N08361013112  N08361013113  N08361013114  N07360110873  N06361010536-R  N07361211812  N07361211813  N08361013175  N08360111987  N08360111988  N08360111989  N08360111990  N06361010533  N07360110796  N07360110852  N07360110860  N07360110868  N07360110876  N06361010531  N07360210951  N07360210964  N07360110866  N07360811397  N07360110898  N07360210944  N07360811380  N07360811384  N07360811386  N07360811398  N07360811401  N07360811402  N07360811403  N07360811427  N08360811403  N07360711362  N07360711365  N07360811387  N09360213549  N06361010534  N07360110863; and   Serial Numbers for the International Units: N06360910392  N06360910400  N06360910426  N06361110580  N07361111637  N07361111638  N09360313710  N06360910352  N07360110878  N07360411101  N07360611242  N07360611243  N07360611244  N07361211794  N07361211795  N07361211796  N07361211797  N07361211798  N08360111890  N08360111891  N08360111893  N08360111894  N08360612578  N08360612579  N08360612580  N08360612581  N08360612582  N08360612583  N08360612593  N08360712711  N08360712712  N08360712713  N09360313786  N09360313787  N07360311089  N07360411096  N07360411105  N07360411111  N07360411129  N07360511161  N07361011506  N07361011507  N07361011523  N07361011524  N07361011525  N07361011526  N07361011527  N07361011528  N07361011529  N07361011530  N07361011531  N07361011532  N07361011533  N07361011534  N07361011535  N07361011537  N07361011538  N07361011539  N07361011540  N07361011541  N07361011542  N07361011543  N07361011544  N07360210959  N08360812800  N08360812801  N08360812804  N08360812809  N08360812810  N08360812811  N08360812812  N08360812813  N08360812816  N08360812817  N08360812818  N08360812819  N08360812820  N08360812823  N08360812824  N09360213600  N09360313696  N09360313727  N09360313728  N09360313729  N09360313731  N09360313847  N09360413942  N09360413943  N09360413944  N09360413945  N09360413946  N09360414002  N09360414003  N06360610265  N07360210963  N07360210966  N07360210967  N07360210968  N07360210970  N09360113452  N07360110885  N07360110891  N07360110909  N07360210972  N07360311090  N07360311091  N07360411095  N07360411107  N07360411108  N08360312183  E360-0408-183  E360-0408-191  E360-0408-195  E360-0408-202  E360-0508-003  E360-0508-006  E360-0508-012  E360-0508-017  E360-0508-018  E360-0508-019  E360-0508-020  E360-0508-021  E360-0508-023  E360-0508-024  E360-0508-025  E360-0508-026  E360-0508-027  E360-0508-028  E360-0508-029  E360-0508-030  E360-0508-031  E360-0508-032  E360-0508-033  E360-0508-034  E360-0508-035  E360-0508-036  E360-0508-037  E360-0508-038  E360-0508-039  E360-0508-041  E360-0508-042  E360-0508-043  E360-0508-044  E360-0508-045  E360-0508-047  E360-0508-048  E360-0508-049  E360-0508-050  E360-0508-052  E360-0508-053  E360-0508-054  E360-0508-056  E360-0508-057  E360-0508-058  E360-0508-059  E360-0508-060  E360-0508-061  E360-0508-062  E360-0508-063  E360-0508-064  E360-0508-065  E360-0508-066  E360-0508-070  E360-0808-126  E360-0808-172  E360-0908-056  E360-0908-109  E360-0908-118  E360-0908-131  E360-0908-137  E360-0908-138  E360-0908-148  E360-0908-153  E360-0908-154  E360-0908-164  E360-0908-167  E360-0908-172  E360-0908-174  E360-0908-183  E360-0908-187  E360-0908-188  E360-0908-189  E360-1008-001  E360-1008-004  E360-1008-005  E360-1008-006  E360-1008-008  E360-1008-009  E360-1008-012  E360-1008-014  E360-1008-015  E360-1008-018  E360-1008-019  E360-1008-027  E360-1008-029  E360-1008-030  E360-1008-031  E360-1008-034  E360-1008-035  E360-1008-036  E360-1008-038  E360-1008-045  E360-1008-046  E360-1008-050  E360-1008-051  E360-1008-052  E360-1008-054  E360-1008-055  E360-1008-057  E360-1008-061  E360-1008-062  E360-1008-067  E360-1008-070  E360-1008-071  E360-1008-072  E360-1008-073  E360-1008-077  E360-1008-078  E360-1008-079  E360-1008-080  E360-1008-081  E360-1008-083  E360-1008-084  E360-1008-085  E360-1008-086  E360-1008-087  E360-1008-089  E360-1008-091  E360-1008-092  E360-1008-095  E360-1008-096  E360-1008-097  E360-1008-098  E360-1008-099  E360-1008-100  E360-1008-101  E360-1008-102  E360-1008-103  E360-1008-104  E360-1008-106  E360-1008-107  E360-1008-108  E360-1008-109  E360-1008-110  E360-1008-112  E360-1008-113  E360-1008-114  E360-1008-115  E360-1008-117  E360-1008-119  E360-1008-120  E360-1008-121  E360-1008-122  E360-1008-123  E360-1008-124  E360-1008-125  E360-1008-126  E360-1008-129  E360-1008-130  E360-1008-131  E360-1008-132  E360-1008-137  E360-1008-140  E360-1008-144  E360-1008-145  E360-1008-146  E360-1008-148  E360-1008-151  E360-1008-152  E360-1008-155  E360-1008-156  E360-1008-159  E360-1008-160  E360-1008-161  E360-1108-008  E360-1108-014  E360-1108-015  E360-1108-016  E360-1108-017  E360-1108-018  E360-1108-019  E360-1108-020  E360-1108-021  E360-1108-022  E360-1108-023  E360-1108-024  E360-1108-025  E360-1108-026  E360-1108-027  E360-1108-028  E360-1108-031  E360-1108-032  E360-1108-033  E360-1108-034  E360-1108-036  E360-1108-037  E360-1108-044  E360-1108-045  E360-1108-047  E360-1108-048  E360-1108-049  E360-1108-050  E360-1108-051  E360-1108-052  E360-1108-055  E360-1108-056  E360-1108-057  E360-1108-058  E360-1108-059  E360-1108-060  E360-1108-061  E360-1108-062  E360-1108-063  E360-1108-064  E360-1108-065  E360-1108-066  E360-1108-067  E360-1108-068  E360-1108-072  E360-1108-073  E360-1108-074  E360-1108-075  E360-1108-076  E360-1108-077  E360-1108-078  E360-1108-079  E360-1108-081  E360-1108-084  E360-1108-085  E360-1108-087  E360-1108-088  E360-1108-092  E360-1108-096  E360-1108-098  E360-1108-100  E360-1108-101  E360-1108-157  E360-1108-163  E360-1108-164  E360-1108-167  E360-1108-171  E360-1108-173  E360-1108-178  E360-1108-180  E360-1108-181  E360-1108-182  E360-1108-186  E360-1108-192  E360-1208-019  E360-1208-020  E360-1208-021  E360-1208-023  E360-1208-025  E360-1208-026  E360-1208-027  E360-1208-030  E360-1208-031  E360-1208-037  E360-1208-041  E360-1208-044  E360-1208-052  E360-1208-054  E360-1208-058  E360-1208-059  E360-1208-061  E360-1208-062  E360-1208-069  E360-1208-072  E360-1208-078  E360-1208-085  E360-1208-087  E360-1208-090  E360-1208-093  E360-1208-094  E360-1208-099  E360-1208-101  E360-1208-102  E360-1208-103  E360-1208-104  E360-1208-105  E360-1208-106  E360-1208-107  E360-1208-108  E360-1208-109  E360-1208-110  E360-1208-111  E360-1208-112  E360-1208-113  E360-1208-114  E360-1208-115  E360-1208-116  E360-1208-117  E360-1208-118  E360-1208-119  E360-1208-120  E360-1208-121  E360-1208-122  E360-1208-123  E360-1208-124  E360-1208-125  E360-1208-126  E360-1208-127  E360-1208-129  E360-1208-130  N07361111693  N07361111694  N07361111695  N07361111696  N07361111697  N07361111722  N07361111723  N07361111724  N07361111730  N07361111731  N07361111732  N07361111736  N07361111737  N07361111738  N07361111739  N07361111740  N07361111741  N07361111742  N07361111743  N07361111744  N07361111745  N07361111746  N07361111747  N07361111748  N07361111750  N07361111754  N07361111755  N07361111756  N07361111757  N07361111758  N07361111759  N07361111760  N07361111761  N07361111762  N07361111763  N07361111764  N07361111765  N07361111766  N07361111767  N07361111768  N08360912976  N08360913020  N08360913021  N08360913022  N08360913023  N08360913024  N08360913025  N08360913026  N08360913027  N08360913028  N08360913029  N08360913030  N08360913031  N08360913032  N08360913033  N08360913034  N08360913035  N08360913036  N08360913037  N08360913038  N08360913039  N08360913040  N08360913041  N08360913042  N08360913043  N08360913044  N08360913045  N08360913046  N08360913047  N08360913048  N08360913049  N08360913050  N08360913051  N08360913052  N08360913055  N08360913056  N08360913057  N08360913058  N08360913059  N08360913060  N08360913061  N08360913062  N08360913063  N08360913065  N08360913066  N08360913067  N08360913068  N08360913069  N08360913070  N08360913071  N08360913072  N08360913073  N08360913074  N08360913075  N08360913076  N08360913077  N08360913078  N08360913079  N08360913080  N08360913081  N05361110106  N06360310212  N06360710293  N06360710296  N06360710305  N06360710306  N06361010541  N08360111957  N08360111958  N08360812838  N06361210660  N09360413997  N06360910373  N06360910398  N06361010516  N07360911465  N07361211775  N08360212066  N08360212067  N08360212068  N08360212069  N08360612521  N08360612522  N08360612523  N08360612524  N08360612525  N08360612526  N08360712675  N09360413993  N06360610250  N06361010478  N06361110629  N07360110829  N07360110830  N07360210961  N07360210962  N07361011554  N07361111728  N07361111749  N07361211824  N07360811404  N08360913064  N07360210945  N06360910380  N06360910379  N07360210936  N06360910387  N07360611245  N07360611246  N07360611292  N07360711318  N07360711321  N07360711322  N08361013124  N08361013125  N08361013126  N08361013127  N08361013128  N08361013129  N08361013130  N08361013131  N08361013137  N08361013138  N08361013142  N08361013143  N08361013144  N08361013145  N08361013146  N08361013147  N09360413983  N09360413984  N09360413985  N09360413986  N09360413987  N09360413988  N09360413989  N09360413990  N07361111682  N06360810331  N06360810334  N06361010473  N06361010481  N07360711317  N07360811412  N07360811414  N07360811416  N07360911469  N07360911471  N07360911472  N07360911473  N07360911474  N07360911475  N07360911476  N07360911485  N07360911486  N07360911487  N07360911488  N07361011589  N07361011590  N07361011591  N08360913000  N08360913001  N08361013105  N08361013106  N09360113411  N09360113412  N09360113413  N09360113414  N09360213551  N09360213552  N09360213554  N09360213555  N06360810326  N06360810332  N06360810335  N06361110582  N07360110889  N07360110890  N07361211772  N07361211773  N08360111876  N08360111886  N08360111887  N08360111888  N08360111889  N08360111892  N08360111895  N08360111896  N08360111897  N08360111898  N08360111899  N08360312155  N08360312157  N08360312158  N08360312159  N08360312160  N08360312161  N08360312162  N08360312163  N08360312182  N08360612586  N08360612587  N08360612588  N08360612589  N08360612590  N08360712715  N08360712716  N08360712718  N08360912995  N08360912996  N08360912997  N08360913002  N08360913005  N08360913007  N08360913008  N08360913009  N08360913010  N08361113248  N08361113249  N08361113250  N08361113251  N08361113252  N08361213384  N08361213385  N08361213386  N08361213387  N08361213388  N08361213389  N08361213390  N08361213391  N08361213392  N08361213393  N08361213394  N08361213395  N08361213396  N08361213397  N08361213398  N08361213399  N08361213400  N06361210647  N06361210658  N06360210210*  N06360610256  N06360610262  N06361010526  N06361010527  N07360110914 N07360511139  N07360511160  N07360811373  N07360811395  N07360811407  N07361011516  N07361111621  N07361111622  N07361111623  N07361111628  N08360312131  N08360512378  N08360512401  N08360612562  N08360612563  N08360612564  N08360612565  N08360612566  N08360612567  N08360612568  N08360612569  N08360612570  N08360612571  N08360612572  N08360612573  N08360612574  N08360612575  N08360612576  N08360612577  N08360812825  N08360812828  N08360812829  N08360812830  N08360812831  N08360812832  N08360812833  N08360812834  N08360812835  N08360812836  N08360812837  N08360812839  N08360812840  N08360812841  N08360812842  N08360812843  N08360812844  N08360812851  N08360812859  N08360812860  N08360812861  N08360812868  N08360812869  N08360812870  N08360812871  N08360812872  N08360812875  N08360812876  N08360812877  N08360812878  N08360912896  N08360912897  N08360912898  N08360912899  N08360912900  N08360912901  N08360912902  N08360912903  N08360912904  N08360912905  N08360912906  N08360912907  N08360912908  N08360912909  N08360912913  N08360912914  N08360912916  N08360912920  N08360912921  N08360912926  N08360912927  N08360912957  N08361013161  N08361013162  N08361113183  N08361113184  N09360113500  N09360113501  N08360312174  N08360612529  N08360612537  N08360612538  N08360612539  N08360612540  N08360612541  N08360612542  N08360612543  N08360612544  N08360612545  N08360612546  N08360612547  N08361013160  N08361013164  N07360611276  N07360611294  N07360611296  N07360611300  N07360611304  N07360611305  N07360611306  N07360611308  N07360611309  N07360611310  N07360611311  N07360611312  N07360611313  N07360611314  N07360611315  N07360711329  N07360711330  N07360711332  N07360711333  N07360711334  N07360711335  N07360711336  N07360711337  N07360711338  N07360711339  N07360711340  N07360711342  N07360711343  N07360711344  N07360711345  N07360711346  N07360711347  N07360711348  N07360711353  N07360811377  N07360811383  N07360811388  N06360910374  N06360910376  N07360110913  N07361011515  N08360412243  N08360412244  N08360612637  N08360612645  N08360712762  N08360712763  N08360912960  N08361013094  N09360213567  N06360910434  N06360910439  N06360910445  N06361010465  N06361010491  N06361010537  N06361010538  N06361010539  N06361010540  N06361210634  N07361211810  N06361110618  N08360312227  N09360313709  N06360210205  N06360210208  N06360210209  N06360210214  N06360310211  N06360310218  N06360310218  N06360310220  N06360610253  N07360110831  N07360110832  N07360110834  N07360110835  N07360110893  N07360210965  N07360411102  N07360411104  N07360411106  N07360411110  N07360411112  N07360411113  N07360411114  N07360411115  N07360411116  N07360411117  N07360411118  N07360411119  N07360411120  N07360411121  N07360411122  N07360411123  N07360411124  N07360411125  N07360411126  N07360511130  N07360511142  N07360511143  N07360511144  N07360511145  N07360511146  N07360511147  N07360511148  N07360511149  N07360511150  N07360511151  N07360511152  N07360511153  N07360511154  N07360511155  N07360511156  N07360511163  N07360511164  N07360511165  N07360511166  N07360511167  N07360511168  N07360511169  N07360511170  N07360511171  N07360511172  N07360511173  N07360511174  N07360511175  N07360511176  N07360511177  N07360511185  N07360511186  N07360511187  N07360511190  N07360511191  N07360511192  N07360511193  N07360511194  N07360511195  N07360511200  N07360511201  N07360611257  N07360611258  N07360611259  N07360611260  N07360611261  N07360611262  N07360611263  N07360611264  N07360611265  N07360611266  N07360611267  N07360611268  N07360611269  N07360611270  N07360611271  N07360611272  N07360611273  N07360611274  N07360611275  N08360111877  N08360111878  N08360111879  N08360111880  N08360111881  N08360111882  N08360111883  N08360111884  N08360111885  N08360111910  N08360111911  N08360111912  N08360111913  N08360111914  N08360111915  N08360111916  N08360111917  N08360111929  N08360111930  N08360111931  N08360111932  N08360111933  N08360111934  N08360111935  N08360111936  N08360111937  N08360111946  N08360111947  N08360111948  N08360111949  N08360111950  N08360111951  N08360111952  N08360111953  N08360111954  N08360111955  N08360111956  N08360212056  N08360212057  N08360212058  N08360212059  N08360212060  N08360212072  N08360212091  N08360212092  N08360212093  N08360212102  N08360212103  N08360212104  N08360212105  N08360212106  N08360212107  N08360212108  N08360212109  N08360212110  N08360212111  N08360212112  N08360212113  N08360212114  N08360212115  N08360212116  N08360312164  N08360312165  N08360312177  N08360312178  N08360312179  N08360312180  N08360312181  N08360312184  N08360312185  N08360312186  N08360312187  N08360312188  N08360312189  N08360312190  N08360312191  N08360312192  N08360312193  N08360312194  N08360312195  N08360312196  N08360312197  N08360312198  N08360312199  N08360312200  N08360312201  N08360312202  N08360312203  N08360312204  N08360312205  N08360312206  N08360312207  N08360312208  N08360312209  N08360312210  N08360312211  N08360312212  N08360312213  N08360312214  N08360312215  N08360312216  N08360312217  N08360312218  N08360312219  N08360312220  N08360312221  N08360312222  N08360312223  N08360312224  N08360312225  N08360312226  N08360512376  N08360512377  N08360512379  N08360512380  N08360512387  N08360512388  N08360512389  N08360512390  N08360512391  N08360512392  N08360512393  N08360512425  N08360512426  N08360512427  N08360512428  N08360512429  N08360512430  N08360512431  N08360512432  N08360512433  N08360512434  N08360512435  N08360512438  N08360512442  N08360512444  N08360512446  N08360512449  N08360512450  N08360512451  N08360512452  N08360512453  N08360512458  N08360512459  N08360512460  N08360512461  N08360512462  N08360512465  N08360512466  N08360512491  N08360512497  N08360512498  N08360512499  N08360512500  N08360512501  N08360512502  N08360512503  N08360512504  N08360512505  N08360512506  N08360512507  N08360512508  N08360512509  N08360512510  N08360512511  N08360512512  N08360512513  N08360512514  N08360512515  N08360512516  N08360512517  N08360612596  N08360612597  N08360612598  N08360612599  N08360612600  N08360612601  N08360612602  N08360612603  N08360612604  N08360612605  N08360612606  N08360612607  N08360612608  N08360612609  N08360612610  N08360612611  N08360612612  N08360612613  N08360612614  N08360612615  N08360612616  N08360612617  N08360612618  N08360612619  N08360612620  N08360612621  N08360612622  N08360612623  N08360612624  N08360612625  N08360612626  N08360612627  N08360612628  N08360612629  N08360612630  N08360612631  N08360612632  N08360612633  N08360612634  N08360612635  N08360812854  N08360812855  N08360812856  N08360812857  N09360113472  N09360113473  N09360113474  N09360113475  N09360113476  N09360113477  N09360113478  N09360113479  N09360113480  N09360113481  N09360113482  N09360113483  N09360113484  N09360113485  N09360113486  N09360113487  N09360113488  N09360113489  N09360113490  N09360113491  N09360113509  N09360113510  N09360113511  N09360113512  N09360113513  N09360113514  N09360113515  N09360113516  N09360113517  N09360113518  N09360113519  N09360113520  N09360113521  N09360113522  N09360113523  N09360113524  N09360113525  N09360113526  N09360113527  N09360113528  N09360113529  N09360113530  N09360113531  N09360113532  N09360113533  N09360113534  N09360113535  N09360113536  N09360113537  N09360113538  N09360113539  N09360113540  N09360113541  N09360113542  N09360113543  N09360113544  N09360113545  N09360113546  N09360113547  N09360113548  N09360313746  N09360413866  N09360413867  N09360413868  N09360413869  N09360413870  N09360413871  N09360413872  N09360413873  N09360413874  N09360413875  N09360413876  N09360413877  N09360413878  N09360413879  N09360413880  N09360413881  N09360413882  N09360413883  N09360413884  N09360413885  N09360413886  N09360413887  N09360413888  N09360413889  N09360413890  N09360413891  N09360413892  N09360413893  N09360413894  N09360413895  N09360413897  N09360413898  N09360413899  N09360413900  N09360413907  N09360413908  N09360413909  N09360413910  N09360413911  N09360413912  N09360413913  N09360413914  N09360413915  N09360413916  N09360413917  N09360413918  N09360413919  N09360413920  N09360413921  N09360413922  N09360413923  N09360413924  N09360413925  N09360413926  N09360413927  N09360413928  N09360413929  N09360413930  N09360413931  N09360413932  N09360413933  N09360413934  N09360413935  N09360413950  N09360413951  N09360413952  N09360413953  N09360413954  N09360413955  N09360413956  N09360413957  N09360413958  N09360413959  N09360413960  N09360413961  N09360413962  N09360413963  N09360413964  N09360413965  N09360413966  N09360413967  N09360413968  N09360413969  N09360413970  N09360413972  N09360413973  N09360413974  N09360413975  N09360413976  N09360413977  N09360413978  N06361210716  N07360110836  N07360110837  N07360110894  N07360110910  N07360110915  N06360210206  N06360510236  N06360610254  N06360610264  N06360610266  N06360610274  N06360710303  N06360710304  N06360810328  N06361210664  N06361210666  N06361210667  N06361210674  N06361210675  N06361210676  N06361210689  N07360110838  N07360110839  N07360110841  N07360110874  N07360411128  N07360511157  N07360511158  N07360511159  N07360611240  N07360611241  N07360611298  N07360811391  N07360811392  N07360811418  N07360811419  N07360811422  N07360811424  N07360811425  N07360811426  N07360811458  N07360811462  N07361011512  N07361011513  N07361011514  N07361111629  N07361111630  N07361111631  N07361111639  N07361111640  N07361111642  N07361211781  N07361211815  N07361211816  N07361211817  N07361211818  N07361211819  N07361211820  N08360111906  N08360111907  N08360111908  N08360111909  N08360212080  N08360212081  N08360312153  N08360412228  N08360412229  N08360412230  N08360412231  N08360712657  N08360712744  N08360712745  N08360712747  N08360712748  N08360712749  N08360912910  N08360912911  N08360912912  N08360912958  N08360912959  N08360912978  N08360913012  N08360913013  N08360913014  N08360913015  N08360913016  N08360913017  N08360913018  N08360913019  N09360113441  N09360113442  N09360313726  N09360313782  N09360313783  N09360313784  N09360313785  N09360313789  N08360111959  N08360111960  N08360111962  N08360212074  N08360212075  N06360510237  N06360710297  N06360710298  N06360710299  N06360710300  N06360910358  N06360910394  N06360910396  N06361110565  N06361110566  N06361110567  N06361110574  N06361110583  N06361110584  N06361110585  N06361110586  N06361110587  N06361110588  N06361110589  N06361110590  N06361110591  N06361110592  N06361110593  N06361110594  N06361110595  N06361110596  N06361110597  N06361110598  N06361110599  N06361110600  N06361110601  N06361110602  N06361110603  N06361110604  N06361110605  N06361110606  N06361110607  N06361110608  N06361110609  N06361110610  N06361110611  N06361110612  N06361110613  N06361110614  N06361110615  N06361110616  N06361110617  N07360511137  N07360611233  N07360611234  N07361011518  N07361011519  N08360812866  N08360812874  N08360913053  N08360913054  N08361013157  N08361013176  N08361013177  N09360113453  N09360113454  N07360911467  N07361011520  N07361011521  N07361011522  N08361013091  N08361013092  N08361013093  N06360210198  N06360210200  N073601111669  N07360310980  N07360310984  N07360310987  N07360310994  N07360311009  N07360311012  N07360311032  N07360511203  N07360511205  N07360511211  N07360511231  N07360511232  N07360611250  N07360811390  N07360811442  N07360811463  N07360811463  N07361011553  N07361011553  N07361111624  N07361111625  N07361111625  N07361111625  N07361211802  N07361211808  N07361211809  N07361211860  N07361211861  N08360111922  N08360111972  N08360212012  N08360212037  N08360212062  N08360412265  N08360512467  N08360712663  N08360712664  N08360712665  N08360712666  N08360712667  N08360712669  N08360812788  N08360812793  N08360812808  N08360912883  N08361213342  N06360610269  N07360110842  N07360110843  N07360110844  N07360110845  N07360110846  N07360110847  N07360110848  N07360310980  N07360310981  N07360310982  N07360310984  N07360310985  N07360310986  N07360310987  N07360310988  N07360310989  N07360310990  N07360310991  N07360310992  N07360310993  N07360310994  N07360310995  N07360310996  N07360310997  N07360310998  N07360310999  N07360311000  N07360311001  N07360311002  N07360311003  N07360311004  N07360311005  N07360311006  N07360311007  N07360311008  N07360311009  N07360311010  N07360311011  N07360311012  N07360311013  N07360311014  N07360311015  N07360311016  N07360311017  N07360311018  N07360311019  N07360311020  N07360311021  N07360311022  N07360311023  N07360311024  N07360311025  N07360311026  N07360311027  N07360311028  N07360311029  N07360311030  N07360311031  N07360311031  N07360311032  N07360311034  N07360311035  N07360311036  N07360311037  N07360311038  N07360311039  N07360311040  N07360311041  N07360311042  N07360311043  N07360311044  N07360311045  N07360311046  N07360311047  N07360311048  N07360311049  N07360311050  N07360311051  N07360311052  N07360311053  N07360311054  N07360311055  N07360311056  N07360311057  N07360311058  N07360311059  N07360311060  N07360311061  N07360311062  N07360311063  N07360311064  N07360311065  N07360311066  N07360311067  N07360311068  N07360311069  N07360311070  N07360311073  N07360311074  N07360311075  N07360311076  N07360311077  N07360311078  N07360311079  N07360311080  N07360311081  N07360311082  N07360311092  N07360511182  N07360511203  N07360511204  N07360511205  N07360511206  N07360511207  N07360511208  N07360511209  N07360511210  N07360511211  N07360511212  N07360511213  N07360511214  N07360511215  N07360511216  N07360511217  N07360511218  N07360511219  N07360511220  N07360511221  N07360511222  N07360511223  N07360511224  N07360511225  N07360511226  N07360511227  N07360511228  N07360511229  N07360511230  N07360611247  N07360611248  N07360611249  N07360611250  N07360611251  N07360611252  N07360611253  N07360611254  N07360611255  N07360611256  N07360611287  N07360611288  N07360611289  N07360611290  N07360611291  N07360811374  N07360811376  N07360811378  N07360811379  N07360811381  N07360811382  N07360811385  N07360811389  N07360811390  N07360811393  N07360811394  N07360811399  N07360811400  N07360811405  N07360811415  N07360811417  N07360811429  N07360811433  N07360811434  N07360811435  N07360811436  N07360811437  N07360811438  N07360811439  N07360811440  N07360811441  N07360811442  N07360811443  N07360811444  N07360811445  N07361011493  N07361011494  N07361011495  N07361011496  N07361011497  N07361011498  N07361011499  N07361011500  N07361011501  N07361011502  N07361011504  N07361011505  N07361011508  N07361011510  N07361011511  N07361011545  N07361011546  N07361011547  N07361011548  N07361011549  N07361011550  N07361011551  N07361011552  N07361011553  N07361111668  N07361111669  N07361111670  N07361111671  N07361111672  N07361111673  N07361111674  N07361111675  N07361111676  N07361111677  N07361111716  N07361111717  N07361111718  N07361111719  N07361111720  N07361111721  N07361111727  N07361111733  N07361111734  N07361111735  N07361211776  N07361211777  N07361211782  N07361211809  N07361211814  N07361211859  N07361211860  N07361211861  N07361211862  N07361211873  N08360111919  N08360111920  N08360111921  N08360111922  N08360111923  N08360111924  N08360111925  N08360111926  N08360111927  N08360111928  N08360111938  N08360111940  N08360111941  N08360111942  N08360111943  N08360111944  N08360111971  N08360111972  N08360111973  N08360111974  N08360111976  N08360111977  N08360111978  N08360113456  N08360113456  N08360113458  N08360113458  N08360113459  N08360113459  N08360113460  N08360113460  N08360211993  N08360211994  N08360211995  N08360211996  N08360211997  N08360211998  N08360211999  N08360212000  N08360212001  N08360212002  N08360212003  N08360212005  N08360212006  N08360212007  N08360212008  N08360212009  N08360212010  N08360212011  N08360212012  N08360212013  N08360212014  N08360212015  N08360212016  N08360212017  N08360212018  N08360212019  N08360212020  N08360212027  N08360212028  N08360212029  N08360212030  N08360212031  N08360212032  N08360212033  N08360212034  N08360212035  N08360212036  N08360212037  N08360212038  N08360212039  N08360212040  N08360212041  N08360212042  N08360212043  N08360212044  N08360212045  N08360212046  N08360212053  N08360212054  N08360212055  N08360212061  N08360212062  N08360212063  N08360212064  N08360212065  N08360212070  N08360212071  N08360212073  N08360212077  N08360212078  N08360212079  N08360212094  N08360212095  N08360212096  N08360212097  N08360212098  N08360212099  N08360212100  N08360212101  N08360312117  N08360312118  N08360312119  N08360312120  N08360312121  N08360312122  N08360312123  N08360312124  N08360312125  N08360312126  N08360312127  N08360312128  N08360312129  N08360312130  N08360312134  N08360312135  N08360312136  N08360312137  N08360312138  N08360312139  N08360312140  N08360312141  N08360312142  N08360312143  N08360312144  N08360312145  N08360312146  N08360312147  N08360312148  N08360412232  N08360412233  N08360412234  N08360412235  N08360412236  N08360412237  N08360412238  N08360412239  N08360412251  N08360412253  N08360412257  N08360412258  N08360412259  N08360412260  N08360412263  N08360412264  N08360412265  N08360412271  N08360412272  N08360412273  N08360512408  N08360512409  N08360512410  N08360512411  N08360512412  N08360512416  N08360512417  N08360512418  N08360512419  N08360512420  N08360512421  N08360512422  N08360512423  N08360512424  N08360512436  N08360512437  N08360512439  N08360512440  N08360512441  N08360512443  N08360512445  N08360512447  N08360512448  N08360512454  N08360512455  N08360512456  N08360512457  N08360512463  N08360512464  N08360512467  N08360512468  N08360512470  N08360512478  N08360512479  N08360512480  N08360512481  N08360512482  N08360512484  N08360512489  N08360512490  N08360712663  N08360712664  N08360712665  N08360712666  N08360712667  N08360712668  N08360712669  N08360712670  N08360712671  N08360712672  N08360712677  N08360712678  N08360712679  N08360712680  N08360712681  N08360712682  N08360712683  N08360712684  N08360712685  N08360712686  N08360812787  N08360812789  N08360812790  N08360812791  N08360812792  N08360812794  N08360812795  N08360812796  N08360812797  N08360812798  N08360812799  N08360812805  N08360812806  N08360812807  N08360812814  N08360812815  N08360812821  N08360812822  N08360812826  N08360812827  N08360812847  N08360812848  N08360812849  N08360812850  N08360812858  N08360812862  N08360812863  N08360812864  N08360812865  N08360812867  N08360912883  N08360912884  N08360912885  N08360912886  N08360912887  N08360912888  N08360912889  N08360912890  N08360912891  N08360912892  N08361013095  N08361013096  N08361013097  N08361013098  N08361013099  N08361013100  N08361013101  N08361013102  N08361013115  N08361013116  N08361013117  N08361013118  N08361013119  N08361013120  N08361013121  N08361013122  N08361013123  N08361013132  N08361013133  N08361013134  N08361013135  N08361013136  N08361013139  N08361013140  N08361013141  N08361113196  N08361113197  N08361113198  N08361113199  N08361113200  N08361113201  N08361113202  N08361113203  N08361113204  N08361113205  N08361113206  N08361113207  N08361113208  N08361113222  N08361113223  N08361113224  N08361113232  N08361113233  N08361113234  N08361113235  N08361113253  N08361113254  N08361113255  N08361113256  N08361113257  N08361113262  N08361113263  N08361113264  N08361113268  N08361213282  N08361213291  N08361213296  N08361213297  N08361213298  N08361213299  N08361213300  N08361213303  N08361213304  N08361213305  N08361213306  N08361213307  N08361213308  N08361213309  N08361213310  N08630212071  N09360113430  N09360113431  N09360113432  N09360113433  N09360113436  N09360113437  N09360113438  N09360113439  N09360113440  N09360113449  N09360113450  N09360113455  N09360113456  N09360113458  N09360113459  N09360113460  N09360213553  N09360213562  N09360213563  N09360213564  N09360213565  N09360213568  N09360213578  N09360213579  N09360213580  N09360213581  N09360213582  N09360213583  N09360213584  N09360213585  N09360213586  N09360213587  N09360213588  N09360213589  N09360213590  N09360213591  N09360213593  N09360213594  N09360213595  N09360213596  N09360213598  N09360213601  N09360213602  N09360213604  N09360213605  N09360213607  N09360213608  N09360213612  N09360213613  N09360213615  N09360213618  N09360213619  N09360213620  N09360213644  N09360213647  N09360213648  N09360213649  N09360213650  N09360213651  N09360213652  N09360213653  N09360213654  N09360213655  N09360213656  N09360213675  N09360213676  N09360213677  N09360213678  N09360213679  N09360213680  N09360213681  N09360213682  N09360213683  N09360213684  N09360213685  N09360213686  N09360213690  N09360213691  N09360213692  N09360213693  N09360213694  N09360213695  N09360313721  N09360313722  N09360313723  N09360313724  N09360313732  N09360313733  N09360313741  N09360313760  N09360313770  N09360313775  N09360313776  N09360313777  N09360313778  N09360313780  N09360313781  N09360413902  N09360413903  N09360413904  N09360413905  N09360413906  N09360413936  N09360413937  N09360413938  N09360413939  N09360413940  N09360413941  N09360413947  N09360413948  N09360413949  N09360413971  N09360414014  N09360414029  N09360414030  N09360414031  N09360414032  N09360414033  N09360414035  N09360414036  N09360414037  N07360110888  N08360212083  N08360212086  N08360212087  N08360312170  N08360312171  N08360412245  N08360412246  N08360412248  N08360412250  N08360412266  N08360412365  N08361213280  N08361213281  N09360413991  N09360413998  N09360413999  N09360414001  N09360414015  N08360412241  N07360511162  N06360410228  N06360810325  N07360210955  N07360210956  N07360611239  N07360711316  N07360711323  N07360711325  N07361111643  N07361111644  N07361111647  N07361111658  N07361111659  N07361111660  N07361111662  N07361111663  N07361111664  N07361111665  N07361111666  N07361111667  N07361111678  N07361111679  N07361111685  N08360212047  N08360212089  N08360212090  N08360312173  N08360412267  N08360512404  N08360512405  N08360512413  N08360712673  N08360712674  N08360912961  N08360912963  N08360912970  N08360912977  N07361211780  N08360212076  N07361011517  N06360410230  N08360412368  N08360412369  N08360412374  N08360912969  N08360912973  N08360912974  N08360912975  N08361113187  N08361113188  N08361213285  N08361213286  N08361213287  N08361213288  N08361213289  N08361213290  N08361013156  N08361113240  N08361113241  N08361113242  N08361113243  N09360414000  N06360910390  N06361010535  N06361110568  N06361210630  N07360611295  N07360611297  N08360111961  N08360212082  N08360212084  N08360212085  N08360412375  N08360512473  N08360812852  N08360812853  N06360510234  N06360810327  N06360810329  N06360810330  N06361210670  N06361210671  N06361210672  N06361210673  N07360411103  N07360511179  N07360511180  N07361211774  N08360412262  N08360912966  N06360610246  N06360610261  N06361110573  N06361110575  N06361110576  N06361110577  N06361110579  N07360110901  N07360110902  N07360110903  N07360110904  N07360110905  N07360110906  N07360110907  N07360110908  N07360911470  N07361111686  N07361211799  N07361211800  N07361211801  N08360612591  N08360612592  N08360612594  N08360612595  N08360712714  N08360712717  N08360912928  N08360912929  N08360912930  N08360912998  N08360912999  N08360913003  N08360913004  N08360913006  N08361113189  N08361113190  N08361113191  N08361113192  N08361113193  N08361113194  N08361113195  N09360113406  N09360113407  N09360113408  N09360113409  N09360113410  N09360313697  N09360313698  N09360313699  N09360313700  N09360313701  N09360313702  N09360313703  N09360313704  N09360313705  N09360313706  N09360313707  N09360313708  N09360514039  N09360514040  N09360514041  N09360514042  N08360111945  N08360211991  N08360211992  N06360410224  N06360410229  N06360610240  N06360610241  N06360610242  N06360610243  N06360610244  N06360610255  N06360610263  N06360610267  N06360810333  N06360910391  N07360711352  N08360412252  N08360412254  N08360412255  N08360412256  N09360413979  N09360413980  N09360413981  N09360413982  N09360414004  N09360414005  N09360414006  N09360414007  N09360414008  N09360414009  N09360414010  N09360414011  N09360414012  N09360414013  N09360414016  N09360414017  N09360414018  N09360414019  N09360414020  N09360414021  N09360414022  N09360414023  N09360414024  N09360414025  N09360414026  N09360414027  N09360414028  N09360414034  N07361011509  N08360111918  N08360111969  N08360111970  N08360111979  N08360111980  N08360111981  N08360111982  N08360111983  N08360111984  N08360111985  N08360111986  N08360612527  N08360712761  N09360213687  N09360213688  N09360213689  N09360213560  N08360812768  N08360812769  N08360812770  N08360812771  N08360812772  N08361013150  N08361013152  N08361013155  N08361013170  N08361013171  N08361013172  N08361013173  N08361013174  N08361213343  N08361213344  N08361213345  N08361213346  N08361213347  N08360312172  N08360312175  N08360612560  N08360712656  N09360213570  N09360213572  N09360213574  N09360213575  N09360213603  N09360213609  N06360610275  N06360910381  N06360910382  N06360910383  N06360910384  N06360910385  N06360910386  N07361111641  N08360412367  N08360512394  N08360812779  N08360812802  N08360812803  N08360912893  N08360912894  N08360912895  N07360711365  N07360811451  N07360811452  N07360811453  N07360811454  N07360811455  N07360811456  N07360811457  N08360111905  N08360312133  N08360312154  N08360912962  N08360912965  N08360912968  N08360912990  N08360912991  N08360912992  N08360912993  N08360912994  N09360113457  N09360113461  N09360113462  N09360113463  N09360113465  N09360113466  N09360113467  N09360113468  N09360113469  N09360113470  N09360113471  N09360113492  N07361111687  N07361111729  N08361013086  N08361013087  N08361013088  N08361013089  N08361013090  N08361013148  N08361013149  N08361013151  N08361013154  N08361213292  N08361213293  N08361213294  N08361213348  N08361213349  N06360210202  N06360610239  N06360610252  N06361010517  N07361111661  N07361211826  N07361211827  N07361211828  N07361211829  N07361211830  N08360111900  N08360111901  N08360111902  N08360111903  N08360111904  N08361213350  N08361213351  N08361213352  N08361213353  N08361213354  N08361213355  N08361213356  N08361213357  N08361213358  N08361213359  N08361213360  N08361213361  N08361213362  N08361213363  N08361213364  N08361213365  N08361213366  N08361213367  N08361213368  N09360213592  N09360213597  N09360213599  N09360213611  N09360213614  N09360213621  N09360213622  N09360213623  N09360213624  N09360213625  N09360213626  N09360213627  N09360213628  N09360213629  N09360213630  N09360213631  N09360213632  N09360213633  N09360213634  N09360213666  N09360213667  N09360213668  N09360213669  N09360213670  N09360213671  N09360213672  N09360213673  N09360213674  N06360710301  N06360710302  N07360611280  N07361211811  N08360312156  N08360512406  N08360512407  N08360812785  N08360812786  N09360213616  N09360213617  N06360410227  N06360710288  N06360710289  N06360810320  N06360810322  N06360810324  N06360910356  N06360910357  N06361010524  N06361010525  N07360210939  N07360210940  N07360210941  N07360210942  N07360210943  N07361111751  N07361111752  N07361111753  N07361211778  N07361211779  N07361211838  N07361211839  N07361211840  N07361211841  N08360312149  N08360312150  N08360312151  N08360312152  N08360312168  N08360312169  N08360412370  N08360412371  N08360412372  N08360412373  N08360712739  N08360712740  N08360712741  N08360712742  N08360712743  N08361113185  N08361113186  N09360113434  N09360113435  N09360113443  N09360113444  N09360113445  N07360210958  N07360511132  N07360511133  N07360511140  N07360511141  N07360711320  N07360711326  N07360711327  N07360711371  N07360811409  N07360811421  N07360811423  N07360911478  N07361011588  N07361011592  N07361011593  N07361011594  N07361011595  N07361011596  N07361011597  N07361011598  N07361011599  N07361011600  N07361011601  N07361011602  N07361011603 N07361011604  N07361011606  N07361111704  N07361111705  N07361111714  N07361111715  N08360512483  N08360512485  N08360512486  N08360512487  N08360512488  N08360612531  N08360612532  N08360612534  N08360612535  N08360612536  N06360610270  N06360910393  N06361210632  N06360810337  N06360910349  N06360910353  N06360910355  N07361111726  N08360212048  N08360212049  N08360212050  N08360212051  N08360212052  N08360512469  N08360512471  N08360512472  E360-0108-062  E360-0108-079  E360-0108-083  E360-0108-084  E360-0108-089  E360-0208-013  E360-0208-014  E360-0208-015  E360-0208-017  E360-0208-023  E360-0208-032  E360-0208-045  E360-0208-052  E360-0208-056  E360-0208-060  E360-0208-066  E360-0208-067  E360-0208-073  E360-0208-074  E360-0208-078  E360-0208-079  E360-0208-080  E360-0208-082  E360-0208-083  E360-0208-091  E360-0208-092  E360-0208-097  E360-0208-101  E360-0208-102  E360-0208-103  E360-0208-105  E360-0208-109  E360-0208-117  E360-0208-118  E360-0208-121  E360-0208-132  E360-0208-136  E360-0208-139  E360-0208-140  E360-0208-141  E360-0208-142  E360-0208-143  E360-0208-145  E360-0208-147  E360-0208-148  E360-0208-152  E360-0208-153  E360-0208-154  E360-0208-157  E360-0208-160  E360-0208-161  E360-0208-163  E360-0208-165  E360-0208-166  E360-0208-168  E360-0208-170  E360-0208-172  E360-0208-173  E360-0208-175  E360-0208-178  E360-0208-179  E360-0208-180  E360-0208-181  E360-0208-182  E360-0208-184  E360-0208-186  E360-0208-187  E360-0208-188  E360-0208-191  E360-0208-192  E360-0208-193  E360-0208-194  E360-0208-197  E360-0208-198  E360-0208-199  E360-0208-200  E360-0208-201  E360-0208-202  E360-0208-203  E360-0208-204  E360-0208-206  E360-0208-207  E360-0208-208  E360-0208-210  E360-0208-212  E360-0208-214  E360-0208-218  E360-0208-220  E360-0208-221  E360-0208-222  E360-0208-223  E360-0208-224  E360-0208-226  E360-0208-228  E360-0208-230  E360-0208-231  E360-0208-232  E360-0208-233  E360-0208-242  E360-0208-243  N06360410232  N06361010515  N06361110623  N06361110624  N06361110625  N06361110626  N06361110627  N06361110628  N07360511134  N07360511136  N07360511138  N07360711328  N07360711341  N07360711349  N07360811375  N07360811461  N07361011605  N07361011607  N07361011608  N07361011609  N07361011610  N07361011614  N07361111702  N07361111706  N07361111707  N07361211770  N07361211771  N07361211822  N07361211823  N07361211842  N07361211843  N07361211844  N07361211845  N07361211846  N07361211847  N07361211848  N07361211849  N07361211850  N07361211851  N08360312132  N08360312167  N08360412240  N08360412242  N08360412247  N08360512403  N08360612530  N08360612533  N08360612559  N08360812773  N08360812774  N08360812775  N08360812776  N08360812777  N08360812778  N08360812780  N08360812781  N08360812782  N08360812783  N08360812784  N08360912931  N08360912932  N08360912933  N08360912934  N08360912935  N08360912936  N08360912937  N08360912938  N08360912939  N08360912940  N08360912941  N08360912942  N08360912943  N08360912944  N08360912979  N08360912980  N08360912981  N08360912982  N08360912983  N08360912984  N08360912985  N08360912986  N08360912987  N08360912988  N08360912989  N08360913011  N06360210140  N06361210668  N07360110912  N07360110916  N07360110917  N07360110918  N07360110919  N07360210920  N07360210921  N07360210923  N07360210924  N07360210925  N07360210946  N07360210947  N07360210948  N07360210949  N07360210950  N07360511181  N07360511183  N07360511184  N07360511188  N07360511189  N07360511196  N07360511197  N07360511198  N07360511199  N07360511202  N07360711331  N07360711351  N07360711354  N07360711355  N07360711356  N07360711357  N07360711358  N07360711359  N07360711366  N07360711367  N07361111680  N07361111681  N07361111683  N07361111684  N07361111689  N07361111690  N07361111691  N07361111692  N07361111692*  N07361111698  N07361111699  N08360212088  N08360412249  N08360412269  N08360412270  N08360412274  N08360412275  N08360412276  N08360412277  N08360412278  N08360412279  N08360412280  N08360412281  N08360412282  N08360412283  N08360412284  N08360412285  N08360412286  N08360412287  N08360412288  N08360412289  N08360412290  N08360412291  N08360412292  N08360412293  N08360412294  N08360412295  N08360412296  N08360412297  N08360412298  N08360412299  N08360412300  N08360412301  N08360412302  N08360412303  N08360412304  N08360412305  N08360412306  N08360412307  N08360412308  N08360412309  N08360412310  N08360412311  N08360412312  N08360412313  N08360412314  N08360412315  N08360412316  N08360412317  N08360412318  N08360412319  N08360412320  N08360412321  N08360412322  N08360412323  N08360412324  N08360412325  N08360412326  N08360412327  N08360412328  N08360412329  N08360412330  N08360412331  N08360412332  N08360412333  N08360412334  N08360412335  N08360412336  N08360412337  N08360412338  N08360412339  N08360412340  N08360412341  N08360412342  N08360412343  N08360412344  N08360412345  N08360412346  N08360412347  N08360412348  N08360412349  N08360412350  N08360412351  N08360412352  N08360412353  N08360412354  N08360412355  N08360412356  N08360412357  N08360412358  N08360412359  N08360412360  N08360412361  N08360412362  N08360412363  N08360412364  N08361113209  N08361113210  N08361113211  N08361113212  N08361113213  N08361113214  N08361113215  N08361113216  N08361113217  N08361113218  N08361113219  N08361113220  N08361113221  N08361113225  N08361113226  N08361113227  N08361113228  N08361113229  N08361113230  N08361113231  N08361113236  N08361113237  N08361113238  N08361113239  N08361113244  N08361113245  N08361113246  N08361113247  N08361113258  N08361113259  N08361113260  N08361113261  N08361113265  N08361113266  N08361113267  N08361113269  N08361113270  N08361113271  N08361113272  N08361113273  N08361113274  N08361113275  N08361113276  N08361113277  N08361113278  N08361113279  N09360113415  N09360113416  N09360113417  N09360113418  N09360113419  N09360113420  N09360113421  N09360313725  N09360313744  N09360313745  N09360313747  N09360313748  N09360313749  N09360313750  N09360313751  N09360313752  N09360313753  N09360313754  N09360313755  N09360313756  N09360313757  N09360313758  N09360313761  N09360313762  N09360313763  N09360313764  N09360313765  N09360313769  N09360313771  N09360313773  N09360313774  N09360313779  N09360313788  N09360313790  N09360313791  N09360313792  N09360313793  N09360313794  N09360313795  N09360313796  N09360313797  N09360313798  N09360313799  N09360313800  N09360313801  N09360313802  N09360313803  N09360313804  N09360313805  N09360313806  N09360313807  N09360313808  N09360313809  N09360313810  N09360313811  N09360313812  N09360313813  N09360313814  N09360313815  N09360313816  N09360313817  N09360313818  N09360313819  N09360313820  N09360313821  N09360313822  N09360313823  N09360313824  N09360313825  N09360313826  N09360313827  N09360313828  N09360313829  N09360313830  N09360313831  N09360313833  N09360313834  N09360313835  N09360313836  N09360313837  N09360313838  N09360313839  N09360313840  N09360313841  N09360313842  N09360313843  N09360313844  N09360313845  N09360313846  N09360313848  N09360313849  N09360313850  N09360313851  N09360313852  N09360313853  N09360313854  N09360313855  N09360313856  N09360313857  N09360313858  N09360313859  N09360313860  N09360313861  N09360313862  N09360313863  N09360313864  N09360313865  N08360111939  N08360111963  N08360111964  N08360111965  N08360111966  N08360111967  N08360111968  N08360111975  N03631110581  N06360310221  N06360610276  N06360810338  N06360910395  N07360210971  N07360210973  N07360210974  N07360210975  N07360210976  N07360210977  N07360210978  N07360210979  N07360310983  N07360311071  N07360311072  N07360311083  N07360311084  N07360311085  N07360311086  N07360311087  N07360311088  N08360612636  N08360612638  N08360612639  N08360612640  N08360612641  N08360612642  N08360612643  N08360612644  N09360113422  N09360113423  N09360113424  N09360113425  N09360113426  N09360113427  N09360113428  N09360113429  N09360113464  N09360113493  N09360113497  N09360113498  N09360113499  N09360113502  N09360113503  N09360113504  N09360113505  N09360113506  N09360113507  N09360113508  N06360810339  N07361111645  N07361111646  N08360312176  N08360612584  N08360612585  N09360213566  N06361110569  N06361210640  N06361210641  N06361210642  N06361210643  N07361211821  N08360212021  N08360212022  N08360212023  N08360212024  N08360212025  N08360212026  N08360612548  N08360612549  N08360612550  N08360612551  N08360612552  N08360612553  N08360612554  N08360612555  N08360612556  N08360612557  N08360612558  N09360213557  N09360213558  N09360213559  N09360213569  N09360213571  N09360213573  N09360213576  N09360213577  N09360413901  N06360310215  N06360410233  N06360810336  N07361111688  N08360412261  N08360412268  N08360512414  N08360512415  N08360612518  N08360612519  N08360612520  N08360812845  N08360812846  N08360912879  N08360912880  N08360912881  N08360912882  N08360912964  N08360912971  N08360912972  N09360113403  N09360113404  N09360113405  N09360113446  N09360113447  N09360113448  N06360210199  N06360610260  N06361010549  N06361110619  N06361110620  N06361110621  N06361110622  N06361210635  N06361210637  N06361210639  N06361210644  N06361210645  N06361210648  N06361210649  N06361210650  N06361210651  N06361210652  N06361210653  N06361210654  N06361210655  N06361210656  N06361210657  N06361210659  N06361210661  N06361210662  N06361210677  N06361210678  N06361210679  N06361210680  N06361210681  N06361210682  N06361210683  N06361210684  N06361210685  N06361210686  N06361210687  N06361210688  N08361213301  N08361213302  N08361213311  N08361213312  N08361213313  N08361213314  N08361213315  N08361213316  N08361213317  N08361213318  N08361213319  N08361213320  N08361213321  N08361213322  N08361213323  N08361213324  N08361213325  N08361213326  N08361213327  N08361213328  N08361213329  N08361213330  N08361213331  N08361213332  N08361213333  N08361213334  N08361213335  N08361213336  N08361213337  N08361213338  N08361213339  N08361213340  N08361213341  N09360213635  N09360213636  N09360213637  N09360213638  N09360213639  N09360213640  N09360213641  N09360213642  N09360213643  N09360213645  N09360213646  N09360213657  N09360213658  N09360213659  N09360213660  N09360213661  N09360213662  N09360213663  N09360213664  N09360213665  N06360410231  N06360510235  N06360610247  N06360610248  N06360610249  N06360610251  N06360610272  N06360610273  N06360610277  N06360610278  N06360610279  N06360610280  N06360610281  N06360710282  N06360710283  N06360710284  N06360710285  N06360710286  N06360710287  N06360710290  N06360710291  N06360710292  N06360810315  N06360910341  N06360910342  N06360910343  N06360910344  N06360910345  N06360910346  N06360910347  N06360910348  N06360910359  N07360110814  N07360110815  N07360110816  N07360110817  N07360110879  N07360110880  N07360110892  N07360110895  N07360110896  N07360110897  N07360110899  N07360110900  N07360411097  N07360411098  N07360411099  N07360411100  N07360511178  N07360611237  N07360611238  N07360711350  N07360911466  N07360911468  N07360911477  N07360911479  N07360911480  N07360911481  N07360911482  N07360911483  N07360911484  N07361011536  N07361211837  N07361211852  N07361211853  N07361211854  N07361211855  N07361211856  N07361211857  N08360912922  N08360912923  N08361013103  N08361013104  N09360113494  N09360113495  N09360113496  N09360413992  N09360413994  N09360413996  N07360411093  N07361011503  N07361011555  N07361011556  N07361011557  N07361011558  N07361011559  N07361011560  N07361011561  N07361011562  N07361011563  N07361011564  N07361011565  N07361011566  N07361011567  N07361011568  N07361011569  N07361011570  N07361011571  N07361011572  N07361011573  N07361011574  N07361011575  N07361011576  N07361011577  N07361011578  N07361011579  N07361011580  N07361011581  N07361011582  N07361011583  N07361011584  N07361011585  N07361011586  N07361011587  N08360912915  N08360912917  N08360912918  N08360912945  N08360912946  N08360912947  N08360912948  N08360912951  N08360912952  N08360912953  N08360912954  N08360912956  N08360912967  N08361013082  N08361013083  N08361013084  N08361013085  N08361013158  N08361013159  N06360210101  N06360210102  N06360210103  N06360210104  N06360210105  N06360210106  N06360210107  N06360210108  N06360210109  N06360210110  N06360210111  N06360210112  N06360210113  N06360210114  N06360210115  N06360210116  N06360210117  N06360210118  N06360210119  N06360210120  N06360210121  N06360210122  N06360210123  N06360210124  N06360210125  N06360210126  N06360210127  N06360210128  N06360210129  N06360210130  N06360210131  N06360210132  N06360210133  N06360210134  N06360210135  N06360210136  N06360210137  N06360210138  N06360210139  N06360210141  N06360210142  N06360210143  N06360210144  N06360210145  N06360210146  N06360210147  N06360210148  N06360210149  N06360210150  N06360210151  N06360210152  N06360210153  N06360210154  N06360210155  N06360210156  N06360210157  N06360210158  N06360210159  N06360210160  N06360210161  N06360210162  N06360210163  N06360210164  N06360210165  N06360210166  N06360210167  N06360210168  N06360210169  N06360210170  N06360210171  N06360210172  N06360210173  N06360210174  N06360210175  N06360210176  N06360210177  N06360210178  N06360210179  N06360210180  N06360210181  N06360210182  N06360210183  N06360210184  N06360210185  N06360210186  N06360210187  N06360210188  N06360210189  N06360210190  N06360210191  N06360210192  N06360210193  N06360210194  N06360210195  N06360210196  N06360210197  N06360210203  N06361110578  N06361210636  N06361210646  N06361210663  N06361210665  N06361210690  N06361210691  N06361210692  N06361210694  N06361210695  N06361210696  N06361210697  N06361210698  N06361210699  N06361210700  N06361210701  N06361210702  N06361210703  N06361210704  N06361210705  N06361210707  N06361210708  N06361210709  N06361210710  N06361210711  N06361210712  N06361210713  N06361210714  N06361210715  N06361210717  N06361210718  N06361210719  N06361210720  N06361210721  N06361210722  N06361210723  N06361210724  N06361210725  N06361210726  N06361210727  N06361210728  N06361210729  N06361210730  N06361210731  N06361210732  N06361210733  N06361210734  N06361210735  N06361210736  N06361210737  N06361210738  N06361210739  N06361210740  N06361210741  N06361210742  N06361210743  N06361210744  N06361210745  N06361210746  N06361210747  N06361210748  N06361210749  N06361210750  N06361210751  N06361210752  N06361210753  N06361210754  N06361210755  N06361210756  N06361210757  N06361210758  N06361210759  N07360110760  N07360110761  N07360110762  N07360110763  N07360110764  N07360110765  N07360110766  N07360110767  N07360110768  N07360110769  N07360110770  N07360110771  N07360110772  N07360110773  N07360110774  N07360110775  N07360110776  N07360110777  N07360110778  N07360110779  N07360110780  N07360110781  N07360110782  N07360110783  N07360110784  N07360110785  N07360110786  N07360110787  N07360110788  N07360110789  N07360110790  N07360110791  N07360110792  N07360110793  N07360110794  N07360110795  N07360110798  N07360110799  N07360110800  N07360110801  N07360110802  N07360110803  N07360110804  N07360110805  N07360110806  N07360110807  N07360110808  N07360110809  N07360110810  N07360110811  N07360110812  N07360110813  N07360110818  N07360110819  N07360110820  N07360110821  N07360110822  N07360110823  N07360210927  N07360210928  N07360210929  N07360210930  N07360210931  N07360210932  N07360210933  N07360210934  N07360210935  N07360210957  N07360210960  N07360210969  N07360411094  N07360811408  N07360811420  N07360811428  N07360811446  N07360811447  N07360811448  N07360811449  N07360811450  N07360811459  N07360811460  N07361011611  N07361011612  N07361011613  N07361011615  N07361011616  N07361011617  N07361011618  N07361011619  N07361211803  N07361211804  N07361211805  N07361211806  N07361211807  N08360212004  N08360512492  N08360512493  N08360512494  N08360512495  N08360512496  N08360712676  N08360712719  N08360712720  N08360712721  N08360712722  N08360712723  N08360712724  N08360712725  N08360712727  N08360712728  N08360712729  N08360712730  N08360712731  N08360712732  N08360712733  N08360712734  N08360712735  N08360712736  N08360712737  N08360712738  N09360313711  N09360313712  N09360313713  N09360313714  N09360313715  N09360313716  N09360313717  N09360313718  N09360313719  N09360313720  N06360910397  N07361111708  N07361111709  N07361111710  N08360712726  N08360712746  N08360712750  N08360712751  N08360712752  N08360712753  N08360712754  N08360712755  N08360712756  N08360712757  N08360712758  N08360712759  N08360712760  N08360712764  N08360712765  N08360712766  N08360712767  N08360912924  N08360912925  N08360912949  N08360912950  N08360912955  N08361013163  N08361213401  N08361213402  N07360511131  N07360511135  N07360611301  N07360611302  N07360611303  N07360611307  N07360811430  N07360811431  N07360811432  N07361111700  N07361111701  N07361111703  N08360512474  N08360512475  N08360512476  N08360512477  N06360910369  N07360110877  N08360912919  N09360213556  N09360213561  N09360213606  N09360213610  N07360210922 and  N07360110840.  
Recalling Firm/
Manufacturer
Newport Medical Instruments Inc
1620 Sunflower Ave
Costa Mesa CA 92626
For Additional Information Contact
714-427-5811
Manufacturer Reason
for Recall
In December 2008, Newport initiated a field correction to upgrade e360 ventilators with software versions 3.0 or higher to software version 3.3. As a result of this upgrade, some customers experienced unexpected device alert alarms. Newport's investigation into these complaints confirmed that the 3.3 software responds too quickly to certain situations, giving a false device alert. Newport condu
FDA Determined
Cause 2
Software design
ActionNewport Medical initally sent out a Field Correction Notice to a limited number consignees dated December 17, 2008 and then the notice was expanded to include all e360 consignees dated June 15, 2009. The firm notified consignees that they have revised the e360 software to version 3.8. A customer verification form was attached to verify upgrade completion. For further questions, contact Newport Customer Technical Support Department at 1-800-451-3111 or +1-714-427-5811 Extension 500.
Quantity in Commerce4,117 units (214 Domestically; 3,903 Internationally)
DistributionWorldwide Distribution -- US (including AL, AZ, CA, CO, FL, GA, IL, IN, KY, MA, MD, ME, MI, NJ, NM, NY, OH, OK, OR, PA, UT and VA), Albania, Algeria, Argentina, Armenia, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Brunei, Bulgaria, Canada, Chile, Colombia, Czech Republic, Dominican Republic, Ecuador, Egypt, England, Finland, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Japan, Jordan, Kazahkstan, Kenya, Korea, Kuwait, Lebanon, Liechtenstein, Lithuania, Malaysia, Mauritius, Mexico, Morocco, Nambia, Nepal, New Zealand, Nicaragua, Nigeria, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, Tunisia, Turkey, AUE, Ukraine, United Kingdom, Yemen Republic and Zimbabwe.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.