• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall BHM/MediMan Combi Sling

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall BHM/MediMan Combi Slingsee related information
Date Initiated by FirmApril 07, 2009
Date PostedAugust 31, 2009
Recall Status1 Terminated 3 on January 07, 2011
Recall NumberZ-1990-2009
Recall Event ID 51999
Product Classification Combi Sling - Product Code FNG
ProductBHM Combi Sling, 100% Polyester Shell/100% Polyester Strap; soft polyester net fabric sling for ceiling and floor patient lifts; BHM, Magog, Canada, made in China. Part 626002: Combi Deluxe - Medium, capacity 68-113 kg, 150-250 lbs; Part 626002M: Combi Mesh Deluxe - Medium, capacity 68-113 kg, 150-250 lbs; Part 626002C Combi Sling Deluxe Small, capacity 20-68 kg, 45-150 lbs; Part 626002C-M Combi Sling Mesh Deluxe Small, capacity 20-68 kg, 45-150 lbs; Part 626003: Combi Deluxe - Large, capacity 113-272 kg, 250-600 lbs; Part 626003M: Combi Mesh Deluxe - Large, capacity 113-272 kg, 250-600 lbs; and Part 626003X: Combi Oversize Heavy Duty, capacity 113-272 kg, 250-600 lbs. Between 2004 and 2006, the Combi Slings were labeled as BHM Medical Inc. (800) 868-0441, 100% Polyester Shell/100% Nylon Strap, made in Canada. Prior to 2004, the slings were labeled Medi-Man, 100% Polyester Shell, Rehabilitation Products Inc., Mississauga, Ontario, Canada L5T 1X7, made in Canada. The accessory is intended to be used with patient lift for the transfer of patient in hospitals, nursing homes, or other health care facilities by trained caregivers.
Code Information All lots of BHM/Medi-man Combi Slings released prior to 2/1/09: Part Numbers 626002, 626002M, Part 626002C, Part 626002C-M, Part 626003, Part 626003M and Part 626003X.
FEI Number 3002655016
Recalling Firm/
Manufacturer
B.H.M. Medical, Inc.
2001 Tanguay Street
Magog Canada
For Additional Information ContactMs. Traci Giovenco
800-323-1245 Ext. 54743
Manufacturer Reason
for Recall
Premature failure of the stitching at the junction of the shoulder strap and sling body of the BHM/Medi-Man Combi Sling used with patient lifts.
FDA Determined
Cause 2
Device Design
ActionArjo sent "Urgent Device Recall Customer Notification" cover letters dated April 15, 2009 to the end user accounts who received the affected BHM/Medi-Man Combi Slings, to the attention of the Administrator/Risk Manager, advising them of the potential for possible stitching failure that could lead to a hazardous situation. Enclosed with the letter was a copy of the BHM Medical Field Safety Notice, FSN 10/2009/ca, dated April 3, 2009 which was sent to their BHM customers in the U.S. on April 7, 2009. The FSN identified the part numbers of the BHM/Medi-Man Combi Slings, described the potential health hazard, and listed the actions to take. The accounts were instructed to inspect and assess their slings' suitability for use per the Slings Maintenance Instructions. If the sling is showing this kind of breakage, the account was instructed to remove it from use and discard it immediately. If the inspection reveals no deterioration of the stitching, the sling is safe to use. The accounts were requested to complete and return the enclosed FSN receipt form within 10 days, acknowledging the receipt and understanding of the letter, and indicating the number of slings in use prior to inspection, the number of slings in use after the inspection, and the number of slings rejected. The accounts will be contacted for the replacement of their eligible old slings (in use up to two years).
Quantity in Commerce1,396 slings
DistributionNationwide
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-