| | Class 2 Device Recall Stryker ExFen 5.0" extended fenestration catheter set for PainPump |  |
| Date Initiated by Firm | June 16, 2009 |
| Date Posted | August 19, 2009 |
| Recall Status1 |
Terminated 3 on October 27, 2010 |
| Recall Number | Z-1738-2009 |
| Recall Event ID |
52452 |
| 510(K)Number | K042405 K043466 |
| Product Classification |
Pump, infusion - Product Code FRN
|
| Product | Stryker ExFen 5.0" extended fenestration catheter set for PainPump, sterile, Stryker Instruments, Kalamazoo, MI; REF 500-50. |
| Code Information |
Lot Numbers: 2008011401 through 2008091101. |
Recalling Firm/ Manufacturer |
Stryker Instruments Div. of Stryker Corporation 4100 E. Milham Ave. Kalamazoo MI 49001
|
| For Additional Information Contact | 800-800-4236 Ext. 3808 |
Manufacturer Reason for Recall | The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Stryker Instruments issued a recall letter dated June 16, 2009 and instructed Consignees to cease using and destroy the product and to return a business reply form for credit via fax to Stryker Instruments Regulatory Department, 1-866-521-2762.
For further questions, contact Stryker Instruments at 1-800-800-4236 extension 3808. |
| Quantity in Commerce | 290 |
| Distribution | Worldwide Distribution -- Including the United States, Argentina, Australia, Brazil, Canada, England, France, Germany, Hong Kong, Italy, Japan, Latin America, Malaysia, Mexico, Netherlands, New Zealand, Singapore, South Africa, Spain, Sweden, Switzerland and Taiwan. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|