Class 2 Device Recall Stryker 5.0 catheter set for PainPump2 and AutoFuser PainPump |
|
Date Initiated by Firm |
June 16, 2009 |
Date Posted |
August 19, 2009 |
Recall Status1 |
Terminated 3 on October 27, 2010 |
Recall Number |
Z-1772-2009 |
Recall Event ID |
52452 |
510(K)Number |
K043466 K042405
|
Product Classification |
Pump, infusion - Product Code FRN
|
Product |
Stryker 5.0 catheter set for PainPump2 and AutoFuser PainPump, sterile, Stryker Instruments, Kalamazoo, MI; REF 0537-005-000. |
Code Information |
Lot Numbers: 8136753340 through 9131936639. |
Recalling Firm/ Manufacturer |
Stryker Instruments Div. of Stryker Corporation 4100 E. Milham Ave. Kalamazoo MI 49001
|
For Additional Information Contact |
800-800-4236 Ext. 3808
|
Manufacturer Reason for Recall |
The catheter may break when removed from the body, or the catheter may have kinks or be occluded such that the patient will not receive the intended quantity of medication.
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
Stryker Instruments issued a recall letter dated June 16, 2009 and instructed Consignees to cease using and destroy the product and to return a business reply form for credit via fax to Stryker Instruments Regulatory Department, 1-866-521-2762.
For further questions, contact Stryker Instruments at 1-800-800-4236 extension 3808. |
Quantity in Commerce |
89 |
Distribution |
Worldwide Distribution -- Including the United States, Argentina, Australia, Brazil, Canada, England, France, Germany, Hong Kong, Italy, Japan, Latin America, Malaysia, Mexico, Netherlands, New Zealand, Singapore, South Africa, Spain, Sweden, Switzerland and Taiwan. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|