 Class 2 Device Recall Outlook Pump IV Set w/Universal Spike, Ultrasite Inj Site, 1.2um Filter, Spin Lock Connector, 120 insee related information
Date Initiated by FirmNovember 24, 2009
Date PostedJanuary 26, 2010
Recall Status1 Terminated 3 on February 14, 2011
Recall NumberZ-0694-2010
Recall Event ID 53955
510(K)NumberK904518 
Product Classification Intravascular administration set - Product Code FPA
ProductOutlook Pump IV Set w/Universal Spike, Ultrasite Inj Site, 1.2um Filter, Spin Lock Connector, 120 in US3112F
Code Information Product Code 474001. Lot numbers: 61055314 exp 4/30/2012, 61060763 exp 5/31/2012, 61063024 exp 7/31/2012
FEI Number 2523676
Recalling Firm/
Manufacturer
B. Braun Medical, Inc.
901 Marcon Boulevard
Allentown PA 18109-9512
For Additional Information ContactScott Haff
610-266-0500
Manufacturer Reason
for Recall
System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe recalling firm issued a Product Removal Notice via US Postal Service Certified Mail with registered return receipt or FedEx to all customers and distributors in receipt of the affected product. The notices were sent 11/30/09 and 12/1/09. Direct customers also received letters addressed to Material Management that informed them of the recall and provided instruction for disposition of affected product in their inventory. Distributors received the direct customer recall letter along with a cover letter outlining their responsibilities to execute the recall and notify their customers. Questions are directed to the company's Customer Support Departments: US customers please call (800) 227-2862. Canadian customers, please call (800) 624-2920. International customers, please call (610) 691-5400.
Quantity in Commerce17088 units
DistributionThe products were shipped to distributors and medical facilities nationwide. The product was also shipped to Afghanistan, Canada and LU.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = FPA
-
-