Class 2 Device Recall Outlook Pump Add On Set for Epidural Administration w/Microbore tubing & Sliding Spin Lock Connector

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Outlook Pump Add On Set for Epidural Administration w/Microbore tubing & Sliding Spin Lock Connectorsee related information
Date Initiated by FirmNovember 24, 2009
Date PostedJanuary 26, 2010
Recall Status1 Terminated 3 on February 14, 2011
Recall NumberZ-0712-2010
Recall Event ID 53955
510(K)NumberK904518 
Product Classification Intravascular administration set - Product Code FPA
ProductOutlook Pump Add On Set for Epidural Administration w/Microbore tubing & Sliding Spin Lock Connector, 88 in
Code Information Product Code V7402. Lot number 61034231 exp 5/31/2012
Recalling Firm/
Manufacturer
B. Braun Medical, Inc.
901 Marcon Boulevard
Allentown PA 18109-9512
For Additional Information ContactScott Haff
610-266-0500
Manufacturer Reason
for Recall
System error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe recalling firm issued a Product Removal Notice via US Postal Service Certified Mail with registered return receipt or FedEx to all customers and distributors in receipt of the affected product. The notices were sent 11/30/09 and 12/1/09. Direct customers also received letters addressed to Material Management that informed them of the recall and provided instruction for disposition of affected product in their inventory. Distributors received the direct customer recall letter along with a cover letter outlining their responsibilities to execute the recall and notify their customers. Questions are directed to the company's Customer Support Departments: US customers please call (800) 227-2862. Canadian customers, please call (800) 624-2920. International customers, please call (610) 691-5400.
Quantity in Commerce1008 units
DistributionThe products were shipped to distributors and medical facilities nationwide. The product was also shipped to Afghanistan, Canada and LU.
Total Product Life CycleTPLC Device Report