| | Class 2 Device Recall AortaScan AMI 9700 |  |
| Date Initiated by Firm | August 23, 2010 |
| Date Posted | October 04, 2010 |
| Recall Status1 |
Terminated 3 on September 08, 2011 |
| Recall Number | Z-0004-2011 |
| Recall Event ID |
56411 |
| 510(K)Number | K082456 |
| Product Classification |
System, imaging, pulsed echo, ultrasonic - Product Code IYO
|
| Product | AortaScan AMI 9700
The AortaScan AMI 9700 is an ultrasound device that projects ultrasound energy into the mid-abdomen to obtain an image of the abdominal aorta for aortic diameter measurements |
| Code Information |
Serial numbers: A7001001, A7001005, A7001040, A7001041, A7001042, A7001043, A7001044, A7001045, A7001046, A7001047, A7001048, A7001049, A7001050, A7001051, A7001052, A7001053, A7001054, A7001055, A7001056, A7001057, A7001062, A7001065, A7001066, A7001067, A7001069, A7001071, A7001089, A7001090, A7001091, A7001092, A7001093, A7001094, A7001095, A7001096, A7001097, A7001098, A7001099, A7001100, A7001101, A7001102, A7001103, A7001106, A7001107, A7001108, A7001109, A7001110, A7001111, A7001112, A7001113, A7001114, A7001115, A7001116, A7001117, A7001118, A7001119, A7001120, A7001121, A7001122, A7001123, A7001124, A7001125, A7001127, A7001128, A7001129, A7001130, A7001131, A7001132, A7001133, A7001134, A7001135, A7001136, A7001138, and A7001139. |
Recalling Firm/ Manufacturer |
Verathon, Inc. 20001 N Creek Pkwy Bothell WA 98011-8218
|
| For Additional Information Contact | Verathon Medical Customer Care 800-331-2313 |
Manufacturer Reason for Recall | The AortaScan AMI 9700 devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function. |
FDA Determined Cause 2 | Software design |
| Action | Verathon sent an URGENT Medical Device Recall Notification letter dated August 12, 2010 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Verathon provided an addendum to the User Manual with instructions on how to reset the instrument. Customers were advised to:
" Add the enclosed addendum to their current Users Manual.
" Update the application software of the device to v2.7.0 or higher.
If the customers had questions or needed support for the software update process, they were to contact Verathon Medical Customer Care at (425) 867-1348 or (800) 331-2313 or
cservice@verathon.com. |
| Quantity in Commerce | 73 units |
| Distribution | Worldwide Distribution - USA and the countries of Australia, Canada, China, Colombia, European Union, United Kingdom, Saudi Arabia, and Singapore. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
| 510(K) Database | 510(K)s with Product Code = IYO
|
|
|
|