Class 2 Device Recall DEPUY ASR 300 ACETABULAR CUP SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall DEPUY ASR 300 ACETABULAR CUP SYSTEMsee related information
Date Initiated by FirmAugust 23, 2010
Date PostedMarch 24, 2011
Recall Status1 Terminated 3 on August 29, 2013
Recall NumberZ-1797-2011
Recall Event ID 57177
510(K)NumberK073413 
Product Classification Prothesis, hip, semi-constrained (metal uncemented acetabular component - Product Code KWA
ProductASR 300 Size 48, DEPUY ASR 300 ACETABULAR CUP SYSTEM. The DePuy ASR 300 Acetabular Cup System is part of a modular system designed to replace the natural articular surface of the hip joint in total hip replacement. The acetabular cup is designed as cobalt-chrome molybdenum (CoCrMo) alloy one-piece cup. The outer surface of the cup has a porous coating with the addition of a hydroxyapatite (HA) coating. The cups feature three spikes for adjunct fixation and are available in ten sizes. Prescription Use, Sterile The DePuy ASR 300 Acetabular Cup System is part of a modular system designed to replace the natural articular surface of the hip joint in total hip replacement. The acetabular cup is designed as cobalt-chrome molybdenum (CoCrMo) alloy one-piece cup. The outer surface of the cup has a porous coating with the addition of a hydroxyapatite (HA) coating. The cups feature three spikes for adjunct fixation and are available in ten sizes. The device is indicated for use in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, slipped capital femoral epiphysis and disability due to previous fusion, where bone stock is inadequate for other reconstruction techniques. Porous-coated ASR 300 Acetabular Cups are indicated for cementless application. The device is part of a modular system for use in total hip replacement in which the acetabular component articulates with a femoral component. The DePuy ASR 300 Acetabular Cup System is compatible with ASR femoral components.
Code Information Part Number 999830748, Size 48, All lots.
FEI Number 1818910
Recalling Firm/
Manufacturer
Depuy Orthopaedics, Inc.
700 Orthopaedic Dr
Warsaw IN 46582-3994
For Additional Information ContactASR Helpline
888-627-2677
Manufacturer Reason
for Recall
As part of our ongoing post-market surveillance of all products, DePuy is continually evaluating data from a variety of sources including national joint replacement registries, published literature, company sponsored clinical trials, internal complaints data and unpublished clinical research reports. DePuy Orthopaedics issued a Field Safety Notice in March 2010 after receiving new data from the
FDA Determined
Cause 2
Device Design
ActionDePuy Orthopaedics issued a Field Safety Notice in March 2010 to all affected customers. The Notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to share the Field Safety Notice with their organization and any organization where the ASR products may have been transferred. Customer were instructed not to implant the ASR devices. Patients who received the ASR System should be informed of this recall and instructed to return for a follow up visit. Patients with radiographic changes indicative of product failure should be addressed according to normal procedures. All other patients should be followed according to the April 22, 2010, and May 25, 2010 UK Medicines and Healthcare products Regulatory Agency (MHRA) Device Alerts. Follow up for all patients implanted with ASR acetabular cups at least annually for five years postoperatively. Beyond five years, follow up in accordance with locally agreed protocols. For patients who are symptomatic or implanted with a cup angle greater than 45 degrees, particulary where a small component has been implanted.: " Consider measuring cobalt and chromium ion levels in whle blood and/or performing cross sectional imaging including MRI or ultrasound scans. " If metal ion levels in whole blood are elevated above 7 parts per billion (ppb) for either metal ion, a second test should be performed three months after the first in order to identify patients who require closer surveillance, which may include cross sectional imaging " If MRI or ultrasound scan reveals soft tissue reactions, fluid collections or tissue masses, then revision surgery should be considered. DePuy representatives were to assist with returns of any remaining inventory. For questions regarding this recall call 574-372-7333.
Quantity in Commerce353
DistributionWorldwide
Total Product Life CycleTPLC Device Report