Class 2 Device Recall Brilliance Big Bore CT XRay System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Brilliance Big Bore CT XRay Systemsee related information
Date Initiated by FirmMarch 07, 2011
Date PostedMay 19, 2011
Recall Status1 Terminated 3 on February 22, 2013
Recall NumberZ-2207-2011
Recall Event ID 58510
510(K)NumberK033357 
Product Classification Computed Tomography X-Ray System - Product Code JAK
ProductBrilliance Big Bore Computed Tomography X-Ray System, Model Number: 728243. The units subject to correction are equipped with software version: 3.5.17030. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Code Information The following Serial Numbers of the Brilliance CT Big Bore system are subject to recall/software correction: 7006, 7060, 7154, 7298, and 7349.
FEI Number 1525965
Recalling Firm/
Manufacturer
Philips Medical Systems (Cleveland) Inc
595 Miner Road
Cleveland OH 44143-2131
For Additional Information Contact
440-483-7000
Manufacturer Reason
for Recall
A program software error in the algorithm produces image artifacts that could possibly lead to a misdiagnosis of patients who have small metal implants.
FDA Determined
Cause 2
Other
ActionA 'Field Safety Notice - Urgent - Medical Device Correction' notification was sent to the firm's customers via Certified mail on 3/8/2011. The notification letter explains the nature of the Metal Artifact Reduction/Metal Artifact Correction software error; the Brilliance CT X-Ray systems subject to correction and how to identify them; and the circumstances under which this error can occur. It also asks that the user stop running the MAR application on any of the affected Brilliance systems identified with the recalled serial numbers and software versions alluded to in the medical device correction notification. Lastly, the notification states that Philips Medical will provide a solution that will fix the issue and the notification lists contact names; telephone numbers; and E-mail addresses which the customer can contact for further information concerning this device issue, or customer support assistance.
Quantity in Commerce5 units
DistributionNationwide Distribution -- DC, IL, IN, MI, MO, NY, OH, OR, PA, TX, and VT.
Total Product Life CycleTPLC Device Report