• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 066 Date of Entry 05/25/2026 
FR Recognition Number 1-203
Standard
ISO  10079-1 Fourth edition 2022-03 [Including AMD1:2026]
Medical suction equipment - Part 1: Electrically powered suction equipment [Including Amendment 1 (2026)]
Scope/Abstract
This document specifies safety and performance requirements for electrically powered medical and surgical suction equipment. It applies to equipment used in health care facilities such as hospitals, for domiciliary care of patients and for field use and transport use.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized because it is scientifically and technically valid and not in conflict with existing regulatory policies.
Transition Period
FDA recognition of ISO 10079-1 Fourth edition 2022-03 [Rec# 1-157] will be superseded by recognition of ISO 10079-1 Fourth edition 2022-03 [Including AMD1:2026] [Rec# 1-203]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 1-157] until July 2, 2028. After this transition period, declarations of conformity to [Rec# 1-157] will not be accepted.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§868.1040 Algesimeter, Powered Class 2 BSI
§870.5050 Apparatus, Suction, Patient Care Class 2 DWM
§870.5910 Esophageal Thermal Regulation And Gastric Suctioning Device Class 2 PLA
§874.5350 Device, Antichoke, Suction Class 3 EWT
§878.4683 Negative Pressure Wound Therapy Non-Powered Suction Apparatus Class 2 OKO
§878.4780 Pump, Portable, Aspiration (Manual Or Powered) Class 2 BTA
§878.4780 Apparatus, Suction, Ward Use, Portable, Ac-Powered Class 2 JCX
§878.4780 Negative Pressure Wound Therapy Powered Suction Pump Class 2 OMP
§878.4780 Wound Drain Catheter System Class 2 OTK
§878.5040 System, Suction, Lipoplasty Class 2 MUU
§878.5040 System, Suction, Lipoplasty For Removal Class 2 QPB
Relevant FDA Guidance and/or Supportive Publications*
1. Guidance for Industry and FDA Reviewers/Staff: Guidance Document for Powered Suction Pump 510(k)s, issued September 1998.

2. ISO 10079-4 First edition 2021-08 Medical suction equipment - Part 4: General requirements.

3. ANSI/AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)].

4. ISO 5356-1 Third edition 2004-05 Anaesthetic and respiratory equipment - Conical connectors - Part 1: Cones and sockets.

5. ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.

6. IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION Medical devices - Part 1: Application of usability engineering to medical devices.

7. ISO 5359 Fourth edition 2014-10 Anaesthetic and respiratory equipment - Low-pressure hose assemblies for use with medical gases [Including AMENDMENT 1 (2017)].

8. ISO 4135 Fourth edition 2022-01 Anaesthetic and respiratory equipment - Vocabulary.

9. Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT) - Class II Special Controls Guidance for Industry and FDA Staff, issued November 2010.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Sidra Mirza
  FDA/OC/CDRH/OPEQ/OHTI/DHTIC/
  301-796-6471
  sidra.mirza@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Anesthesiology
*These are provided as examples and others may be applicable.
-
-