| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | SYNCHROMED(TM) INFUSION SYSTEM |
| Generic Name | Pump, infusion, implanted, programmable |
| Applicant | Medtronic, Inc. 7000 Central Ave., NE Minneapolis, MN 55432 |
| PMA Number | P860004 |
| Supplement Number | S010 |
| Date Received | 04/10/1989 |
| Decision Date | 06/26/1989 |
| Product Code |
LKK |
| Advisory Committee |
General Hospital |
| Supplement Type | Normal 180 Day Track |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
|
|