| Device | SYNCHROMED II INFUSION SYSTEM |
| Generic Name | Pump, infusion, implanted, programmable |
| Applicant | Medtronic, Inc. 7000 Central Ave., NE Minneapolis, MN 55432 |
| PMA Number | P860004 |
| Supplement Number | S138 |
| Date Received | 08/10/2010 |
| Decision Date | 03/03/2011 |
| Product Code |
LKK |
| Advisory Committee |
General Hospital |
| Supplement Type | 135 Review Track For 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | Yes |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement APPROVAL OF THE FOLLOWING MANUFACTURING PROCESS CHANGES RELATED TO THE BASELINE TEST OF THE IMPLANTABLE SYNCHROMED II INFUSION PUMP, MODEL 8637: 1) AUTOMATION OF THE TRANSFER OF BASELINE TEST DATA; 2) CHANGE THE DUTY CYCLE AND LOWER THE SPECIFICATION LIMIT OF THE DYNAMIC CURRENT TEST; AND 3) AN INCREASE TO THE TIME TO MEASURE THE MICRO AWAKE CURRENT. |