| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | HIRESOLUTION BIONIC EAR SYSTEM |
| Generic Name | Implant, cochlear |
| Applicant | Advanced Bionics 28515 Westinghouse Place Valencia, CA 91355 |
| PMA Number | P960058 |
| Supplement Number | S058 |
| Date Received | 05/07/2007 |
| Decision Date | 07/02/2007 |
| Product Code |
MCM |
| Advisory Committee |
Ear Nose & Throat |
| Supplement Type | Normal 180 Day Track |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement APPROVAL FOR THREE ?VACUUM BAKE AIR FLOW ENHANCEMENTS? TO THE HIRES90K COCHLEAR IMPLANT, AS FOLLOWS: 1) THE BOTTOM COVER SHIELD INSIDE THE HIRES90K CASE WAS MODIFIED FROM HAVING A ?V?-SHAPED CHANNEL TO HAVING A LARGER ?U?-SHAPED CHANNEL. 2) HOLES WERE ADDED TO THE INSULATING KAPTON DISC. 3) SLOTS WERE IMPLEMENTED INSIDE THE FEEDTHROUGH CASE. THE DEVICE, AS MODIFIED, WILL BE MARKETED UNDER THE TRADE NAME HIRESOLUTION BIONIC EAR SYSTEM AND IS INDICATED FOR INDIVIDUALS AGED 12-MONTHS AND OLDER WITH SEVERE-TO-PROFOUND SENSORINEURAL HEARING LOSS TO RESTORE A LEVEL OF AUDITORY SENSATION VIA ELECTRICAL STIMULATION OF THE AUDITORY NERVE. |
|
|