| Device | Novoste™ Beta-Cath™ System |
| Generic Name | Intravascular radiation delivery system |
| Applicant | Best Vascular, Inc. 4350 International Blvd. Norcross, GA 30093 |
| PMA Number | P000018 |
| Supplement Number | S051 |
| Date Received | 05/05/2023 |
| Decision Date | 09/26/2024 |
| Product Code |
MOU |
| Advisory Committee |
Cardiovascular |
| Supplement Type | Normal 180 Day Track |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement approval for a reduction in the existing minimum dose rate of the Jacketed Radiation Source Trains (JRST) of the Beta-Cath 3.5Fr System from 64.4 mGy/s to 58.5 mGy/s. |