Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceCONTAK RENEWAL 1
Generic NameImplantable pulse generator, pacemaker (non-CRT)
ApplicantBoston Scientific Corp
4100 Hamline Ave. N
St Paul, MN 55112
PMA NumberP010012
Supplement NumberS076
Date Received09/26/2005
Decision Date11/10/2005
Product Code LWP 
Advisory Committee Cardiovascular
Supplement TypeReal-Time Process
Supplement Reason Change Design/Components/Specifications/Material
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
APPROVAL TO REPLACE THE EXISTING POLYIMIDE TUBING THAT PROVIDES INSULATION IN THE PULSE GENERATOR HEADER ASSEMBLY WITH POLYETHERETHERKE-TONE (PEEK) TUBING.
-
-