| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | CareLink SmartSync Claria Amplia Compia App, CareLink SmartSync Cobalt, Crome App, CareLink SmartSync Viva Brava Evera A |
| Generic Name | Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) |
| Applicant | Medtronic Cardiac Rhythm Disease Management 8200 Coral Sea St. N.E Mounds View, MN 55112 |
| PMA Number | P010031 |
| Supplement Number | S867 |
| Date Received | 09/16/2024 |
| Decision Date | 11/14/2024 |
| Product Code |
NIK |
| Advisory Committee |
Cardiovascular |
| Supplement Type | Real-Time Process |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement approval for software changes to the CareLink SmartSync Device Manager to address instances of inappropriate display of the device electrical reset message, and to correct the functionality of the abort button |
|
|