| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | Trim Change |
| Generic Name | Stimulator, spinal-cord, totally implanted for pain relief |
| Applicant | ABBOTT MEDICAL 6901 Preston Rd. Plano, TX 75024 |
| PMA Number | P010032 |
| Supplement Number | S129 |
| Date Received | 06/02/2017 |
| Decision Date | 06/28/2017 |
| Product Code |
LGW |
| Advisory Committee |
Neurology |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Software update to the Automated Test Equipment (ATE) that will set the value of Open Threshold and Switch Threshold for all Proclaim, Infinity, and Proclaim DRG devices to a constant value of 253 during manufacturing instead of calculating the trim value using a formula. |
|
|