| Device | Precision Montage MRI Spinal Cord Stimulator System |
| Generic Name | Stimulator, spinal-cord, totally implanted for pain relief |
| Applicant | Boston Scientific Corp 25155 Rye Canyon Loop Valencia, CA 91355 |
| PMA Number | P030017 |
| Supplement Number | S308 |
| Date Received | 12/12/2017 |
| Decision Date | 03/12/2018 |
| Product Code |
LGW |
| Advisory Committee |
Neurology |
| Supplement Type | Real-Time Process |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Approval for an update to the Precision Montage and Precision Montage MRI Spinal Cord Stimulator (SCS) System firmware to adjust the temperature sensor calibration with an aim to improve the accuracy of the sensor reading, associated changes to the Printed Circuit Board Assembly and device level Automated Test Equipment software, and rework of the IPG devices that failed the device level temperature sensor test. |