| Device | TAXUS EXPRESS2 MONORAIL AND OVER-THE-WIRE CORONARY STENT SYSTEMS |
| Generic Name | Coronary drug-eluting stent |
| Applicant | Boston Scientific Corp One Scimed Pl. Maple Grove, MN 55311 |
| PMA Number | P030025 |
| Supplement Number | S002 |
| Date Received | 04/05/2004 |
| Decision Date | 05/05/2004 |
| Withdrawal Date | 12/06/2018 |
| Product Code |
NIQ |
| Advisory Committee |
Cardiovascular |
| Supplement Type | Special (Immediate Track) |
| Expedited Review Granted? | No |
| Combination Product | Yes |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement APPROVAL TO IMPLEMENT THE FOLLOWING: 1) THE ADDITION OF AN ADDITIONAL IN-PROCESS INSPECTION, 2) MODIFICATION OF A CURRENT IN-PROCESS INSPECTION, AND 3) MODIFICATION OF A MANUFACTURING PROCESS TO ADDRESS COMPLAINTS RELATED TO FAILURE OF THE DELIVERY SYSTEM BALLOON TO PROPERLY DEFLATE FOLLOWING STENT DEPLOYMENT. |