Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceFUJI COMPUTED RADIOGRAPHY MAMMOGRAPHY SUITE
Generic NameFull field digital, system, x-ray, mammographic
Regulation Number892.1715
ApplicantFujifilm Medical System U.S.A., Inc.
419 W. Ave.
Stamford, CT 06902
PMA NumberP050014
Supplement NumberS001
Date Received10/28/2008
Decision Date01/09/2009
Reclassified Date12/06/2010
Product Code MUE 
Advisory Committee Radiology
Supplement TypeReal-Time Process
Supplement Reason Change Design/Components/Specifications/Material
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
APPROVAL FOR: 1) SOFTWARE VERSION 5.0 WITH 12-BIT OUTPUT ENABLED. CURRENTLY FOR MAMMOGRAPHY,ONLY 10-BIT OUTPUT IS ENABLED SO THIS CHANGE WOULD GIVE THE END USER THE OPTION OF EITHER10-BIT OR 12-BIT OUTPUT FOR MAMMOGRAPHY IMAGES; AND2) LABELING CHANGES MADE TO THE FCRM FUJI CR FOR MAMMOGRAPHY USER GUIDE 1STEDITION JULY, 2006 TO INCLUDE:A) INFORMATION IN CHAPTERS 3 AND 4 WERE REVERSED SO THAT EXPOSURE FACTORS AREDISCUSSED BEFORE AUTOMATIC EXPOSURE CONTROL (AEC) CALIBRATION;B) MORE SPECIFIC X-RAY EXPOSURE INSTRUCTIONS KVP RANGES, AND RECOMMENDED TARGETAND FILTER SELECTION HAVE BEEN ADDED TO HELP FCRM USERS SELECT EXPOSURE SETTINGSBASED ON CONTRAST-TO-NOISE RATIO (CNR); ANDC) INSTRUCTIONS ARE PROVIDED DURING THE SETUP OF THE AEC SYSTEM SO THAT EXPOSURETECHNIQUES WILL BE BASED ON THE X-RAY EXPOSURE SETTINGS RECOMMENDED IN CHAPTER 3.
-
-