| Device | PROMETRA PROGRAMMABLE IMPLANTABLE PUMP |
| Generic Name | Pump, infusion, implanted, programmable |
| Applicant | Infusyn Therapeutics, LLC 121 Shelley Dr. 2h Hackettstown, NJ 07840 |
| PMA Number | P080012 |
| Supplement Number | S032 |
| Date Received | 10/26/2015 |
| Decision Date | 11/24/2015 |
| Product Code |
LKK |
| Advisory Committee |
General Hospital |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | Yes |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement TO MODIFY THE MANUFACTURING PROCEDURE FOR THE PATIENT THERAPY CONTROLLER (PTC) TO REROUTE THE PTC'S POWER CABLE WITHIN THE ASSEMBLY TO PREVENT IT FROM DAMAGE. ADDITIONALLY, UPDATE THE FINAL RELEASE TEST TO ADD A "BOLUS TEST" TO FURTHER VERIFY THE OPERATION OF THE PTC PRIOR TO RELEASE. |
|
|