| Device | Pipeline Flex Embolization Device and Pipeline Flex Embolization Device with Shield Technology |
| Generic Name | Intracranial aneurysm flow diverter |
| Applicant | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular 9775 Toledo Way Irvine, CA 92618 |
| PMA Number | P100018 |
| Supplement Number | S033 |
| Date Received | 07/12/2021 |
| Decision Date | 07/27/2021 |
| Product Code |
OUT |
| Advisory Committee |
Neurology |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Modifications to hypotube tensile strength test of Pipeline Flex Embolization Device and Pipeline Flex Embolization Device with Shield Technology, using an in-process manufacturing acceptance criteria of >= 3.0 N for the test, and shifting the hypotube tensile strength test from the final inspection step to an earlier subassembly step. |