| Device | Elecsys Anti-HBC Immunoassay test system |
| Generic Name | Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen) |
| Regulation Number | 866.3172 |
| Applicant | Roche Diagnostics Corp. 9115 Hague Rd. Indianapolis, IN 46250 |
| PMA Number | P100032 |
| Supplement Number | S014 |
| Date Received | 02/21/2017 |
| Decision Date | 08/16/2017 |
| Reclassified Date | 10/20/2025 |
| Product Code |
LOM |
| Advisory Committee |
Microbiology |
| Supplement Type | Normal 180 Day Track |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Approval for 1) the inclusion of a second antigen source for the recombinant hepatitis B core antigen used in Reagent 1 of the test kit; 2) an update in the device by changing the standardization traceability to the World Health Organization Standard NIBSC 95/522; 3) extension of the reagent rackpack onboard stability from 4 weeks to 8 weeks; 4) addition of K3-EDTA plasma as a specimen type; and 5) modification of the device name. |