| Device | Promus Premier Everolimus-Eluting Platinum Chromium Coronary Stent System |
| Generic Name | Coronary drug-eluting stent |
| Applicant | Boston Scientific Corp One Scimed Pl. Maple Grove, MN 55311 |
| PMA Number | P110010 |
| Supplement Number | S144 |
| Date Received | 08/04/2017 |
| Decision Date | 08/31/2017 |
| Product Code |
NIQ |
| Advisory Committee |
Cardiovascular |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | Yes |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Update the BSC2000-2 cycle conducted in Chamber 2 at the BSC Coventry, RI facility with new Programmable Logic Controller (PLC) software, an additional vacuum pump booster, and a new validation documentation structure used to support the system. |