| Device | Medtronic MiniMed 630G System (MM630G) |
| Generic Name | Automated insulin dosing , threshold suspend |
| Applicant | Medtronic Minimed 18000 Devonshire St. Northridge, CA 91325 |
| PMA Number | P150001 |
| Supplement Number | S105 |
| Date Received | 02/13/2023 |
| Decision Date | 03/13/2023 |
| Product Code |
OZO |
| Advisory Committee |
Clinical Chemistry |
| Supplement Type | 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Manufacturing change to expand the welding parameters used when manufacturing the pump case assembly. The pump case assembly is a component of the MiniMed 630G, MiniMed 670G, and MiniMed 770G systems. |