| Device | Surpass Elite Flow Diverter System |
| Generic Name | Intracranial aneurysm flow diverter |
| Applicant | Stryker Neurovascular 47900 Bayside Pkwy. Fremont, CA 94538 |
| PMA Number | P170024 |
| Supplement Number | S012 |
| Date Received | 05/01/2023 |
| Decision Date | 05/18/2024 |
| Product Code |
OUT |
| Advisory Committee |
Neurology |
| Supplement Type | Normal 180 Day Track |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement the Surpass Elite Flow Diverter System that was developed through design modifications to the approved Surpass Evolve Flow Diverter System (P170024/S003). The design modifications consist of changes to the delivery system resheath pad and distal protection assembly, the stent braid to include a 5.25 mm diameter stent model and picks per inch specifications to be the same for all stent diameters, and manufacturing process for the stent braid and delivery system. |