Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceProspera SCS System, Resilience Surgical Lead
Generic NameStimulator, spinal-cord, totally implanted for pain relief
ApplicantBiotronik Nro, Inc.
6024 Jean Rd.
Lake Oswego, OR 97035
PMA NumberP210037
Supplement NumberS003
Date Received12/14/2023
Decision Date03/13/2024
Product Code LGW 
Advisory Committee Neurology
Supplement TypeReal-Time Process
Supplement Reason Change Design/Components/Specifications/Material
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
Approval for an updated Power Delivery Controller software change to bypass the gas gauge within the charger and instead determine internal battery State of Charge based on measurements from the actual battery voltage via the battery charger integrated circuit
-
-